Volumn 3, Issue 4, 1998, Pages 255-260

Validation of the A and D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects' characteristics

Author keywords

Ambulatory monitoring; BHS protocol; Blood pressure; Device; Validation

Indexed keywords

MERCURY;

EID: 0031774809     PISSN: 13595237     EISSN: None     Source Type: Journal    
DOI: None     Document Type: Article
Times cited : (125)

References (21)
  • 1
    • 0027210887 scopus 로고
    • Automated ambulatory blood pressure and self-measured blood pressure monitoring devices: Their role in the diagnosis and management of hypertension
    • American College of Physicians. Automated ambulatory blood pressure and self-measured blood pressure monitoring devices: their role in the diagnosis and management of hypertension. Ann Intern Med 1993; 118:889-892.
    • (1993) Ann Intern Med , vol.118 , pp. 889-892
  • 2
    • 0027230992 scopus 로고
    • Ambulatory blood pressure monitoring can improve cost-effective management of hypertension
    • Krakoff LR. Ambulatory blood pressure monitoring can improve cost-effective management of hypertension. Am J Hypertens 1993; 6 (suppl 6):220S-224S.
    • (1993) Am J Hypertens , vol.6 , Issue.6 SUPPL.
    • Krakoff, L.R.1
  • 3
    • 0026473183 scopus 로고
    • Ambulatory blood pressure monitoring: A critical review of the current methods to handle outliers
    • Berardi L, Chau NP, Chanudet X, Vilar J, Larroque P. Ambulatory blood pressure monitoring: a critical review of the current methods to handle outliers. J Hypertens 1992; 10:1243-1248.
    • (1992) J Hypertens , vol.10 , pp. 1243-1248
    • Berardi, L.1    Chau, N.P.2    Chanudet, X.3    Vilar, J.4    Larroque, P.5
  • 4
    • 0030981349 scopus 로고    scopus 로고
    • Ambulatory blood pressure monitoring editing criteria: Is a standardization needed?
    • Winnicki M, Canali C, Mormino P, Palatini P. Ambulatory blood pressure monitoring editing criteria: is a standardization needed? Am J Hypertens 1997; 10:419-427.
    • (1997) Am J Hypertens , vol.10 , pp. 419-427
    • Winnicki, M.1    Canali, C.2    Mormino, P.3    Palatini, P.4
  • 5
    • 0028931962 scopus 로고
    • Investigation of ambulatory blood pressure monitoring data editing criteria
    • Lee DR, Farmer AJ, Swift CG, Jackson SHD. Investigation of ambulatory blood pressure monitoring data editing criteria. J Hum Hypertens 1995; 9:195-198.
    • (1995) J Hum Hypertens , vol.9 , pp. 195-198
    • Lee, D.R.1    Farmer, A.J.2    Swift, C.G.3    Jackson, S.H.D.4
  • 6
    • 0028818147 scopus 로고
    • State of the market. A review of ambulatory blood pressure monitoring devices
    • O'Brien E, Atkins N, Staessen J. State of the market. A review of ambulatory blood pressure monitoring devices. Hypertension 1995; 26:835-842.
    • (1995) Hypertension , vol.26 , pp. 835-842
    • O'Brien, E.1    Atkins, N.2    Staessen, J.3
  • 7
    • 0027479012 scopus 로고
    • National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers
    • White WB, Berson AS, Robbins C, Jamieson MJ, Prisant LM, Roccella E, et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993; 21:504-509.
    • (1993) Hypertension , vol.21 , pp. 504-509
    • White, W.B.1    Berson, A.S.2    Robbins, C.3    Jamieson, M.J.4    Prisant, L.M.5    Roccella, E.6
  • 8
    • 0025071770 scopus 로고
    • The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems
    • O'Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, O'Malley K, et al. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8:607-619.
    • (1990) J Hypertens , vol.8 , pp. 607-619
    • O'Brien, E.1    Petrie, J.2    Littler, W.3    De Swiet, M.4    Padfield, P.L.5    O'Malley, K.6
  • 9
    • 0001849046 scopus 로고
    • The British Hypertension Society protocol for the evaluation of blood pressure measuring devices
    • O'Brien, Petrie J, Littler W, de Swiet M, Padfield PL, Altman DG, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11 (suppl 2):S43-S63.
    • (1993) J Hypertens , vol.11 , Issue.2 SUPPL.
    • O'Brien1    Petrie, J.2    Littler, W.3    De Swiet, M.4    Padfield, P.L.5    Altman, D.G.6
  • 11
    • 0000065032 scopus 로고
    • Validation of the accuracy of the A&D TM-2420 model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance
    • Palatini P, Penzo M, Canali C, Pessina AC. Validation of the accuracy of the A&D TM-2420 model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance. J Ambul Monitor 1991; 4:281-288.
    • (1991) J Ambul Monitor , vol.4 , pp. 281-288
    • Palatini, P.1    Penzo, M.2    Canali, C.3    Pessina, A.C.4
  • 12
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinic measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinic measurement. Lancet 1986; 1:307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 13
    • 0001816125 scopus 로고
    • Performance reliability of the del Mar Avionics non-invasive ambulatory blood pressure instrument in clinical use
    • Kennedy HL, Padgett NE, Horan MJ. Performance reliability of the Del Mar Avionics non-invasive ambulatory blood pressure instrument in clinical use. Ambul Electrocardiol 1979; 4:13-17.
    • (1979) Ambul Electrocardiol , vol.4 , pp. 13-17
    • Kennedy, H.L.1    Padgett, N.E.2    Horan, M.J.3
  • 14
    • 0025873385 scopus 로고
    • Assessment of ambulatory blood pressure recorders: Accuracy and clinical performance
    • White WB. Assessment of ambulatory blood pressure recorders: accuracy and clinical performance. Clin Invest Med 1991; 14:202-211.
    • (1991) Clin Invest Med , vol.14 , pp. 202-211
    • White, W.B.1
  • 15
    • 0025873917 scopus 로고
    • Accuracy of the Space Labs 90207 determined by the British Hypertension Society protocol
    • O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Space Labs 90207 determined by the British Hypertension Society protocol. J Hypertens 1991; 9 (suppl 5):S25-S31.
    • (1991) J Hypertens , vol.9 , Issue.5 SUPPL.
    • O'Brien, E.1    Mee, F.2    Atkins, N.3    O'Malley, K.4
  • 16
    • 0026395370 scopus 로고
    • Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society protocol
    • O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society protocol. J Hypertens 1991; 9 (suppl 5):S9-S15.
    • (1991) J Hypertens , vol.9 , Issue.5 SUPPL.
    • O'Brien, E.1    Mee, F.2    Atkins, N.3    O'Malley, K.4
  • 17
    • 0025800399 scopus 로고
    • Accuracy of the del Mar Avionics Pressurometer IV determined by the British Hypertension Society protocol
    • O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society protocol. J Hypertens 1991; 9 (suppl 5):S1-S7.
    • (1991) J Hypertens , vol.9 , Issue.5 SUPPL.
    • O'Brien, E.1    Mee, F.2    Atkins, N.3    O'Malley, K.4
  • 19
    • 0025352835 scopus 로고
    • The accuracy and precision of the Accutracker ambulatory blood pressure monitor
    • Appel LJ, Whelton PK, Seidler AJ, Patel AR, Klag MJ. The accuracy and precision of the Accutracker ambulatory blood pressure monitor. Am J Epidemiol 1990; 132:343-354.
    • (1990) Am J Epidemiol , vol.132 , pp. 343-354
    • Appel, L.J.1    Whelton, P.K.2    Seidler, A.J.3    Patel, A.R.4    Klag, M.J.5
  • 20
    • 0000966943 scopus 로고    scopus 로고
    • Validation of the SpaceLabs 90207 automatic non-invasive blood pressure monitor in elderly subjects
    • Iqbal P, Fotherby MD, Potter JF. Validation of the SpaceLabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996; 1:367-373.
    • (1996) Blood Press Monit , vol.1 , pp. 367-373
    • Iqbal, P.1    Fotherby, M.D.2    Potter, J.F.3
  • 21
    • 0000874806 scopus 로고    scopus 로고
    • Accuracy of the SpaceLabs 90207 ambulatory blood pressure monitor in children and adolescents
    • Belsha CW, Wells TG, Rice HB, Neaville WA, Berry PL. Accuracy of the SpaceLabs 90207 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit 1996; 1:127-133.
    • (1996) Blood Press Monit , vol.1 , pp. 127-133
    • Belsha, C.W.1    Wells, T.G.2    Rice, H.B.3    Neaville, W.A.4    Berry, P.L.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.