Telemetry guided pacemaker programming: Impact of output amplitude and the use of low threshold leads on projected pacemaker longevity

Author keywords

Battery telemetry; Low threshold electrode; Pacemaker programming; Projected pacemaker longevity; Safety margin

Indexed keywords

ARTICLE; ARTIFICIAL HEART PACEMAKER; CLINICAL ARTICLE; COST EFFECTIVENESS ANALYSIS; ELECTRIC POTENTIAL; HEART PACING; HUMAN; PACEMAKER ELECTRODE; SAFETY; SYSTEM ANALYSIS; TELEMETRY;

EID: 0031723992     PISSN: 01478389     EISSN: None     Source Type: Journal    
DOI: 10.1111/j.1540-8159.1998.tb01123.x     Document Type: Article
Times cited : (9)

References (50)
  • 1
    • 0024722618 scopus 로고
    • From the editor... Pacemaker longevity
    • Furman S. From the editor... Pacemaker longevity. PACE 1989; 12:1437-1438.
    • (1989) PACE , vol.12 , pp. 1437-1438
    • Furman, S.1
  • 2
    • 0029684941 scopus 로고    scopus 로고
    • Pacemaker miniaturization: A good trend?
    • Cazeau S, Ritter P, Lazarus A, et al. Pacemaker miniaturization: A good trend? PACE 1996; 19: 1-3.
    • (1996) PACE , vol.19 , pp. 1-3
    • Cazeau, S.1    Ritter, P.2    Lazarus, A.3
  • 3
    • 0029690331 scopus 로고    scopus 로고
    • Should the race for pulse generator miniaturization be stopped?
    • Daubert JC. Should the race for pulse generator miniaturization be stopped? PACE 1996; 19:4-6.
    • (1996) PACE , vol.19 , pp. 4-6
    • Daubert, J.C.1
  • 4
    • 0343236196 scopus 로고
    • Development of low-stimulation-threshold, low-polarization electrodes
    • SS Barold, J Mugica (eds.): Mount Kisco, New York, Futura Publishing Company
    • Mond HG. Development of low-stimulation-threshold, low-polarization electrodes. In SS Barold, J Mugica (eds.): New Perspectives in Cardiac Pacing. Mount Kisco, New York, Futura Publishing Company, 1991, pp. 133-162.
    • (1991) New Perspectives in Cardiac Pacing , pp. 133-162
    • Mond, H.G.1
  • 5
    • 0344650511 scopus 로고
    • The porous titanium steroid eluting electrode: A double blind study assessing the stimulation threshold effects of steroid
    • Mond H, Stokes K, Helland J, et al. The porous titanium steroid eluting electrode: A double blind study assessing the stimulation threshold effects of steroid. PACE 1988; 11:214-219.
    • (1988) PACE , vol.11 , pp. 214-219
    • Mond, H.1    Stokes, K.2    Helland, J.3
  • 6
    • 0024204992 scopus 로고
    • Clinical experience with new leads
    • Mugica J, Duconge B, Henry L, et al. Clinical experience with new leads. PACE 1988; 11:1745-1752.
    • (1988) PACE , vol.11 , pp. 1745-1752
    • Mugica, J.1    Duconge, B.2    Henry, L.3
  • 7
    • 0024273775 scopus 로고
    • Composite electrode tips containing externally placed drug releasing collars
    • Brewer G, Mathivanar R, Skalsky M, et al. Composite electrode tips containing externally placed drug releasing collars. PACE 1988; 11:1760-1769.
    • (1988) PACE , vol.11 , pp. 1760-1769
    • Brewer, G.1    Mathivanar, R.2    Skalsky, M.3
  • 8
    • 0024379433 scopus 로고
    • A comparison of platinized grooved electrode performance with ring-tip electrodes
    • Midei MG, Jones BR, Brinker JA. A comparison of platinized grooved electrode performance with ring-tip electrodes. PACE 1989; 12:752-756.
    • (1989) PACE , vol.12 , pp. 752-756
    • Midei, M.G.1    Jones, B.R.2    Brinker, J.A.3
  • 9
    • 0024450644 scopus 로고
    • Lower chronic stimulation in the carbon-tip than in the platinum-tip endocardial electrode: A randomized study
    • Thuesen L, Jensen PJ, Vejby-Christensen H. Lower chronic stimulation in the carbon-tip than in the platinum-tip endocardial electrode: A randomized study. PACE 1989; 12:1592-1595.
    • (1989) PACE , vol.12 , pp. 1592-1595
    • Thuesen, L.1    Jensen, P.J.2    Vejby-Christensen, H.3
  • 10
    • 0025045994 scopus 로고
    • Stimulation characteristics of a steroid-eluting electrode compared with three conventional electrodes
    • Klein HH, Steinberger J, Knake W. Stimulation characteristics of a steroid-eluting electrode compared with three conventional electrodes. PACE 1990; 13:134-137.
    • (1990) PACE , vol.13 , pp. 134-137
    • Klein, H.H.1    Steinberger, J.2    Knake, W.3
  • 11
    • 0025696084 scopus 로고
    • Drug-eluting collar - A new approach to reducing threshold
    • Wilson A, Cowling R, Mathivanar et al. Drug-eluting collar - A new approach to reducing threshold. PACE 1990; 13:1876-1878.
    • (1990) PACE , vol.13 , pp. 1876-1878
    • Wilson, A.1    Cowling, R.2    Mathivanar3
  • 12
    • 0025649796 scopus 로고
    • Steroid tipped leads versus porous platinum permanent pacemaker leads: A controlled study
    • Wish M, Swartz J, Cohen A, et al. Steroid tipped leads versus porous platinum permanent pacemaker leads: A controlled study. PACE 1990; 13:1887-1890.
    • (1990) PACE , vol.13 , pp. 1887-1890
    • Wish, M.1    Swartz, J.2    Cohen, A.3
  • 13
    • 2242428032 scopus 로고
    • The impact of steroid eluting leads on long-term pacing in the atrium and ventricle
    • Sutton R, Guneri S. The impact of steroid eluting leads on long-term pacing in the atrium and ventricle. Eur JCPE 1991; 1:10-15.
    • (1991) Eur JCPE , vol.1 , pp. 10-15
    • Sutton, R.1    Guneri, S.2
  • 14
    • 0026512433 scopus 로고
    • The electrode-tissue interface: The revolutionary role of steroid elution
    • Mond HG, Stokes KB. The electrode-tissue interface: The revolutionary role of steroid elution. PACE 1992; 15:95-107.
    • (1992) PACE , vol.15 , pp. 95-107
    • Mond, H.G.1    Stokes, K.B.2
  • 15
    • 0027315354 scopus 로고
    • Do steroid-eluting electrodes really have better performance than other state-of-the-art designs?
    • Santini M, De Seta F. Do steroid-eluting electrodes really have better performance than other state-of-the-art designs? PACE 1993; 16:722-728.
    • (1993) PACE , vol.16 , pp. 722-728
    • Santini, M.1    De Seta, F.2
  • 16
    • 0027215372 scopus 로고
    • A randomized comparison of a bipolar steroid-eluting electrode and a bipolar microporous platinum electrode: Implications for long-term programming
    • Gillis AM, Rothschild JM, Hillier K, et al. A randomized comparison of a bipolar steroid-eluting electrode and a bipolar microporous platinum electrode: Implications for long-term programming. PACE 1993; 16:964-970.
    • (1993) PACE , vol.16 , pp. 964-970
    • Gillis, A.M.1    Rothschild, J.M.2    Hillier, K.3
  • 17
    • 0027310480 scopus 로고
    • Acute and long term ventricular stimulation thresholds with a new iridium oxide-coated electrode
    • Del Bufalo AGA, Schlaepfer J, Fromer M, et al. Acute and long term ventricular stimulation thresholds with a new iridium oxide-coated electrode. PACE 1993; 16:1240-1244.
    • (1993) PACE , vol.16 , pp. 1240-1244
    • Del Bufalo, A.G.A.1    Schlaepfer, J.2    Fromer, M.3
  • 18
    • 0027932703 scopus 로고
    • A new steroid eluting screw-in electrode
    • Fröhlig G, Schwaab B, Schwerdt H, et al. A new steroid eluting screw-in electrode. PACE 1994; 17:1134-1142.
    • (1994) PACE , vol.17 , pp. 1134-1142
    • Fröhlig, G.1    Schwaab, B.2    Schwerdt, H.3
  • 19
    • 0028071593 scopus 로고
    • Clinical experience with a new lead combining active fixation with steroid elution
    • Payne GE, Dunham DG, Ashok CR, et al. Clinical experience with a new lead combining active fixation with steroid elution. PACE 1994; 17:1665-1671.
    • (1994) PACE , vol.17 , pp. 1665-1671
    • Payne, G.E.1    Dunham, D.G.2    Ashok, C.R.3
  • 20
    • 0027959723 scopus 로고
    • Comparison of threshold values between steroid and nonsteroid unipolar membrane leads
    • Svensson O, Karlsson JE, Binner L, et al. Comparison of threshold values between steroid and nonsteroid unipolar membrane leads. PACE 1994; 17:2008-2011.
    • (1994) PACE , vol.17 , pp. 2008-2011
    • Svensson, O.1    Karlsson, J.E.2    Binner, L.3
  • 21
    • 0028894430 scopus 로고
    • Programming of the pacing impulse in pacemakers connected to steroid lead systems
    • Schuchert A, Cappato R, Kuck KH, et al. Programming of the pacing impulse in pacemakers connected to steroid lead systems. PACE 1995; 18:318-322.
    • (1995) PACE , vol.18 , pp. 318-322
    • Schuchert, A.1    Cappato, R.2    Kuck, K.H.3
  • 22
    • 0024344413 scopus 로고
    • Do electrode and lead differences for permanent cardiac pacing translate into clinically demonstrable differences?
    • Bourke JP, Howell L, Murray A, et al. Do electrode and lead differences for permanent cardiac pacing translate into clinically demonstrable differences? PACE 1989; 12:1419-1425.
    • (1989) PACE , vol.12 , pp. 1419-1425
    • Bourke, J.P.1    Howell, L.2    Murray, A.3
  • 23
    • 0026529384 scopus 로고
    • The safety and efficacy of chronic ventricular pacing at 1.6 Volts using a steroid eluting lead
    • Stamato NJ, O'Toole MF, Fetter JG, et al. The safety and efficacy of chronic ventricular pacing at 1.6 Volts using a steroid eluting lead. PACE 1992; 15:248-251.
    • (1992) PACE , vol.15 , pp. 248-251
    • Stamato, N.J.1    O'Toole, M.F.2    Fetter, J.G.3
  • 24
    • 0030984196 scopus 로고    scopus 로고
    • Chronic ventricular pacing using an output amplitude of 1.0 Volts
    • Schwaab B, Schwerdt H, Heisel A, et al. Chronic ventricular pacing using an output amplitude of 1.0 Volts. PACE 1997; 20:2171-2178.
    • (1997) PACE , vol.20 , pp. 2171-2178
    • Schwaab, B.1    Schwerdt, H.2    Heisel, A.3
  • 25
    • 7144231801 scopus 로고
    • Pulse generator circuitry
    • KA Ellenbogen, GN Kay, BL Wilkoff (eds.): Philadelphia, PA, W.B. Saunders
    • Ireland JR, Kay GN. Pulse generator circuitry. In KA Ellenbogen, GN Kay, BL Wilkoff (eds.): Clinical Cardiac Pacing. Philadelphia, PA, W.B. Saunders, 1995, pp. 112-126.
    • (1995) Clinical Cardiac Pacing , pp. 112-126
    • Ireland, J.R.1    Kay, G.N.2
  • 26
    • 0002816078 scopus 로고
    • Basic concepts
    • Mount Kisco, New York, Futura Publishing Company
    • Furman S. Basic concepts. In: A Practice of Cardiac Pacing. 3rd ed. Mount Kisco, New York, Futura Publishing Company, 1993, pp. 29-89.
    • (1993) A Practice of Cardiac Pacing. 3rd Ed. , pp. 29-89
    • Furman, S.1
  • 27
    • 0028267795 scopus 로고
    • Methodological and clinical aspects of minimizing power consumption in pacemakers providing telemetry
    • Schwerdt H, Fröhlig G, Kindermann M, et al. Methodological and clinical aspects of minimizing power consumption in pacemakers providing telemetry. Herz Kreislauf 1994; 26:119-124.
    • (1994) Herz Kreislauf , vol.26 , pp. 119-124
    • Schwerdt, H.1    Fröhlig, G.2    Kindermann, M.3
  • 28
    • 0019596453 scopus 로고
    • Stimulation and sensing thresholds for cardiac pacing: Electrophysiologic and technical aspects
    • Barold SS, Ong LS, Heinle RA. Stimulation and sensing thresholds for cardiac pacing: Electrophysiologic and technical aspects. Prog Cardiovasc Dis 1981; 24:1-24.
    • (1981) Prog Cardiovasc Dis , vol.24 , pp. 1-24
    • Barold, S.S.1    Ong, L.S.2    Heinle, R.A.3
  • 29
    • 0014092571 scopus 로고
    • Myocardial stimulation: The electrochemistry of electrode-tissue coupling
    • Mansfield PB. Myocardial stimulation: The electrochemistry of electrode-tissue coupling. Am J Physiol 1967; 212:1475-1488.
    • (1967) Am J Physiol , vol.212 , pp. 1475-1488
    • Mansfield, P.B.1
  • 30
    • 84989483270 scopus 로고
    • Pacemaker output programming for maximum safety and maximum longevity
    • Astrinsky EA, Furman S. Pacemaker output programming for maximum safety and maximum longevity. Clin Prog Pacing Electrophysiol 1983; 1:51-58.
    • (1983) Clin Prog Pacing Electrophysiol , vol.1 , pp. 51-58
    • Astrinsky, E.A.1    Furman, S.2
  • 31
    • 28144446706 scopus 로고
    • The longevity of miniaturized cardiac pacemakers
    • Irnich W. The longevity of miniaturized cardiac pacemakers. Herzschrittmacher 1989; 9:164-171.
    • (1989) Herzschrittmacher , vol.9 , pp. 164-171
    • Irnich, W.1
  • 32
    • 28144440391 scopus 로고
    • Permanent intracardiac stimulation for complete heart block and Adam-Stokes syndrome
    • Lagergren H. Permanent intracardiac stimulation for complete heart block and Adam-Stokes syndrome. Acta Chir Scand 1966; 132:663-677.
    • (1966) Acta Chir Scand , vol.132 , pp. 663-677
    • Lagergren, H.1
  • 33
    • 0019569730 scopus 로고
    • Power pulse generators, electrodes and longevity
    • Tyers GFO, Brownlee RR. Power pulse generators, electrodes and longevity. Prog Cardiovasc Dis 1981; 23:421-434.
    • (1981) Prog Cardiovasc Dis , vol.23 , pp. 421-434
    • Tyers, G.F.O.1    Brownlee, R.R.2
  • 34
    • 0017409169 scopus 로고
    • A contribution for the selection of cardiac pacemakers with different batteries
    • Irnich W, Gebhardt U, Effert S. A contribution for the selection of cardiac pacemakers with different batteries. Biomed Eng 1977; 22:106-114.
    • (1977) Biomed Eng , vol.22 , pp. 106-114
    • Irnich, W.1    Gebhardt, U.2    Effert, S.3
  • 36
    • 9044243404 scopus 로고    scopus 로고
    • Initial experience with 1.5 mm2, high impedance, steroid-eluting pacing electrodes
    • Fogel R, Pirzada F, Casavant D, et al. Initial experience with 1.5 mm2, high impedance, steroid-eluting pacing electrodes. PACE 1996; 19:188-196.
    • (1996) PACE , vol.19 , pp. 188-196
    • Fogel, R.1    Pirzada, F.2    Casavant, D.3
  • 37
    • 0025362505 scopus 로고
    • Energy-conserving programming of VVI pacemakers: A telemetry-supported, long-term, follow-up study
    • Klein HH, Knake W. Energy-conserving programming of VVI pacemakers: A telemetry-supported, long-term, follow-up study. Clin Cardiol 1990;: 409-413.
    • (1990) Clin Cardiol , pp. 409-413
    • Klein, H.H.1    Knake, W.2
  • 38
    • 0027958217 scopus 로고
    • Influence of internal current and pacing current on pacemaker longevity
    • Schuchert A, Kuck KH. Influence of internal current and pacing current on pacemaker longevity. PACE 1994; 17:13-17.
    • (1994) PACE , vol.17 , pp. 13-17
    • Schuchert, A.1    Kuck, K.H.2
  • 39
    • 0031018579 scopus 로고    scopus 로고
    • Energy parameters in cardiac pacing should be abandoned
    • Barold SS, Stokes K, Byrd CL, et al. Energy parameters in cardiac pacing should be abandoned. PACE 1997; 20:112-121.
    • (1997) PACE , vol.20 , pp. 112-121
    • Barold, S.S.1    Stokes, K.2    Byrd, C.L.3
  • 40
    • 0014696241 scopus 로고
    • Alteration of pacemaker threshold by drug and physiological factors
    • Preston TA, Judge RD. Alteration of pacemaker threshold by drug and physiological factors. Ann NY Acad Sci 1969; 167:686-692.
    • (1969) Ann NY Acad Sci , vol.167 , pp. 686-692
    • Preston, T.A.1    Judge, R.D.2
  • 41
    • 0014696243 scopus 로고
    • Physiological changes in threshold
    • Sowton E, Barr I. Physiological changes in threshold. Ann N Y Acad Sci 1969; 167:679-685.
    • (1969) Ann N Y Acad Sci , vol.167 , pp. 679-685
    • Sowton, E.1    Barr, I.2
  • 42
    • 0005000069 scopus 로고    scopus 로고
    • Does doubling the charge threshold give an adequate safety margin for pacing with active fixation atrial leads at low voltages?
    • Kusch O, Schwaab B, Fröhlig G, et al. Does doubling the charge threshold give an adequate safety margin for pacing with active fixation atrial leads at low voltages? (abstract) PACE 1997; 20:1152.
    • (1997) PACE , vol.20 , pp. 1152
    • Kusch, O.1    Schwaab, B.2    Fröhlig, G.3
  • 43
    • 9344227490 scopus 로고
    • Three microporous active fixation leads in atrial position
    • Schwaab B, Fröhlig G, Schwerdt H, et al. Three microporous active fixation leads in atrial position. (abstract) PACE 1995; 18:865.
    • (1995) PACE , vol.18 , pp. 865
    • Schwaab, B.1    Fröhlig, G.2    Schwerdt, H.3
  • 44
    • 7144248532 scopus 로고
    • Output voltage below the battery voltage - It is worth it?
    • Schwaab B, Fröhlig G, Schwerdt H, et al. Output voltage below the battery voltage - It is worth it? (Abstract) PACE 1995; 18:865.
    • (1995) PACE , vol.18 , pp. 865
    • Schwaab, B.1    Fröhlig, G.2    Schwerdt, H.3
  • 45
    • 0024379812 scopus 로고
    • Survival rates after pacemaker implantation: A study of patients paced for sick sinus syndrome and atrioventricular block
    • Zanini R, Facchinetti A, Gallo G, et al. Survival rates after pacemaker implantation: A study of patients paced for sick sinus syndrome and atrioventricular block. PACE 1989; 12:1065-1069.
    • (1989) PACE , vol.12 , pp. 1065-1069
    • Zanini, R.1    Facchinetti, A.2    Gallo, G.3
  • 46
    • 0023708514 scopus 로고
    • Survival rate and causes of death in patients with pacemakers: Dependence on symptoms leading to pacemaker implantation
    • Müller CH, Cernin J, Glogar D, et al. Survival rate and causes of death in patients with pacemakers: Dependence on symptoms leading to pacemaker implantation. Eur Heart J 1988; 9:1003-1009.
    • (1988) Eur Heart J , vol.9 , pp. 1003-1009
    • Müller, C.H.1    Cernin, J.2    Glogar, D.3
  • 47
    • 0030075055 scopus 로고    scopus 로고
    • From the editor... Long-term survivors of pacemaker implant
    • Furman S. From the editor... Long-term survivors of pacemaker implant. PACE 1996; 19:134-135.
    • (1996) PACE , vol.19 , pp. 134-135
    • Furman, S.1
  • 48
    • 0005671953 scopus 로고    scopus 로고
    • Patient survival after pacemaker implantation
    • Furman S. Patient survival after pacemaker implantation. PACE 1996; 19:397.
    • (1996) PACE , vol.19 , pp. 397
    • Furman, S.1
  • 49
    • 0023275942 scopus 로고
    • Evaluation of pacemaker pulse generator and patient longevity in patients aged 1 day to 20 years
    • Serwer A, Mericle JM. Evaluation of pacemaker pulse generator and patient longevity in patients aged 1 day to 20 years. Am J Cardiol 1987; 59:824-827
    • (1987) Am J Cardiol , vol.59 , pp. 824-827
    • Serwer, A.1    Mericle, J.M.2
  • 50
    • 0029792462 scopus 로고    scopus 로고
    • The impact of pulse generator longevity on the long-term costs of cardiac pacing
    • Gillis AM, MacQuarrie D, Wilson S. The impact of pulse generator longevity on the long-term costs of cardiac pacing. PACE 1996; 19:1459-1468.
    • (1996) PACE , vol.19 , pp. 1459-1468
    • Gillis, A.M.1    MacQuarrie, D.2    Wilson, S.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.