Volumn 51, Issue 1, 2014, Pages 54-67

A combined laboratory and field evaluation of the Cholestech LDX and CardioChek PA point-of-care testing lipid and glucose analysers

Author keywords

Cholesterol; Glucose; Lipids; Point of care testing

Indexed keywords

ASCORBIC ACID; CHOLESTEROL; GLUCOSE; HIGH DENSITY LIPOPROTEIN CHOLESTEROL; LIPID;

EID: 84899498212     PISSN: 00045632     EISSN: 17581001     Source Type: Journal    
DOI: 10.1177/0004563213482890     Document Type: Article
Times cited : (29)

References (45)
  • 1
    • 84899504900 scopus 로고    scopus 로고
    • World Health Organization, See, (last checked July
    • World Health Organization. Health topics: Cardiovascular diseases, See http://www.who.int/topics/cardiovascular_diseases/en/ (last checked July 2012).
    • (2012) Health Topics: Cardiovascular Diseases
  • 3
    • 0034912564 scopus 로고    scopus 로고
    • The impact of diabetes mellitus on mortality from all causes and coronary heart disease in women: 20 years of follow-up
    • Hu FB, Stampfer MJ, Solomon CG, et al. The impact of diabetes mellitus on mortality from all causes and coronary heart disease in women: 20 years of follow-up. Arch Intern Med 2001; 161: 1717-1723.
    • (2001) Arch Intern Med , vol.161 , pp. 1717-1723
    • Hu, F.B.1    Stampfer, M.J.2    Solomon, C.G.3
  • 4
    • 9244240769 scopus 로고    scopus 로고
    • Trends in cardiovascular complications of diabetes
    • Fox CS, Coady S, Sorlie PD, et al. Trends in cardiovascular complications of diabetes. JAMA 2004; 292: 2495-2499.
    • (2004) JAMA , vol.292 , pp. 2495-2499
    • Fox, C.S.1    Coady, S.2    Sorlie, P.D.3
  • 5
    • 0032515775 scopus 로고    scopus 로고
    • Risk factors for coronary artery disease in noninsulin dependent diabetes mellitus: United Kingdom prospective diabetes study (UKPDS: 23)
    • Turner RC, Millns H, Neil HAW, et al. Risk factors for coronary artery disease in noninsulin dependent diabetes mellitus: United Kingdom prospective diabetes study (UKPDS: 23). BMJ 1998; 316: 823-828.
    • (1998) BMJ , vol.316 , pp. 823-828
    • Turner, R.C.1    Millns, H.2    Neil, H.A.W.3
  • 6
    • 0025833955 scopus 로고
    • UK prospective diabetes study VIII: Study design, progress and performance
    • UKPDS Group
    • UKPDS Group. UK prospective diabetes study VIII: study design, progress and performance. Diabetologia 1991; 34: 877-890.
    • (1991) Diabetologia , vol.34 , pp. 877-890
  • 7
    • 11944263528 scopus 로고
    • Onset of NIDDM occurs at least 4-7 years before clinical diagnosis
    • Harris MI, Klein R, Welborn TA, et al. Onset of NIDDM occurs at least 4-7 years before clinical diagnosis. Diabetes Care 1992; 15: 815-819.
    • (1992) Diabetes Care , vol.15 , pp. 815-819
    • Harris, M.I.1    Klein, R.2    Welborn, T.A.3
  • 8
    • 84899577040 scopus 로고    scopus 로고
    • NHS Health Check: Vascular Risk Assessment and Management
    • Department of Health. Putting Prevention First, London, UK: Department of Health, See, last checked July 2012)
    • Department of Health. Putting Prevention First. NHS Health Check: Vascular Risk Assessment and Management. Best Practice Guidance. London, UK: Department of Health, 2009. See www.dh.gov.uk/en/Publicationsandstatistics/Publications/PublicationsPolicyAndGuidance/DH_097489 (last checked July 2012).
    • (2009) Best Practice Guidance
  • 9
    • 84899555432 scopus 로고    scopus 로고
    • Delivering the NHS Health Check: A Practical Guide to Implementation
    • Department of Health, London, UK: DOH, See, fileticket1/4usgYCuRswjg%3d&tabid1/436 (last checked July 2012)
    • Department of Health. Delivering the NHS Health Check: A Practical Guide to Implementation. A Practical Guide to Point of Care Testing. London, UK: DOH, 2009. See www.improvement.nhs.uk/NHSHealthCheck/LinkClick.aspx? fileticket1/4usgYCuRswjg%3d&tabid1/436 (last checked July 2012).
    • (2009) A Practical Guide to Point of Care Testing
  • 10
    • 0023728740 scopus 로고
    • Current status of blood cholesterol measurement in clinical laboratories in the US
    • National Cholesterol Education Program Laboratory Standardization Panel
    • National Cholesterol Education Program Laboratory Standardization Panel. Current status of blood cholesterol measurement in clinical laboratories in the US. Clin Chem 1988; 34: 193-201.
    • (1988) Clin Chem , vol.34 , pp. 193-201
  • 11
    • 0003407166 scopus 로고
    • National Cholesterol Education Program Laboratory Standardization Panel, National Institute for Health Publication No 90-2964. Washington: US Department of Health and Human Services
    • National Cholesterol Education Program Laboratory Standardization Panel. Recommendations for Improving Cholesterol Measurement. National Institute for Health Publication No 90-2964. Washington: US Department of Health and Human Services, 1990.
    • (1990) Recommendations For Improving Cholesterol Measurement
  • 12
    • 0028853551 scopus 로고
    • National Cholesterol Education Program recommendations for measurement of high-density lipoprotein cholesterol: Executive summary
    • Warnick GR and Wood PD. National Cholesterol Education Program recommendations for measurement of high-density lipoprotein cholesterol: executive summary. Clin Chem 1995; 41: 1427-1433.
    • (1995) Clin Chem , vol.41 , pp. 1427-1433
    • Warnick, G.R.1    Wood, P.D.2
  • 14
    • 84899574495 scopus 로고    scopus 로고
    • PTS Panels: CHOL+HDL+GLU panel test strips insert
    • Polymer Technology Systems Inc
    • Polymer Technology Systems Inc. PTS Panels: CHOL+HDL+GLU panel test strips insert. Polymer Technology Systems Inc, 2007.
    • (2007) Polymer Technology Systems Inc
  • 15
    • 33748598000 scopus 로고    scopus 로고
    • Ascorbic acid interference in the measurement of serum biochemical parameters: In vivo and in vitro studies
    • Martinello F and da Silva LE. Ascorbic acid interference in the measurement of serum biochemical parameters: in vivo and in vitro studies. Clin Biochem 2006; 39: 396-403.
    • (2006) Clin Biochem , vol.39 , pp. 396-403
    • Martinello, F.1    da Silva, L.E.2
  • 16
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM and Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 17
    • 0031434669 scopus 로고    scopus 로고
    • Evaluation of lipid profile measurements obtained from the Cholestech LDX analyzer
    • Bard RL, Kaminsky LA, Whaley MH, et al. Evaluation of lipid profile measurements obtained from the Cholestech LDX analyzer. J Cardiopulm Rehabil 1997; 17: 413-418.
    • (1997) J Cardiopulm Rehabil , vol.17 , pp. 413-418
    • Bard, R.L.1    Kaminsky, L.A.2    Whaley, M.H.3
  • 18
    • 0028179710 scopus 로고
    • Evaluation of the Cholestech L.D.X. desktop analyser for cholesterol, HDL-cholesterol, and triacylglycerols in heparinized venous blood
    • Cobbaert C, Boerma GJ and Lindemans J. Evaluation of the Cholestech L.D.X. desktop analyser for cholesterol, HDL-cholesterol, and triacylglycerols in heparinized venous blood. Eur J Clin Chem Clin Biochem 1994; 32: 391-394.
    • (1994) Eur J Clin Chem Clin Biochem , vol.32 , pp. 391-394
    • Cobbaert, C.1    Boerma, G.J.2    Lindemans, J.3
  • 19
    • 0028074601 scopus 로고
    • An evaluation in the diabetes clinic of the Cholestech LDX analyser
    • Cummings MH, Watts GF, Mazurkiewicz JC, et al. An evaluation in the diabetes clinic of the Cholestech LDX analyser. Practical Diabetes 1994; 11: 151-153.
    • (1994) Practical Diabetes , vol.11 , pp. 151-153
    • Cummings, M.H.1    Watts, G.F.2    Mazurkiewicz, J.C.3
  • 20
    • 33645815482 scopus 로고    scopus 로고
    • Agreement between laboratory results and on-site pathology testing using Bayer DCA2000+ and Cholestech LDX point-ofcare methods in remote Australian Aboriginal commnities
    • Shemesh T, Rowley KG, Shepard M, et al. Agreement between laboratory results and on-site pathology testing using Bayer DCA2000+ and Cholestech LDX point-ofcare methods in remote Australian Aboriginal commnities. Clin Chim Acta 2006; 367: 69-76.
    • (2006) Clin Chim Acta , vol.367 , pp. 69-76
    • Shemesh, T.1    Rowley, K.G.2    Shepard, M.3
  • 21
    • 79952997223 scopus 로고    scopus 로고
    • Point of care testing is appropriate for national health service health check
    • Jain A, Persaud JW, Rao N, et al. Point of care testing is appropriate for national health service health check. Ann Clin Biochem 2011; 48: 159-165.
    • (2011) Ann Clin Biochem , vol.48 , pp. 159-165
    • Jain, A.1    Persaud, J.W.2    Rao, N.3
  • 23
    • 0031974915 scopus 로고    scopus 로고
    • Analytic and clinical performance of two compact cholesterol-testing devices
    • Volles DF, McKenney JM, Miller WG, et al. Analytic and clinical performance of two compact cholesterol-testing devices. Pharmacotherapy 1998; 18: 184-192.
    • (1998) Pharmacotherapy , vol.18 , pp. 184-192
    • Volles, D.F.1    McKenney, J.M.2    Miller, W.G.3
  • 24
    • 76749088885 scopus 로고    scopus 로고
    • Assessing agreement between point of care and laboratory results for lipid testing from a clinical perspective
    • Gialamas A, Laurence CO, Yelland LN, et al. Assessing agreement between point of care and laboratory results for lipid testing from a clinical perspective. Clin Biochem 2010; 43: 515-518.
    • (2010) Clin Biochem , vol.43 , pp. 515-518
    • Gialamas, A.1    Laurence, C.O.2    Yelland, L.N.3
  • 26
    • 0036151043 scopus 로고    scopus 로고
    • Inaccuracy of lipid measurements with the portable Cholestech LDX analyzer in patients with hypercholesterolemia
    • Stein JH, Carlsson CM, Papcke-Benson K, et al. Inaccuracy of lipid measurements with the portable Cholestech LDX analyzer in patients with hypercholesterolemia. Clin Chem 2002; 48: 284-290.
    • (2002) Clin Chem , vol.48 , pp. 284-290
    • Stein, J.H.1    Carlsson, C.M.2    Papcke-Benson, K.3
  • 27
    • 77953106059 scopus 로고    scopus 로고
    • Lipid profile in ambulatory subjects using 3 point-of-care devices and comparison with reference methods
    • Mendez-Gonzalez J, Bonet-Marques R and OrdonezLlanos J. Lipid profile in ambulatory subjects using 3 point-of-care devices and comparison with reference methods. Point of Care 2009; 8: 102-107.
    • (2009) Point of Care , vol.8 , pp. 102-107
    • Mendez-Gonzalez, J.1    Bonet-Marques, R.2    Ordonezllanos, J.3
  • 28
    • 23344435343 scopus 로고    scopus 로고
    • Performance of the CardioChek PA and Cholestech LDX point-of-care analysers compared to clinical diagnostic laboratory methods for the measurement of lipids
    • Panz VR, Raal FJ, Paiker J, et al. Performance of the CardioChek PA and Cholestech LDX point-of-care analysers compared to clinical diagnostic laboratory methods for the measurement of lipids. Cardiovasc J S Afr 2005; 16: 112-117.
    • (2005) Cardiovasc J S Afr , vol.16 , pp. 112-117
    • Panz, V.R.1    Raal, F.J.2    Paiker, J.3
  • 29
    • 38649095980 scopus 로고    scopus 로고
    • Comparative performance of two point-of-care analysers for lipid testing
    • Shephard MDS, Mazzachi BC and Shephard AK. Comparative performance of two point-of-care analysers for lipid testing. Clin Lab 2007; 53: 561-566.
    • (2007) Clin Lab , vol.53 , pp. 561-566
    • Shephard, M.D.S.1    Mazzachi, B.C.2    Shephard, A.K.3
  • 30
    • 42949173366 scopus 로고    scopus 로고
    • Comparison of two point-of-care lipid analyzers for use in global cardiovascular risk assessments
    • Dale RA, Jensen LH and Krantz MJ. Comparison of two point-of-care lipid analyzers for use in global cardiovascular risk assessments. Ann Pharmacother 2008; 42: 633-639.
    • (2008) Ann Pharmacother , vol.42 , pp. 633-639
    • Dale, R.A.1    Jensen, L.H.2    Krantz, M.J.3
  • 31
    • 3042601409 scopus 로고    scopus 로고
    • Cholesterol: Point-of-care testing
    • Taylor JR and Lopez LM. Cholesterol: point-of-care testing. Ann Pharmacother 2004; 38: 1252-1257.
    • (2004) Ann Pharmacother , vol.38 , pp. 1252-1257
    • Taylor, J.R.1    Lopez, L.M.2
  • 32
    • 84857777622 scopus 로고    scopus 로고
    • Pointof-care testing for the analysis of lipid panels: Primary care diagnostic technology update
    • Pluddemann A, Thompson M, Price CP, et al. Pointof-care testing for the analysis of lipid panels: primary care diagnostic technology update. Br J Gen Pract 2012; 62: e224-e226.
    • (2012) Br J Gen Pract , vol.62
    • Pluddemann, A.1    Thompson, M.2    Price, C.P.3
  • 33
    • 44249099664 scopus 로고    scopus 로고
    • Lipid external quality assessment: Commutability between external quality assessment and clinical specimens
    • Cramb R, French J, Mackness M, et al. Lipid external quality assessment: commutability between external quality assessment and clinical specimens. Clin Chem 2008; 45: 260-265.
    • (2008) Clin Chem , vol.45 , pp. 260-265
    • Cramb, R.1    French, J.2    Mackness, M.3
  • 34
    • 33745204481 scopus 로고    scopus 로고
    • Historical milestones in measurement of HDL-cholesterol:Impact on clinical and laboratory practice
    • Langlois MR and Blaton VH. Historical milestones in measurement of HDL-cholesterol:Impact on clinical and laboratory practice. Clin Chim Acta 2006; 369: 168-178.
    • (2006) Clin Chim Acta , vol.369 , pp. 168-178
    • Langlois, M.R.1    Blaton, V.H.2
  • 35
    • 0034868731 scopus 로고    scopus 로고
    • Evolution of methods for measurement of HDL-cholesterol: From ultracentrifugation to homogeneous assays
    • Warnick GR, Nauck M and Rifai N. Evolution of methods for measurement of HDL-cholesterol: from ultracentrifugation to homogeneous assays. Clin Chem 2001; 47: 1579-1596.
    • (2001) Clin Chem , vol.47 , pp. 1579-1596
    • Warnick, G.R.1    Nauck, M.2    Rifai, N.3
  • 36
    • 0017672728 scopus 로고
    • Simultaneously obtained skin-puncture serum, skinpuncture plasma, and venous serum compared, and effects of warming the skin before puncture
    • Blumfield TA, Hertelendy WG and Ford SH. Simultaneously obtained skin-puncture serum, skinpuncture plasma, and venous serum compared, and effects of warming the skin before puncture. Clin Chem 1977; 23: 1705-1710.
    • (1977) Clin Chem , vol.23 , pp. 1705-1710
    • Blumfield, T.A.1    Hertelendy, W.G.2    Ford, S.H.3
  • 37
    • 36649016885 scopus 로고    scopus 로고
    • Glucose: A simple molecule that is not simple to quantify
    • Gambino R. Glucose: a simple molecule that is not simple to quantify. Clin Chem 2007; 53: 2040-2041.
    • (2007) Clin Chem , vol.53 , pp. 2040-2041
    • Gambino, R.1
  • 38
    • 43449138195 scopus 로고    scopus 로고
    • Stabilization of glucose in blood specimens: Mechanism of delay in fluoride inhibition of glycolysis
    • Mikesh LM and Bruns DE. Stabilization of glucose in blood specimens: mechanism of delay in fluoride inhibition of glycolysis. Clin Chem 2008; 54: 930-932.
    • (2008) Clin Chem , vol.54 , pp. 930-932
    • Mikesh, L.M.1    Bruns, D.E.2
  • 39
    • 0025215075 scopus 로고
    • Reagent-strip glucose methods and haematocrit
    • Labib M, Digger T and Perks D. Reagent-strip glucose methods and haematocrit. Lancet 1990; 335: 973-974.
    • (1990) Lancet , vol.335 , pp. 973-974
    • Labib, M.1    Digger, T.2    Perks, D.3
  • 40
    • 84899509706 scopus 로고
    • How reliable are blood glucose meter measurements?
    • Stringer MS and Mackenzie JM. How reliable are blood glucose meter measurements? Clin Biochem Revs 1990; 11: 134.
    • (1990) Clin Biochem Revs , vol.11 , pp. 134
    • Stringer, M.S.1    Mackenzie, J.M.2
  • 41
    • 0026098817 scopus 로고
    • The effect of haematocrit on reagent strip test for glucose
    • Weiner K. The effect of haematocrit on reagent strip test for glucose. Diab Med 1991; 8: 172-175.
    • (1991) Diab Med , vol.8 , pp. 172-175
    • Weiner, K.1
  • 42
    • 0025269008 scopus 로고
    • Reflectance meter type and management of gestational diabetes
    • Harmelin D, Robin H, Molyneaux L, et al. Reflectance meter type and management of gestational diabetes. Lancet 1990; 335: 735-736.
    • (1990) Lancet , vol.335 , pp. 735-736
    • Harmelin, D.1    Robin, H.2    Molyneaux, L.3
  • 43
    • 33748614566 scopus 로고    scopus 로고
    • Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore
    • Martinello F and da Silva LE. Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore. Clin Chim Acta 2006; 373: 108-116.
    • (2006) Clin Chim Acta , vol.373 , pp. 108-116
    • Martinello, F.1    da Silva, L.E.2
  • 44
    • 33748598000 scopus 로고    scopus 로고
    • Ascorbic acid interference in the measurement of serum biochemical parameters: In vivo and in vitro studies
    • Martinello F and da Silva LE. Ascorbic acid interference in the measurement of serum biochemical parameters: in vivo and in vitro studies. Clin Biochem 2006; 39: 396-403.
    • (2006) Clin Biochem , vol.39 , pp. 396-403
    • Martinello, F.1    da Silva, L.E.2
  • 45
    • 84899535592 scopus 로고    scopus 로고
    • Document DB2010 (02) Management and Use of IVD Point of Care Test Devices
    • Medicines and Healthcare products Regulatory Agency
    • Medicines and Healthcare products Regulatory Agency. Document DB2010 (02) Management and Use of IVD Point of Care Test Devices. MHRA, 2010. www.mhra.gov.uk/Publications/Safetyguidance/Device/Bulletins/CON071082.
    • (2010) MHRA


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.