Volumn 86, Issue 6, 2014, Pages 497-503

Effect of an automatic triggering and cycling system on comfort and patient-ventilator synchrony during pressure support ventilation

Author keywords

Artificial; Mechanical ventilation; Respiration; Respiratory mechanics

Indexed keywords

ADULT; ARTICLE; BREATHING; CHRONIC OBSTRUCTIVE LUNG DISEASE; ENDOTRACHEAL TUBE; EXPIRATORY FLOW; FEMALE; HUMAN; INSPIRATORY CAPACITY; LUNG FUNCTION TEST; MALE; MEDICAL TECHNOLOGY; MIDDLE AGED; MOUTH BREATHING; NORMAL HUMAN; PATIENT COMFORT; PEAK INSPIRATORY FLOW; PILOT STUDY; PRESSURE SUPPORT VENTILATION; PRIORITY JOURNAL; QUANTITATIVE ANALYSIS; RESPIRATORY AIRFLOW; SIMULATOR; TIDAL VOLUME; VENTILATED PATIENT; YOUNG ADULT; ADULT RESPIRATORY DISTRESS SYNDROME; ARTIFICIAL VENTILATION; BREATHING MECHANICS; BREATHING PATTERN; BREATHING RATE; HUMAN EXPERIMENT; PATIENT VENTILATOR SYNCHRONY; RESPIRATORY THERAPEUTIC DEVICE;

EID: 84893102958     PISSN: 00257931     EISSN: 14230356     Source Type: Journal    
DOI: 10.1159/000353256     Document Type: Article
Times cited : (11)

References (25)
  • 1
    • 33749237011 scopus 로고    scopus 로고
    • Patient-ventilator asynchrony during assisted mechanical ventilation
    • DOI 10.1007/s00134-006-0301-8
    • Thille AW, Rodriguez P, Cabello B, Lellouche F, Brochard L: Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med 2006; 32:1515-1522. (Pubitemid 44484340)
    • (2006) Intensive Care Medicine , vol.32 , Issue.10 , pp. 1515-1522
    • Thille, A.W.1    Rodriguez, P.2    Cabello, B.3    Lellouche, F.4    Brochard, L.5
  • 2
    • 77449139718 scopus 로고    scopus 로고
    • Patient-ventilator dyssynchrony: Clinical significance and implications for practice
    • Mellott KG, Grap MJ, Munro CL, Sessler CN, Wetzel PA: Patient-ventilator dyssynchrony: clinical significance and implications for practice. Crit Care Nurse 2009; 29:41-55.
    • (2009) Crit Care Nurse , vol.29 , pp. 41-55
    • Mellott, K.G.1    Grap, M.J.2    Munro, C.L.3    Sessler, C.N.4    Wetzel, P.A.5
  • 3
    • 0031468718 scopus 로고    scopus 로고
    • Patient-ventilator trigger asynchrony in prolonged mechanical ventilation
    • Chao DC, Scheinhorn DJ, Stearn-Hassen-pflug M: Patient-ventilator trigger asynchrony in prolonged mechanical ventilation. Chest 1997; 112:1592-1599. (Pubitemid 28007768)
    • (1997) Chest , vol.112 , Issue.6 , pp. 1592-1599
    • Chao, D.C.1
  • 5
    • 4544292361 scopus 로고    scopus 로고
    • Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation
    • DOI 10.1378/chest.126.3.851
    • Vitacca M, Bianchi L, Zanotti E, Vianello A, Barbano L, Porta R, Clini E: Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation. Chest 2004; 3:851-859. (Pubitemid 39258041)
    • (2004) Chest , vol.126 , Issue.3 , pp. 851-859
    • Vitacca, M.1    Bianchi, L.2    Zanotti, E.3    Vianello, A.4    Barbano, L.5    Porta, R.6    Clini, E.7
  • 6
    • 33846591902 scopus 로고    scopus 로고
    • Using ventilator graphics to identify patient-ventilator asynchrony
    • Nilsestuen JO, Hargett KD: Using ventilator graphics to identify patient-ventilator asynchrony. Respir Care 2005; 50:202-234.
    • (2005) Respir Care , vol.50 , pp. 202-234
    • Nilsestuen, J.O.1    Hargett, K.D.2
  • 8
    • 35448934139 scopus 로고    scopus 로고
    • Automatic detection of ineffective triggering and double triggering during mechanical ventilation
    • DOI 10.1007/s00134-007-0767-z
    • Mulqueeny Q, Ceriana P, Carlucci A, Fan-fulla F, Delmastro M, Nava S: Automatic detection of ineffective triggering and double triggering during mechanical ventilation. Intensive Care Med 2007; 33: 2014-2018. (Pubitemid 47632261)
    • (2007) Intensive Care Medicine , vol.33 , Issue.11 , pp. 2014-2018
    • Mulqueeny, Q.1    Ceriana, P.2    Carlucci, A.3    Fanfulla, F.4    Delmastro, M.5    Nava, S.6
  • 9
    • 78651396296 scopus 로고    scopus 로고
    • Monitoring of patient-ventilator interaction at the bedside
    • De Wit M: Monitoring of patient-ventilator interaction at the bedside. Respir Care 2011; 56:61-72.
    • (2011) Respir Care , vol.56 , pp. 61-72
    • De Wit, M.1
  • 10
    • 34447627194 scopus 로고    scopus 로고
    • Clinical review: Patient-ventilator interaction in chronic obstructive pulmonary disease
    • Jolliet P, Tassaux D: Clinical review: patient-ventilator interaction in chronic obstructive pulmonary disease. Crit Care 2006; 10:236-241.
    • (2006) Crit Care , vol.10 , pp. 236-241
    • Jolliet, P.1    Tassaux, D.2
  • 13
    • 84893145875 scopus 로고    scopus 로고
    • Respironics Inc Pittsburgh Respironics
    • Respironics Inc: Digital Auto-Trak. Pittsburgh, Respironics, 2000.
    • (2000) Digital Auto-Trak
  • 15
    • 0344033601 scopus 로고    scopus 로고
    • Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support
    • DOI 10.1007/s00134-003-1703-5
    • Kondili E, Prinianakis G, Georgopoulos D: Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support. Intensive Care Med 2003;29:1950-1959. (Pubitemid 37487129)
    • (2003) Intensive Care Medicine , vol.29 , Issue.11 , pp. 1950-1959
    • Prinianakis, G.1    Kondili, E.2    Georgopoulos, D.3
  • 16
    • 70350116968 scopus 로고    scopus 로고
    • Increases in endotracheal tube resistance are unpredictable relative to duration of intubation
    • Wilson AM, Gray DM, Thomas JG: Increases in endotracheal tube resistance are unpredictable relative to duration of intubation. Chest 2009;136:1006-1013.
    • (2009) Chest , vol.136 , pp. 1006-1013
    • Wilson, A.M.1    Gray, D.M.2    Thomas, J.G.3
  • 17
    • 0032717999 scopus 로고    scopus 로고
    • Influence of ventilator settings in determining respiratory frequency during mechanical ventilation
    • Laghi F, Karamchandani K, Tobin MJ: Influence of ventilator settings in determining respiratory frequency during mechanical ventilation. Am J Respir Crit Care Med 1999; 160: 1766-1770.
    • (1999) Am J Respir Crit Care Med , vol.160 , pp. 1766-1770
    • Laghi, F.1    Karamchandani, K.2    Tobin, M.J.3
  • 18
    • 0030925973 scopus 로고    scopus 로고
    • Comparison of pressure- and flow-triggered pressure-support ventilation on weaning parameters in patients recovering from acute respiratory failure
    • DOI 10.1097/00003246-199705000-00008
    • Tütüncü AS, Cakar N, Camci E, Esen F, Telci L, Akpir K: Comparison of pressure-and flow-triggered pressure-support ventilation on weaning parameters in patients recovering from acute respiratory failure. Crit Care Med 1997;25:756-760. (Pubitemid 27249150)
    • (1997) Critical Care Medicine , vol.25 , Issue.5 , pp. 756-760
    • Tutuncu, A.S.1    Cakar, N.2    Camci, E.3    Esen, F.4    Telci, L.5    Akpir, K.6
  • 19
    • 79951598000 scopus 로고    scopus 로고
    • Patient-ventilation interaction during noninvasive ventilation
    • Hess DR: Patient-ventilation interaction during noninvasive ventilation. Respir Care 2011;56:153-167.
    • (2011) Respir Care , vol.56 , pp. 153-167
    • Hess, D.R.1
  • 20
    • 49749123812 scopus 로고    scopus 로고
    • Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: A bench study
    • Ferreira JC, Chipman DW, Kacmarek RM: Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study. Intensive Care Med 2008; 34:1669-1675.
    • (2008) Intensive Care Med , vol.34 , pp. 1669-1675
    • Ferreira, J.C.1    Chipman, D.W.2    Kacmarek, R.M.3
  • 21
    • 68849129113 scopus 로고    scopus 로고
    • Bilevel vs. ICU ventilators providing noninvasive ventilation: Effect of system leaks: A COPD lung model comparison
    • Ferreira JC, Chipman DW, Hill NS, Kacmarek RM: Bilevel vs. ICU ventilators providing noninvasive ventilation: effect of system leaks: a COPD lung model comparison. Chest 2009;136:448-456.
    • (2009) Chest , vol.136 , pp. 448-456
    • Ferreira, J.C.1    Chipman, D.W.2    Hill, N.S.3    Kacmarek, R.M.4
  • 22
    • 29844448981 scopus 로고    scopus 로고
    • Bedside waveforms interpretation as a tool to identify patient-ventilator asynchronies
    • DOI 10.1007/s00134-005-2828-5
    • Georgopoulos D, Kondili E, Prinianakis G: Bedside waveforms interpretation as a tool to identify patient-ventilator asynchronies. Intensive Care Med 2006;32:34-47. (Pubitemid 43039375)
    • (2006) Intensive Care Medicine , vol.32 , Issue.1 , pp. 34-47
    • Georgopoulos, D.1    Prinianakis, G.2    Kondili, E.3
  • 23
    • 33846020584 scopus 로고    scopus 로고
    • Modulation and treatment of patient-ventilator dyssynchrony
    • DOI 10.1097/MCC.0b013e328011278d, PII 0007519820070200000014
    • Kondili E, Xirouchaki N, Georgopoulos D: Modulation and treatment of patient-ventilator dyssynchrony. Curr Opin Crit Care 2007; 13:84-89. (Pubitemid 46052174)
    • (2007) Current Opinion in Critical Care , vol.13 , Issue.1 , pp. 84-89
    • Kondili, E.1    Xirouchaki, N.2    Georgopoulos, D.3
  • 24
    • 34547103990 scopus 로고    scopus 로고
    • Performance of noninvasive ventilation modes on ICU ventilators during pressure support: A bench model study
    • DOI 10.1007/s00134-007-0713-0
    • Vignaux L, Tassaux D, Jolliet P: Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med 2007; 33:1444-1451. (Pubitemid 47101714)
    • (2007) Intensive Care Medicine , vol.33 , Issue.8 , pp. 1444-1451
    • Vignaux, L.1    Tassaux, D.2    Jolliet, P.3
  • 25
    • 44449120506 scopus 로고    scopus 로고
    • Evaluation of the user interface simplicity in the modern generation of mechanical ventilators
    • Uzawa Y, Yamada Y, Suzukawa M: Evaluation of the user interface simplicity in the modern generation of mechanical ventilators. Respir Care 2008; 53:329-337.
    • (2008) Respir Care , vol.53 , pp. 329-337
    • Uzawa, Y.1    Yamada, Y.2    Suzukawa, M.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.