Volumn 23, Issue 12, 2013, Pages 1166-1173
|
Accuracy of acoustic respiration rate monitoring in pediatric patients
|
Author keywords
apnea; capnography; MeSH; monitoring physiologic; oximetry; postanesthesia nursing; respiration rate
|