Defibrillation threshold testing does not predict clinical outcomes during long-term follow-up: A meta-analysis

Author keywords

defibrillation threshold testing; implantable cardioverter defibrillators; mortality; ventricular arrhythmias

Indexed keywords

ARTICLE; DEFIBRILLATION; FOLLOW UP; HEART VENTRICLE ARRHYTHMIA; HUMAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; META ANALYSIS (TOPIC); MORTALITY; OUTCOME ASSESSMENT;

EID: 84886773338     PISSN: 01478389     EISSN: 15408159     Source Type: Journal    
DOI: 10.1111/pace.12218     Document Type: Article
Times cited : (11)

References (31)
  • 2
    • 0037149716 scopus 로고    scopus 로고
    • Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
    • Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002; 346: 877-883.
    • (2002) N Engl J Med , vol.346 , pp. 877-883
    • Moss, A.J.1    Zareba, W.2    Hall, W.J.3    Klein, H.4    Wilber, D.J.5    Cannom, D.S.6    Daubert, J.P.7
  • 5
    • 84865738536 scopus 로고    scopus 로고
    • Clinical evaluation of defibrillation testing in an unselected population of 2,120 consecutive patients undergoing first implantable cardioverter- defibrillator implant
    • Brignole M, Occhetta E, Bongiorni MG, Proclemer A, Favale S, Iacopino S, Calo L, et al. Clinical evaluation of defibrillation testing in an unselected population of 2,120 consecutive patients undergoing first implantable cardioverter-defibrillator implant. J Am Coll Cardiol 2012; 60: 981-987.
    • (2012) J Am Coll Cardiol , vol.60 , pp. 981-987
    • Brignole, M.1    Occhetta, E.2    Bongiorni, M.G.3    Proclemer, A.4    Favale, S.5    Iacopino, S.6    Calo, L.7
  • 6
    • 71849086249 scopus 로고    scopus 로고
    • Intraoperative defibrillation threshold testing during implantable cardioverter-defibrillator insertion: Do we really need it?
    • Calvi V, Dugo D, Capodanno D, Arancio R, Di Grazia A, Liotta C, Puzzangara E, et al. Intraoperative defibrillation threshold testing during implantable cardioverter-defibrillator insertion: Do we really need it? Am Heart J 2010; 159: 98-102.
    • (2010) Am Heart J , vol.159 , pp. 98-102
    • Calvi, V.1    Dugo, D.2    Capodanno, D.3    Arancio, R.4    Di Grazia, A.5    Liotta, C.6    Puzzangara, E.7
  • 7
    • 84871256608 scopus 로고    scopus 로고
    • A randomized-controlled pilot study comparing icd implantation with and without intraoperative defibrillation testing in patients with heart failure and severe left ventricular dysfunction: A substudy of the raft trial
    • Healey JS, Gula LJ, Birnie DH, Sterns L, Connolly SJ, Sapp J, Crystal E, et al. A randomized-controlled pilot study comparing icd implantation with and without intraoperative defibrillation testing in patients with heart failure and severe left ventricular dysfunction: A substudy of the raft trial. J Cardiovasc Electrophysiol 2012; 23: 1313-1316.
    • (2012) J Cardiovasc Electrophysiol , vol.23 , pp. 1313-1316
    • Healey, J.S.1    Gula, L.J.2    Birnie, D.H.3    Sterns, L.4    Connolly, S.J.5    Sapp, J.6    Crystal, E.7
  • 8
    • 79955436789 scopus 로고    scopus 로고
    • Defibrillation threshold testing fails to show clinical benefit during long-term follow-up of patients undergoing cardiac resynchronization therapy defibrillator implantation
    • Michowitz Y, Lellouche N, Contractor T, Bourke T, Wiener I, Buch E, Boyle N, et al. Defibrillation threshold testing fails to show clinical benefit during long-term follow-up of patients undergoing cardiac resynchronization therapy defibrillator implantation. Europace 2011; 13: 683-688.
    • (2011) Europace , vol.13 , pp. 683-688
    • Michowitz, Y.1    Lellouche, N.2    Contractor, T.3    Bourke, T.4    Wiener, I.5    Buch, E.6    Boyle, N.7
  • 9
    • 33644778037 scopus 로고    scopus 로고
    • Intraoperative testing of the implantable cardioverter-defibrillator: How much is enough?
    • Pires LA, Johnson KM,. Intraoperative testing of the implantable cardioverter-defibrillator: How much is enough? J Cardiovasc Electrophysiol 2006; 17: 140-145.
    • (2006) J Cardiovasc Electrophysiol , vol.17 , pp. 140-145
    • Pires, L.A.1    Johnson, K.M.2
  • 12
    • 0031861803 scopus 로고    scopus 로고
    • Changes in cerebral oxygen uptake and cerebral electrical activity during defibrillation threshold testing
    • de Vries JW, Bakker PF, Visser GH, Diephuis JC, van Huffelen AC,. Changes in cerebral oxygen uptake and cerebral electrical activity during defibrillation threshold testing. Anesth Analg 1998; 87: 16-20.
    • (1998) Anesth Analg , vol.87 , pp. 16-20
    • De Vries, J.W.1    Bakker, P.F.2    Visser, G.H.3    Diephuis, J.C.4    Van Huffelen, A.C.5
  • 13
    • 34547887245 scopus 로고    scopus 로고
    • Defibrillation testing at the time of implantation of cardioverter defibrillator in the clinical practice: A nation-wide survey
    • Brignole M, Raciti G, Bongiorni MG, De Martino G, Favale S, Gasparini M, Luise R, et al. Defibrillation testing at the time of implantation of cardioverter defibrillator in the clinical practice: A nation-wide survey. Europace 2007; 9: 540-543.
    • (2007) Europace , vol.9 , pp. 540-543
    • Brignole, M.1    Raciti, G.2    Bongiorni, M.G.3    De Martino, G.4    Favale, S.5    Gasparini, M.6    Luise, R.7
  • 14
    • 0037197741 scopus 로고    scopus 로고
    • Efficacy and temporal stability of reduced safety margins for ventricular defibrillation: Primary results from the low energy safety study (less)
    • Gold MR, Higgins S, Klein R, Gilliam FR, Kopelman H, Hessen S, Payne J, et al. Efficacy and temporal stability of reduced safety margins for ventricular defibrillation: Primary results from the low energy safety study (less). Circulation 2002; 105: 2043-2048.
    • (2002) Circulation , vol.105 , pp. 2043-2048
    • Gold, M.R.1    Higgins, S.2    Klein, R.3    Gilliam, F.R.4    Kopelman, H.5    Hessen, S.6    Payne, J.7
  • 15
    • 44649093349 scopus 로고    scopus 로고
    • High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems
    • Day JD, Olshansky B, Moore S, Brown S, Stolen KQ, Lerew DR,. High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems. Europace 2008; 10: 347-350.
    • (2008) Europace , vol.10 , pp. 347-350
    • Day, J.D.1    Olshansky, B.2    Moore, S.3    Brown, S.4    Stolen, K.Q.5    Lerew, D.R.6
  • 16
    • 3242794231 scopus 로고    scopus 로고
    • Is defibrillation testing required for defibrillator implantation?
    • Strickberger SA, Klein GJ,. Is defibrillation testing required for defibrillator implantation? J Am Coll Cardiol 2004; 44: 88-91.
    • (2004) J Am Coll Cardiol , vol.44 , pp. 88-91
    • Strickberger, S.A.1    Klein, G.J.2
  • 18
    • 80051666300 scopus 로고    scopus 로고
    • Defibrillation testing during implantable cardioverter-defibrillator implantation in italian current practice: The assessment of long-term induction clinical value (alive) project
    • Stefano B, Pietro RR, Maurizio G, Maurizio L, Renato M, Pietro R, Alessandro P, et al. Defibrillation testing during implantable cardioverter-defibrillator implantation in italian current practice: The assessment of long-term induction clinical value (alive) project. Am Heart J 2011; 162: 390-397.
    • (2011) Am Heart J , vol.162 , pp. 390-397
    • Stefano, B.1    Pietro, R.R.2    Maurizio, G.3    Maurizio, L.4    Renato, M.5    Pietro, R.6    Alessandro, P.7
  • 19
    • 65649083360 scopus 로고    scopus 로고
    • Primary prevention implantation of cardioverter defibrillator without defibrillation threshold testing: 2-year follow-up
    • Bianchi S, Ricci RP, Biscione F, Sgreccia F, Di Belardino N, Rossi P, Giuli S, et al. Primary prevention implantation of cardioverter defibrillator without defibrillation threshold testing: 2-year follow-up. Pacing Clin Electrophysiol 2009; 32: 573-578.
    • (2009) Pacing Clin Electrophysiol , vol.32 , pp. 573-578
    • Bianchi, S.1    Ricci, R.P.2    Biscione, F.3    Sgreccia, F.4    Di Belardino, N.5    Rossi, P.6    Giuli, S.7
  • 20
    • 84865738536 scopus 로고    scopus 로고
    • Clinical evaluation of defibrillation testing in an unselected population of 2,120 consecutive patients undergoing first implantable cardioverter- defibrillator implant
    • Brignole M, Occhetta E, Bongiorni MG, Proclemer A, Favale S, Iacopino S, Calo L, et al. Clinical evaluation of defibrillation testing in an unselected population of 2,120 consecutive patients undergoing first implantable cardioverter-defibrillator implant. J Am Coll Cardiol 2012; 60: 981-987.
    • (2012) J Am Coll Cardiol , vol.60 , pp. 981-987
    • Brignole, M.1    Occhetta, E.2    Bongiorni, M.G.3    Proclemer, A.4    Favale, S.5    Iacopino, S.6    Calo, L.7
  • 21
    • 39549083219 scopus 로고    scopus 로고
    • Comparison of outcomes in patients undergoing defibrillation threshold testing at the time of implantable cardioverter-defibrillator implantation versus no defibrillation threshold testing
    • Hall B, Jeevanantham V, Levine E, Daubert J, McNitt S, Hall F, Hall FM, et al. Comparison of outcomes in patients undergoing defibrillation threshold testing at the time of implantable cardioverter-defibrillator implantation versus no defibrillation threshold testing. Cardiol J 2007; 14: 463-469.
    • (2007) Cardiol J , vol.14 , pp. 463-469
    • Hall, B.1    Jeevanantham, V.2    Levine, E.3    Daubert, J.4    McNitt, S.5    Hall, F.6    Hall, F.M.7
  • 22
    • 20244387585 scopus 로고    scopus 로고
    • Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion?
    • Russo AM, Sauer W, Gerstenfeld EP, Hsia HH, Lin D, Cooper JM, Dixit S, et al. Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm 2005; 2: 456-461.
    • (2005) Heart Rhythm , vol.2 , pp. 456-461
    • Russo, A.M.1    Sauer, W.2    Gerstenfeld, E.P.3    Hsia, H.H.4    Lin, D.5    Cooper, J.M.6    Dixit, S.7
  • 27
    • 76849091527 scopus 로고    scopus 로고
    • Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients
    • Sweeney MO, Sherfesee L, DeGroot PJ, Wathen MS, Wilkoff BL,. Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients. Heart Rhythm 2010; 7: 353-360.
    • (2010) Heart Rhythm , vol.7 , pp. 353-360
    • Sweeney, M.O.1    Sherfesee, L.2    Degroot, P.J.3    Wathen, M.S.4    Wilkoff, B.L.5
  • 29
    • 17744402108 scopus 로고    scopus 로고
    • Probability of successful defibrillation at multiples of the defibrillation energy requirement in patients with an implantable defibrillator
    • Strickberger SA, Daoud EG, Davidson T, Weiss R, Bogun F, Knight BP, Bahu M, et al. Probability of successful defibrillation at multiples of the defibrillation energy requirement in patients with an implantable defibrillator. Circulation 1997; 96: 1217-1223.
    • (1997) Circulation , vol.96 , pp. 1217-1223
    • Strickberger, S.A.1    Daoud, E.G.2    Davidson, T.3    Weiss, R.4    Bogun, F.5    Knight, B.P.6    Bahu, M.7
  • 31
    • 84873098223 scopus 로고    scopus 로고
    • (accessed November 12, 2012)
    • Shockless Implant Evaluation (SIMPLE). Available at: http://www. clinicaltrials.gov/ct2/show/NCT00800384?term=NCT00800384&rank=1. (accessed November 12, 2012).
    • Shockless Implant Evaluation (SIMPLE)


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.