A clinical comparison between a non-invasive blood pressure monitor using high definition oscillometry (Memodiagnostic MD 15/90 Pro) and invasive arterial blood pressure measurement in anaesthetized dogs

Author keywords

Anaesthesia; Arterial blood pressure; Dog; High definition oscillometry; Non invasive blood pressure measurement

Indexed keywords

ANIMALIA; CANIS FAMILIARIS;

EID: 84881537315     PISSN: 14672987     EISSN: 14672995     Source Type: Journal    
DOI: 10.1111/vaa.12035     Document Type: Article
Times cited : (30)

References (35)
  • 1
    • 77955855663 scopus 로고    scopus 로고
    • Agreement between directly measured blood pressure and pressures obtained with three veterinary specific oscillometric units in cats
    • Acierno MJ, Seaton D, Mitchell MA et al. (2010) Agreement between directly measured blood pressure and pressures obtained with three veterinary specific oscillometric units in cats. J Am Vet Med Assoc 237, 402-406.
    • (2010) J Am Vet Med Assoc , vol.237 , pp. 402-406
    • Acierno, M.J.1    Seaton, D.2    Mitchell, M.A.3
  • 2
    • 0000004140 scopus 로고
    • Measurement in medicine: the analysis of method comparison studies
    • Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. The Statistician 32, 307-317.
    • (1983) The Statistician , vol.32 , pp. 307-317
    • Altman, D.G.1    Bland, J.M.2
  • 3
    • 84881507666 scopus 로고    scopus 로고
    • Anon () American National Standards Institute/Association for the Advancement of medical instrumentation. ANSI/AAMI SP10:2002 & ANSIAAMI SP10/A1:2003 & ANSI/AAMI SP10:2002 & ANSIAAMI SP10/A1:2003 & ANSI/AAMI SP102002/A2:2006. AAMI/Washington.
    • Anon (2006) American National Standards Institute/Association for the Advancement of medical instrumentation. ANSI/AAMI SP10:2002 & ANSIAAMI SP10/A1:2003 & ANSI/AAMI SP10:2002 & ANSIAAMI SP10/A1:2003 & ANSI/AAMI SP102002/A2:2006. AAMI/Washington.
    • (2006)
  • 4
    • 84881501751 scopus 로고    scopus 로고
    • Clinical evaluation of accuracy of the Mindray Beneview T5 in measuring non-invasive blood pressure in anaesthetized dogs. Abstracts presented at the Association of Veterinary Anaesthetists Spring meeting, 30-31st March 2010, Cambridge, UK. Vet Anaesth Analg 37, 9.
    • Aprea F, Girard N, Corletto F (2010) Clinical evaluation of accuracy of the Mindray Beneview T5 in measuring non-invasive blood pressure in anaesthetized dogs. Abstracts presented at the Association of Veterinary Anaesthetists Spring meeting, 30-31st March 2010, Cambridge, UK. Vet Anaesth Analg 37, 9.
    • (2010)
    • Aprea, F.1    Girard, N.2    Corletto, F.3
  • 5
    • 84881532420 scopus 로고    scopus 로고
    • Evaluation of HDO (High Definition Oscillometry) - a new non-invasive blood pressure technique- in comparison to invasive measurement (HSE - Hydro Sachs Electronics) in anaesthetized dogs. Research abstract program of the 2009 ACVIM Forum & Canadian Veterinary Medical Association Convention.Poster Presentations
    • Baumgartner C, Erhardt W, Faltenmeyer CH et al. (2009) Evaluation of HDO (High Definition Oscillometry) - a new non-invasive blood pressure technique- in comparison to invasive measurement (HSE - Hydro Sachs Electronics) in anaesthetized dogs. Research abstract program of the 2009 ACVIM Forum & Canadian Veterinary Medical Association Convention.Poster Presentations. J Vet Intern Med 23, 680.
    • (2009) J Vet Intern Med , vol.23 , pp. 680
    • Baumgartner, C.1    Erhardt, W.2    Faltenmeyer, C.H.3
  • 6
    • 34347375489 scopus 로고    scopus 로고
    • Agreement between methods of measurement with multiple observations per individual
    • Bland JM, Altman DG (2007) Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 17, 571-582.
    • (2007) J Biopharm Stat , vol.17 , pp. 571-582
    • Bland, J.M.1    Altman, D.G.2
  • 7
    • 49549115126 scopus 로고    scopus 로고
    • The risk of death: the confidential enquiry into perioperative small animal fatalities
    • Brodbelt D, Blissitt KJ, Hammond RA et al. (2008) The risk of death: the confidential enquiry into perioperative small animal fatalities. Vet Anaesth Anal 35, 365-373.
    • (2008) Vet Anaesth Anal , vol.35 , pp. 365-373
    • Brodbelt, D.1    Blissitt, K.J.2    Hammond, R.A.3
  • 8
    • 34249341125 scopus 로고    scopus 로고
    • Guidelines for the identification, evaluation and management of systemic hypertension in dogs and cats
    • Brown S, Atkins C, Bagley R et al. (2007) Guidelines for the identification, evaluation and management of systemic hypertension in dogs and cats. J Vet Intern Med 21, 542-558.
    • (2007) J Vet Intern Med , vol.21 , pp. 542-558
    • Brown, S.1    Atkins, C.2    Bagley, R.3
  • 9
    • 78649408142 scopus 로고    scopus 로고
    • Questions conclusions in study of blood pressure measurement system
    • Carr AP (2010) Questions conclusions in study of blood pressure measurement system. J Am Vet Med Assoc 237, 1018-1020.
    • (2010) J Am Vet Med Assoc , vol.237 , pp. 1018-1020
    • Carr, A.P.1
  • 10
    • 38349070676 scopus 로고    scopus 로고
    • Clinical evaluation of the Surgivet V60046, a non invasive blood pressure monitor in anaesthetized dogs
    • Deflandre CJ, Hellebrekers LJ (2008) Clinical evaluation of the Surgivet V60046, a non invasive blood pressure monitor in anaesthetized dogs. Vet Anaesth Analg 35, 13-21.
    • (2008) Vet Anaesth Analg , vol.35 , pp. 13-21
    • Deflandre, C.J.1    Hellebrekers, L.J.2
  • 11
    • 85031133810 scopus 로고    scopus 로고
    • Importance of blood pressure measurement-techniques
    • In: (4th edn). Egner B (ed.). VetVerlag VBS. Buchhandel and Seminar, Babenhausen, Germany. pp. -.
    • Egner B, Carr A, Brown S (2007) Importance of blood pressure measurement-techniques. In: Essential facts of blood pressure in dogs and cats (4th edn). Egner B (ed.). VetVerlag VBS. Buchhandel and Seminar, Babenhausen, Germany. pp. 51-63.
    • (2007) Essential facts of blood pressure in dogs and cats , pp. 51-63
    • Egner, B.1    Carr, A.2    Brown, S.3
  • 12
    • 0029331209 scopus 로고
    • Comparison of Direct and Indirect Blood Pressure Measurements in Anesthetized Dogs
    • Gains MJ, Grodecki KM, Jacobs RM et al. (1995) Comparison of Direct and Indirect Blood Pressure Measurements in Anesthetized Dogs. Can J Vet Res 59, 238-240.
    • (1995) Can J Vet Res , vol.59 , pp. 238-240
    • Gains, M.J.1    Grodecki, K.M.2    Jacobs, R.M.3
  • 13
    • 0032952234 scopus 로고    scopus 로고
    • Complications and mortality associated with anaesthesia in dogs and cats
    • Gaynor JS, Dunlop CI, Wagner AE et al. (1999) Complications and mortality associated with anaesthesia in dogs and cats. J Am Anim Hosp Assoc 35, 13-17.
    • (1999) J Am Anim Hosp Assoc , vol.35 , pp. 13-17
    • Gaynor, J.S.1    Dunlop, C.I.2    Wagner, A.E.3
  • 14
    • 33746430032 scopus 로고    scopus 로고
    • Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs
    • Haberman C, Kang CW, Morgan JD et al. (2006) Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs. Can J Vet Res 70, 211-217.
    • (2006) Can J Vet Res , vol.70 , pp. 211-217
    • Haberman, C.1    Kang, C.W.2    Morgan, J.D.3
  • 15
    • 0024544132 scopus 로고
    • Understanding natural frequency and damping and how they relate to the measurement of blood pressure
    • Kleinman B (1989) Understanding natural frequency and damping and how they relate to the measurement of blood pressure. J Clin Monit 5, 137-147.
    • (1989) J Clin Monit , vol.5 , pp. 137-147
    • Kleinman, B.1
  • 17
    • 77954702618 scopus 로고    scopus 로고
    • Comparison of invasive and non-invasive blood monitoring during clinical anaesthesia in dogs
    • MacFarlane P, Grint N, Dugdale A (2010) Comparison of invasive and non-invasive blood monitoring during clinical anaesthesia in dogs. Vet Res Commun 34, 217-227.
    • (2010) Vet Res Commun , vol.34 , pp. 217-227
    • MacFarlane, P.1    Grint, N.2    Dugdale, A.3
  • 18
    • 0345772869 scopus 로고    scopus 로고
    • Hypotension during anesthesia in dogs and cats: recognition, causes, and treatment
    • Mazzaferro E, Wagner AE (2001) Hypotension during anesthesia in dogs and cats: recognition, causes, and treatment. Compend Contin Educ Prac Vet 8, 728-737.
    • (2001) Compend Contin Educ Prac Vet , vol.8 , pp. 728-737
    • Mazzaferro, E.1    Wagner, A.E.2
  • 19
    • 33748743900 scopus 로고    scopus 로고
    • Accuracy of an oscillometric blood pressure monitor during phenylephrine induced hypertension in dogs
    • McMurphy M, Stoll MR, McCubrey R (2006) Accuracy of an oscillometric blood pressure monitor during phenylephrine induced hypertension in dogs. Am J Vet Res 9, 1541-1545.
    • (2006) Am J Vet Res , vol.9 , pp. 1541-1545
    • McMurphy, M.1    Stoll, M.R.2    McCubrey, R.3
  • 20
    • 77956156800 scopus 로고    scopus 로고
    • Comparison of telemetry and high definition oscillometry for blood pressure measurements in conscious dogs: effects of torcetrapib
    • Meyer O, Jenni R, Greiter-Wilke A et al. (2010) Comparison of telemetry and high definition oscillometry for blood pressure measurements in conscious dogs: effects of torcetrapib. J Am Assoc Lab Anim Sci 49, 464-471.
    • (2010) J Am Assoc Lab Anim Sci , vol.49 , pp. 464-471
    • Meyer, O.1    Jenni, R.2    Greiter-Wilke, A.3
  • 21
    • 77956169583 scopus 로고    scopus 로고
    • Sensitivity of two non invasive blood pressure measurement techniques compared to telemetry in cynomolgus monkeys and beagle dogs
    • Mitchell AZ, McMahon C, Beck TW et al. (2010) Sensitivity of two non invasive blood pressure measurement techniques compared to telemetry in cynomolgus monkeys and beagle dogs. J Pharmacol Toxicol Methods 62, 54-63.
    • (2010) J Pharmacol Toxicol Methods , vol.62 , pp. 54-63
    • Mitchell, A.Z.1    McMahon, C.2    Beck, T.W.3
  • 22
    • 85031141153 scopus 로고    scopus 로고
    • Accuracy of a species and cuff site optimized oscillometric BP in dogs. Accessed September 16, 2012.
    • Muir W (2011) Accuracy of a species and cuff site optimized oscillometric BP in dogs. http://www.cardiocommand.com/petmapweb/study_abstracts.html. Accessed September 16, 2012.
    • (2011)
    • Muir, W.1
  • 23
    • 0001469411 scopus 로고    scopus 로고
    • Contours of pressure and flow waves in arteries
    • In: (6th edn). Penny S (ed.). Oxford University Press, New York, USA. pp. -.
    • Nichols WW, O'Rourke MF (2011) Contours of pressure and flow waves in arteries. In: MacDonald's blood flow in arteries (6th edn). Penny S (ed.). Oxford University Press, New York, USA. pp. 170-200.
    • (2011) MacDonald's blood flow in arteries , pp. 170-200
    • Nichols, W.W.1    O'Rourke, M.F.2
  • 24
    • 34547930364 scopus 로고    scopus 로고
    • Blood pressure and peripheral circulation
    • In: (4th edn). Weinberg RW (ed). Lippincott Williams & Wilkins, Philadelphia, USA. pp. -.
    • Opie LH (2004) Blood pressure and peripheral circulation. In: Heart physiology from cell to circulation (4th edn). Weinberg RW (ed). Lippincott Williams & Wilkins, Philadelphia, USA. pp. 431-459.
    • (2004) Heart physiology from cell to circulation , pp. 431-459
    • Opie, L.H.1
  • 25
    • 35348862405 scopus 로고    scopus 로고
    • Normal values and incidence of cardiorespiratory complications in dogs during general anaesthesia. A review of 1281 cases
    • Redondo JI, Rubio M, Soler G et al. (2007) Normal values and incidence of cardiorespiratory complications in dogs during general anaesthesia. A review of 1281 cases. J Vet Med A, Physiol Pathol Clin Med 54, 470-477.
    • (2007) J Vet Med A, Physiol Pathol Clin Med , vol.54 , pp. 470-477
    • Redondo, J.I.1    Rubio, M.2    Soler, G.3
  • 26
    • 33846895903 scopus 로고    scopus 로고
    • Reliability and survival of arterial catheter: optimal dynamic response)
    • Riachy M, Riachy E, Sleilaty G et al. (2007) Reliability and survival of arterial catheter: optimal dynamic response). Ann Fr Anesth Reanim 26, 119-124.
    • (2007) Ann Fr Anesth Reanim , vol.26 , pp. 119-124
    • Riachy, M.1    Riachy, E.2    Sleilaty, G.3
  • 27
    • 1542426070 scopus 로고    scopus 로고
    • Evaluation of a new oscillometric blood pressure monitor in isoflurane-anesthetized dogs
    • Sawyer D, Guikema A, Siegel E (2004) Evaluation of a new oscillometric blood pressure monitor in isoflurane-anesthetized dogs. Vet Anaesth Analg 31, 27-39.
    • (2004) Vet Anaesth Analg , vol.31 , pp. 27-39
    • Sawyer, D.1    Guikema, A.2    Siegel, E.3
  • 28
    • 80955179357 scopus 로고    scopus 로고
    • Cardiovascular Monitoring
    • In: (7th edn). Miller RD (ed). Churchill Livingstone, Philadelphia, USA. pp. -.
    • Schroeder RA, Barbeito A, Bar-Yosef S et al. (2010) Cardiovascular Monitoring. In: Miller's Anesthesia (7th edn). Miller RD (ed). Churchill Livingstone, Philadelphia, USA. pp. 1267-1328.
    • (2010) Miller's Anesthesia , pp. 1267-1328
    • Schroeder, R.A.1    Barbeito, A.2    Bar-Yosef, S.3
  • 29
    • 84871385015 scopus 로고    scopus 로고
    • Comparison of high definition oscillometric and Doppler ultrasound devices with invasive blood pressure in anaesthetized dogs
    • Seliškar A, Zrimšek P, Sredenšek J et al. (2013) Comparison of high definition oscillometric and Doppler ultrasound devices with invasive blood pressure in anaesthetized dogs. Vet Anaesth Analg 40, 21-27.
    • (2013) Vet Anaesth Analg , vol.40 , pp. 21-27
    • Seliškar, A.1    Zrimšek, P.2    Sredenšek, J.3
  • 30
    • 77954704066 scopus 로고    scopus 로고
    • Evaluation of an indirect oscillometric blood pressure monitor innormotensive and hypotensive anesthetized dogs
    • Shih A, Robertson S, Vigani A et al. (2010) Evaluation of an indirect oscillometric blood pressure monitor innormotensive and hypotensive anesthetized dogs. J Vet Emerg Crit Care 20, 313-318.
    • (2010) J Vet Emerg Crit Care , vol.20 , pp. 313-318
    • Shih, A.1    Robertson, S.2    Vigani, A.3
  • 31
    • 80053327347 scopus 로고    scopus 로고
    • Arterial waveform analysis for the anesthesiologist: past, present, and future concepts
    • Thiele RH, Durieux ME (2011) Arterial waveform analysis for the anesthesiologist: past, present, and future concepts. Anesth Analg 113, 766-776.
    • (2011) Anesth Analg , vol.113 , pp. 766-776
    • Thiele, R.H.1    Durieux, M.E.2
  • 32
    • 0034243518 scopus 로고    scopus 로고
    • Direct blood pressure monitoring
    • Waddell LS (2000) Direct blood pressure monitoring. Clin Tech Small Anim Pract 15, 111-118.
    • (2000) Clin Tech Small Anim Pract , vol.15 , pp. 111-118
    • Waddell, L.S.1
  • 33
    • 0031135887 scopus 로고    scopus 로고
    • Arterial blood pressure monitoring in anesthetized animals
    • Wagner AE, Brodbelt DC (1997) Arterial blood pressure monitoring in anesthetized animals. J Am Vet Med Assoc 9, 1279-1285.
    • (1997) J Am Vet Med Assoc , vol.9 , pp. 1279-1285
    • Wagner, A.E.1    Brodbelt, D.C.2
  • 35
    • 77954504876 scopus 로고    scopus 로고
    • Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetized dogs using ACVIM guidelines
    • Wernick M, Doherr M, Howard J et al. (2010) Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetized dogs using ACVIM guidelines. J Small Anim Pract 51, 318-324.
    • (2010) J Small Anim Pract , vol.51 , pp. 318-324
    • Wernick, M.1    Doherr, M.2    Howard, J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.