Volumn 31, Issue 6, 2013, Pages 991-997

Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty: A biomechanical validation

Author keywords

cement; dual mobility acetabular component; fixation strength; instability; revision total hip arthroplasty

Indexed keywords

BONE CEMENT; POLYETHYLENE;

EID: 84876289014     PISSN: 07360266     EISSN: 1554527X     Source Type: Journal    
DOI: 10.1002/jor.22314     Document Type: Article
Times cited : (28)

References (42)
  • 1
    • 58649122701 scopus 로고    scopus 로고
    • The epidemiology of revision total hip arthroplasty in the United States
    • Bozic KJ, Kurtz SM, Lau E, et al. 2009. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am 91: 128-133.
    • (2009) J Bone Joint Surg Am , vol.91 , pp. 128-133
    • Bozic, K.J.1    Kurtz, S.M.2    Lau, E.3
  • 2
    • 34147150632 scopus 로고    scopus 로고
    • Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030
    • Kurtz S, Ong K, Lau E, et al. 2007. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89: 780-785.
    • (2007) J Bone Joint Surg Am , vol.89 , pp. 780-785
    • Kurtz, S.1    Ong, K.2    Lau, E.3
  • 4
    • 0036788526 scopus 로고    scopus 로고
    • Success rate of modular component exchange for the treatment of an unstable total hip arthroplasty
    • Earll MD, Fehring TK, Griffin WL, et al. 2002. Success rate of modular component exchange for the treatment of an unstable total hip arthroplasty. J Arthroplasty 17: 864-869.
    • (2002) J Arthroplasty , vol.17 , pp. 864-869
    • Earll, M.D.1    Fehring, T.K.2    Griffin, W.L.3
  • 5
    • 43049097831 scopus 로고    scopus 로고
    • Revision total hip arthroplasty for instability: Surgical techniques and principles
    • Parvizi J, Picinic E, Sharkey PF., 2008. Revision total hip arthroplasty for instability: surgical techniques and principles. J Bone Joint Surg Am 90: 1134-1142.
    • (2008) J Bone Joint Surg Am , vol.90 , pp. 1134-1142
    • Parvizi, J.1    Picinic, E.2    Sharkey, P.F.3
  • 6
    • 2442612732 scopus 로고    scopus 로고
    • Cementing a liner into a stable cementless acetabular shell: The double-socket technique
    • Beaule PE, Ebramzadeh E, Le Duff M, et al. 2004. Cementing a liner into a stable cementless acetabular shell: the double-socket technique. J Bone Joint Surg Am 86-A: 929-934.
    • (2004) J Bone Joint Surg Am , vol.86 , pp. 929-934
    • Beaule, P.E.1    Ebramzadeh, E.2    Le Duff, M.3
  • 7
    • 0036714997 scopus 로고    scopus 로고
    • Cementation of a polyethylene liner into a metal shell. Factors related to mechanical stability
    • Bonner KF, Delanois RE, Harbach G, et al. 2002. Cementation of a polyethylene liner into a metal shell. Factors related to mechanical stability. J Bone Joint Surg Am 84-A: 1587-1593.
    • (2002) J Bone Joint Surg Am , vol.84 , pp. 1587-1593
    • Bonner, K.F.1    Delanois, R.E.2    Harbach, G.3
  • 8
    • 34547596446 scopus 로고    scopus 로고
    • Cementation of a polyethylene liner into a metal shell
    • Delanois RE, Seyler TM, Essner A, et al. 2007. Cementation of a polyethylene liner into a metal shell. J Arthroplasty 22: 732-737.
    • (2007) J Arthroplasty , vol.22 , pp. 732-737
    • Delanois, R.E.1    Seyler, T.M.2    Essner, A.3
  • 9
    • 0036086080 scopus 로고    scopus 로고
    • Polyethylene liner cementation into fixed acetabular shells
    • Haft GF, Heiner AD, Callaghan JJ, et al. 2002. Polyethylene liner cementation into fixed acetabular shells. J Arthroplasty 17: 167-170.
    • (2002) J Arthroplasty , vol.17 , pp. 167-170
    • Haft, G.F.1    Heiner, A.D.2    Callaghan, J.J.3
  • 10
    • 0038351046 scopus 로고    scopus 로고
    • A biomechanical analysis of polyethylene liner cementation into a fixed metal acetabular shell
    • Haft GF, Heiner AD, Dorr LD, et al. 2003. A biomechanical analysis of polyethylene liner cementation into a fixed metal acetabular shell. J Bone Joint Surg Am 85-A: 1100-1110.
    • (2003) J Bone Joint Surg Am , vol.85 , pp. 1100-1110
    • Haft, G.F.1    Heiner, A.D.2    Dorr, L.D.3
  • 11
    • 67650745377 scopus 로고    scopus 로고
    • Cementation of a polyethylene liner into a metal acetabular shell: A biomechanical study
    • Hofmann AA, Prince EJ, Drake FT, et al. 2009. Cementation of a polyethylene liner into a metal acetabular shell: a biomechanical study. J Arthroplasty 24: 775-782.
    • (2009) J Arthroplasty , vol.24 , pp. 775-782
    • Hofmann, A.A.1    Prince, E.J.2    Drake, F.T.3
  • 12
    • 0036903342 scopus 로고    scopus 로고
    • Revision of polyethylene acetabular liners with a cemented polyethylene cup: A laboratory study
    • Kummer FJ, Adams MC, Dicesare PE., 2002. Revision of polyethylene acetabular liners with a cemented polyethylene cup: a laboratory study. J Arthroplasty 17: 1055-1057.
    • (2002) J Arthroplasty , vol.17 , pp. 1055-1057
    • Kummer, F.J.1    Adams, M.C.2    Dicesare, P.E.3
  • 13
    • 0031925520 scopus 로고    scopus 로고
    • Technique for acetabular liner revision in a nonmodular metal-backed component
    • LaPorte DM, Mont MA, Pierre-Jacques H, et al. 1998. Technique for acetabular liner revision in a nonmodular metal-backed component. J Arthroplasty 13: 348-350.
    • (1998) J Arthroplasty , vol.13 , pp. 348-350
    • Laporte, D.M.1    Mont, M.A.2    Pierre-Jacques, H.3
  • 14
    • 49549118697 scopus 로고    scopus 로고
    • Cementation of polyethylene liners into well-fixed metal shells at the time of revision total hip arthroplasty
    • Mauerhan DR, Peindl RD, Coley ER, et al. 2008. Cementation of polyethylene liners into well-fixed metal shells at the time of revision total hip arthroplasty. J Arthroplasty 23: 873-878.
    • (2008) J Arthroplasty , vol.23 , pp. 873-878
    • Mauerhan, D.R.1    Peindl, R.D.2    Coley, E.R.3
  • 15
    • 0034862928 scopus 로고    scopus 로고
    • The strength of a cement acetabular locking mechanism
    • Meldrum RD, Hollis JM., 2001. The strength of a cement acetabular locking mechanism. J Arthroplasty 16: 748-752.
    • (2001) J Arthroplasty , vol.16 , pp. 748-752
    • Meldrum, R.D.1    Hollis, J.M.2
  • 16
    • 0141924465 scopus 로고    scopus 로고
    • Cementation of an acetabular liner into a well-fixed acetabular shell during revision total hip arthroplasty
    • Springer BD, Hanssen AD, Lewallen DG., 2003. Cementation of an acetabular liner into a well-fixed acetabular shell during revision total hip arthroplasty. J Arthroplasty 18: 126-130.
    • (2003) J Arthroplasty , vol.18 , pp. 126-130
    • Springer, B.D.1    Hanssen, A.D.2    Lewallen, D.G.3
  • 17
    • 5344258184 scopus 로고    scopus 로고
    • A constrained liner cemented into a secure cementless acetabular shell
    • Callaghan JJ, Parvizi J, Novak CC., et al. 2004. A constrained liner cemented into a secure cementless acetabular shell. J Bone Joint Surg Am 86-A: 2206-2211.
    • (2004) J Bone Joint Surg Am , vol.86 , pp. 2206-2211
    • Callaghan, J.J.1    Parvizi, J.2    Novak, C.C.3
  • 18
    • 26444577041 scopus 로고    scopus 로고
    • The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures
    • Berend KR, Lombardi AV Jr, Mallory TH, et al. 2005. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty 20: 93-102.
    • (2005) J Arthroplasty , vol.20 , pp. 93-102
    • Berend, K.R.1    Lombardi, Jr.A.V.2    Mallory, T.H.3
  • 19
    • 46749140309 scopus 로고    scopus 로고
    • Modes of failure of Osteonics constrained tripolar implants: A retrospective analysis of forty-three failed implants
    • Guyen O, Lewallen DG, Cabanela ME., 2008. Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am 90: 1553-1560.
    • (2008) J Bone Joint Surg Am , vol.90 , pp. 1553-1560
    • Guyen, O.1    Lewallen, D.G.2    Cabanela, M.E.3
  • 20
    • 59649103930 scopus 로고    scopus 로고
    • Use of a dual mobility socket to manage total hip arthroplasty instability
    • Guyen O, Pibarot V, Vaz G, et al. 2009. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res 467: 465-472.
    • (2009) Clin Orthop Relat Res , vol.467 , pp. 465-472
    • Guyen, O.1    Pibarot, V.2    Vaz, G.3
  • 21
    • 34548432931 scopus 로고    scopus 로고
    • Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation
    • Guyen O, Pibarot V, Vaz G, et al. 2007. Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation. J Arthroplasty 22: 849-858.
    • (2007) J Arthroplasty , vol.22 , pp. 849-858
    • Guyen, O.1    Pibarot, V.2    Vaz, G.3
  • 22
    • 71749112733 scopus 로고    scopus 로고
    • Prevention of dislocation in total hip revision surgery using a dual mobility design
    • Philippot R, Adam P, Reckhaus M, et al. 2009. Prevention of dislocation in total hip revision surgery using a dual mobility design. Orthop Traumatol Surg Res 95: 407-413.
    • (2009) Orthop Traumatol Surg Res , vol.95 , pp. 407-413
    • Philippot, R.1    Adam, P.2    Reckhaus, M.3
  • 24
    • 33645879240 scopus 로고    scopus 로고
    • Dual articulation retentive acetabular liners and wear: Surface analysis of 40 retrieved polyethylene implants
    • Adam P, Farizon F, Fessy MH., 2005. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants. Rev Chir Orthop Reparatrice Appar Mot 91: 627-636.
    • (2005) Rev Chir Orthop Reparatrice Appar Mot , vol.91 , pp. 627-636
    • Adam, P.1    Farizon, F.2    Fessy, M.H.3
  • 25
    • 0031926423 scopus 로고    scopus 로고
    • Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases
    • Goetz DD, Capello WN, Callaghan JJ, et al. 1998. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases. J Bone Joint Surg Am 80: 502-509.
    • (1998) J Bone Joint Surg Am , vol.80 , pp. 502-509
    • Goetz, D.D.1    Capello, W.N.2    Callaghan, J.J.3
  • 26
    • 78649330927 scopus 로고    scopus 로고
    • The use of a cemented dual mobility socket to treat recurrent dislocation
    • Hamadouche M, Biau DJ, Huten D, et al. 2010. The use of a cemented dual mobility socket to treat recurrent dislocation. Clin Orthop Relat Res 468: 3248-3254.
    • (2010) Clin Orthop Relat Res , vol.468 , pp. 3248-3254
    • Hamadouche, M.1    Biau, D.J.2    Huten, D.3
  • 27
    • 41949088160 scopus 로고    scopus 로고
    • Dual mobility cemented cups have low dislocation rates in THA revisions
    • Langlais FL, Ropars M, Gaucher F, et al. 2008. Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res 466: 389-395.
    • (2008) Clin Orthop Relat Res , vol.466 , pp. 389-395
    • Langlais, F.L.1    Ropars, M.2    Gaucher, F.3
  • 28
    • 0242574966 scopus 로고    scopus 로고
    • The use of a constrained acetabular component to treat instability after total hip arthroplasty
    • Shrader MW, Parvizi J, Lewallen DG., 2003. The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am 85-A: 2179-2183.
    • (2003) J Bone Joint Surg Am , vol.85 , pp. 2179-2183
    • Shrader, M.W.1    Parvizi, J.2    Lewallen, D.G.3
  • 29
    • 0024000543 scopus 로고
    • Wear, creep, and frictional heating of femoral implant articulating surfaces and the effect on long-term performance-Part II, Friction, heating, and torque
    • Davidson JA, Schwartz G, Lynch G, et al. 1988. Wear, creep, and frictional heating of femoral implant articulating surfaces and the effect on long-term performance-Part II, Friction, heating, and torque. J Biomed Mater Res 22: 69-91.
    • (1988) J Biomed Mater Res , vol.22 , pp. 69-91
    • Davidson, J.A.1    Schwartz, G.2    Lynch, G.3
  • 30
    • 33846815186 scopus 로고    scopus 로고
    • Unconstrained tripolar hip implants: Effect on hip stability
    • Guyen O, Chen QS, Bejui-Hugues J, et al. 2007. Unconstrained tripolar hip implants: effect on hip stability. Clin Orthop Relat Res 455: 202-208.
    • (2007) Clin Orthop Relat Res , vol.455 , pp. 202-208
    • Guyen, O.1    Chen, Q.S.2    Bejui-Hugues, J.3
  • 31
    • 0033384592 scopus 로고    scopus 로고
    • Early polyethylene wear and osteolysis in cementless total hip arthroplasty: The influence of femoral head size and polyethylene thickness
    • Lee PC, Shih CH, Chen WJ, et al. 1999. Early polyethylene wear and osteolysis in cementless total hip arthroplasty: the influence of femoral head size and polyethylene thickness. J Arthroplasty 14: 976-981.
    • (1999) J Arthroplasty , vol.14 , pp. 976-981
    • Lee, P.C.1    Shih, C.H.2    Chen, W.J.3
  • 32
    • 84855520269 scopus 로고    scopus 로고
    • Unexpected failure of highly cross-linked polyethylene acetabular liner
    • Waewsawangwong W, Goodman SB., 2012. Unexpected failure of highly cross-linked polyethylene acetabular liner. J Arthroplasty 27: e321-e324.
    • (2012) J Arthroplasty , vol.27
    • Waewsawangwong, W.1    Goodman, S.B.2
  • 33
    • 9644258640 scopus 로고    scopus 로고
    • Fixation strength of an all-metal acetabular component cemented into an acetabular shell: A biomechanical analysis
    • Ebramzadeh E, Beaule PE, Culwell JL, et al. 2004. Fixation strength of an all-metal acetabular component cemented into an acetabular shell: a biomechanical analysis. J Arthroplasty 19: 45-49.
    • (2004) J Arthroplasty , vol.19 , pp. 45-49
    • Ebramzadeh, E.1    Beaule, P.E.2    Culwell, J.L.3
  • 34
    • 77955575155 scopus 로고    scopus 로고
    • Failed metal-on-metal hip arthroplasties: A spectrum of clinical presentations and operative findings
    • Browne JA, Bechtold CD, Berry DJ, et al. 2010. Failed metal-on-metal hip arthroplasties: a spectrum of clinical presentations and operative findings. Clin Orthop Relat Res 468: 2313-2320.
    • (2010) Clin Orthop Relat Res , vol.468 , pp. 2313-2320
    • Browne, J.A.1    Bechtold, C.D.2    Berry, D.J.3
  • 35
    • 0032440129 scopus 로고    scopus 로고
    • Results of cemented metal-backed acetabular components: A 10-year-average follow-up study
    • Chen FS, Di Cesare PE, Kale AA, et al. 1998. Results of cemented metal-backed acetabular components: a 10-year-average follow-up study. J Arthroplasty 13: 867-873.
    • (1998) J Arthroplasty , vol.13 , pp. 867-873
    • Chen, F.S.1    Di Cesare, P.E.2    Kale, A.A.3
  • 36
    • 0026738353 scopus 로고
    • Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis
    • Nizard RS, Sedel L, Christel P, et al. 1992. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res 282: 53-63.
    • (1992) Clin Orthop Relat Res , vol.282 , pp. 53-63
    • Nizard, R.S.1    Sedel, L.2    Christel, P.3
  • 37
    • 0023617819 scopus 로고
    • Long-term results and survivorship of the McKee-Farrar total hip prosthesis
    • Visuri T., 1987. Long-term results and survivorship of the McKee-Farrar total hip prosthesis. Arch Orthop Trauma Surg 106: 368-374.
    • (1987) Arch Orthop Trauma Surg , vol.106 , pp. 368-374
    • Visuri, T.1
  • 38
    • 0031593621 scopus 로고    scopus 로고
    • The Burch-Schneider anti-protrusio cage in revision total hip arthroplasty: Indications, principles and long-term results
    • Gill TJ, Sledge JB, Muller ME., 1998. The Burch-Schneider anti-protrusio cage in revision total hip arthroplasty: indications, principles and long-term results. J Bone Joint Surg Br 80: 946-953.
    • (1998) J Bone Joint Surg Br , vol.80 , pp. 946-953
    • Gill, T.J.1    Sledge, J.B.2    Muller, M.E.3
  • 39
    • 0033852227 scopus 로고    scopus 로고
    • The Kerboull acetabular reinforcement device in major acetabular reconstructions
    • Kerboull M, Hamadouche M, Kerboull L., 2000. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res 378: 155-168.
    • (2000) Clin Orthop Relat Res , vol.378 , pp. 155-168
    • Kerboull, M.1    Hamadouche, M.2    Kerboull, L.3
  • 40
    • 83055178606 scopus 로고    scopus 로고
    • Revision total hip arthroplasty using a reconstruction cage device and a cemented dual mobility cup
    • Schneider L, Philippot R, Boyer B, et al. 2011. Revision total hip arthroplasty using a reconstruction cage device and a cemented dual mobility cup. Orthop Taumatol Surg Res 97: 807-813.
    • (2011) Orthop Taumatol Surg Res , vol.97 , pp. 807-813
    • Schneider, L.1    Philippot, R.2    Boyer, B.3
  • 41
    • 10344237551 scopus 로고    scopus 로고
    • Clinical validation of a structural porous tantalum biomaterial for adult reconstruction
    • Bobyn JD, Poggie RA, Krygier JJ, et al. 2004. Clinical validation of a structural porous tantalum biomaterial for adult reconstruction. J Bone Joint Surg Am 86-A: 123-129.
    • (2004) J Bone Joint Surg Am , vol.86 , pp. 123-129
    • Bobyn, J.D.1    Poggie, R.A.2    Krygier, J.J.3
  • 42
    • 3343023377 scopus 로고    scopus 로고
    • Interfacial frictional behavior: Cancellous bone, cortical bone, and a novel porous tantalum biomaterial
    • Zhang Y, Ahn PB, Fitzpatrick DC, et al. 1999. Interfacial frictional behavior: cancellous bone, cortical bone, and a novel porous tantalum biomaterial. J Musculoskel Res 3: 245-251.
    • (1999) J Musculoskel Res , vol.3 , pp. 245-251
    • Zhang, Y.1    Ahn, P.B.2    Fitzpatrick, D.C.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.