Accuracy of tidal volume, compliance, and resistance measurements on neonatal ventilator displays: An in vitro assessment

Author keywords

compliance; lung function; neonatal ventilators; resistance; tidal volume

Indexed keywords

ARTICLE; BREATHING MECHANICS; HUMAN; IN VITRO STUDY; LUNG COMPLIANCE; LUNG DISEASE; LUNG MODEL; LUNG RESISTANCE; NEWBORN INTENSIVE CARE; PEAK INSPIRATORY FLOW; POSITIVE END EXPIRATORY PRESSURE; PRIORITY JOURNAL; TIDAL VOLUME; VENTILATOR;

EID: 84864282228     PISSN: 15297535     EISSN: 19473893     Source Type: Journal    
DOI: 10.1097/PCC.0b013e3182455586     Document Type: Article
Times cited : (12)

References (31)
  • 1
    • 0032147673 scopus 로고    scopus 로고
    • Accuracy of volume measurements in mechanically ventilated newborns: A comparative study of commercial devices
    • Roske K, Foitzik B, Wauer RR, et al: Accuracy of volume measurements in mechanically ventilated newborns: A comparative study of commercial devices. J Clin Monit Comput 1998; 14:413-420
    • (1998) J Clin Monit Comput , vol.14 , pp. 413-420
    • Roske, K.1    Foitzik, B.2    Wauer, R.R.3
  • 2
    • 0027450228 scopus 로고
    • Long-term pulmonary consequences of respiratory distress syndrome in preterm infants treated with exogenous surfactant
    • Abbasi S, Bhutani VK, Gerdes JS: Long-term pulmonary consequences of respiratory distress syndrome in preterm infants treated with exogenous surfactant. J Pediatr 1993; 122:446-452 (Pubitemid 23067231)
    • (1993) Journal of Pediatrics , vol.122 , Issue.3 , pp. 446-452
    • Abbasi, S.1    Bhutani, V.K.2    Gerdes, J.S.3
  • 3
    • 0034958832 scopus 로고    scopus 로고
    • In vitro assessment of equipment and software to assess tidal breathing parameters in infants
    • DOI 10.1183/09031936.01.17101000
    • Schmalisch G, Foitzik B, Wauer RR, et al: In vitro assessment of equipment and software to assess tidal breathing parameters in infants. Eur Respir J 2001; 17:100-107 (Pubitemid 32600563)
    • (2001) European Respiratory Journal , vol.17 , Issue.1 , pp. 100-107
    • Schmalisch, G.1    Foitzik, B.2    Wauer, R.R.3    Stocks, J.4
  • 4
    • 0025116931 scopus 로고
    • Pulmonary mechanics and energetics of normal, non-ventilated low birthweight infants
    • Abbasi S, Bhutani VK: Pulmonary mechanics and energetics of normal, non-ventilated low birthweight infants. Pediatr Pulmonol 1990; 8:89-95
    • (1990) Pediatr Pulmonol , vol.8 , pp. 89-95
    • Abbasi, S.1    Bhutani, V.K.2
  • 5
    • 0025981585 scopus 로고
    • Pulmonary mechanics in preterm neonates with respiratory failure treated with high-frequency oscillatory ventilation compared with conventional mechanical ventilation
    • Abbasi S, Bhutani VK, Spitzer AR, et al: Pulmonary mechanics in preterm neonates with respiratory failure treated with high-frequency oscillatory ventilation compared with conventional mechanical ventilation. Pediatrics 1991; 87:487-493
    • (1991) Pediatrics , vol.87 , pp. 487-493
    • Abbasi, S.1    Bhutani, V.K.2    Spitzer, A.R.3
  • 6
    • 0028394933 scopus 로고
    • Pulmonary function testing in the critically ill neonate, Part II: Methodology
    • Cullen JA, Greenspan JS, Antunes MJ, et al: Pulmonary function testing in the critically ill neonate, Part II: Methodology. Neonatal Netw 1994; 13:7-13
    • (1994) Neonatal Netw , vol.13 , pp. 7-13
    • Cullen, J.A.1    Greenspan, J.S.2    Antunes, M.J.3
  • 8
    • 0025172232 scopus 로고
    • Measuring on-line compliance in ventilated infants using hot wire anemometry
    • Baboolal R, Kirpalani H: Measuring on-line compliance in ventilated infants using hot wire anemometry. Crit Care Med 1990; 18:1070-1072 (Pubitemid 20364043)
    • (1990) Critical Care Medicine , vol.18 , Issue.10 , pp. 1070-1072
    • Baboolal, R.1    Kirpalani, H.2
  • 9
    • 0023810275 scopus 로고
    • Evaluation of neonatal pulmonary mechanics and energetics: A two factor least mean square analysis
    • Bhutani VK, Sivieri EM, Abbasi S, et al: Evaluation of neonatal pulmonary mechanics and energetics: A two factor least mean square analysis. Pediatr Pulmonol 1988; 4:150-158
    • (1988) Pediatr Pulmonol , vol.4 , pp. 150-158
    • Bhutani, V.K.1    Sivieri, E.M.2    Abbasi, S.3
  • 10
    • 0035009705 scopus 로고    scopus 로고
    • The infant lung function model: A mechanical analogue to test infant lung function equipment
    • DOI 10.1183/09031936.01.17407550
    • Frey U, Reinmann B, Stocks J: The infant lung function model: A mechanical analogue to test infant lung function equipment. Eur Respir J 2001; 17:755-764 (Pubitemid 32492567)
    • (2001) European Respiratory Journal , vol.17 , Issue.4 , pp. 755-764
    • Frey, U.1    Reinmann, B.2    Stocks, J.3
  • 11
    • 0001875190 scopus 로고    scopus 로고
    • Plethysmographic assessment of functional residual capacity and airway resistance
    • Stocks JF, Sly PD, Tepper RS, et al (Eds). New York, NY, John Wiley and Sons
    • Stocks JF, Marchal R, Kraemer P, et al: Plethysmographic assessment of functional residual capacity and airway resistance. In: Stocks Infant Respiratory Function Testing. Stocks JF, Sly PD, Tepper RS, et al (Eds). New York, NY, John Wiley and Sons, 1996, pp 191-240
    • (1996) Stocks Infant Respiratory Function Testing , pp. 191-240
    • Stocks, J.F.1    Marchal, R.2    Kraemer, P.3
  • 12
    • 0343523177 scopus 로고
    • The action of adiabatic effects on the compliance of an artificial thorax
    • Hill DW, Moore V: The action of adiabatic effects on the compliance of an artificial thorax. Br J Anaesth 1965; 37:19-22
    • (1965) Br J Anaesth , vol.37 , pp. 19-22
    • Hill, D.W.1    Moore, V.2
  • 13
    • 0015293317 scopus 로고
    • An airway resistance for use in an artificial lung
    • Burton GW, Fox DE: An airway resistance for use in an artificial lung. Br J Anaesth 1972; 44:1253-1262
    • (1972) Br J Anaesth , vol.44 , pp. 1253-1262
    • Burton, G.W.1    Fox, D.E.2
  • 15
    • 0036852034 scopus 로고    scopus 로고
    • Accuracy of displayed values of tidal volume in the pediatric intensive care unit
    • Castle RA, Dunne CJ, Mok Q, et al: Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 2002; 30:2566-2574 (Pubitemid 35351880)
    • (2002) Critical Care Medicine , vol.30 , Issue.11 , pp. 2566-2574
    • Castle, R.A.1    Dunne, C.J.2    Mok, Q.3    Wade, A.M.4    Stocks, J.5
  • 17
    • 0027654065 scopus 로고
    • The effects of bedside pulmonary mechanics testing during infant mechanical ventilation: A retrospective analysis
    • Rosen WC, Mammel MC, Fisher JB, et al: The effects of bedside pulmonary mechanics testing during infant mechanical ventilation: A retrospective analysis. Pediatr Pulmonol 1993; 16:147-152
    • (1993) Pediatr Pulmonol , vol.16 , pp. 147-152
    • Rosen, W.C.1    Mammel, M.C.2    Fisher, J.B.3
  • 18
    • 0034122861 scopus 로고    scopus 로고
    • Laboratory assessment of the Bird T-Bird VS ventilator performance using a model lung
    • Wong LS, McGuire NM: Laboratory assessment of the Bird T-Bird VS ventilator performance using a model lung. Br J Anaesth 2000; 84:811-817 (Pubitemid 30312059)
    • (2000) British Journal of Anaesthesia , vol.84 , Issue.6 , pp. 811-817
    • Wong, L.S.S.1    McGuire, N.M.2
  • 19
    • 33847196138 scopus 로고    scopus 로고
    • Performance of neonatal ventilators in volume targeted ventilation mode
    • Sharma A, Milner AD, Greenough A: Performance of neonatal ventilators in volume targeted ventilation mode. Acta Paediatr 2007; 96:176-180
    • (2007) Acta Paediatr , vol.96 , pp. 176-180
    • Sharma, A.1    Milner, A.D.2    Greenough, A.3
  • 20
    • 0023795362 scopus 로고
    • Some infant ventilators do not limit peak inspiratory pressure reliably during active expiration
    • Kirpalani H, Santos-Lyn R, Roberts R: Some infant ventilators do not limit peak inspiratory pressure reliably during active expiration. Crit Care Med 1988; 16:880-883
    • (1988) Crit Care Med , vol.16 , pp. 880-883
    • Kirpalani, H.1    Santos-Lyn, R.2    Roberts, R.3
  • 22
    • 0017076099 scopus 로고
    • Evaluation of the lung ventilator performance analyser
    • Chakrabarti MK, Sykes MK: Evaluation of the Lung Ventilator Performance Analyser. Anaesthesia 1976; 31:521-528
    • (1976) Anaesthesia , vol.31 , pp. 521-528
    • Chakrabarti, M.K.1    Sykes, M.K.2
  • 23
    • 0024075477 scopus 로고
    • Construction of linear resistance units for a model lung
    • Mecklenburgh JS: Construction of linear resistance units for a model lung. Med Biol Eng Comput 1988; 26:552-554
    • (1988) Med Biol Eng Comput , vol.26 , pp. 552-554
    • Mecklenburgh, J.S.1
  • 24
    • 0018921891 scopus 로고
    • The construction of linear resistances for the testing of ventilators
    • Saklad M, Weyerhaeuser R: The construction of linear resistances for the testing of ventilators. Anesthesiology 1980; 52:71-73 (Pubitemid 10212388)
    • (1980) Anesthesiology , vol.52 , Issue.1 , pp. 71-73
    • Saklad, M.1    Weyerhaeuser, R.2
  • 25
    • 0036708193 scopus 로고    scopus 로고
    • Are tidal volume measurements in neonatal pressure-controlled ventilation accurate?
    • DOI 10.1002/ppul.10150
    • Chow LC, Vanderhal A, Raber J, et al: Are tidal volume measurements in neonatal pressure-controlled ventilation accurate? Pediatr Pulmonol 2002; 34:196-202 (Pubitemid 34971072)
    • (2002) Pediatric Pulmonology , vol.34 , Issue.3 , pp. 196-202
    • Chow, L.C.1    Vanderhal, A.2    Raber, J.3    Sola, A.4
  • 26
    • 0034088680 scopus 로고    scopus 로고
    • Resistance of pediatric and neonatal endotracheal tubes: Influence of flow rate, size, and shape
    • Manczur T, Greenough A, Nicholson GP, et al: Resistance of pediatric and neonatal endotracheal tubes: Influence of flow rate, size, and shape. Crit Care Med 2000; 28:1595-1598 (Pubitemid 30324546)
    • (2000) Critical Care Medicine , vol.28 , Issue.5 , pp. 1595-1598
    • Manczur, T.1    Greenough, A.2    Nicholson, G.P.3    Rafferty, G.F.4
  • 28
    • 0018069090 scopus 로고
    • Tracheal tubes and connectors used in neonates: Dimensions and resistance to breathing
    • Hatch DJ: Tracheal tubes and connectors used in neonates-dimensions and resistance to breathing. Br J Anaesth 1978; 50:959-964 (Pubitemid 9063446)
    • (1978) British Journal of Anaesthesia , vol.50 , Issue.9 , pp. 959-964
    • Hatch, D.J.1
  • 29
    • 0343893748 scopus 로고    scopus 로고
    • Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome
    • Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med 2000; 342:1301-1308
    • (2000) The Acute Respiratory Distress Syndrome Network. N Engl J Med , vol.342 , pp. 1301-1308
  • 30
    • 0026813535 scopus 로고
    • Variability of dynamic compliance measurements in spontaneously breathing and ventilated newborn infants
    • Ratjen FA, Wiesemann HG: Variability of dynamic compliance measurements in spontaneously breathing and ventilated newborn infants. Pediatr Pulmonol 1992; 12:73-80
    • (1992) Pediatr Pulmonol , vol.12 , pp. 73-80
    • Ratjen, F.A.1    Wiesemann, H.G.2
  • 31
    • 0029919515 scopus 로고    scopus 로고
    • Intrasubject variability of repeated pulmonary function measurements in preterm ventilated infants
    • DOI 10.1002/(SICI)1099-0496(19 9601)21:1<35::AID-PPUL6>3. 0.CO;2-P
    • Gonzalez A, Tortorolo L, Gerhardt T, et al: Intrasubject variability of repeated pulmonary function measurements in preterm ventilated infants. Pediatr Pulmonol 1996; 21:35-41 (Pubitemid 26104993)
    • (1996) Pediatric Pulmonology , vol.21 , Issue.1 , pp. 35-41
    • Gonzalez, A.1    Tortorolo, L.2    Gerhardt, T.3    Rojas, M.4    Everett, R.5    Bancalari, E.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.