3D elbow kinematics with monoplanar fluoroscopy: In silico evaluation

Author keywords

[No Author keywords available]

Indexed keywords

AUTOMATED TECHNIQUES; BONE SEGMENTS; DISTANCE MAP; ELBOW JOINTS; ELBOW KINEMATICS; ESTIMATION ERRORS; IMAGE FEATURES; IN-PLANE; IN-SILICO; IN-VITRO; IN-VIVO; INTER QUARTILE RANGES; JOINT MOTION ANALYSIS; LONG BONE; MEASUREMENT ACCURACY; OPTIMIZATION ALGORITHMS; OUT-OF-PLANE; PHYSIOLOGICAL CONDITION; POSE ESTIMATION; POSE PARAMETERS; SMALL DISPERSION; VALIDATION STUDY;

EID: 80052691218     PISSN: 16876172     EISSN: 16876180     Source Type: Journal    
DOI: 10.1155/2010/142989     Document Type: Article
Times cited : (13)

References (44)
  • 2
    • 0033201375 scopus 로고    scopus 로고
    • Model-based method for the reconstruction of total knee replacement kinematics
    • DOI 10.1109/42.811310
    • Zuffi S., Leardini A., Catani F., Fantozzi S., Cappello A., A model-based method for the reconstruction of total knee replacement kinematics IEEE Transactions on Medical Imaging 1999 18 10 981 991 (Pubitemid 32211113)
    • (1999) IEEE Transactions on Medical Imaging , vol.18 , Issue.10 , pp. 981-991
    • Zuffi, S.1    Leardini, A.2    Catani, F.3    Fantozzi, S.4    Cappello, A.5
  • 3
    • 57049181892 scopus 로고    scopus 로고
    • A 3D kinematic estimation of knee prosthesis using X-ray projection images: Clinical assessment of the improved algorithm for fluoroscopy images
    • Hirokawa S., Abrar Hossain M., Kihara Y., Ariyoshi S., A 3D kinematic estimation of knee prosthesis using X-ray projection images: clinical assessment of the improved algorithm for fluoroscopy images Medical and Biological Engineering and Computing 2008 46 12 1253 1262
    • (2008) Medical and Biological Engineering and Computing , vol.46 , Issue.12 , pp. 1253-1262
    • Hirokawa, S.1    Abrar Hossain, M.2    Kihara, Y.3    Ariyoshi, S.4
  • 4
  • 6
    • 10444220215 scopus 로고    scopus 로고
    • In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics
    • DOI 10.1016/j.jbiomech.2004.02.042, PII S002192900400082X
    • Dennis D. A., Mahfouz M. R., Komistek R. D., Hoff W., In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics Journal of Biomechanics 2005 38 2 241 253 (Pubitemid 39642944)
    • (2005) Journal of Biomechanics , vol.38 , Issue.2 , pp. 241-253
    • Dennis, D.A.1    Mahfouz, M.R.2    Komistek, R.D.3    Hoff, W.4
  • 7
    • 20444475413 scopus 로고    scopus 로고
    • Soft tissue artifact compensation in knee kinematics by double anatomical landmark calibration: Performance of a novel method during selected motor tasks
    • DOI 10.1109/TBME.2005.846728
    • Cappello A., Stagni R., Fantozzi S., Leardini A., Soft tissue artifact compensation in knee kinematics by double anatomical landmark calibration: performance of a novel method during selected motor tasks IEEE Transactions on Biomedical Engineering 2005 52 6 992 998 (Pubitemid 40813831)
    • (2005) IEEE Transactions on Biomedical Engineering , vol.52 , Issue.6 , pp. 992-998
    • Cappello, A.1    Stagni, R.2    Fantozzi, S.3    Leardini, A.4
  • 8
    • 0030175447 scopus 로고    scopus 로고
    • Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy
    • DOI 10.1109/10.495283
    • Banks S. A., Hodge W. A., Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy IEEE Transactions on Biomedical Engineering 1996 43 6 638 649 (Pubitemid 26158677)
    • (1996) IEEE Transactions on Biomedical Engineering , vol.43 , Issue.6 , pp. 638-649
    • Banks, S.A.1    Hodge, W.A.2
  • 12
    • 0031543693 scopus 로고    scopus 로고
    • In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties
    • DOI 10.1016/S0883-5403(97)90026-7
    • Banks S. A., Markovich G. D., Hodge W. A., In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties The Journal of Arthroplasty 1997 12 3 297 304 (Pubitemid 29021330)
    • (1997) Journal of Arthroplasty , vol.12 , Issue.3 , pp. 297-304
    • Banks, S.A.1    Markovich, G.D.2    Hodge, W.A.3
  • 13
    • 0344012667 scopus 로고    scopus 로고
    • A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images
    • Mahfouz M. R., Hoff W. A., Komistek R. D., Dennis D. A., A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images IEEE Transactions on Medical Imaging 2003 22 12 1561 1574
    • (2003) IEEE Transactions on Medical Imaging , vol.22 , Issue.12 , pp. 1561-1574
    • Mahfouz, M.R.1    Hoff, W.A.2    Komistek, R.D.3    Dennis, D.A.4
  • 16
    • 0035063689 scopus 로고    scopus 로고
    • Determination of hip joint separation and the forces generated due to impact loading conditions
    • DOI 10.1016/S0021-9290(00)00239-6, PII S0021929000002396
    • Dennis D. A., Komistek R. D., Northcut E. J., Ochoa J. A., Ritchie A., In vivo determination of hip joint separation and the forces generated due to impact loading conditions Journal of Biomechanics 2001 34 5 623 629 (Pubitemid 32328250)
    • (2001) Journal of Biomechanics , vol.34 , Issue.5 , pp. 623-629
    • Dennis, D.A.1    Komistek, R.D.2    Northcut, E.J.3    Ochoa, J.A.4    Ritchie, A.5
  • 17
    • 0036796360 scopus 로고    scopus 로고
    • In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty
    • Komistek R. D., Dennis D. A., Ochoa J. A., Haas B. D., Hammill C., In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty The Journal of Bone and Joint Surgery. American 2002 84 10 1836 1841 (Pubitemid 35176478)
    • (2002) Journal of Bone and Joint Surgery - Series A , vol.84 , Issue.10 , pp. 1836-1841
    • Komistek, R.D.1    Dennis, D.A.2    Ochoa, J.A.3    Haas, B.D.4    Hammill, C.5
  • 18
    • 0032189583 scopus 로고    scopus 로고
    • Three-dimensional determination of femoral-tibial contact positions under in vivo conditions using fluoroscopy
    • DOI 10.1016/S0268-0033(98)00009-6, PII S0268003398000096
    • Hoff W. A., Komistek R. D., Dennis D. A., Gabriel S. M., Walker S. A., Three-dimensional determination of femoral-tibial contact positions under in vivo conditions using fluoroscopy Clinical Biomechanics 1998 13 7 455 472 (Pubitemid 28442179)
    • (1998) Clinical Biomechanics , vol.13 , Issue.7 , pp. 455-472
    • Hoff, W.A.1    Komistek, R.D.2    Dennis, D.A.3    Gabriel, S.M.4    Walker, S.A.5
  • 21
    • 37249006719 scopus 로고    scopus 로고
    • Image-based RSA: Roentgen stereophotogrammetric analysis based on 2D-3D image registration
    • DOI 10.1016/j.jbiomech.2007.07.002, PII S0021929007003107
    • de Bruin P. W., Kaptein B. L., Stoel B. C., Reiber J. H. C., Rozing P. M., Valstar E. R., Image-based RSA: roentgen stereophotogrammetric analysis based on 2D-3D image registration Journal of Biomechanics 2008 41 1 155 164 (Pubitemid 350265895)
    • (2008) Journal of Biomechanics , vol.41 , Issue.1 , pp. 155-164
    • De Bruin, P.W.1    Kaptein, B.L.2    Stoel, B.C.3    Reiber, J.H.C.4    Rozing, P.M.5    Valstar, E.R.6
  • 22
    • 0035354206 scopus 로고    scopus 로고
    • In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs: Application to knee kinematics
    • DOI 10.1109/42.929617, PII S0278006201044081
    • You B.-M., Siy P., Anderst W., Tashman S., In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs: application to knee kinematics IEEE Transactions on Medical Imaging 2001 20 6 514 525 (Pubitemid 32657855)
    • (2001) IEEE Transactions on Medical Imaging , vol.20 , Issue.6 , pp. 514-525
    • You, B.-M.1    Siy, P.2    Anderst, W.3    Tashman, S.4
  • 23
    • 57849144293 scopus 로고    scopus 로고
    • Validation of three-dimensional model-based tibio-femoral tracking during running
    • Anderst W., Zauel R., Bishop J., Demps E., Tashman S., Validation of three-dimensional model-based tibio-femoral tracking during running Medical Engineering Physics 2009 31 1 10 16
    • (2009) Medical Engineering Physics , vol.31 , Issue.1 , pp. 10-16
    • Anderst, W.1    Zauel, R.2    Bishop, J.3    Demps, E.4    Tashman, S.5
  • 24
    • 23644455465 scopus 로고    scopus 로고
    • Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy
    • DOI 10.1115/1.1933949
    • Fregly B. J., Rahman H. A., Banks S. A., Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy Journal of Biomechanical Engineering 2005 127 4 692 699 (Pubitemid 41116929)
    • (2005) Journal of Biomechanical Engineering , vol.127 , Issue.4 , pp. 692-699
    • Fregly, B.J.1    Rahman, H.A.2    Banks, S.A.3
  • 25
    • 0031172670 scopus 로고    scopus 로고
    • Parameters for modeling the upper extremity
    • DOI 10.1016/S0021-9290(97)00011-0, PII S0021929097000110
    • Veeger H. E. J., Yu B., An K.-N., Rozendal R. H., Parameters for modeling the upper extremity Journal of Biomechanics 1997 30 6 647 652 (Pubitemid 27214199)
    • (1997) Journal of Biomechanics , vol.30 , Issue.6 , pp. 647-652
    • Veeger, H.E.J.1    Yu, B.2    An, K.-N.3    Rozendal, R.H.4
  • 28
    • 0029089828 scopus 로고
    • Loose-hinge total elbow arthroplasty: An experimental study of the effects of implant alignment on three-dimensional elbow kinematics
    • Schuind F., ODriscoll S., Korinek S., An K. N., Morrey B. F., Loose-hinge total elbow arthroplasty: an experimental study of the effects of implant alignment on three-dimensional elbow kinematics The Journal of Arthroplasty 1995 10 5 670 678
    • (1995) The Journal of Arthroplasty , vol.10 , Issue.5 , pp. 670-678
    • Schuind, F.1    Odriscoll, S.2    Korinek, S.3    An, K.N.4    Morrey, B.F.5
  • 29
    • 0032893788 scopus 로고    scopus 로고
    • A marker-based measurement procedure for unconstrained wrist and elbow motions
    • DOI 10.1016/S0021-9290(99)00036-6, PII S0021929099000366
    • Schmidt R., Disselhorst-Klug C., Silny J., Rau G., A marker-based measurement procedure for unconstrained wrist and elbow motions Journal of Biomechanics 1999 32 6 615 621 (Pubitemid 29219710)
    • (1999) Journal of Biomechanics , vol.32 , Issue.6 , pp. 615-621
    • Schmidt, R.1    Disselhorst-Klug, C.2    Silny, J.3    Rau, G.4
  • 31
    • 21744442284 scopus 로고    scopus 로고
    • Soft tissue artefact assessment in humeral axial rotation
    • DOI 10.1016/j.gaitpost.2004.04.001
    • Cutti A. G., Paolini G., Troncossi M., Cappello A., Davalli A., Soft tissue artefact assessment in humeral axial rotation Gait Posture 2005 21 3 341 349 (Pubitemid 41583199)
    • (2005) Gait and Posture , vol.21 , Issue.3 , pp. 341-349
    • Cutti, A.G.1    Paolini, G.2    Troncossi, M.3    Cappello, A.4    Davalli, A.5
  • 32
    • 13844256386 scopus 로고    scopus 로고
    • Quantification of soft tissue artefact in motion analysis by combining 3D fluoroscopy and stereophotogrammetry: A study on two subjects
    • DOI 10.1016/j.clinbiomech.2004.11.012, PII S0268003304002839
    • Stagni R., Fantozzi S., Cappello A., Leardini A., Quantification of soft tissue artefact in motion analysis by combining 3D fluoroscopy and stereophotogrammetry: a study on two subjects Clinical Biomechanics 2005 20 3 320 329 (Pubitemid 40247688)
    • (2005) Clinical Biomechanics , vol.20 , Issue.3 , pp. 320-329
    • Stagni, R.1    Fantozzi, S.2    Cappello, A.3    Leardini, A.4
  • 33
    • 0029288672 scopus 로고
    • Recovering the position and orientation of free-form objects from image contours using 3D distance maps
    • Lavallee S., Szeliski R., Recovering the position and orientation of free-form objects from image contours using 3D distance maps IEEE Transactions on Pattern Analysis and Machine Intelligence 1995 17 4 378 390
    • (1995) IEEE Transactions on Pattern Analysis and Machine Intelligence , vol.17 , Issue.4 , pp. 378-390
    • Lavallee, S.1    Szeliski, R.2
  • 35
    • 0000169232 scopus 로고
    • An algorithm for least-squares estimation of nonlinear parameters
    • Marquardt D. W., An algorithm for least-squares estimation of nonlinear parameters SIAM Journal on Applied Mathematics 1963 11 2 431 441
    • (1963) SIAM Journal on Applied Mathematics , vol.11 , Issue.2 , pp. 431-441
    • Marquardt, D.W.1
  • 37
    • 0020787054 scopus 로고
    • A hierarchical data structure for multidimensional digital images
    • Yau M.-M., Srihari S. N., A hierarchical data structure for multidimensional digital images Communications of the ACM 1983 26 7 504 515
    • (1983) Communications of the ACM , vol.26 , Issue.7 , pp. 504-515
    • Yau, M.-M.1    Srihari, S.N.2
  • 38
    • 36249029277 scopus 로고    scopus 로고
    • Effects of radiograph projection parameter uncertainty on TKA kinematics from model-image registration
    • DOI 10.1016/j.jbiomech.2007.06.011, PII S0021929007002734
    • Chouteau J., Lerat J. L., Testa R., Moyen B., Banks S. A., Effects of radiograph projection parameter uncertainty on TKA kinematics from model-image registration Journal of Biomechanics 2007 40 16 3744 3747 (Pubitemid 350122509)
    • (2007) Journal of Biomechanics , vol.40 , Issue.16 , pp. 3744-3747
    • Chouteau, J.1    Lerat, J.L.2    Testa, R.3    Moyen, B.4    Banks, S.A.5
  • 39
    • 0033430134 scopus 로고    scopus 로고
    • Correction for geometric image distortion in the x-ray imaging chain: Local technique versus global technique
    • DOI 10.1118/1.598800
    • Gronenschild E., Correction for geometric image distortion in the x-ray imaging chain: local technique versus global technique Medical Physics 1999 26 12 2602 2616 (Pubitemid 30007706)
    • (1999) Medical Physics , vol.26 , Issue.12 , pp. 2602-2616
    • Gronenschild, E.1
  • 40
    • 0031455395 scopus 로고    scopus 로고
    • The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain
    • DOI 10.1118/1.598101
    • Gronenschild E., The accuracy and reproducibility of a global method to correct for geometric image distortion in the X-ray imaging chain Medical Physics 1997 24 12 1875 1888 (Pubitemid 28013502)
    • (1997) Medical Physics , vol.24 , Issue.12 , pp. 1875-1888
    • Gronenschild, E.1
  • 41
    • 0037321795 scopus 로고    scopus 로고
    • A global method based on thin-plate splines for correction of geometric distortion: An application to fluoroscopic images
    • DOI 10.1118/1.1538228
    • Fantozzi S., Cappello A., Leardini A., A global method based on thin-plate splines for correction of geometric distortion: an application to fluoroscopic images Medical Physics 2003 30 2 124 131 (Pubitemid 36222847)
    • (2003) Medical Physics , vol.30 , Issue.2 , pp. 124-131
    • Fantozzi, S.1    Cappello, A.2    Leardini, A.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.