Volumn 30, Issue 9, 2011, Pages 1678-1690

Sensitivity of photon-counting based K-Edge Imaging in X-ray computed tomography

Author keywords

Computed tomography (CT); photon counting; spectral K edge imaging

Indexed keywords

ATOMIC NUMBERS; ATTENUATION COEFFICIENT; CDTE PIXEL DETECTORS; COMPUTED TOMOGRAPHY; CONTRAST MATERIAL; CT SIMULATION; EDGE ENERGY; ENERGY RESOLUTIONS; ENERGY THRESHOLDS; HEAVY ELEMENTS; K-EDGE IMAGING; OBJECT SIZE; PHOTON COUNTING; PHOTON RATES; PHYSICAL EFFECTS; PULSE PILE-UP; SCATTERED X-RAYS; SIGNAL LOSS; SIGNAL TO NOISE; SNR VALUES; SPECTRAL RESPONSE; SPECTRAL RESPONSE FUNCTIONS; TRANSMISSION MEASUREMENTS; X RAY SPECTRUM; X-RAY COMPUTED TOMOGRAPHY; X-RAY DETECTOR; X-RAY TUBE VOLTAGES;

EID: 80052279399     PISSN: 02780062     EISSN: None     Source Type: Journal    
DOI: 10.1109/TMI.2011.2142188     Document Type: Article
Times cited : (108)

References (27)
  • 1
    • 34547137886 scopus 로고    scopus 로고
    • K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors
    • DOI 10.1088/0031-9155/52/15/020, PII S0031915507383851, 020
    • E. Roessl and R. Proksa, "-edge imaging in X-ray computed tomography using multi-bin photon counting detectors," Phys. Med. Biol., vol. 52, no. 15, pp. 4679-4696, Aug. 2007. (Pubitemid 47103149)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.15 , pp. 4679-4696
    • Roessl, E.1    Proksa, R.2
  • 3
    • 0017108614 scopus 로고
    • Energy-selective reconstructions in X-ray computerized tomography
    • R. E. Alvarez and A. Macovski, "Energy-selective reconstructions in X-ray computerized tomography," Phys. Med. Biol., vol. 21, no. 5, pp. 733-744, 1976.
    • (1976) Phys. Med. Biol. , vol.21 , Issue.5 , pp. 733-744
    • Alvarez, R.E.1    MacOvski, A.2
  • 4
    • 0017693719 scopus 로고
    • Selective iodine imaging using K-edge energies in computerized X-ray tomography
    • DOI 10.1118/1.594357
    • S. J. Riederer and C. A. Mistretta, "Selective iodine imaging using edge energies in computerized X-ray tomography," Med. Phys., vol. 4, no. 6, pp. 474-481, 1977. (Pubitemid 8242279)
    • (1977) Medical Physics , vol.4 , Issue.6 , pp. 474-481
    • Riederer, S.J.1    Mistretta, C.A.2
  • 6
    • 63649121535 scopus 로고    scopus 로고
    • Cramér-Rao lower bound of basis image noise in multiple-energy X-ray imaging
    • Mar
    • E. Roessl and C. Herrmann, "Cramér-Rao lower bound of basis image noise in multiple-energy X-ray imaging," Phys. Med. Biol., vol. 54, no. 5, pp. 1307-1318, Mar. 2009.
    • (2009) Phys. Med. Biol. , vol.54 , Issue.5 , pp. 1307-1318
    • Roessl, E.1    Herrmann, C.2
  • 7
    • 78650911265 scopus 로고    scopus 로고
    • Sufficient statistics as a generalization of binning in spectral X-ray imaging
    • Jan
    • A. S. Wang and N. J. Pelc, "Sufficient statistics as a generalization of binning in spectral X-ray imaging," IEEE Trans. Med. Imag., vol. 30, no. 1, pp. 84-93, Jan. 2011.
    • (2010) IEEE Trans. Med. Imag. , vol.30 , Issue.1 , pp. 84-93
    • Wang, A.S.1    Pelc, N.J.2
  • 8
    • 0003205588 scopus 로고
    • Fundamentals of statistical signal processing
    • Upper Saddle River, NJ: Prentice Hall Prentice Hall Signal Processing Series
    • S. M. Kay, "Fundamentals of statistical signal processing," in Estimation Theory. Upper Saddle River, NJ: Prentice Hall, 1993, vol. 1, Prentice Hall Signal Processing Series.
    • (1993) Estimation Theory , vol.1
    • Kay, S.M.1
  • 10
    • 38649124058 scopus 로고    scopus 로고
    • Optimal energy threshold arrangement in photon-counting spectral x-ray imaging
    • DOI 10.1109/NSSMIC.2006.354276, 4179389, 2006 IEEE Nuclear Science Symposium - Conference Record
    • E. Roessl and R. Proksa, "Optimal energy threshold arrangement in photon-counting spectral X-ray imaging," in Proc. IEEE Nucl. Sci. Symp. Conf. Rec., 2006, vol. 3, pp. 1950-1954. (Pubitemid 351164522)
    • (2007) IEEE Nuclear Science Symposium Conference Record , vol.3 , pp. 1950-1954
    • Roessl, E.1    Proksa, R.2
  • 14
    • 0141675877 scopus 로고
    • Photon, electron, proton and neutron interaction data for body tissues
    • Photon, electron, proton and neutron interaction data for body tissues Int. Commission Radiation Units and Measurements (ICRU), 1992, 46.
    • (1992) Int. Commission Radiation Units and Measurements (ICRU) , pp. 46
  • 15
    • 0025756532 scopus 로고
    • Semiempirical model for generating tungsten target X-ray spectra
    • D. M. Tucker, G. T. Barnes, and D. P. Chakraborty, "Semiempirical model for generating tungsten target X-ray spectra," Med. Phys., vol. 18, no. 2, pp. 211-218, 1991.
    • (1991) Med. Phys. , vol.18 , Issue.2 , pp. 211-218
    • Tucker, D.M.1    Barnes, G.T.2    Chakraborty, D.P.3
  • 16
    • 45849108387 scopus 로고    scopus 로고
    • Testing an energy-dispersive counting-mode detector with hard X-rays from a synchrotron source
    • C. Bäumer, G. Martens, B. Menser, E. Roessl, J. P. Schlomka, R. Steadman, and G. Zeitler, "Testing an energy-dispersive counting-mode detector with hard X-rays from a synchrotron source," IEEE Trans. Nucl. Sci., vol. 55, no. 3, pp. 1785-1790, 2008.
    • (2008) IEEE Trans. Nucl. Sci. , vol.55 , Issue.3 , pp. 1785-1790
    • Bäumer, C.1    Martens, G.2    Menser, B.3    Roessl, E.4    Schlomka, J.P.5    Steadman, R.6    Zeitler, G.7
  • 17
    • 80052278258 scopus 로고    scopus 로고
    • [Online]. Available: http://www.imp.uni-erlangen.de/forbild
  • 19
    • 0028140493 scopus 로고
    • X-ray tube potential, filtration, and detector considerations in dual- energy chest radiography
    • DOI 10.1118/1.597373
    • D. M. Gauntt and G. T. Barnes, "X-ray tube potential, filtration, and detector considerations in dual-energy chest radiography," Med. Phys., vol. 21, no. 2, pp. 203-218, Feb. 1994. (Pubitemid 24068785)
    • (1994) Medical Physics , vol.21 , Issue.2 , pp. 203-218
    • Gauntt, D.M.1    Barnes, G.T.2
  • 20
    • 0024070365 scopus 로고
    • An algorithm for noise suppression in dual energy CT material density images
    • Sep
    • W. A. Kalender, E. Klotz, and L. Kostaridou, "An algorithm for noise suppression in dual energy CT material density images," IEEE Trans. Med. Imag., vol. 7, no. 3, pp. 218-224, Sep. 1988.
    • (1988) IEEE Trans. Med. Imag. , vol.7 , Issue.3 , pp. 218-224
    • Kalender, W.A.1    Klotz, E.2    Kostaridou, L.3
  • 21
    • 67649610804 scopus 로고    scopus 로고
    • Optimal image-based weighting for energy-resolved CT
    • Jul
    • T. G. Schmidt, "Optimal image-based weighting for energy-resolved CT," Med. Phys., vol. 36, no. 7, pp. 3018-3027, Jul. 2009.
    • (2009) Med. Phys. , vol.36 , Issue.7 , pp. 3018-3027
    • Schmidt, T.G.1
  • 22
    • 35148835355 scopus 로고    scopus 로고
    • Investigation of the use of photon counting X-ray detectors with energy discrimination capability for material decomposition in micro-computed tomography
    • 1
    • E. C. Frey, X.Wang, Y. Du, K. Taguchi, J. Xu, and B. M. W. Tsui, "Investigation of the use of photon counting X-ray detectors with energy discrimination capability for material decomposition in micro-computed tomography," in Proc. SPIE Med. Imag. Conf., 2007, vol. 6510, p. 65100A, 1.
    • (2007) Proc. SPIE Med. Imag. Conf , vol.6510
    • Frey, E.C.1    Wang, X.2    Du, Y.3    Taguchi, K.4    Xu, J.5    Tsui, B.M.W.6
  • 23
    • 78149475974 scopus 로고    scopus 로고
    • An analytical model of the effects of pulse pile-up on the energy spectrum recorded by energy resolved photon counting X-ray detectors
    • K. Taguchi, E. C. Frey, X.Wang, J. S. Iwanczyk, and W. C. Barber, "An analytical model of the effects of pulse pile-up on the energy spectrum recorded by energy resolved photon counting X-ray detectors," Med. Phys., vol. 37, no. 8, pp. 3957-3969, 2010.
    • (2010) Med. Phys. , vol.37 , Issue.8 , pp. 3957-3969
    • Taguchi, K.1    Frey, E.C.2    Wang, X.3    Iwanczyk, J.S.4    Barber, W.C.5
  • 24
    • 79551662219 scopus 로고    scopus 로고
    • Modeling the performance of a photon counting X-ray detector for CT: Energy response and pulse pile-up effects
    • K. Taguchi, M. Zhang, E. C. Frey, X.Wang, J. S. Iwanczyk, E. Nygard, N. E. Hartsough, B. M. W. Tsui, and W. C. Barber, "Modeling the performance of a photon counting X-ray detector for CT: Energy response and pulse pile-up effects," Med. Phys., vol. 38, no. 2, pp. 1089-1102, 2011.
    • (2011) Med. Phys. , vol.38 , Issue.2 , pp. 1089-1102
    • Taguchi, K.1    Zhang, M.2    Frey, E.C.3    Wang, X.4    Iwanczyk, J.S.5    Nygard, E.6    Hartsough, N.E.7    Tsui, B.M.W.8    Barber, W.C.9
  • 25
    • 67649231480 scopus 로고    scopus 로고
    • Edge-on semiconductor X-ray detectors-towards high-rate counting computed tomography
    • E. Roessl, J. P. Schlomka, and R. Proksa, "Edge-on semiconductor X-ray detectors-towards high-rate counting computed tomography," in Proc. IEEE Nucl. Sci. Symp. Conf. Rec., 2008, pp. 1748-1751.
    • (2008) Proc. IEEE Nucl. Sci. Symp. Conf. Rec , pp. 1748-1751
    • Roessl, E.1    Schlomka, J.P.2    Proksa, R.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.