Volumn 35, Issue 8, 2011, Pages 765-772

In vitro evaluation of a compliant inflow cannula reservoir to reduce suction events with extracorporeal rotary ventricular assist device support

Author keywords

Compliant; Extracorporeal circulation; Heart failure; Suction event; Venous reservoir; Ventricular assist device

Indexed keywords

ARTIFICIAL ORGANS; CONTROL SYSTEMS; FIBER OPTIC SENSORS;

EID: 80051614000     PISSN: 0160564X     EISSN: 15251594     Source Type: Journal    
DOI: 10.1111/j.1525-1594.2011.01311.x     Document Type: Article
Times cited : (10)

References (31)
  • 1
    • 77949491740 scopus 로고    scopus 로고
    • Clinical management of continuous-flow left ventricular assist devices in advanced heart failure
    • Slaughter MS, Pagani FD, Rogers JG, etal. Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transplant 2010;29:S1-39.
    • (2010) J Heart Lung Transplant , vol.29
    • Slaughter, M.S.1    Pagani, F.D.2    Rogers, J.G.3
  • 2
    • 0014985060 scopus 로고
    • Left ventricular bypass pump for cardiac assistance
    • DeBakey ME. Left ventricular bypass pump for cardiac assistance. Am J Cardiol 1971;27:3-11.
    • (1971) Am J Cardiol , vol.27 , pp. 3-11
    • DeBakey, M.E.1
  • 3
    • 33750500333 scopus 로고    scopus 로고
    • Left ventricular assist device and drug therapy for the reversal of heart failure
    • Birks EJ, Tansley PD, Hardy J, etal. Left ventricular assist device and drug therapy for the reversal of heart failure. N Engl J Med 2006;355:1873-84.
    • (2006) N Engl J Med , vol.355 , pp. 1873-1884
    • Birks, E.J.1    Tansley, P.D.2    Hardy, J.3
  • 4
    • 56449096749 scopus 로고    scopus 로고
    • Current axial-flow devices-the HeartMate II and Jarvik 2000 left ventricular assist devices
    • John R. Current axial-flow devices-the HeartMate II and Jarvik 2000 left ventricular assist devices. Semin Thorac Cardiovasc Surg 2008;20:264-72.
    • (2008) Semin Thorac Cardiovasc Surg , vol.20 , pp. 264-272
    • John, R.1
  • 5
    • 56449088083 scopus 로고    scopus 로고
    • Continuous-flow rotary left ventricular assist devices with "3rd generation" design
    • Pagani FD. Continuous-flow rotary left ventricular assist devices with "3rd generation" design. Semin Thorac Cardiovasc Surg 2008;20:255-63.
    • (2008) Semin Thorac Cardiovasc Surg , vol.20 , pp. 255-263
    • Pagani, F.D.1
  • 6
    • 42649142928 scopus 로고    scopus 로고
    • Bridge to decision using the Levitronix CentriMag short-term ventricular assist device
    • De Robertis F, Rogers P, Amrani M, etal. Bridge to decision using the Levitronix CentriMag short-term ventricular assist device. J Heart Lung Transplant 2008;27:474-8.
    • (2008) J Heart Lung Transplant , vol.27 , pp. 474-478
    • De Robertis, F.1    Rogers, P.2    Amrani, M.3
  • 8
    • 0027164750 scopus 로고
    • Postcardiotomy shock: clinical evaluation of the BVS 5000 biventricular support system
    • Guyton RA, Schonberger JPAM, Everts PAM, etal. Postcardiotomy shock: clinical evaluation of the BVS 5000 biventricular support system. Ann Thorac Surg 1993;56:346-56.
    • (1993) Ann Thorac Surg , vol.56 , pp. 346-356
    • Guyton, R.A.1    Schonberger, J.P.A.M.2    Everts, P.A.M.3
  • 9
    • 73349129701 scopus 로고    scopus 로고
    • Advanced heart failure treated with continuous-flow left ventricular assist device
    • Slaughter MS, Rogers JG, Milano CA, etal. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med 2009;361:2241-51.
    • (2009) N Engl J Med , vol.361 , pp. 2241-2251
    • Slaughter, M.S.1    Rogers, J.G.2    Milano, C.A.3
  • 10
    • 20544444028 scopus 로고    scopus 로고
    • Initial experience with the abiomed AB5000 ventricular assist device system
    • Samuels LE, Holmes EC, Garwood P, Ferdinand F. Initial experience with the abiomed AB5000 ventricular assist device system. Ann Thorac Surg 2005;80:309-12.
    • (2005) Ann Thorac Surg , vol.80 , pp. 309-312
    • Samuels, L.E.1    Holmes, E.C.2    Garwood, P.3    Ferdinand, F.4
  • 11
    • 55949133476 scopus 로고    scopus 로고
    • Early experience with the Levitronix Centrimag device for extra-corporeal membrane oxygenation following lung transplantation
    • Khan NU, Al-aloul M, Shah R, Yonan N. Early experience with the Levitronix Centrimag device for extra-corporeal membrane oxygenation following lung transplantation. Eur J Cardiothorac Surg 2008;34:1262-4.
    • (2008) Eur J Cardiothorac Surg , vol.34 , pp. 1262-1264
    • Khan, N.U.1    Al-aloul, M.2    Shah, R.3    Yonan, N.4
  • 12
    • 80051654108 scopus 로고    scopus 로고
    • Use of echocardiography to optimize left ventricular assist devices
    • Kirkpatrick JN, etal. Use of echocardiography to optimize left ventricular assist devices. US Cardiology 2010;7:11-5.
    • (2010) US Cardiology , vol.7 , pp. 11-15
    • Kirkpatrick, J.N.1
  • 13
    • 77957261537 scopus 로고    scopus 로고
    • Atrial versus ventricular cannulation for a rotary ventricular assist device
    • Timms D, Gregory S, Hsu PL, etal. Atrial versus ventricular cannulation for a rotary ventricular assist device. Artif Organs 2010;34:714-20.
    • (2010) Artif Organs , vol.34 , pp. 714-720
    • Timms, D.1    Gregory, S.2    Hsu, P.L.3
  • 14
    • 34648857562 scopus 로고    scopus 로고
    • Numerical comparison of hemodynamics with atrium to aorta and ventricular apex to aorta VAD support
    • Korakianitis T, Shi Y. Numerical comparison of hemodynamics with atrium to aorta and ventricular apex to aorta VAD support. ASAIO J 2007;53:537-48.
    • (2007) ASAIO J , vol.53 , pp. 537-548
    • Korakianitis, T.1    Shi, Y.2
  • 15
    • 34047225873 scopus 로고    scopus 로고
    • Mathematical modeling of ventricular suction induced by a rotary ventricular assist device. In Proceedings of the 2006 American Control Conference. Minneapolis, Minnesota, USA
    • Yu YC, Porter J. Mathematical modeling of ventricular suction induced by a rotary ventricular assist device. In Proceedings of the 2006 American Control Conference. Minneapolis, Minnesota, USA, 2006.
    • (2006)
    • Yu, Y.C.1    Porter, J.2
  • 16
    • 34250771040 scopus 로고    scopus 로고
    • Suction due to left ventricular assist: implications for device control and management
    • Reesink K, Dekker A, Van der Nagel T, etal. Suction due to left ventricular assist: implications for device control and management. Artif Organs 2007;31:542-9.
    • (2007) Artif Organs , vol.31 , pp. 542-549
    • Reesink, K.1    Dekker, A.2    Van der Nagel, T.3
  • 18
    • 79952798514 scopus 로고    scopus 로고
    • Right ventricular failure-a continuing problem in patients with left ventricular assist device support
    • John R, Lee S, Eckman P. Right ventricular failure-a continuing problem in patients with left ventricular assist device support. J Cardiovasc Trans Res 2010;3:604-11.
    • (2010) J Cardiovasc Trans Res , vol.3 , pp. 604-611
    • John, R.1    Lee, S.2    Eckman, P.3
  • 19
    • 77957267367 scopus 로고    scopus 로고
    • Left ventricle afterload impedance control by an axial flow ventricular assist device: a potential tool for ventricular recovery
    • Moscato F, Arabia M, Colacino FM, Naiyanetr P, Danieli GA, Schima H. Left ventricle afterload impedance control by an axial flow ventricular assist device: a potential tool for ventricular recovery. Artif Organs 2010;34:736-44.
    • (2010) Artif Organs , vol.34 , pp. 736-744
    • Moscato, F.1    Arabia, M.2    Colacino, F.M.3    Naiyanetr, P.4    Danieli, G.A.5    Schima, H.6
  • 20
    • 38349148970 scopus 로고    scopus 로고
    • Acute in vivo evaluation of an implantable continuous flow biventricular assist system
    • Saeed D, Ootaki Y, Ootaki C, etal. Acute in vivo evaluation of an implantable continuous flow biventricular assist system. ASAIO J 2008;54:20-4.
    • (2008) ASAIO J , vol.54 , pp. 20-24
    • Saeed, D.1    Ootaki, Y.2    Ootaki, C.3
  • 21
    • 77952009833 scopus 로고    scopus 로고
    • Development of an inlet pressure sensor for control in a left ventricular assist device
    • Fritz B, Cysyk J, Newswanger R, Weiss W, Rosenborg G. Development of an inlet pressure sensor for control in a left ventricular assist device. ASAIO J 2010;56:180-5.
    • (2010) ASAIO J , vol.56 , pp. 180-185
    • Fritz, B.1    Cysyk, J.2    Newswanger, R.3    Weiss, W.4    Rosenborg, G.5
  • 22
    • 58149231147 scopus 로고    scopus 로고
    • Intraoperative evaluation of the HeartMate II flow estimator
    • Slaughter MS, etal. Intraoperative evaluation of the HeartMate II flow estimator. J Heart Lung Transplant 2009;28:39-43.
    • (2009) J Heart Lung Transplant , vol.28 , pp. 39-43
    • Slaughter, M.S.1
  • 23
    • 37749039827 scopus 로고    scopus 로고
    • Physiological control of a rotary left ventricular assist device. In 26th Chinese Control Conference. Zhangjiajie, Hunan, China
    • Yi W. Physiological control of a rotary left ventricular assist device. In 26th Chinese Control Conference. Zhangjiajie, Hunan, China, 2007.
    • (2007)
    • Yi, W.1
  • 25
    • 0028604435 scopus 로고
    • Hydrodynamic response of a VAD with a compliant cannula: results of in vitro tests
    • Antoranz JC, Del Canizo JF, Desco MM. Hydrodynamic response of a VAD with a compliant cannula: results of in vitro tests. Int J Artif Organs 1994;17:635-42.
    • (1994) Int J Artif Organs , vol.17 , pp. 635-642
    • Antoranz, J.C.1    Del Canizo, J.F.2    Desco, M.M.3
  • 28
    • 77953704021 scopus 로고    scopus 로고
    • Long-term biventricular support with the heartware implantable continuous flow pump
    • Stepanenko A, Hennig E, Potapov EV. Long-term biventricular support with the heartware implantable continuous flow pump. J Heart Lung Transplant 2010;29:822-4.
    • (2010) J Heart Lung Transplant , vol.29 , pp. 822-824
    • Stepanenko, A.1    Hennig, E.2    Potapov, E.V.3
  • 29
    • 67650093608 scopus 로고    scopus 로고
    • Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device
    • Pagani FD, Miller LW, Russel SD, etal. Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device. J Am Coll Cardiol 2009;54:312-21.
    • (2009) J Am Coll Cardiol , vol.54 , pp. 312-321
    • Pagani, F.D.1    Miller, L.W.2    Russel, S.D.3
  • 31
    • 33745202386 scopus 로고    scopus 로고
    • Exhaustion of the Frank-Starling mechanism in conscious dogs with heart failure induced by chronic coronary microembolization
    • Gill RM, Jones BD, Corbly AK, etal. Exhaustion of the Frank-Starling mechanism in conscious dogs with heart failure induced by chronic coronary microembolization. Life Sci 2006;79:536-44.
    • (2006) Life Sci , vol.79 , pp. 536-544
    • Gill, R.M.1    Jones, B.D.2    Corbly, A.K.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.