Flow-regulated extracorporeal arteriovenous tubing loop for cardiac output measurements by ultrasound velocity dilution: Validation in post-cardiac surgery intensive care unit patients

Author keywords

[No Author keywords available]

Indexed keywords

ACCEPTABLE LIMIT; BODY TEMPERATURE; CARDIAC OUTPUT; CARDIAC SURGERY; MEASUREMENT SETS; PERCENTAGE ERROR; PULMONARY ARTERY; RADIAL ARTERY; THERMODILUTION; ULTRASOUND VELOCITY;

EID: 78349252680     PISSN: 10582916     EISSN: None     Source Type: Journal    
DOI: 10.1097/MAT.0b013e3181effdf8     Document Type: Article
Times cited : (14)

References (31)
  • 1
    • 0032169990 scopus 로고    scopus 로고
    • Use of an extracorporeal arteriovenous tubing loop to measure cardiac output in intensive care unit patients by ultrasound velocity dilution
    • Eremenko A, Balykov I, Chaus N, et al: Use of an extracorporeal arteriovenous tubing loop to measure cardiac output in intensive care unit patients by ultrasound velocity dilution. ASAIO J 44: M462-M464, 1998.
    • (1998) ASAIO J , vol.44
    • Eremenko, A.1    Balykov, I.2    Chaus, N.3
  • 2
    • 73949146232 scopus 로고    scopus 로고
    • Cardiac output measurement by arterial pressure waveform analysis during optimization of biventricular pacing after cardiac surgery
    • Booth JH, Quinn T, Richmond ME, et al: Cardiac output measurement by arterial pressure waveform analysis during optimization of biventricular pacing after cardiac surgery. ASAIO J 55: 587-591, 2009.
    • (2009) ASAIO J , vol.55 , pp. 587-591
    • Booth, J.H.1    Quinn, T.2    Richmond, M.E.3
  • 3
    • 0014964972 scopus 로고
    • Catheterization of the heart in man with use of a flow-directed ballon-tipped catheter
    • Swan HJ, Ganz W, Forrester J, et al: Catheterization of the heart in man with use of a flow-directed ballon-tipped catheter. N Engl J Med 283: 447-451, 1970.
    • (1970) N Engl J Med , vol.283 , pp. 447-451
    • Swan, H.J.1    Ganz, W.2    Forrester, J.3
  • 4
    • 14844335085 scopus 로고    scopus 로고
    • Pulmonary artery rupture induced by pulmonary artery catheter: A case report and review of the literature
    • Abreu AR, Campos MA, Krieger BP: Pulmonary artery rupture induced by pulmonary artery catheter: A case report and review of the literature. J Intensive Care Med 19: 291-296, 2004.
    • (2004) J Intensive Care Med , vol.19 , pp. 291-296
    • Abreu, A.R.1    Campos, M.A.2    Krieger, B.P.3
  • 5
    • 0037413466 scopus 로고    scopus 로고
    • A randomized, controlled trial of the use of pulmonaryartery catheters in high-risk surgical patients
    • Sandham JD, Hull RD, Brant RF, et al: A randomized, controlled trial of the use of pulmonaryartery catheters in high-risk surgical patients. N Engl J Med 348: 5-14, 2003.
    • (2003) N Engl J Med , vol.348 , pp. 5-14
    • Sandham, J.D.1    Hull, R.D.2    Brant, R.F.3
  • 6
    • 0027219058 scopus 로고
    • A new method of measuring cardiac output in man using lithium dilution
    • Linton RAF, Band DM, Haire KM: A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 71: 262-266, 1993.
    • (1993) Br J Anaesth , vol.71 , pp. 262-266
    • Raf, L.1    Band, D.M.2    Haire, K.M.3
  • 7
    • 2942549668 scopus 로고    scopus 로고
    • Measurement of cardiac output by transpulmonary arterial thermodilution using a long radial artery catheter A comparison with intermittent pulmonary artery thermodilution
    • Orme RM, Pigott DW, Mihm FG: Measurement of cardiac output by transpulmonary arterial thermodilution using a long radial artery catheter. A comparison with intermittent pulmonary artery thermodilution. Anaesthesia 59: 590-594, 2004.
    • (2004) Anaesthesia , vol.59 , pp. 590-594
    • Orme, R.M.1    Pigott, D.W.2    Mihm, F.G.3
  • 8
    • 61749084801 scopus 로고    scopus 로고
    • Theory and in-vitro validation of a new extracorporeal approach for hemodynamic assessment in pediatric and neonatal ICU patients
    • Krivitski NM, Kislukhin VV, Thuramalla NV: Theory and in-vitro validation of a new extracorporeal approach for hemodynamic assessment in pediatric and neonatal ICU patients. Pediatr Crit Care Med J 9: 423-428, 2008.
    • (2008) Pediatr Crit Care Med J , vol.9 , pp. 423-428
    • Krivitski, N.M.1    Kislukhin, V.V.2    Thuramalla, N.V.3
  • 9
    • 0028808026 scopus 로고
    • Novel method to measure access flow during hemodialysis by ultrasound dilution technique
    • Krivitski NM: Novel method to measure access flow during hemodialysis by ultrasound dilution technique. ASAIO J 41: M741-M745, 1995.
    • (1995) ASAIO J , vol.41
    • Krivitski, N.M.1
  • 10
    • 0032601383 scopus 로고    scopus 로고
    • Extracorporeal recording of mouse hemodynamic parameters by ultrasound velocity dilution
    • Krivitski NM, Starostin D, Smith TL: Extracorporeal recording of mouse hemodynamic parameters by ultrasound velocity dilution. ASAIO J 45: 32-36, 1999.
    • (1999) ASAIO J , vol.45 , pp. 32-36
    • Krivitski, N.M.1    Starostin, D.2    Smith, T.L.3
  • 11
    • 0024264701 scopus 로고
    • Blood flow and blood volume determinations in aorta and in coronary circulation by density dilution
    • Moser M, Kenner T: Blood flow and blood volume determinations in aorta and in coronary circulation by density dilution. Basic Res Cardiol 83: 577-589, 1988.
    • (1988) Basic Res Cardiol , vol.83 , pp. 577-589
    • Moser, M.1    Kenner, T.2
  • 12
    • 0344466424 scopus 로고
    • Distribution spaces of intravenously injected saline, plasma and red cell concentrate
    • Moser M, Kenner T: Distribution spaces of intravenously injected saline, plasma and red cell concentrate. Pflugers Arch 394S: R48, 1982.
    • (1982) Pflugers Arch 394S , vol.R48
    • Moser, M.1    Kenner, T.2
  • 14
    • 0033003166 scopus 로고    scopus 로고
    • A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    • Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15: 85-91, 1999.
    • (1999) J Clin Monit Comput , vol.15 , pp. 85-91
    • Critchley, L.A.1    Critchley, J.A.2
  • 15
    • 0025215247 scopus 로고
    • Power and sample size calculations: A review and computer program
    • Dupont WD, Plummer WD: Power and sample size calculations: A review and computer program. Control Clin Trials 11: 116-128, 1990.
    • (1990) Control Clin Trials , vol.11 , pp. 116-128
    • Dupont, W.D.1    Plummer, W.D.2
  • 17
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1: 307-310, 1986.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 18
    • 0030250734 scopus 로고    scopus 로고
    • Validation of a new method to measure cardiac output during extracorporeal detoxification
    • Nikiforov YV, Kisluchine VV, Chaus NI: Validation of a new method to measure cardiac output during extracorporeal detoxification. ASAIO J 42: M903-M905, 1996.
    • (1996) ASAIO J , vol.42
    • Nikiforov, Y.V.1    Kisluchine, V.V.2    Chaus, N.I.3
  • 19
    • 0030250242 scopus 로고    scopus 로고
    • Validation of a novel ultrasound dilution method to measure cardiac output during hemodialysis
    • Kisloukhine VV, Dean DA: Validation of a novel ultrasound dilution method to measure cardiac output during hemodialysis. ASAIO J 42: M906-M907, 1996.
    • (1996) ASAIO J , vol.42
    • Kisloukhine, V.V.1    Dean, D.A.2
  • 20
    • 34250893209 scopus 로고    scopus 로고
    • Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs
    • Cooper ES, Muir WW: Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs. Crit Care Med 35: 1724-1729, 2007.
    • (2007) Crit Care Med , vol.35 , pp. 1724-1729
    • Cooper, E.S.1    Muir, W.W.2
  • 21
    • 34247551188 scopus 로고    scopus 로고
    • Comparison of uncali-brated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements
    • Sander M, Spies CD, Grubitzsch H, et al: Comparison of uncali-brated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care 10: R164, 2006.
    • (2006) Crit Care , vol.10
    • Sander, M.1    Spies, C.D.2    Grubitzsch, H.3
  • 22
    • 33646913835 scopus 로고    scopus 로고
    • Freeman SA: Lithium: Clinical considerations in internal medicine
    • Freeman MP, Freeman SA: Lithium: Clinical considerations in internal medicine. Am J Med 119: 478-481, 2006.
    • (2006) Am J Med , vol.119 , pp. 478-481
    • Freeman, M.P.1
  • 24
    • 0036242474 scopus 로고    scopus 로고
    • Pulse contour analysis versus thermodilution in cardiac surgery patients
    • Rauch H, Muller M, Fleischer F, et al: Pulse contour analysis versus thermodilution in cardiac surgery patients. Acta Anaesthesiol Scand 46: 424-429, 2002.
    • (2002) Acta Anaesthesiol Scand , vol.46 , pp. 424-429
    • Rauch, H.1    Muller, M.2    Fleischer, F.3
  • 25
    • 0142212422 scopus 로고    scopus 로고
    • Pulse contour analysis for cardiac output monitoring in cardiac surgery for congenital heart disease
    • Mahajan A, Shabanie A, Turner J, et al: Pulse contour analysis for cardiac output monitoring in cardiac surgery for congenital heart disease. Anesth Analg 97: 1283-1288, 2003.
    • (2003) Anesth Analg , vol.97 , pp. 1283-1288
    • Mahajan, A.1    Shabanie, A.2    Turner, J.3
  • 26
    • 33748925557 scopus 로고    scopus 로고
    • Pulse-contour-derived cardiac output is unreliable in septic shock
    • Berthelse PG: Pulse-contour-derived cardiac output is unreliable in septic shock. Acta Anaesthesiol Scand 50:1169-1170, 2006.
    • (2006) Acta Anaesthesiol Scand , vol.50 , pp. 1169-1170
    • Berthelse, P.G.1
  • 27
    • 12144256186 scopus 로고    scopus 로고
    • Comparison of cardiac output determined by bioimpedance, thermodilution, and the Fick method
    • Engoren M, Barbee D: Comparison of cardiac output determined by bioimpedance, thermodilution, and the Fick method. Am J Crit Care 14: 40-45, 2005.
    • (2005) Am J Crit Care , vol.14 , pp. 40-45
    • Engoren, M.1    Barbee, D.2
  • 28
    • 75149142459 scopus 로고    scopus 로고
    • Cardiac outputmeasurement using an ultrasound dilution method: A validation study in ventilated piglets
    • De Boode WP, van Heijst AF, Hopman JC, et al: Cardiac outputmeasurement using an ultrasound dilution method: A validation study in ventilated piglets. Pediatr Crit Care Med 11: 103-108, 2010.
    • (2010) Pediatr Crit Care Med , vol.11 , pp. 103-108
    • De Boode, W.P.1    Van Heijst, A.F.2    Hopman, J.C.3
  • 29
    • 70349294953 scopus 로고    scopus 로고
    • Validation of novel ultra-sound dilution cardiac output method for pediatric and neonatal patients
    • Gleed R, Smith T, Callahan M, et al: Validation of novel ultra-sound dilution cardiac output method for pediatric and neonatal patients. Intensive Care Med 32: S172, 2006.
    • (2006) Intensive Care Med , vol.32
    • Gleed, R.1    Smith, T.2    Callahan, M.3
  • 30
    • 70349265727 scopus 로고    scopus 로고
    • A novel method tomeasure cardiac output in the pediatric ICU: Animal validation and preliminary clinical study
    • Schulenberg A, Harmon W, Rubenstein J, et al: A novel method tomeasure cardiac output in the pediatric ICU: Animal validation and preliminary clinical study. Crit Care Med 34: A12, 2007.
    • (2007) Crit Care Med , vol.34
    • Schulenberg, A.1    Harmon, W.2    Rubenstein, J.3
  • 31
    • 78349296234 scopus 로고    scopus 로고
    • Validation of ultrasound dilutioncardiac output measurements in pediatric patients
    • Crittendon I, Dreyer W, Kim J: Validation of ultrasound dilutioncardiac output measurements in pediatric patients. Crit Care Med 37: A90, 2009.
    • (2009) Crit Care Med , vol.37
    • Crittendon, I.1    Dreyer, W.2    Kim, J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.