Combining scatter reduction and correction to improve image quality in cone-beam computed tomography (CBCT)

Author keywords

cone beam computed tomography (CBCT); image artifact; image quality; scatter correction

Indexed keywords

DATA REDUCTION; IMAGE ENHANCEMENT; IMAGE QUALITY; IMAGE RECONSTRUCTION; IMAGE REGISTRATION; IMAGE SEGMENTATION; PHANTOMS; QUALITY CONTROL; RIGID STRUCTURES; ROTATION;

EID: 78149462521     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3497272     Document Type: Article
Times cited : (81)

References (44)
  • 1
    • 33845938235 scopus 로고    scopus 로고
    • Assessment of Myocardial Microvascular Function: New Opportunities in Fast Computed Tomography
    • DOI 10.1016/j.tcm.2006.10.002, PII S105017380600171X
    • E. Daghini and L. O. Lerman, " Assessment of myocardial microvascular function: New opportunities in fast computed tomography.," Trends Cardiovasc. Med. TCMDEQ 1050-1738 17 (1), 14-19 (2007). 10.1016/j.tcm.2006.10.002 (Pubitemid 46038767)
    • (2007) Trends in Cardiovascular Medicine , vol.17 , Issue.1 , pp. 14-19
    • Daghini, E.1    Lerman, L.O.2
  • 2
    • 33646259255 scopus 로고    scopus 로고
    • Technology insight: Cardiac CT angiography
    • NCPCC4 1743-4297 (),. 10.1038/ncpcardio0541
    • M. Poon, " Technology insight: Cardiac CT angiography.," Nat. Clin. Pract. Cardiovasc. Med. NCPCC4 1743-4297 3 (5), 265-275 (2006). 10.1038/ncpcardio0541
    • (2006) Nat. Clin. Pract. Cardiovasc. Med. , vol.3 , Issue.5 , pp. 265-275
    • Poon, M.1
  • 3
    • 44449104318 scopus 로고    scopus 로고
    • Cardiac CT and CT coronary angiography: Technology and application
    • DOI 10.1136/hrt.2007.116392
    • W. T. Roberts, J. J. Bax, and L. C. Davies, " Cardiac CT and CT coronary angiography: Technology and application.," Heart ZZZZZZ 1355-6037 94 (6), 781-792 (2008). 10.1136/hrt.2007.116392 (Pubitemid 351752298)
    • (2008) Heart , vol.94 , Issue.6 , pp. 781-792
    • Roberts, W.T.1    Bax, J.J.2    Davies, L.C.3
  • 4
    • 34047117764 scopus 로고    scopus 로고
    • CT fluoroscopy: Novel application for the treatment of ventricular pathologies
    • DOI 10.1007/s00234-007-0208-2
    • K. Murphy, D. A. Nussbaum, and P. Gailloud, " CT fluoroscopy: Novel application for the treatment of ventricular pathologies.," Neuroradiology ZZZZZZ 0028-3940 49 (4), 373-378 (2007). 10.1007/s00234-007-0208-2 (Pubitemid 46510893)
    • (2007) Neuroradiology , vol.49 , Issue.4 , pp. 373-378
    • Murphy, K.1    Nussbaum, D.A.2    Gailloud, P.3
  • 5
    • 33847662456 scopus 로고    scopus 로고
    • Four-dimensional measurement of lung tumor displacement using 256-multi-slice CT-scanner
    • DOI 10.1016/j.lungcan.2006.11.011, PII S0169500206006155
    • S. Mori, M. Endo, S. Komatsu, T. Yashiro, S. Kandatsu, and M. Baba, " Four-dimensional measurement of lung tumor displacement using 256-multi-slice CT-scanner.," Lung Cancer ZZZZZZ 0169-5002 56 (1), 59-67 (2007). 10.1016/j.lungcan.2006.11.011 (Pubitemid 46356637)
    • (2007) Lung Cancer , vol.56 , Issue.1 , pp. 59-67
    • Mori, S.1    Endo, M.2    Komatsu, S.3    Yashiro, T.4    Kandatsu, S.5    Baba, M.6
  • 7
    • 34548508702 scopus 로고    scopus 로고
    • Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging
    • DOI 10.1259/dmfr/19164138
    • S. H. Bartling, O. Majdani, R. Gupta, T. Rodt, C. Dullin, P. F. Fitzgerald, and H. Becker, " Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging.," Dentomaxillofac Radiol. ZZZZZZ 0250-832X 36 (6), 317-327 (2007). 10.1259/dmfr/19164138 (Pubitemid 47372497)
    • (2007) Dentomaxillofacial Radiology , vol.36 , Issue.6 , pp. 317-327
    • Bartling, S.H.1    Majdani, O.2    Gupta, R.3    Rodt, T.4    Dullin, C.5    Fitzgerald, P.F.6    Becker, H.7
  • 8
    • 42749091654 scopus 로고    scopus 로고
    • High resolution imaging of craniofacial bone specimens by flat-panel volumetric computed tomography
    • ZZZZZZ 1010-5182 (),. 10.1016/j.jcms.2008.02.006
    • H. Schaaf, P. Streckbein, M. Obert, B. Goertz, P. Christophis, H. P. Howaldt, and H. Traupe, " High resolution imaging of craniofacial bone specimens by flat-panel volumetric computed tomography.," J. Craniomaxillofac Surg. ZZZZZZ 1010-5182 36 (4), 234-238 (2008). 10.1016/j.jcms.2008.02.006
    • (2008) J. Craniomaxillofac Surg. , vol.36 , Issue.4 , pp. 234-238
    • Schaaf, H.1    Streckbein, P.2    Obert, M.3    Goertz, B.4    Christophis, P.5    Howaldt, H.P.6    Traupe, H.7
  • 10
    • 12844254281 scopus 로고    scopus 로고
    • Volume CT with a flat-panel detector on a mobile, isocentric C-arm: Pre-clinical investigation in guidance of minimally invasive surgery
    • DOI 10.1118/1.1836331
    • J. H. Siewerdsen, D. J. Moseley, S. Burch, S. K. Bisland, A. Bogaards, B. C. Wilson, and D. A. Jaffray, " Volume CT with a flat-panel detector on a mobile, isocentric C-arm: Pre-clinical investigation in guidance of minimally invasive surgery.," Med. Phys. MPHYA6 0094-2405 32 (1), 241-254 (2005). 10.1118/1.1836331 (Pubitemid 40165536)
    • (2005) Medical Physics , vol.32 , Issue.1 , pp. 241-254
    • Siewerdsen, J.H.1    Moseley, D.J.2    Burch, S.3    Bisland, S.K.4    Bogaards, A.5    Wilson, B.C.6    Jaffray, D.A.7
  • 11
    • 42449098553 scopus 로고    scopus 로고
    • The use of portable intraoperative computed tomography scanning for real-time image guidance: A pilot cadaver study
    • DOI 10.2500/ajr.2008.22.3152
    • S. Das, P. A. Maeso, R. E. Figueroa, B. A. Senior, J. M. Delgaudio, M. J. Sillers, R. J. Schlosser, and S. E. Kountakis, " The use of portable intraoperative computed tomography scanning for real-time image guidance: A pilot cadaver study.," Am. J. Rhinol. ZZZZZZ 1050-6586 22 (2), 166-169 (2008). 10.2500/ajr.2008.22.3152 (Pubitemid 351570559)
    • (2008) American Journal of Rhinology , vol.22 , Issue.2 , pp. 166-169
    • Das, S.1    Maeso, P.A.2    Figueroa, R.E.3    Senior, B.A.4    Delgaudio, J.M.5    Sillers, M.J.6    Schlosser, R.J.7    Kountakis, S.E.8
  • 13
    • 34548814455 scopus 로고    scopus 로고
    • Computational Challenges for Image-Guided Radiation Therapy: Framework and Current Research
    • DOI 10.1016/j.semradonc.2007.07.004, PII S1053429607000616, Image-Guided Radiation Therapy
    • L. Xing, J. Siebers, and P. Keall, " Computational challenges for image-guided radiation therapy: Framework and current research.," Semin. Radiat. Oncol. ZZZZZZ 1053-4296 17 (4), 245-257 (2007). 10.1016/j.semradonc. 2007.07.004 (Pubitemid 47446284)
    • (2007) Seminars in Radiation Oncology , vol.17 , Issue.4 , pp. 245-257
    • Xing, L.1    Siebers, J.2    Keall, P.3
  • 14
    • 0031512175 scopus 로고    scopus 로고
    • Computerized scatter correction in diagnostic radiology
    • CTPHAF 0010-7514 (),. 10.1080/001075197182469
    • K. P. Maher and J. F. Malone, " Computerized scatter correction in diagnostic radiology.," Contemp. Phys. CTPHAF 0010-7514 38 (2), 131-148 (1997). 10.1080/001075197182469
    • (1997) Contemp. Phys. , vol.38 , Issue.2 , pp. 131-148
    • Maher, K.P.1    Malone, J.F.2
  • 15
    • 66449132276 scopus 로고    scopus 로고
    • Scatter correction for cone-beam CT in radiation therapy
    • MPHYA6 0094-2405 (),. 10.1118/1.3130047
    • L. Zhu, Y. Xie, J. Wang, and L. Xing, " Scatter correction for cone-beam CT in radiation therapy.," Med. Phys. MPHYA6 0094-2405 36 (6), 2258-2268 (2009). 10.1118/1.3130047
    • (2009) Med. Phys. , vol.36 , Issue.6 , pp. 2258-2268
    • Zhu, L.1    Xie, Y.2    Wang, J.3    Xing, L.4
  • 16
    • 0026689155 scopus 로고
    • Grids or air gaps for scatter reduction in digital radiography: A model calculation
    • MPHYA6 0094-2405 (),. 10.1118/1.596836
    • U. Neitzel, " Grids or air gaps for scatter reduction in digital radiography: A model calculation.," Med. Phys. MPHYA6 0094-2405 19 (2), 475-481 (1992). 10.1118/1.596836
    • (1992) Med. Phys. , vol.19 , Issue.2 , pp. 475-481
    • Neitzel, U.1
  • 17
    • 0033849841 scopus 로고    scopus 로고
    • Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography)
    • DOI 10.1118/1.1286590
    • J. H. Siewerdsen and D. A. Jaffray, " Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography).," Med. Phys. MPHYA6 0094-2405 27 (8), 1903-1914 (2000). 10.1118/1.1286590 (Pubitemid 30666365)
    • (2000) Medical Physics , vol.27 , Issue.8 , pp. 1903-1914
    • Siewerdsen, J.H.1    Jaffray, D.A.2
  • 18
    • 0018893475 scopus 로고
    • Performance characteristics of improved antiscatter grids
    • DOI 10.1118/1.594752
    • J. A. Sorenson, L. T. Niklason, and D. F. Knutti, " Performance characteristics of improved antiscatter grids.," Med. Phys. MPHYA6 0094-2405 7 (5), 525-528 (1980). 10.1118/1.594752 (Pubitemid 10052112)
    • (1980) Medical Physics , vol.7 , Issue.5 , pp. 525-528
    • Sorenson, J.A.1    Niklason, L.T.2    Knutti, D.F.3
  • 19
    • 11244314035 scopus 로고    scopus 로고
    • The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors
    • DOI 10.1118/1.1819789
    • J. H. Siewerdsen, " The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors.," Med. Phys. MPHYA6 0094-2405 31 (12), 3506-3520 (2004). 10.1118/1.1819789 (Pubitemid 40065003)
    • (2004) Medical Physics , vol.31 , Issue.12 , pp. 3506-3520
    • Siewerdsen, J.H.1    Moseley, D.J.2    Bakhtiar, B.3    Richard, S.4    Jaffray, D.A.5
  • 20
    • 33748309392 scopus 로고    scopus 로고
    • Magnitude and effects of x-ray scatter in a 256-slice CT scanner
    • DOI 10.1118/1.2239366
    • M. Endo, S. Mori, T. Tsunoo, and H. Miyazaki, " Magnitude and effects of x-ray scatter in a 256-slice CT scanner.," Med. Phys. MPHYA6 0094-2405 33 (9), 3359-3368 (2006). 10.1118/1.2239366 (Pubitemid 44327439)
    • (2006) Medical Physics , vol.33 , Issue.9 , pp. 3359-3368
    • Endo, M.1    Mori, S.2    Tsunoo, T.3    Miyazaki, H.4
  • 21
    • 34347396603 scopus 로고    scopus 로고
    • Compensators for dose and scatter management in cone-beam computed tomography
    • DOI 10.1118/1.2740466
    • S. A. Graham, D. J. Moseley, J. H. Siewerdsen, and D. A. Jaffray, " Compensators for dose and scatter management in cone-beam computed tomography.," Med. Phys. MPHYA6 0094-2405 34 (7), 2691-2703 (2007). 10.1118/1.2740466 (Pubitemid 47019791)
    • (2007) Medical Physics , vol.34 , Issue.7 , pp. 2691-2703
    • Graham, S.A.1    Moseley, D.J.2    Siewerdsen, J.H.3    Jaffray, D.A.4
  • 22
    • 0034914838 scopus 로고    scopus 로고
    • Scatter correction in digital radiography using interpolated local sampling of aperture signals
    • AUPMDI 0158-9938 (),. 10.1007/BF03178350
    • K. P. Maher, " Scatter correction in digital radiography using interpolated local sampling of aperture signals.," Australas. Phys. Eng. Sci. Med. AUPMDI 0158-9938 24 (2), 79-85 (2001). 10.1007/BF03178350
    • (2001) Australas. Phys. Eng. Sci. Med. , vol.24 , Issue.2 , pp. 79-85
    • Maher, K.P.1
  • 23
    • 0023727721 scopus 로고
    • Dual-energy x-ray projection imaging: Two sampling schemes for the correction of scattered radiation
    • MPHYA6 0094-2405 (),. 10.1118/1.596188
    • F. C. Wagner, A. Macovski, and D. G. Nishimura, " Dual-energy x-ray projection imaging: Two sampling schemes for the correction of scattered radiation.," Med. Phys. MPHYA6 0094-2405 15 (5), 732-748 (1988). 10.1118/1.596188
    • (1988) Med. Phys. , vol.15 , Issue.5 , pp. 732-748
    • Wagner, F.C.1    MacOvski, A.2    Nishimura, D.G.3
  • 24
    • 0022413012 scopus 로고
    • Quantitative digital subtraction angiography: Two scanning techniques for correction of scattered radiation and veiling glare. Work in progress
    • RADLAX 0033-8419 ().
    • C. G. Shaw and D. B. Plewes, " Quantitative digital subtraction angiography: Two scanning techniques for correction of scattered radiation and veiling glare. Work in progress.," Radiology RADLAX 0033-8419 157 (1), 247-253 (1985).
    • (1985) Radiology , vol.157 , Issue.1 , pp. 247-253
    • Shaw, C.G.1    Plewes, D.B.2
  • 25
    • 2442689175 scopus 로고    scopus 로고
    • X-ray scatter correction algorithm for cone beam CT imaging
    • DOI 10.1118/1.1711475
    • R. Ning, X. Tang, and D. Conover, " X-ray scatter correction algorithm for cone beam CT imaging.," Med. Phys. MPHYA6 0094-2405 31 (5), 1195-1202 (2004). 10.1118/1.1711475 (Pubitemid 38656792)
    • (2004) Medical Physics , vol.31 , Issue.5 , pp. 1195-1202
    • Ning, R.1    Tang, X.2    Conover, D.3
  • 26
    • 33845562622 scopus 로고    scopus 로고
    • Scatter correction method for X-ray CT using primary modulation: Theory and preliminary results
    • DOI 10.1109/TMI.2006.884636
    • L. Zhu, N. R. Bennett, and R. Fahrig, " Scatter correction method for x-ray CT using primary modulation: Theory and preliminary results.," IEEE Trans. Med. Imaging ITMID4 0278-0062 25 (12), 1573-1587 (2006). 10.1109/TMI.2006.884636 (Pubitemid 44936707)
    • (2006) IEEE Transactions on Medical Imaging , vol.25 , Issue.12 , pp. 1573-1587
    • Zhu, L.1    Bennett, N.R.2    Fahrig, R.3
  • 27
    • 43149096724 scopus 로고    scopus 로고
    • The built-in capacity of CT D'OR's static ring for scatter correction
    • PSISDG 0277-786X.
    • H. Schlattl and C. Hoeschen, " The built-in capacity of CT D'OR's static ring for scatter correction.," Proc. SPIE PSISDG 0277-786X 6913, 6913-6952 (2008).
    • (2008) Proc. SPIE , vol.6913 , pp. 6913-6952
    • Schlattl, H.1    Hoeschen, C.2
  • 28
    • 31044451240 scopus 로고    scopus 로고
    • A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT
    • DOI 10.1118/1.2148916
    • J. H. Siewerdsen, M. J. Daly, B. Bakhtiar, D. J. Moseley, S. Richard, H. Keller, and D. A. Jaffray, " A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT.," Med. Phys. MPHYA6 0094-2405 33 (1), 187-197 (2006). 10.1118/1.2148916 (Pubitemid 43121048)
    • (2006) Medical Physics , vol.33 , Issue.1 , pp. 187-197
    • Siewerdsen, J.H.1    Daly, M.J.2    Bakhtiar, B.3    Moseley, D.J.4    Richard, S.5    Keller, H.6    Jaffray, D.A.7
  • 29
    • 0023212881 scopus 로고
    • Scatter estimation for a digital radiographic system using convolution filtering
    • DOI 10.1118/1.596126
    • L. A. Love and R. A. Kruger, " Scatter estimation for a digital radiographic system using convolution filtering.," Med. Phys. MPHYA6 0094-2405 14 (2), 178-185 (1987). 10.1118/1.596126 (Pubitemid 17063679)
    • (1987) Medical Physics , vol.14 , Issue.2 , pp. 178-185
    • Love, L.A.1    Kruger, R.A.2
  • 30
    • 33646460131 scopus 로고    scopus 로고
    • Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography
    • MPHYA6 0094-2405 (),. 10.1118/1.2188076
    • M. Kachelriess, K. Sourbelle, and W. A. Kalender, " Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography.," Med. Phys. MPHYA6 0094-2405 33 (5), 1269-1274 (2006). 10.1118/1.2188076
    • (2006) Med. Phys. , vol.33 , Issue.5 , pp. 1269-1274
    • Kachelriess, M.1    Sourbelle, K.2    Kalender, W.A.3
  • 31
    • 34547098832 scopus 로고    scopus 로고
    • A new method for x-ray scatter correction: First assessment on a cone-beam CT experimental setup
    • DOI 10.1088/0031-9155/52/15/018, PII S0031915507451545, 018
    • J. Rinkel, L. Gerfault, F. Est̀ve, and J. M. Dinten, " A new method for x-ray scatter correction: First assessment on a cone-beam CT experimental setup.," Phys. Med. Biol. PHMBA7 0031-9155 52 (15), 4633-4652 (2007). 10.1088/0031-9155/52/15/018 (Pubitemid 47103147)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.15 , pp. 4633-4652
    • Rinkel, J.1    Gerfault, L.2    Esteve, F.3    Dinten, J.-M.4
  • 32
    • 33748518881 scopus 로고    scopus 로고
    • Combining deterministic and Monte Carlo calculations for fast estimation of scatter intensities in CT
    • DOI 10.1088/0031-9155/51/18/008, PII S0031915506243999, 008
    • Y. Kyriakou, T. Riedel, and W. A. Kalender, " Combining deterministic and Monte Carlo calculations for fast estimation of scatter intensities in CT.," Phys. Med. Biol. PHMBA7 0031-9155 51 (18), 4567-4586 (2006). 10.1088/0031-9155/51/18/008 (Pubitemid 44362515)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.18 , pp. 4567-4586
    • Kyriakou, Y.1    Riedel, T.2    Kalender, W.A.3
  • 33
    • 67649601017 scopus 로고    scopus 로고
    • An analytical approach to estimating the first order scatter in heterogeneous medium. II. A practical application
    • MPHYA6 0094-2405 (),. 10.1118/1.3152115
    • W. Yao and K. W. Leszczynski, " An analytical approach to estimating the first order scatter in heterogeneous medium. II. A practical application.," Med. Phys. MPHYA6 0094-2405 36 (7), 3157-3167 (2009). 10.1118/1.3152115
    • (2009) Med. Phys. , vol.36 , Issue.7 , pp. 3157-3167
    • Yao, W.1    Leszczynski, K.W.2
  • 34
    • 61349173033 scopus 로고    scopus 로고
    • Noise suppression in scatter correction for cone-beam CT
    • MPHYA6 0094-2405 (),. 10.1118/1.3063001
    • L. Zhu, J. Wang, and L. Xing, " Noise suppression in scatter correction for cone-beam CT.," Med. Phys. MPHYA6 0094-2405 36 (3), 741-752 (2009). 10.1118/1.3063001
    • (2009) Med. Phys. , vol.36 , Issue.3 , pp. 741-752
    • Zhu, L.1    Wang, J.2    Xing, L.3
  • 35
    • 78149430702 scopus 로고    scopus 로고
    • Feasibility of a pre-object-grid to reduce scatter and improve image quality in cone-beam computed tomography (CBCT)
    • MPHYA6 0094-2405,. 10.1118/1.3181293
    • J. Y. Jin, L. Liu, J. Kim, N. Wen, B. Movsas, and I. J. Chetty, " Feasibility of a pre-object-grid to reduce scatter and improve image quality in cone-beam computed tomography (CBCT).," Med. Phys. MPHYA6 0094-2405 36, 2475 (2009). 10.1118/1.3181293
    • (2009) Med. Phys. , vol.36 , pp. 2475
    • Jin, J.Y.1    Liu, L.2    Kim, J.3    Wen, N.4    Movsas, B.5    Chetty, I.J.6
  • 38
    • 0033796113 scopus 로고    scopus 로고
    • Fast reconstruction with uniform noise properties in half scan computed tomography
    • MPHYA6 0094-2405 (),. 10.1118/1.1288239
    • X. Pan, " Fast reconstruction with uniform noise properties in half scan computed tomography.," Med. Phys. MPHYA6 0094-2405 27 (9), 2031-2036 (2000). 10.1118/1.1288239
    • (2000) Med. Phys. , vol.27 , Issue.9 , pp. 2031-2036
    • Pan, X.1
  • 39
    • 51849124226 scopus 로고    scopus 로고
    • Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization
    • PHMBA7 0031-9155 (),. 10.1088/0031-9155/53/17/021
    • E. Y. Sidky and X. Pan, " Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization.," Phys. Med. Biol. PHMBA7 0031-9155 53 (17), 4777-4807 (2008). 10.1088/0031-9155/53/17/ 021
    • (2008) Phys. Med. Biol. , vol.53 , Issue.17 , pp. 4777-4807
    • Sidky, E.Y.1    Pan, X.2
  • 40
    • 58149141914 scopus 로고    scopus 로고
    • Iterative image reconstruction for CBCT using edge-preserving prior
    • MPHYA6 0094-2405 (),. 10.1118/1.3036112
    • J. Wang, T. Li, and L. Xing, " Iterative image reconstruction for CBCT using edge-preserving prior.," Med. Phys. MPHYA6 0094-2405 36 (1), 252-260 (2009). 10.1118/1.3036112
    • (2009) Med. Phys. , vol.36 , Issue.1 , pp. 252-260
    • Wang, J.1    Li, T.2    Xing, L.3
  • 41
    • 33645712892 scopus 로고    scopus 로고
    • Compressed sensing
    • IETTAW 0018-9448,. 10.1109/TIT.2006.871582
    • D. L. Donoho, " Compressed sensing.," IEEE Trans. Inf. Theory IETTAW 0018-9448 52, 1289-1306 (2006). 10.1109/TIT.2006.871582
    • (2006) IEEE Trans. Inf. Theory , vol.52 , pp. 1289-1306
    • Donoho, D.L.1
  • 42
    • 0035104021 scopus 로고    scopus 로고
    • Total variation norm for three-dimensional iterative reconstruction in limited view angle tomography
    • DOI 10.1088/0031-9155/46/3/318, PII S0031915501164493
    • M. Persson, D. Bone, and H. Elmqvist, " Total variation norm for three dimensional iterative reconstruction in limited view angle tomography.," Phys. Med. Biol. PHMBA7 0031-9155 46, 853-866 (2001). 10.1088/0031-9155/46/3/318 (Pubitemid 32219728)
    • (2001) Physics in Medicine and Biology , vol.46 , Issue.3 , pp. 853-866
    • Persson, M.1    Bone, D.2    Elmqvist, H.3
  • 44
    • 77952078726 scopus 로고    scopus 로고
    • A patient set-up protocol based on partially blocked cone-beam CT
    • TCRTBS 1533-0346 ().
    • L. Zhu, J. Wang, Y. Xie, J. Starman, R. Fahrig, and L. Xing, " A patient set-up protocol based on partially blocked cone-beam CT.," Technol. Cancer Res. Treat. TCRTBS 1533-0346 9 (2), 191-198 (2010).
    • (2010) Technol. Cancer Res. Treat. , vol.9 , Issue.2 , pp. 191-198
    • Zhu, L.1    Wang, J.2    Xie, Y.3    Starman, J.4    Fahrig, R.5    Xing, L.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.