메뉴 건너뛰기




Volumn 113, Issue 4, 2010, Pages 859-869

Cerebral aneurysm clips in the 3-tesla magnetic field: Laboratory investigation

Author keywords

3 Tesla magnetic resonance imaging; Aneurysm clip; Deflection angle; Torque

Indexed keywords

FERROMAGNETIC MATERIAL; TITANIUM; ALLOY;

EID: 77957842589     PISSN: 00223085     EISSN: 19330693     Source Type: Journal    
DOI: 10.3171/2010.3.JNS091346     Document Type: Article
Times cited : (14)

References (17)
  • 1
    • 77957838240 scopus 로고    scopus 로고
    • F 2052, Standard test method for measurement of magnetically induced displacement force on medical devices in the magnetic resonance environment
    • American Society for Testing and Materials (ASTM) Designation: 13.01. West Conshohocken, PA: ASTM International
    • American Society for Testing and Materials (ASTM) Designation: F 2052, Standard test method for measurement of magnetically induced displacement force on medical devices in the magnetic resonance environment, in Annual Book of ASTM Standards, Section 13, Medical Devices and Services, Volume 13.01. West Conshohocken, PA: ASTM International, 2008
    • (2008) Annual Book of ASTM Standards, Section 13, Medical Devices and Services
  • 2
    • 77957832385 scopus 로고    scopus 로고
    • F 2213, Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment
    • American Society for Testing and Materials (ASTM) Designation: 13.01. West Conshohocken, PA: ASTM International
    • American Society for Testing and Materials (ASTM) Designation: F 2213, Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment, in Annual Book of ASTM Standards, Section 13, Medical Devices and Services, Volume 13.01. West Conshohocken, PA: ASTM International, 2008
    • (2008) Annual Book of ASTM Standards, Section 13, Medical Devices and Services
  • 4
    • 0000098425 scopus 로고
    • Intracranial aneurysm of the internal carotid artery: Cured by operation
    • Dandy WE: Intracranial aneurysm of the internal carotid artery: cured by operation. Ann Surg 107:654-659, 1938
    • (1938) Ann Surg , vol.107 , pp. 654-659
    • Dandy, W.E.1
  • 6
    • 0032901612 scopus 로고    scopus 로고
    • Aneurysm clips: Effects of long-term and multiple exposures to a 1.5-T MR system
    • Kanal E, Shellock FG: Aneurysm clips: effects of long-term and multiple exposures to a 1.5-T MR system. Radiology 210:563-565, 1999
    • (1999) Radiology , vol.210 , pp. 563-565
    • Kanal, E.1    Shellock, F.G.2
  • 7
    • 0027193340 scopus 로고
    • MR imaging of patients with intracranial aneurysm clips
    • Kanal E, Shellock FG: MR imaging of patients with intracranial aneurysm clips. Radiology 187:612-614, 1993
    • (1993) Radiology , vol.187 , pp. 612-614
    • Kanal, E.1    Shellock, F.G.2
  • 8
    • 0033946472 scopus 로고    scopus 로고
    • Aneurysm clips: Evaluation of magnetic field interactions with an 8.0 T MR system
    • Kangarlu A, Shellock FG: Aneurysm clips: evaluation of magnetic field interactions with an 8.0 T MR system. J Magn Reson Imaging 12:107-111, 2000
    • (2000) J Magn Reson Imaging , vol.12 , pp. 107-111
    • Kangarlu, A.1    Shellock, F.G.2
  • 9
    • 0027173880 scopus 로고
    • Placement of a ferromagnetic intracerebral aneurysm clip in a magnetic field with a fatal outcome
    • Klucznik RP, Carrier DA, Pyka R, Haid RW: Placement of a ferromagnetic intracerebral aneurysm clip in a magnetic field with a fatal outcome. Radiology 187:855-856, 1993
    • (1993) Radiology , vol.187 , pp. 855-856
    • Klucznik, R.P.1    Carrier, D.A.2    Pyka, R.3    Haid, R.W.4
  • 11
    • 0020614309 scopus 로고
    • Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging
    • New PF, Rosen BR, Brady TJ, Buonanno FS, Kistler JP, Burt CT, et al: Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology 147:139-148, 1983
    • (1983) Radiology , vol.147 , pp. 139-148
    • New, P.F.1    Rosen, B.R.2    Brady, T.J.3    Buonanno, F.S.4    Kistler, J.P.5    Burt, C.T.6
  • 12
    • 0029856590 scopus 로고    scopus 로고
    • Motion and image artifacts of various intracranial aneurysm clips in a magnetic field
    • Ooka K, Shibuya M, Suzuki Y: Motion and image artifacts of various intracranial aneurysm clips in a magnetic field. Acta Neurochir (Wien) 138:1241-1245, 1996
    • (1996) Acta Neurochir (Wien) , vol.138 , pp. 1241-1245
    • Ooka, K.1    Shibuya, M.2    Suzuki, Y.3
  • 13
    • 0031812967 scopus 로고    scopus 로고
    • Yasargil aneurysm clips: Evaluation of interactions with a 1.5-T MR system
    • Shellock FG, Kanal E: Yasargil aneurysm clips: evaluation of interactions with a 1.5-T MR system. Radiology 207:587-591, 1998
    • (1998) Radiology , vol.207 , pp. 587-591
    • Shellock, F.G.1    Kanal, E.2
  • 14
    • 0031884231 scopus 로고    scopus 로고
    • Spetzler titanium aneurysm clips: Compatibility at MR imaging
    • Shellock FG, Shellock VJ: Spetzler titanium aneurysm clips: compatibility at MR imaging. Radiology 206:838-841, 1998
    • (1998) Radiology , vol.206 , pp. 838-841
    • Shellock, F.G.1    Shellock, V.J.2
  • 15
    • 0037347512 scopus 로고    scopus 로고
    • Aneurysm clips: Evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems
    • Shellock FG, Tkach JA, Ruggieri PM, Masaryk TJ, Rasmussen PA: Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems. AJNR Am J Neuroradiol 24:463-471, 2003
    • (2003) AJNR Am J Neuroradiol , vol.24 , pp. 463-471
    • Shellock, F.G.1    Tkach, J.A.2    Ruggieri, P.M.3    Masaryk, T.J.4    Rasmussen, P.A.5
  • 17
    • 37349050654 scopus 로고    scopus 로고
    • Investigation of radiofrequency-induced temperature elevation of aneurysm clips in a 3.0-tesla magnetic resonance environment
    • Watanabe A, Seguchi T, Koyama J, Aoyama T, Miyahara T, Kakizawa Y, et al: Investigation of radiofrequency-induced temperature elevation of aneurysm clips in a 3.0-tesla magnetic resonance environment. Neurosurgery 61:1062-1066, 2007
    • (2007) Neurosurgery , vol.61 , pp. 1062-1066
    • Watanabe, A.1    Seguchi, T.2    Koyama, J.3    Aoyama, T.4    Miyahara, T.5    Kakizawa, Y.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.