Unexpected finding of a fractured metal prosthetic femoral head in a nonmodular implant during revision total hip arthroplasty

Author keywords

Failure; Femoral head; Fracture; Metal fatigue; Prosthesis; Total hip arthroplasty; Weld

Indexed keywords

ACETABULUM; ADULT; ARTICLE; BONE RADIOGRAPHY; CASE REPORT; FEMUR HEAD PROSTHESIS; HUMAN; INCIDENTAL FINDING; MALE; MECHANICAL STRESS; METAL IMPLANTATION; OSTEOLYSIS; OSTEOTOMY; PREOPERATIVE EVALUATION; PROSTHESIS FAILURE; SCANNING ELECTRON MICROSCOPY; TOTAL HIP PROSTHESIS; TROCHANTERIC OSTEOTOMY;

EID: 77953122650     PISSN: 08835403     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.arth.2009.02.022     Document Type: Article
Times cited : (8)

References (5)
  • 1
    • 0031444317 scopus 로고    scopus 로고
    • Fatigue fracture of a forged cobalt-chromium-molybdenum femoral component inserted with cement. A report of ten cases
    • Woolson S.T., Milbauer J.P., Bobyn J.D., et al. Fatigue fracture of a forged cobalt-chromium-molybdenum femoral component inserted with cement. A report of ten cases. J Bone Joint Surg Am 1997, 79:1842.
    • (1997) J Bone Joint Surg Am , vol.79 , pp. 1842
    • Woolson, S.T.1    Milbauer, J.P.2    Bobyn, J.D.3
  • 2
    • 29244474600 scopus 로고    scopus 로고
    • Late fatigue fracture of a modern cemented cobalt chrome stem for total hip arthroplasty: a report of 10 cases
    • Della Valle A.G., Becksac B., Anderson J., et al. Late fatigue fracture of a modern cemented cobalt chrome stem for total hip arthroplasty: a report of 10 cases. J Arthroplasty 2005, 20:1084.
    • (2005) J Arthroplasty , vol.20 , pp. 1084
    • Della Valle, A.G.1    Becksac, B.2    Anderson, J.3
  • 3
    • 0028343874 scopus 로고
    • Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants
    • Gilbert J.L., Buckley C.A., Jacobs J.J., et al. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants. J Bone Joint Surg Am 1994, 76:110.
    • (1994) J Bone Joint Surg Am , vol.76 , pp. 110
    • Gilbert, J.L.1    Buckley, C.A.2    Jacobs, J.J.3
  • 4
    • 33947589230 scopus 로고    scopus 로고
    • Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup
    • Hwang D.S., Kim Y.M., Lee C.H. Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty 2007, 22:468.
    • (2007) J Arthroplasty , vol.22 , pp. 468
    • Hwang, D.S.1    Kim, Y.M.2    Lee, C.H.3
  • 5
    • 0029902177 scopus 로고    scopus 로고
    • Ceramic femoral head fractures in total hip arthroplasty
    • Fritsch E.W., Gleitz M. Ceramic femoral head fractures in total hip arthroplasty. Clin Orthop Relat Res 1996, 328:129.
    • (1996) Clin Orthop Relat Res , vol.328 , pp. 129
    • Fritsch, E.W.1    Gleitz, M.2