Volumn 37, Issue 5, 2010, Pages 2179-2193

Measurement of ionization chamber absorbed dose kQ factors in megavoltage photon beams

Author keywords

Absorbed dose calibration coefficients; Ion recombination; Ionization chambers; KQ factors; Polarity; Reference dosimetry

Indexed keywords

CALIBRATION; DOSIMETRY; IONIZATION CHAMBERS; IONS; PARTICLE BEAMS; PHOTOELECTRONS; PHOTONS; SCANNING;

EID: 77952340515     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3375895     Document Type: Article
Times cited : (128)

References (50)
  • 1
    • 0032874911 scopus 로고    scopus 로고
    • AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    • DOI 10.1118/1.598691
    • P. Almond, P. J. Biggs, B. M. Coursey, W. F. Hanson, M. S. Huq, R. Nath, and D. W. O. Rogers, " AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Report of AAPM Radiation Therapy Committee Task Group No. 51.," Med. Phys. MPHYA6 0094-2405 26, 1847-1870 (1999). 10.1118/1.598691 (Pubitemid 29442268)
    • (1999) Medical Physics , vol.26 , Issue.9 , pp. 1847-1870
    • Almond, P.R.1    Biggs, P.J.2    Coursey, B.M.3    Hanson, W.F.4    Huq, M.S.5    Nath, R.6    Rogers, D.W.O.7
  • 4
    • 85027228166 scopus 로고    scopus 로고
    • Measurements of kQ factors, ion recombination, polarity and sleeve effects of Farmer type chambers calibrated in terms of absorbed dose to water at NPL
    • Sydney, Australia
    • R. A. S. Thomas, S. Duane, and M. Simon, " Measurements of kQ factors, ion recombination, polarity and sleeve effects of Farmer type chambers calibrated in terms of absorbed dose to water at NPL.," presented at Proceedings of the World Congress on Medical Physics, Sydney, Australia, 2003.
    • (2003) Proceedings of the World Congress on Medical Physics
    • Thomas, R.A.S.1    Duane, S.2    Simon, M.3
  • 7
    • 0033002724 scopus 로고    scopus 로고
    • Absorbed dose beam quality correction factors k(Q) for the NE2571 chamber in a 5 MV and a 10 MV photon beam
    • DOI 10.1088/0031-9155/44/3/002, PII S0031915599953666
    • Q for the NE2571 chamber in a 5 MV and a 10 MV photon beam.," Phys. Med. Biol. PHMBA7 0031-9155 44, 647-663 (1999). 10.1088/0031-9155/44/3/002 (Pubitemid 29133059)
    • (1999) Physics in Medicine and Biology , vol.44 , Issue.3 , pp. 647-663
    • Palmans, H.1    Mondelaers, W.2    Thierens, H.3
  • 8
    • 0037037131 scopus 로고    scopus 로고
    • Beam quality of high-energy photon beams at the Ghent University linear accelerator
    • PHMBA7 0031-9155. 10.1088/0031-9155/47/17/101
    • H. Palmans, W. Mondelaers, and H. Thierens, " Beam quality of high-energy photon beams at the Ghent University linear accelerator.," Phys. Med. Biol. PHMBA7 0031-9155 47, L15-L18 (2002). 10.1088/0031-9155/47/17/ 101
    • (2002) Phys. Med. Biol. , vol.47
    • Palmans, H.1    Mondelaers, W.2    Thierens, H.3
  • 9
    • 0029031216 scopus 로고
    • Q, for ionization chambers used in photon and electron dosimetry
    • PHMBA7 0031-9155. 10.1088/0031-9155/40/7/003
    • Q, for ionization chambers used in photon and electron dosimetry.," Phys. Med. Biol. PHMBA7 0031-9155 40, 1177-1190 (1995). 10.1088/0031-9155/40/7/003
    • (1995) Phys. Med. Biol. , vol.40 , pp. 1177-1190
    • Guerra, A.S.1    Laitano, R.F.2    Pimpinella, M.3
  • 10
    • 0034500295 scopus 로고    scopus 로고
    • Absorbed-dose beam quality conversion factors for cylindrical chambers in high energy photon beams
    • DOI 10.1118/1.1328081
    • J. Seuntjens, C. Ross, K. Shortt, and D. Rogers, " Absorbed-dose beam quality conversion factors for cylindrical chambers in high energy photon beams.," Med. Phys. MPHYA6 0094-2405 27, 2763-2779 (2000). 10.1118/1.1328081 (Pubitemid 32063456)
    • (2000) Medical Physics , vol.27 , Issue.12 , pp. 2763-2779
    • Seuntjens, J.P.1    Ross, C.K.2    Shortt, K.R.3    Rogers, D.W.O.4
  • 11
    • 36048968264 scopus 로고    scopus 로고
    • Q factors for NE 2561 and NE 2571 ionization chambers in 5 cm × 5 cm and 10 cm × 10 cm radiotherapy beams of 8 MV and 16 MV photons
    • DOI 10.1088/0031-9155/52/20/011, PII S0031915507562111
    • A. Krauss and R. -P. Kapsch, " Calorimetric determination of kQ factors for NE 2561 and NE 2571 ionization chambers in 5 cm×5 cm and 10 cm×10 cm radiotherapy beams of 8 MV and 16 MV photons.," Phys. Med. Biol. PHMBA7 0031-9155 52, 6243-6259 (2007). 10.1088/0031-9155/52/20/011 (Pubitemid 350093192)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.20 , pp. 6243-6259
    • Krauss, A.1    Kapsch, R.-P.2
  • 15
    • 34249023467 scopus 로고    scopus 로고
    • Direct calibration of ionization chambers in linac photon beams
    • MPHYA6 0094-2405. 10.1118/1.1998523
    • M. R. McEwen and C. K. Ross, " Direct calibration of ionization chambers in linac photon beams.," Med. Phys. MPHYA6 0094-2405 32, 2128 (2005). 10.1118/1.1998523
    • (2005) Med. Phys. , vol.32 , pp. 2128
    • McEwen, M.R.1    Ross, C.K.2
  • 17
    • 33744784872 scopus 로고    scopus 로고
    • 60Co radiation
    • DOI 10.1088/0026-1394/43/3/008, PII S0026139406219492
    • A. Krauss, " The PTB water calorimeter for the absolute determination of absorbed dose to water in C 60 o radiation.," Metrologia MTRGAU 0026-1394 43, 259-259 (2006). 10.1088/0026-1394/43/3/008 (Pubitemid 43830228)
    • (2006) Metrologia , vol.43 , Issue.3 , pp. 259-272
    • Krauss, A.1
  • 18
    • 0033856077 scopus 로고    scopus 로고
    • Comparison of dosimetry calibration factors at the NRCC and the NIST
    • DOI 10.1118/1.599031
    • K. R. Shortt, J. Shobe, and S. R. Domen, " Comparison of dosimetry calibration factors at the NRCC and the NIST.," Med. Phys. MPHYA6 0094-2405 27, 1644-1654 (2000). 10.1118/1.599031 (Pubitemid 30624260)
    • (2000) Medical Physics , vol.27 , Issue.7 , pp. 1644-1654
    • Shortt, K.1    Shobe, J.2    Domen, S.3
  • 19
    • 0034879566 scopus 로고    scopus 로고
    • 60Co and higher energies
    • DOI 10.1088/0031-9155/46/8/307, PII S0031915501201079
    • K. Shortt, C. Ross, J. Seuntjens, F. Delaunay, A. Ostrowsky, P. Gross, and E. Leroy, " Comparison of dosimetric standards of Canada and France for photons at C 60 o and higher energies.," Phys. Med. Biol. PHMBA7 0031-9155 46, 2119-2142 (2001). 10.1088/0031-9155/46/8/307 (Pubitemid 32747067)
    • (2001) Physics in Medicine and Biology , vol.46 , Issue.8 , pp. 2119-2142
    • Shortt, K.1    Ross, C.2    Seuntjens, J.3    Delaunay, F.4    Ostrowsky, A.5    Gross, P.6    Leroy, E.7
  • 20
    • 40049109564 scopus 로고    scopus 로고
    • The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams
    • DOI 10.1118/1.2839329
    • M. R. McEwen, I. Kawrakow, and C. K. Ross, " The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams.," Med. Phys. MPHYA6 0094-2405 35, 950-958 (2008). 10.1118/1.2839329 (Pubitemid 351322676)
    • (2008) Medical Physics , vol.35 , Issue.3 , pp. 950-958
    • McEwen, M.R.1    Kawrakow, I.2    Ross, C.K.3
  • 22
    • 21344472828 scopus 로고    scopus 로고
    • Pre-irradiation effects on ionization chambers used in radiation therapy
    • DOI 10.1088/0031-9155/50/13/N01
    • J. P. McCaffrey, B. Downton, H. Shen, D. Niven, and M. R. McEwen, " Pre-irradiation effects on ionization chambers used in radiation therapy.," Phys. Med. Biol. PHMBA7 0031-9155 50, N121-N133 (2005). 10.1088/0031-9155/50/ 13/N01 (Pubitemid 40905178)
    • (2005) Physics in Medicine and Biology , vol.50 , Issue.13
    • McCaffrey, J.P.1    Downton, B.2    Shen, H.3    Niven, D.4    McEwen, M.5
  • 23
    • 0031830063 scopus 로고    scopus 로고
    • Ion recombination corrections for the NACP parallel-plate chamber in a pulsed electron beam
    • DOI 10.1088/0031-9155/43/8/003, PII S0031915598907357
    • D. T. Burns and M. R. McEwen, " Ion recombination corrections for the NACP parallel-plate chamber in a pulsed electron beam.," Phys. Med. Biol. PHMBA7 0031-9155 43, 2033-2045 (1998). 10.1088/0031-9155/43/8/003 (Pubitemid 28372606)
    • (1998) Physics in Medicine and Biology , vol.43 , Issue.8 , pp. 2033-2045
    • Burns, D.T.1    McEwen, M.R.2
  • 24
    • 33947717382 scopus 로고    scopus 로고
    • S for some specific plane-parallel and cylindrical ionization chambers in pulsed photon and electron beams
    • DOI 10.1088/0031-9155/52/2/N01, PII S0031915507289237, N01
    • G. Bruggmoser, R. Saum, A. Schmachtenberg, F. Schmid, and E. Schüle, " Determination of the recombination correction factor kS for some specific plane-parallel and cylindrical ionization chambers in pulsed photon and electron beams.," Phys. Med. Biol. PHMBA7 0031-9155 52, N35-N50 (2007). 10.1088/0031-9155/52/2/N01 (Pubitemid 46646469)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.2
    • Bruggmoser, G.1    Saum, R.2    Schmachtenberg, A.3    Schmid, F.4    Schule, E.5
  • 25
    • 0018832381 scopus 로고
    • Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiation
    • BJRAAP 0007-1285. 10.1259/0007-1285-53-629-471
    • J. W. Boag and J. Currant, " Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiation.," Br. J. Radiol. BJRAAP 0007-1285 53, 471-478 (1980). 10.1259/0007-1285-53-629-471
    • (1980) Br. J. Radiol. , vol.53 , pp. 471-478
    • Boag, J.W.1    Currant, J.2
  • 26
    • 0021646455 scopus 로고
    • Determining P(ion), the correction factor for recombination losses in an ionization chamber
    • DOI 10.1118/1.595574
    • M. S. Weinhous and J. A. Meli, " Determining Pion, the correction factor for recombination losses in an ionisation chamber.," Med. Phys. MPHYA6 0094-2405 11, 846-849 (1984). 10.1118/1.595574 (Pubitemid 15223743)
    • (1984) Medical Physics , vol.11 , Issue.6 , pp. 846-849
    • Weinhous, M.S.1    Meli, J.A.2
  • 29
    • 0033795155 scopus 로고    scopus 로고
    • IPEM guidelines on dosimeter systems for use as transfer instruments between the UK primary dosimetry standards laboratory (NPL) and radiotherapy centres
    • PHMBA7 0031-9155. 10.1088/0031-9155/45/9/301
    • A. M. Morgan, E. G. A. Aird, R. J. Aukett, S. Duane, N. H. Jenkins, W. P. M. Mayles, C. Moretti, and D. I. Thwaites, " IPEM guidelines on dosimeter systems for use as transfer instruments between the UK primary dosimetry standards laboratory (NPL) and radiotherapy centres.," Phys. Med. Biol. PHMBA7 0031-9155 45, 2445-2457 (2000). 10.1088/0031-9155/45/9/301
    • (2000) Phys. Med. Biol. , vol.45 , pp. 2445-2457
    • Morgan, A.M.1    Aird, E.G.A.2    Aukett, R.J.3    Duane, S.4    Jenkins, N.H.5    Mayles, W.P.M.6    Moretti, C.7    Thwaites, D.I.8
  • 31
    • 0025810529 scopus 로고
    • Induced effects in ionization chamber cables by photon and electron irradiation
    • MPHYA6 0094-2405. 10.1118/1.596657
    • L. L. Campos and L. V. E. Caldas, " Induced effects in ionization chamber cables by photon and electron irradiation.," Med. Phys. MPHYA6 0094-2405 18, 522-526 (1991). 10.1118/1.596657
    • (1991) Med. Phys. , vol.18 , pp. 522-526
    • Campos, L.L.1    Caldas, L.V.E.2
  • 32
    • 7244257424 scopus 로고    scopus 로고
    • The behavior of several microionization chambers in small intensity modulated radiotherapy fields
    • DOI 10.1118/1.1788911
    • M. Stasi, B. Baiotto, G. Barboni, and G. Scielzo, " The behavior of several microionization chambers in small intensity modulated radiotherapy fields.," Med. Phys. MPHYA6 0094-2405 31, 2792-2795 (2004). 10.1118/1.1788911 (Pubitemid 39431006)
    • (2004) Medical Physics , vol.31 , Issue.10 , pp. 2792-2795
    • Stasi, M.1    Baiotto, B.2    Barboni, G.3    Scielzo, G.4
  • 33
    • 46149117041 scopus 로고    scopus 로고
    • Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrode
    • DOI 10.1118/1.2940190
    • S. Agostinelli, S. Garelli, M. Piergentili, and F. Foppiano, " Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrode.," Med. Phys. MPHYA6 0094-2405 35, 3293-3301 (2008). 10.1118/1.2940190 (Pubitemid 351904533)
    • (2008) Medical Physics , vol.35 , Issue.7 , pp. 3293-3301
    • Agostinelli, S.1    Garelli, S.2    Piergentili, M.3    Foppiano, F.4
  • 34
    • 0027175721 scopus 로고
    • Measurement of saturation correction factors of thimble-type ionization chambers in pulsed photon beams
    • PHMBA7 0031-9155. 10.1088/0031-9155/38/6/009
    • K. Derikum and M. Roos, " Measurement of saturation correction factors of thimble-type ionization chambers in pulsed photon beams.," Phys. Med. Biol. PHMBA7 0031-9155 38, 755-763 (1993). 10.1088/0031-9155/38/6/009
    • (1993) Phys. Med. Biol. , vol.38 , pp. 755-763
    • Derikum, K.1    Roos, M.2
  • 35
    • 0034036230 scopus 로고    scopus 로고
    • Saturation current and collection efficiency for ionization chambers in pulsed beams
    • DOI 10.1118/1.598992
    • F. DeBlois, C. Zankowski, and E. B. Podgorsak, " Saturation current and collection efficiency for ionization chambers in pulsed beams.," Med. Phys. MPHYA6 0094-2405 27, 1146-1155 (2000). 10.1118/1.598992 (Pubitemid 30350028)
    • (2000) Medical Physics , vol.27 , Issue.5 , pp. 1146-1155
    • DeBlois, F.1    Zankowski, C.2    Podgorsak, E.B.3
  • 36
    • 85027239896 scopus 로고    scopus 로고
    • personal communication of NPL data
    • R. A S. Thomas, personal communication of NPL data (2008).
    • (2008)
    • Thomas, R.A.S.1
  • 37
    • 11044232519 scopus 로고    scopus 로고
    • Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams
    • DOI 10.1088/0031-9155/49/23/008, PII S0031915504822942
    • M. Berg and O. Noerrevang, " Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams.," Phys. Med. Biol. PHMBA7 0031-9155 49, 5309-5318 (2004). 10.1088/0031-9155/49/23/008 (Pubitemid 40044863)
    • (2004) Physics in Medicine and Biology , vol.49 , Issue.23 , pp. 5309-5318
    • Berg, M.1    Noerrevang, O.2
  • 38
    • 0026665730 scopus 로고
    • The effect of waterproofing sleeves on ionization chamber response
    • PHMBA7 0031-9155. 10.1088/0031-9155/37/6/017
    • C. K. Ross and K. R. Shortt, " The effect of waterproofing sleeves on ionization chamber response.," Phys. Med. Biol. PHMBA7 0031-9155 37, 1403-1411 (1992). 10.1088/0031-9155/37/6/017
    • (1992) Phys. Med. Biol. , vol.37 , pp. 1403-1411
    • Ross, C.K.1    Shortt, K.R.2
  • 39
    • 31844450298 scopus 로고    scopus 로고
    • wall, for thimble ionization chambers
    • DOI 10.1118/1.2161403
    • L. A. Buckley and D. W. O. Rogers, " Wall correction factors, Pwall, for thimble ionization chambers.," Med. Phys. MPHYA6 0094-2405 33, 455-464 (2006). 10.1118/1.2161403 (Pubitemid 43183990)
    • (2006) Medical Physics , vol.33 , Issue.2 , pp. 455-464
    • Buckley, L.A.1    Rogers, D.W.O.2
  • 42
    • 45749111374 scopus 로고    scopus 로고
    • Monte-Carlo-based perturbation and beam quality correction factors for thimble ionization chambers in high-energy photon beams
    • DOI 10.1088/0031-9155/53/11/005, PII S0031915508724370
    • J. Wulff, J. T. Heverhagen, and K. Zink, " Monte-Carlo-based perturbation and beam quality correction factors for thimble ionization chambers in high-energy photon beams.," Phys. Med. Biol. PHMBA7 0031-9155 53, 2823-2836 (2008). 10.1088/0031-9155/53/11/005 (Pubitemid 351868608)
    • (2008) Physics in Medicine and Biology , vol.53 , Issue.11 , pp. 2823-2836
    • Wulff, J.1    Heverhagen, J.T.2    Zink, K.3
  • 43
    • 77952411767 scopus 로고    scopus 로고
    • A Monte-Carlo investigation of the beam quality conversion factor
    • MPHYA6 0094-2405. 10.1118/1.3182372
    • B. Muir and D. W. O. Rogers, " A Monte-Carlo investigation of the beam quality conversion factor.," Med. Phys. MPHYA6 0094-2405 36, 2732-2733 (2009). 10.1118/1.3182372
    • (2009) Med. Phys. , vol.36 , pp. 2732-2733
    • Muir, B.1    Rogers, D.W.O.2
  • 45
    • 0019362260 scopus 로고
    • Sensitivity dependence of an A-150 walled ion chamber to air humidity
    • PHMBA7 0031-9155. 10.1088/0031-9155/26/5/009
    • M. Kristensen and L. Sundbom, " Sensitivity dependence of an A-150 walled ion chamber to air humidity.," Phys. Med. Biol. PHMBA7 0031-9155 26, 903-905 (1981). 10.1088/0031-9155/26/5/009
    • (1981) Phys. Med. Biol. , vol.26 , pp. 903-905
    • Kristensen, M.1    Sundbom, L.2
  • 46
    • 0021829576 scopus 로고
    • Variations in response to radiation of a nylon-walled ionization chamber induced by humidity changes
    • DOI 10.1118/1.595683
    • B. J. Mijnheer, " Variations in response to radiation of a nylon-walled ionization chamber induced by humidity changes.," Med. Phys. MPHYA6 0094-2405 12, 625-626 (1985). 10.1118/1.595683 (Pubitemid 15248420)
    • (1985) Medical Physics , vol.12 , Issue.5 , pp. 625-626
    • Mijnheer, B.J.1
  • 47
    • 1642362277 scopus 로고    scopus 로고
    • Thermal and temporal response of ionization chambers in radiation dosimetry
    • DOI 10.1118/1.1644520
    • I. Das and T. Zhu, " Thermal and temporal response of ionization chambers in radiation dosimetry.," Med. Phys. MPHYA6 0094-2405 31, 573-578 (2004). 10.1118/1.1644520 (Pubitemid 38381020)
    • (2004) Medical Physics , vol.31 , Issue.3 , pp. 573-578
    • Das, I.J.1    Zhu, T.C.2
  • 48
    • 85027225057 scopus 로고    scopus 로고
    • Long-term stability of PTW farmer type ionization chambers
    • PTW, (PTW, Freiburg)
    • PTW, " Long-term stability of PTW farmer type ionization chambers.," PTW Technical Note D165.200.0/2 (PTW, Freiburg, 1998).
    • (1998) PTW Technical Note D165.200.0/2
  • 50
    • 0025183565 scopus 로고
    • Code of practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service
    • DOI 10.1088/0031-9155/35/10/301
    • S. C. Lillicrap, B. Owen, J. R. Williams, and P. C. Williams, " Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service.," Phys. Med. Biol. PHMBA7 0031-9155 35, 1355-1360 (1990). 10.1088/0031-9155/35/10/301 (Pubitemid 20304104)
    • (1990) Physics in Medicine and Biology , vol.35 , Issue.10 , pp. 1355-1360
    • Lillicrap, S.C.1    Owen, B.2    Williams, J.R.3    Williams, P.C.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.