메뉴 건너뛰기




Volumn 23, Issue 12, 2009, Pages 2822-2826

Evaluation of mist production and tissue dissection efficiency using different types of ultrasound shears

Author keywords

Minimally invasive surgery; Mist production; Tissue dissection efficiency; Ultrasound shears

Indexed keywords

ARTICLE; CONTROLLED STUDY; DISSECTION; ECHOGRAPHY; INSTRUMENT; PRIORITY JOURNAL; PRODUCTIVITY; QUANTITATIVE ANALYSIS; ULTRASOUND DISSECTOR; ULTRASOUND SHEAR;

EID: 71449122838     PISSN: 09302794     EISSN: 14322218     Source Type: Journal    
DOI: 10.1007/s00464-009-0512-5     Document Type: Article
Times cited : (11)

References (7)
  • 1
    • 0142090555 scopus 로고    scopus 로고
    • Laparoscopic bowel resection: A comparison of three ultrasonically activated devices
    • 12722994
    • R Bergamaschi Y Yavuz R Marvik 2003 Laparoscopic bowel resection: a comparison of three ultrasonically activated devices JSLS 7 19 22 12722994
    • (2003) JSLS , vol.7 , pp. 19-22
    • Bergamaschi, R.1    Yavuz, Y.2    Marvik, R.3
  • 2
    • 0032951086 scopus 로고    scopus 로고
    • Acute thermal injury to the canine jejunal free flap: Electrocautery versus ultrasonic dissection
    • 1:STN:280:DyaK1M3gsVKgsQ%3D%3D 10190358
    • DW Birch A Park H Shuhaibar 1999 Acute thermal injury to the canine jejunal free flap: electrocautery versus ultrasonic dissection Am Surg 65 334 337 1:STN:280:DyaK1M3gsVKgsQ%3D%3D 10190358
    • (1999) Am Surg , vol.65 , pp. 334-337
    • Birch, D.W.1    Park, A.2    Shuhaibar, H.3
  • 3
    • 0030746690 scopus 로고    scopus 로고
    • Cells are present in the smoke created during laparoscopic surgery
    • DOI 10.1002/bjs.1800840724
    • G Champault N Taffinder M Ziol H Riskalla JM Catheline 1997 Cells are present in the smoke created during laparoscopic surgery Br J Surg 84 993 995 10.1002/bjs.1800840724 1:STN:280:DyaK2szosVCktQ%3D%3D 9240145 (Pubitemid 27310872)
    • (1997) British Journal of Surgery , vol.84 , Issue.7 , pp. 993-995
    • Champault, G.1    Taffinder, N.2    Ziol, M.3    Riskalla, H.4    Catheline, J.M.C.5
  • 4
    • 36248956054 scopus 로고    scopus 로고
    • In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries
    • DOI 10.1007/s00464-007-9345-2
    • RH Clements R Palepu 2007 In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4-mm and 5-mm arteries Surg Endosc 21 2203 2206 10.1007/s00464-007-9345-2 17479325 (Pubitemid 350125944)
    • (2007) Surgical Endoscopy , vol.21 , Issue.12 , pp. 2203-2206
    • Clements, R.H.1    Palepu, R.2
  • 6
    • 43749124249 scopus 로고    scopus 로고
    • Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure v (LV), and plasma trisector (PT)
    • 10.1007/s00464-007-9650-9 1:STN:280:DC%2BD1czgvFWjug%3D%3D 18027042
    • FJ Kim MF Chammas Jr E Gewehr M Morihisa F Caldas E Hayacibara M Baptistussi F Meyer AC Martins 2008 Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT) Surg Endosc 22 1464 1469 10.1007/s00464-007-9650-9 1:STN:280:DC%2BD1czgvFWjug%3D%3D 18027042
    • (2008) Surg Endosc , vol.22 , pp. 1464-1469
    • Kim, F.J.1    Chammas Jr, M.F.2    Gewehr, E.3    Morihisa, M.4    Caldas, F.5    Hayacibara, E.6    Baptistussi, M.7    Meyer, F.8    Martins, A.C.9
  • 7
    • 0036916006 scopus 로고    scopus 로고
    • New integrated ultrasonic surgical system, 'SonoSurg'
    • DOI 10.1046/j.1443-1661.2002.00195.x
    • T Yamakawa S Kitano T Kimura S Matsumoto 2002 New integrated ultrasonic surgical system, Sonosurg Dig Endosc 14 163 166 10.1046/j.1443-1661.2002.00195.x (Pubitemid 36013994)
    • (2002) Digestive Endoscopy , vol.14 , Issue.4 , pp. 163-166
    • Yamakawa, T.1    Kitano, S.2    Kimura, T.3    Matsumoto, S.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.