Volumn 18, Issue 3, 2009, Pages 224-230

Comparison of glucose point-of-care values with laboratory values in critically ill patients

Author keywords

[No Author keywords available]

Indexed keywords

ADULT; AGED; ARTICLE; CATHETERIZATION; COMPARATIVE STUDY; DIAGNOSTIC TEST; GLUCOSE BLOOD LEVEL; HOSPITAL INFORMATION SYSTEM; HUMAN; INSTRUMENTATION; INTENSIVE CARE; INTENSIVE CARE UNIT; MIDDLE AGED; NON PROFIT HOSPITAL; REPRODUCIBILITY; STANDARD; UNITED STATES;

EID: 67649785244     PISSN: 10623264     EISSN: None     Source Type: Journal    
DOI: 10.4037/ajcc2009448     Document Type: Article
Times cited : (28)

References (42)
  • 1
    • 0035829852 scopus 로고    scopus 로고
    • Intensive insulin therapy in critically ill patients
    • Van den Berghe G, Wouters P, Weekers F, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345(19):1359-1367.
    • (2001) N Engl J Med , vol.345 , Issue.19 , pp. 1359-1367
    • Van den Berghe, G.1    Wouters, P.2    Weekers, F.3
  • 2
    • 33845257074 scopus 로고    scopus 로고
    • Evaluation of nursing work effort and perceptions about blood glucose testing in tight glycemic control
    • Aragon D. Evaluation of nursing work effort and perceptions about blood glucose testing in tight glycemic control. Am J Crit Care. 2006;15(4):370-377.
    • (2006) Am J Crit Care , vol.15 , Issue.4 , pp. 370-377
    • Aragon, D.1
  • 3
    • 24944561405 scopus 로고    scopus 로고
    • Intensive versus modified control of blood glucose level in medical intensive care patients: a pilot study
    • Bland DK, Fankhanel Y, Langford E, et al. Intensive versus modified control of blood glucose level in medical intensive care patients: a pilot study. Am J Crit Care. 2005;14(4): 370-376.
    • (2005) Am J Crit Care , vol.14 , Issue.4 , pp. 370-376
    • Bland, D.K.1    Fankhanel, Y.2    Langford, E.3
  • 4
    • 0035170340 scopus 로고    scopus 로고
    • Pilot study of the accuracy of bedside glucometry in the intensive care unit
    • Ray J, Hamielec C, Mastracci T. Pilot study of the accuracy of bedside glucometry in the intensive care unit. Crit Care Med. 2001;29:2205-2207.
    • (2001) Crit Care Med , vol.29 , pp. 2205-2207
    • Ray, J.1    Hamielec, C.2    Mastracci, T.3
  • 5
    • 0028262386 scopus 로고
    • Use of arterial blood with bedside glucose reflectance meters in an intensive care unit: are they accurate?
    • Maser R, Butler M, DeCherney T. Use of arterial blood with bedside glucose reflectance meters in an intensive care unit: are they accurate? Crit Care Med. 1994:22(4);595-599.
    • (1994) Crit Care Med , vol.22 , Issue.4 , pp. 595-599
    • Maser, R.1    Butler, M.2    DeCherney, T.3
  • 6
    • 0025608626 scopus 로고
    • Use of arterial blood for glucose measurement by reflectance
    • Pressly KB, Batteiger TH, Barnett DZ, Woodie ME. Use of arterial blood for glucose measurement by reflectance. Nurs Res. 1990;39(6):371-373.
    • (1990) Nurs Res , vol.39 , Issue.6 , pp. 371-373
    • Pressly, K.B.1    Batteiger, T.H.2    Barnett, D.Z.3    Woodie, M.E.4
  • 7
    • 29144475766 scopus 로고    scopus 로고
    • Reliability of point-of-care testing for glucose measurement in critically ill adults
    • Kanji S, Buffie J, Hutton B, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33(12):2778-2785.
    • (2005) Crit Care Med , vol.33 , Issue.12 , pp. 2778-2785
    • Kanji, S.1    Buffie, J.2    Hutton, B.3
  • 8
    • 34548594005 scopus 로고    scopus 로고
    • Comparison of point-of-care and laboratory analysis methods in critically ill patients
    • Lacara T, Domagtoy C, Lickliter D, et al. Comparison of point-of-care and laboratory analysis methods in critically ill patients. Am J Crit Care. 2007;16(4):336-346.
    • (2007) Am J Crit Care , vol.16 , Issue.4 , pp. 336-346
    • Lacara, T.1    Domagtoy, C.2    Lickliter, D.3
  • 10
    • 77954608476 scopus 로고    scopus 로고
    • G*Power: a priori, post-hoc, and compromise power analysis for MS-DOS Accessed February 16, 2009
    • Faul F, Erdfleder E. G*Power: a priori, post-hoc, and compromise power analysis for MS-DOS. http://www.psycho.uni-duesseldorf.de/aap/projects/gpower. Accessed February 16, 2009.
    • Faul, F.1    Erdfleder, E.2
  • 12
    • 77954604512 scopus 로고    scopus 로고
    • Olympus 604 AU., Instructions [package, insert]., Melville, NY: Olympus, America, Inc
    • Olympus 604 AU Instructions [package insert]. Melville, NY: Olympus, America, Inc; 1999.
    • (1999)
  • 13
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.1    Altman, D.2
  • 14
    • 0028791287 scopus 로고
    • Comparing methods of measurement: why plotting difference against standard method is misleading
    • Bland J, Altman D. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346:1085-1087.
    • (1995) Lancet , vol.346 , pp. 1085-1087
    • Bland, J.1    Altman, D.2
  • 15
    • 0030154483 scopus 로고    scopus 로고
    • Evaluation of instrument error and method of agreement
    • Chatburn R. Evaluation of instrument error and method of agreement. AANA J. 1996;64(3):261-268.
    • (1996) AANA J , vol.64 , Issue.3 , pp. 261-268
    • Chatburn, R.1
  • 16
    • 0142105857 scopus 로고    scopus 로고
    • Measuring agreement between diagnostic devices
    • Flemons W, Littner M. Measuring agreement between diagnostic devices. Chest. 2003;124(4):1535-1542.
    • (2003) Chest , vol.124 , Issue.4 , pp. 1535-1542
    • Flemons, W.1    Littner, M.2
  • 17
    • 0030140806 scopus 로고    scopus 로고
    • Technology assessment in critical care: understanding statistical analyses used to assess agreement between methods of clinical measurement
    • Szaflarski N, Slaughter R. Technology assessment in critical care: understanding statistical analyses used to assess agreement between methods of clinical measurement. Am J Crit Care. 1996;5:207-216.
    • (1996) Am J Crit Care , vol.5 , pp. 207-216
    • Szaflarski, N.1    Slaughter, R.2
  • 18
    • 0038642315 scopus 로고    scopus 로고
    • Comparison of several POCT glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate
    • Puntmann I, Wosniok W, Haeckel R. Comparison of several POCT glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate. Clin Chem Lab Med. 2003;41(6);809-820.
    • (2003) Clin Chem Lab Med , vol.41 , Issue.6 , pp. 809-820
    • Puntmann, I.1    Wosniok, W.2    Haeckel, R.3
  • 19
    • 0034128795 scopus 로고    scopus 로고
    • Effects of pH on glucose measurements with handheld glucose meters and portable glucose analyzer for point-of care testing
    • Tang Z, Du X, Louie R, Kost G. Effects of pH on glucose measurements with handheld glucose meters and portable glucose analyzer for point-of care testing. Arch Pathol Lab Med. 2000;124:577-582.
    • (2000) Arch Pathol Lab Med , vol.124 , pp. 577-582
    • Tang, Z.1    Du, X.2    Louie, R.3    Kost, G.4
  • 20
    • 1642486752 scopus 로고    scopus 로고
    • Comparison of outpatient point of care glucose testing vs venous glucose in the clinical laboratory
    • Velázquez Medina D, Climent C. Comparison of outpatient point of care glucose testing vs venous glucose in the clinical laboratory. P R Health Sci J. 2003;22(4);385-389.
    • (2003) P R Health Sci J , vol.22 , Issue.4 , pp. 385-389
    • Velázquez Medina, D.1    Climent, C.2
  • 22
    • 0032412333 scopus 로고    scopus 로고
    • Comparative accuracy of glucose monitors
    • Fanghänel G, Sanchez-Reyes L, Morales M, et al. Comparative accuracy of glucose monitors. Arch Med Res. 1998;29(4): 325-329.
    • (1998) Arch Med Res , vol.29 , Issue.4 , pp. 325-329
    • Fanghänel, G.1    Sanchez-Reyes, L.2    Morales, M.3
  • 23
    • 0038059846 scopus 로고    scopus 로고
    • Evolution of analytical performance in portable glucose meters in the last decade
    • Böhme P, Rloriot M, Sirveaux M, et al. Evolution of analytical performance in portable glucose meters in the last decade. Diabetes Care. 2003;26(4):1170-1175.
    • (2003) Diabetes Care , vol.26 , Issue.4 , pp. 1170-1175
    • Böhme, P.1    Rloriot, M.2    Sirveaux, M.3
  • 24
    • 0027357270 scopus 로고
    • A comparison of four glucose monitors in a hospital medical surgical setting
    • Harding K. A comparison of four glucose monitors in a hospital medical surgical setting. Clin Nurse Spec. 1993;7(1):13-16.
    • (1993) Clin Nurse Spec , vol.7 , Issue.1 , pp. 13-16
    • Harding, K.1
  • 25
    • 33646920608 scopus 로고    scopus 로고
    • Accuracy of portable blood glucose monitors
    • Aboezz R, Miller DR. Accuracy of portable blood glucose monitors. J Am Pharm Assoc. 2005;45(4):514-516.
    • (2005) J Am Pharm Assoc , vol.45 , Issue.4 , pp. 514-516
    • Aboezz, R.1    Miller, D.R.2
  • 27
    • 0033966742 scopus 로고    scopus 로고
    • Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design
    • Louie RF, Tang Z, Sutton DV, Lee JH, Kost GJ. Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med. 2000;124(2);257-266.
    • (2000) Arch Pathol Lab Med , vol.124 , Issue.2 , pp. 257-266
    • Louie, R.F.1    Tang, Z.2    Sutton, D.V.3    Lee, J.H.4    Kost, G.J.5
  • 28
    • 14944342183 scopus 로고    scopus 로고
    • Capillary versus venous bedside blood glucose estimations
    • Boyd R, Leigh B, Stuart P. Capillary versus venous bedside blood glucose estimations. Emerg Med J. 2005;22(2):177-179.
    • (2005) Emerg Med J , vol.22 , Issue.2 , pp. 177-179
    • Boyd, R.1    Leigh, B.2    Stuart, P.3
  • 29
    • 4644351844 scopus 로고    scopus 로고
    • Correlation of capillary and venous blood glucometry with laboratory determination
    • Kumar G, Sng BL, Kumar S. Correlation of capillary and venous blood glucometry with laboratory determination. Prehosp Emerg Care. 2004;8(4):378-383.
    • (2004) Prehosp Emerg Care , vol.8 , Issue.4 , pp. 378-383
    • Kumar, G.1    Sng, B.L.2    Kumar, S.3
  • 30
    • 3042794161 scopus 로고    scopus 로고
    • College of American Pathologists Comparative analytical costs of central laboratory glucose and bedside glucose testing: a College of American Pathologists Q-Probe study
    • Howanitz P, Jones B; College of American Pathologists. Comparative analytical costs of central laboratory glucose and bedside glucose testing: a College of American Pathologists Q-Probe study. Arch Pathol Lab Med. 2004;128(7): 739-745.
    • (2004) Arch Pathol Lab Med , vol.128 , Issue.7 , pp. 739-745
    • Howanitz, P.1    Jones, B.2
  • 31
    • 35948967141 scopus 로고    scopus 로고
    • Evidenced-based practice is wonderful. . . sort of [editorial]
    • Dracup K, Bryan-Brown C. Evidenced-based practice is wonderful. . . sort of [editorial]. Am J Crit Care. 2006;15: 356-358.
    • (2006) Am J Crit Care , vol.15 , pp. 356-358
    • Dracup, K.1    Bryan-Brown, C.2
  • 32
    • 0027769056 scopus 로고
    • Capillary blood glucose measurements in hospital inpatients using portable glucose meters
    • Phillipou G, Seaborn C, Hooper J, Phillips P. Capillary blood glucose measurements in hospital inpatients using portable glucose meters. Aust N Z J Med. 1993;23:667-671.
    • (1993) Aust N Z J Med , vol.23 , pp. 667-671
    • Phillipou, G.1    Seaborn, C.2    Hooper, J.3    Phillips, P.4
  • 33
    • 0027954883 scopus 로고
    • Intra-operative blood glucose measurements: the effect of haematocrit on glucose test strips
    • Smith EA, Kilpatrick ES. Intra-operative blood glucose measurements: the effect of haematocrit on glucose test strips. Anaesthesia. 1994;49(2):129-132.
    • (1994) Anaesthesia , vol.49 , Issue.2 , pp. 129-132
    • Smith, E.A.1    Kilpatrick, E.S.2
  • 34
    • 0033849382 scopus 로고    scopus 로고
    • Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing
    • Tang Z, Lee JH, Louie RF, Kost GJ. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med. 2000; 124(8):1135-1140.
    • (2000) Arch Pathol Lab Med , vol.124 , Issue.8 , pp. 1135-1140
    • Tang, Z.1    Lee, J.H.2    Louie, R.F.3    Kost, G.J.4
  • 35
    • 0026098817 scopus 로고
    • The effect of haematocrit on reagent strip tests for glucose
    • Wiener K. The effect of haematocrit on reagent strip tests for glucose. Diabet Med. 1991;8(2):172-175.
    • (1991) Diabet Med , vol.8 , Issue.2 , pp. 172-175
    • Wiener, K.1
  • 36
    • 0023888222 scopus 로고
    • Effect of hematocrit concentration on blood glucose value determined on Glucometer II
    • Barreau PB, Buttery JE. Effect of hematocrit concentration on blood glucose value determined on Glucometer II. Diabetes Care. 1988;11(2):116-118.
    • (1988) Diabetes Care , vol.11 , Issue.2 , pp. 116-118
    • Barreau, P.B.1    Buttery, J.E.2
  • 37
    • 0028346374 scopus 로고
    • Variations in sample pH and pO2 affect ExacTech meter glucose measurements
    • Kilpatrick E, Rumley A, Smith A. Variations in sample pH and pO2 affect ExacTech meter glucose measurements. Diabet Med. 1994;11:506-509.
    • (1994) Diabet Med , vol.11 , pp. 506-509
    • Kilpatrick, E.1    Rumley, A.2    Smith, A.3
  • 38
    • 0029196039 scopus 로고
    • Accuracy of fingerstick glucose values in shock patients
    • Sylvain HF, Pokorny ME, English SM, et al. Accuracy of fingerstick glucose values in shock patients. Am J Crit Care. 1995;4(1):44-48.
    • (1995) Am J Crit Care , vol.4 , Issue.1 , pp. 44-48
    • Sylvain, H.F.1    Pokorny, M.E.2    English, S.M.3
  • 39
    • 12944326317 scopus 로고    scopus 로고
    • Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients
    • Kulkarni A, Saxena M, Price G, O'Leary MJ, Jacques T, Myburgh JA. Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med. 2005;31(1):142-145.
    • (2005) Intensive Care Med , vol.31 , Issue.1 , pp. 142-145
    • Kulkarni, A.1    Saxena, M.2    Price, G.3    O'Leary, M.J.4    Jacques, T.5    Myburgh, J.A.6
  • 42
    • 0031938405 scopus 로고    scopus 로고
    • Multicenter study of oxygeninsensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada
    • Kost GJ, Vu H, Lee JH, et al. Multicenter study of oxygeninsensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada. Crit Care Med. 1998;26(3):581-590.
    • (1998) Crit Care Med , vol.26 , Issue.3 , pp. 581-590
    • Kost, G.J.1    Vu, H.2    Lee, J.H.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.