메뉴 건너뛰기




Volumn 91, Issue 3, 2009, Pages 405-414

The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverage

Author keywords

4D CT; AAA; CCC; Density variations; Gating; Lung cancer; Monte Carlo; PBC; Radiotherapy

Indexed keywords

ALGORITHM; ARTICLE; CONTROLLED STUDY; DIAGNOSTIC ACCURACY; DOSIMETRY; LUNG INFLATION; LUNG PARENCHYMA; LUNG TUMOR; LUNG VOLUME; MONTE CARLO METHOD; PHANTOM; PRIORITY JOURNAL; RADIATION DOSE; RADIATION DOSE DISTRIBUTION; STEREOTACTIC BODY RADIATION THERAPY; TREATMENT PLANNING; TUMOR VOLUME;

EID: 65649096221     PISSN: 01678140     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.radonc.2009.01.008     Document Type: Article
Times cited : (135)

References (66)
  • 1
    • 0037100055 scopus 로고    scopus 로고
    • Evaluation of deep inspiration breath-hold lung treatment plans with Monte Carlo dose calculation
    • Yorke E.D., Wang L., Rosenzweig K.E., Mah D., Paoli J.B., and Chui C.S. Evaluation of deep inspiration breath-hold lung treatment plans with Monte Carlo dose calculation. Int J Radiat Oncol Biol Phys 53 (2002) 1058-1070
    • (2002) Int J Radiat Oncol Biol Phys , vol.53 , pp. 1058-1070
    • Yorke, E.D.1    Wang, L.2    Rosenzweig, K.E.3    Mah, D.4    Paoli, J.B.5    Chui, C.S.6
  • 2
    • 17844385102 scopus 로고    scopus 로고
    • Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking
    • Keall P.J., Joshi S., Vedam S.S., Siebers J.V., Kini V.R., and Mohan R. Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking. Med Phys 32 (2005) 942-951
    • (2005) Med Phys , vol.32 , pp. 942-951
    • Keall, P.J.1    Joshi, S.2    Vedam, S.S.3    Siebers, J.V.4    Kini, V.R.5    Mohan, R.6
  • 3
    • 0018964135 scopus 로고
    • Density patterns in the normal lung as determined by computed tomography
    • Rosenblum L.J., Mauceri R.A., Wellenstein D.E., et al. Density patterns in the normal lung as determined by computed tomography. Radiology 137 (1980) 409-416
    • (1980) Radiology , vol.137 , pp. 409-416
    • Rosenblum, L.J.1    Mauceri, R.A.2    Wellenstein, D.E.3
  • 4
    • 15744398473 scopus 로고    scopus 로고
    • Comparison of inhomogeneity correction algorithms in small photon fields
    • Jones A.O., and Das I.J. Comparison of inhomogeneity correction algorithms in small photon fields. Med Phys 32 (2005) 766-776
    • (2005) Med Phys , vol.32 , pp. 766-776
    • Jones, A.O.1    Das, I.J.2
  • 5
    • 7244223251 scopus 로고    scopus 로고
    • Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium
    • Carrasco P., Jornet N., Duch M.A., et al. Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium. Med Phys 31 (2004) 2899-2911
    • (2004) Med Phys , vol.31 , pp. 2899-2911
    • Carrasco, P.1    Jornet, N.2    Duch, M.A.3
  • 7
    • 0035049385 scopus 로고    scopus 로고
    • Comparison of the Batho, ETAR and Monte Carlo dose calculation methods in CT based patient models
    • du Plessis F.C.P., Willemse C.A., Lötter M.G., and Goedhals L. Comparison of the Batho, ETAR and Monte Carlo dose calculation methods in CT based patient models. Med Phys 28 (2001) 582-589
    • (2001) Med Phys , vol.28 , pp. 582-589
    • du Plessis, F.C.P.1    Willemse, C.A.2    Lötter, M.G.3    Goedhals, L.4
  • 8
    • 16444375876 scopus 로고    scopus 로고
    • Monte Carlo versus pencil-beam-/collapsed-cone-dose calculation in a heterogeneous multi-layer phantom
    • Krieger T., and Sauer O.A. Monte Carlo versus pencil-beam-/collapsed-cone-dose calculation in a heterogeneous multi-layer phantom. Phys Med Biol 50 (2005) 859-868
    • (2005) Phys Med Biol , vol.50 , pp. 859-868
    • Krieger, T.1    Sauer, O.A.2
  • 9
    • 0346121669 scopus 로고    scopus 로고
    • Using Monte Carlo simulations to commission photon beam output factors - a feasibility study
    • Ding G.X. Using Monte Carlo simulations to commission photon beam output factors - a feasibility study. Phys Med Biol 48 (2003) 3865-3874
    • (2003) Phys Med Biol , vol.48 , pp. 3865-3874
    • Ding, G.X.1
  • 10
    • 33749419163 scopus 로고    scopus 로고
    • Direct aperture optimization for IMRT using Monte Carlo generated beamlets
    • Bergman A.M., Bush K., Milette M.P., Popescu I.A., Otto K., and Duzenli C. Direct aperture optimization for IMRT using Monte Carlo generated beamlets. Med Phys 33 (2006) 3666-3679
    • (2006) Med Phys , vol.33 , pp. 3666-3679
    • Bergman, A.M.1    Bush, K.2    Milette, M.P.3    Popescu, I.A.4    Otto, K.5    Duzenli, C.6
  • 11
    • 34247337632 scopus 로고    scopus 로고
    • Evaluation of the analytical anisotropic algorithm in an extreme water-lung interface phantom using Monte Carlo dose calculations
    • Gagne I.M., and Zavgorodni S. Evaluation of the analytical anisotropic algorithm in an extreme water-lung interface phantom using Monte Carlo dose calculations. J Appl Clin Med Phys 8 (2007) 33-46
    • (2007) J Appl Clin Med Phys , vol.8 , pp. 33-46
    • Gagne, I.M.1    Zavgorodni, S.2
  • 12
    • 4644325556 scopus 로고    scopus 로고
    • An investigation of accelerator head scatter and output factor in air
    • Ding G.X. An investigation of accelerator head scatter and output factor in air. Med Phys 31 (2004) 2527-2533
    • (2004) Med Phys , vol.31 , pp. 2527-2533
    • Ding, G.X.1
  • 13
    • 0032822016 scopus 로고    scopus 로고
    • Condensed-history monte-carlo simulation for charged particles: what can it do for us?
    • Nahum A.E. Condensed-history monte-carlo simulation for charged particles: what can it do for us?. Radiat Environ Biophys 38 (1999) 163-173
    • (1999) Radiat Environ Biophys , vol.38 , pp. 163-173
    • Nahum, A.E.1
  • 14
    • 0025730909 scopus 로고
    • Monte Carlo techniques in medical radiation physics
    • Andreo P. Monte Carlo techniques in medical radiation physics. Phys Med Biol 36 (1991) 861-920
    • (1991) Phys Med Biol , vol.36 , pp. 861-920
    • Andreo, P.1
  • 15
    • 65649124345 scopus 로고    scopus 로고
    • Monte-Carlo based patient dose computation
    • Mayles P., Nahum A., and Rosenwald J.-C. (Eds), Taylor & Francis [chapter 28]
    • Nahum A. Monte-Carlo based patient dose computation. In: Mayles P., Nahum A., and Rosenwald J.-C. (Eds). Handbook of radiotherapy physics - theory and practice (2007), Taylor & Francis 603-620 [chapter 28]
    • (2007) Handbook of radiotherapy physics - theory and practice , pp. 603-620
    • Nahum, A.1
  • 17
    • 11144340904 scopus 로고    scopus 로고
    • Influence of calculation model on dose distribution in stereotactic radiotherapy for pulmonary targets
    • Haedinger U., Krieger T., Flentje M., and Wulf J. Influence of calculation model on dose distribution in stereotactic radiotherapy for pulmonary targets. Int J Radiat Oncol Biol Phys 61 (2005) 239-249
    • (2005) Int J Radiat Oncol Biol Phys , vol.61 , pp. 239-249
    • Haedinger, U.1    Krieger, T.2    Flentje, M.3    Wulf, J.4
  • 18
    • 16444381290 scopus 로고    scopus 로고
    • The influence of beam model differences in the comparison of dose calculation algorithms for lung cancer treatment planning
    • Chetty I.J., Rosu M., McShan D.L., Fraass B.A., and Ten Haken R.K. The influence of beam model differences in the comparison of dose calculation algorithms for lung cancer treatment planning. Phys Med Biol 50 (2005) 801-815
    • (2005) Phys Med Biol , vol.50 , pp. 801-815
    • Chetty, I.J.1    Rosu, M.2    McShan, D.L.3    Fraass, B.A.4    Ten Haken, R.K.5
  • 19
    • 0029064833 scopus 로고
    • Limitations of a pencil beam approach to photon dose calculations in lung tissue
    • Knöös T., Ahnesjö A., Nilsson P., and Weber L. Limitations of a pencil beam approach to photon dose calculations in lung tissue. Phys Med Biol 40 (1995) 1411-1420
    • (1995) Phys Med Biol , vol.40 , pp. 1411-1420
    • Knöös, T.1    Ahnesjö, A.2    Nilsson, P.3    Weber, L.4
  • 20
    • 0034892226 scopus 로고    scopus 로고
    • Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours
    • Engelsman M., Damen E.M., Koken P.W., van't Veld A.A., Ingen K.M., and Mijnheer B.J. Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours. Radiother Oncol 60 (2001) 299-309
    • (2001) Radiother Oncol , vol.60 , pp. 299-309
    • Engelsman, M.1    Damen, E.M.2    Koken, P.W.3    van't Veld, A.A.4    Ingen, K.M.5    Mijnheer, B.J.6
  • 21
    • 33750574559 scopus 로고    scopus 로고
    • Testing of the analytical anisotropic algorithm for photon dose calculation
    • Van Esch A., Tillikainen L., Pyykkonen J., et al. Testing of the analytical anisotropic algorithm for photon dose calculation. Med Phys 33 (2006) 4130-4148
    • (2006) Med Phys , vol.33 , pp. 4130-4148
    • Van Esch, A.1    Tillikainen, L.2    Pyykkonen, J.3
  • 22
    • 33846882623 scopus 로고    scopus 로고
    • Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations
    • Knöös T., Wieslander E., Cozzi L., et al. Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations. Phys Med Biol 51 (2006) 5785-5807
    • (2006) Phys Med Biol , vol.51 , pp. 5785-5807
    • Knöös, T.1    Wieslander, E.2    Cozzi, L.3
  • 23
    • 33748881814 scopus 로고    scopus 로고
    • Dose distributions in sbrt of lung tumors: comparison between two different treatment planning algorithms and monte-carlo simulation including breathing motions
    • Lax I., Panettieri V., Wennberg B., et al. Dose distributions in sbrt of lung tumors: comparison between two different treatment planning algorithms and monte-carlo simulation including breathing motions. Acta Oncol 45 (2006) 978-988
    • (2006) Acta Oncol , vol.45 , pp. 978-988
    • Lax, I.1    Panettieri, V.2    Wennberg, B.3
  • 24
    • 34547331930 scopus 로고    scopus 로고
    • Sbrt of lung tumours: Monte Carlo simulation with penelope of dose distributions including respiratory motion and comparison with different treatment planning systems
    • Panettieri V., Wennberg B., Gagliardi G., Duch M.A., Ginjaume M., and Lax I. Sbrt of lung tumours: Monte Carlo simulation with penelope of dose distributions including respiratory motion and comparison with different treatment planning systems. Phys Med Biol 52 (2007) 4265-4281
    • (2007) Phys Med Biol , vol.52 , pp. 4265-4281
    • Panettieri, V.1    Wennberg, B.2    Gagliardi, G.3    Duch, M.A.4    Ginjaume, M.5    Lax, I.6
  • 25
    • 34247207650 scopus 로고    scopus 로고
    • On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculations
    • Fogliata A., Vanetti E., Albers D., et al. On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculations. Phys Med Biol 52 (2007) 1363-1385
    • (2007) Phys Med Biol , vol.52 , pp. 1363-1385
    • Fogliata, A.1    Vanetti, E.2    Albers, D.3
  • 26
  • 27
    • 33846256274 scopus 로고    scopus 로고
    • Comparison of 6 MV and 18 MV photons for IMRT treatment of lung cancer
    • Madani I., Vanderstraeten B., Bral S., et al. Comparison of 6 MV and 18 MV photons for IMRT treatment of lung cancer. Radiother Oncol 82 (2007) 63-69
    • (2007) Radiother Oncol , vol.82 , pp. 63-69
    • Madani, I.1    Vanderstraeten, B.2    Bral, S.3
  • 28
    • 0242321252 scopus 로고    scopus 로고
    • Incorporating an improved dose-calculation algorithm in conformal radiotherapy of lung cancer: re-evaluation of dose in normal lung tissue
    • De Jaeger K., Hoogeman M.S., Engelsman M., et al. Incorporating an improved dose-calculation algorithm in conformal radiotherapy of lung cancer: re-evaluation of dose in normal lung tissue. Radiother Oncol 69 (2003) 1-10
    • (2003) Radiother Oncol , vol.69 , pp. 1-10
    • De Jaeger, K.1    Hoogeman, M.S.2    Engelsman, M.3
  • 30
    • 33749826650 scopus 로고    scopus 로고
    • Dose evaluation in lung-equivalent media in high-energy photon external radiotherapy
    • Duch M.A., Carrasco P., Ginjaume M., Jornet N., Ortega X., and Ribas M. Dose evaluation in lung-equivalent media in high-energy photon external radiotherapy. Radiat Prot Dosim 120 (2006) 43-47
    • (2006) Radiat Prot Dosim , vol.120 , pp. 43-47
    • Duch, M.A.1    Carrasco, P.2    Ginjaume, M.3    Jornet, N.4    Ortega, X.5    Ribas, M.6
  • 31
    • 4644266572 scopus 로고    scopus 로고
    • Dosimetric verification of a commercial collapsed cone algorithm in simulated clinical situations
    • Nisbet A., Beange I., Vollmar H.S., Irvine C., Morgan A., and Thwaites D.I. Dosimetric verification of a commercial collapsed cone algorithm in simulated clinical situations. Radiother Oncol 73 (2004) 79-88
    • (2004) Radiother Oncol , vol.73 , pp. 79-88
    • Nisbet, A.1    Beange, I.2    Vollmar, H.S.3    Irvine, C.4    Morgan, A.5    Thwaites, D.I.6
  • 32
    • 0032724373 scopus 로고    scopus 로고
    • Clinical implementation of a Monte Carlo treatment planning system
    • Ma C.M., Mok E., Kapur A., et al. Clinical implementation of a Monte Carlo treatment planning system. Med Phys 26 (1999) 2133-2143
    • (1999) Med Phys , vol.26 , pp. 2133-2143
    • Ma, C.M.1    Mok, E.2    Kapur, A.3
  • 33
    • 0032978977 scopus 로고    scopus 로고
    • Experimental evaluation of interface doses in the presence of air cavities compared with treatment planning algorithms
    • Shahine B.H., Al-Ghazi M.S.A.L., and El-Khatib E. Experimental evaluation of interface doses in the presence of air cavities compared with treatment planning algorithms. Med Phys 26 (1999) 350-355
    • (1999) Med Phys , vol.26 , pp. 350-355
    • Shahine, B.H.1    Al-Ghazi, M.S.A.L.2    El-Khatib, E.3
  • 35
    • 11944272314 scopus 로고    scopus 로고
    • Influence of calculation algorithm on dose distribution in irradiation of non-small cell lung cancer (nsclc) collapsed cone versus pencil beam
    • Koelbl O., Krieger T., Haedinger U., Sauer O., and Flentje M. Influence of calculation algorithm on dose distribution in irradiation of non-small cell lung cancer (nsclc) collapsed cone versus pencil beam. Strahlenther Onkol 180 (2004) 783-788
    • (2004) Strahlenther Onkol , vol.180 , pp. 783-788
    • Koelbl, O.1    Krieger, T.2    Haedinger, U.3    Sauer, O.4    Flentje, M.5
  • 36
    • 34247599929 scopus 로고    scopus 로고
    • Monte Carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator
    • Sterpin E., Tomsej M., De S.B., Reynaert N., and Vynckier S. Monte Carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator. Med Phys 34 (2007) 1665-1677
    • (2007) Med Phys , vol.34 , pp. 1665-1677
    • Sterpin, E.1    Tomsej, M.2    De, S.B.3    Reynaert, N.4    Vynckier, S.5
  • 37
    • 42449161540 scopus 로고    scopus 로고
    • The impact of photon dose calculation algorithms on expected dose distributions in lungs under different respiratory phases
    • Fogliata A., Nicolini G., Vanetti E., Clivio A., Winkler P., and Cozzi L. The impact of photon dose calculation algorithms on expected dose distributions in lungs under different respiratory phases. Phys Med Biol 53 (2008) 2375-2390
    • (2008) Phys Med Biol , vol.53 , pp. 2375-2390
    • Fogliata, A.1    Nicolini, G.2    Vanetti, E.3    Clivio, A.4    Winkler, P.5    Cozzi, L.6
  • 38
    • 39149092356 scopus 로고    scopus 로고
    • Clinical implications of the Anisotropic Analytical Algorithm for IMRT treatment planning and verification
    • Bragg C.M., Wingate K., and Conway J. Clinical implications of the Anisotropic Analytical Algorithm for IMRT treatment planning and verification. Radiother Oncol 86 (2008) 276-284
    • (2008) Radiother Oncol , vol.86 , pp. 276-284
    • Bragg, C.M.1    Wingate, K.2    Conway, J.3
  • 39
    • 38349049217 scopus 로고    scopus 로고
    • Clinical implications of the implementation of advanced treatment planning algorithms for thoracic treatments
    • Morgan A.M., Knöös T., McNee S.G., Evans C.J., and Thwaites D.I. Clinical implications of the implementation of advanced treatment planning algorithms for thoracic treatments. Radiother Oncol 86 (2008) 48-54
    • (2008) Radiother Oncol , vol.86 , pp. 48-54
    • Morgan, A.M.1    Knöös, T.2    McNee, S.G.3    Evans, C.J.4    Thwaites, D.I.5
  • 40
    • 52049120799 scopus 로고    scopus 로고
    • Transition from a simple to a more advanced dose calculation algorithm for radiotherapy of non-small cell lung cancer (NSCLC): implications for clinical implementation in an individualized dose-escalation protocol
    • Elmpt W., Öllers M., Velders M., et al. Transition from a simple to a more advanced dose calculation algorithm for radiotherapy of non-small cell lung cancer (NSCLC): implications for clinical implementation in an individualized dose-escalation protocol. Radiother Oncol 88 (2008) 326-334
    • (2008) Radiother Oncol , vol.88 , pp. 326-334
    • Elmpt, W.1    Öllers, M.2    Velders, M.3
  • 41
    • 44449122864 scopus 로고    scopus 로고
    • Accuracy of dose measurements and calculations within and beyond heterogeneous tissues for 6 MV photon fields smaller than 4 cm produced by cyberknife
    • Wilcox E.E., and Daskalov G.M. Accuracy of dose measurements and calculations within and beyond heterogeneous tissues for 6 MV photon fields smaller than 4 cm produced by cyberknife. Med Phys 35 (2008) 2259-2266
    • (2008) Med Phys , vol.35 , pp. 2259-2266
    • Wilcox, E.E.1    Daskalov, G.M.2
  • 42
    • 65649094635 scopus 로고    scopus 로고
    • ICRU. ICRU report 37, stopping powers for electrons and positrons. Bethesda, MD, USA: International Commission on Radiation Units and measurements ICRU.
    • ICRU. ICRU report 37, stopping powers for electrons and positrons. Bethesda, MD, USA: International Commission on Radiation Units and measurements ICRU.
  • 43
    • 65649110659 scopus 로고    scopus 로고
    • Varian Oncology Systems. Planning reference guide for eclipse algorithms. Helsinki, Finland: Varian Medical System Inc.; 2005.
    • Varian Oncology Systems. Planning reference guide for eclipse algorithms. Helsinki, Finland: Varian Medical System Inc.; 2005.
  • 44
    • 0037038880 scopus 로고    scopus 로고
    • A DICOM-RT-based toolbox for the evaluation and verification of radiotherapy plans
    • Spezi E., Lewis D.G., and Smith C.W. A DICOM-RT-based toolbox for the evaluation and verification of radiotherapy plans. Phys Med Biol 47 (2002) 4223-4232
    • (2002) Phys Med Biol , vol.47 , pp. 4223-4232
    • Spezi, E.1    Lewis, D.G.2    Smith, C.W.3
  • 45
    • 0037572347 scopus 로고    scopus 로고
    • CERR: a computational environment for radiotherapy research
    • Deasy J.O., Blanco A.I., and Clark V.H. CERR: a computational environment for radiotherapy research. Med Phys 30 (2003) 979-985
    • (2003) Med Phys , vol.30 , pp. 979-985
    • Deasy, J.O.1    Blanco, A.I.2    Clark, V.H.3
  • 46
    • 0142182110 scopus 로고    scopus 로고
    • Dose perturbation in the presence of metallic implants: treatment planning system versus Monte Carlo simulations
    • Wieslander E., and Knöös T. Dose perturbation in the presence of metallic implants: treatment planning system versus Monte Carlo simulations. Phys Med Biol 48 (2003) 3295-3305
    • (2003) Phys Med Biol , vol.48 , pp. 3295-3305
    • Wieslander, E.1    Knöös, T.2
  • 47
    • 31844437176 scopus 로고    scopus 로고
    • A direct voxel tracking method for four-dimensional Monte Carlo dose calculations in deforming anatomy
    • Heath E., and Seuntjens J. A direct voxel tracking method for four-dimensional Monte Carlo dose calculations in deforming anatomy. Med Phys 33 (2006) 434-445
    • (2006) Med Phys , vol.33 , pp. 434-445
    • Heath, E.1    Seuntjens, J.2
  • 48
    • 0032785775 scopus 로고    scopus 로고
    • Calculation of a pencil beam kernel from measured photon beam data
    • Storchi P.R., Van Battum L.J., and Woudstra E. Calculation of a pencil beam kernel from measured photon beam data. Phys Med Biol 44 (1999) 2917-2928
    • (1999) Phys Med Biol , vol.44 , pp. 2917-2928
    • Storchi, P.R.1    Van Battum, L.J.2    Woudstra, E.3
  • 49
    • 22844450224 scopus 로고    scopus 로고
    • A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations
    • Ulmer W., Pyyry J., and Kaissl W. A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations. Phys Med Biol 50 (2005) 1767-1790
    • (2005) Phys Med Biol , vol.50 , pp. 1767-1790
    • Ulmer, W.1    Pyyry, J.2    Kaissl, W.3
  • 50
    • 47649106110 scopus 로고    scopus 로고
    • A 3D pencil-beam-based superposition algorithm for photon dose calculation in heterogeneous media
    • Tillikainen L., Helminen H., Torsti T., et al. A 3D pencil-beam-based superposition algorithm for photon dose calculation in heterogeneous media. Phys Med Biol 53 (2008) 3821-3839
    • (2008) Phys Med Biol , vol.53 , pp. 3821-3839
    • Tillikainen, L.1    Helminen, H.2    Torsti, T.3
  • 51
    • 0034104708 scopus 로고    scopus 로고
    • Accurate condensed history Monte Carlo simulation of electron transport. i. EGSnrc, the new EGS4 version
    • Kawrakow I. Accurate condensed history Monte Carlo simulation of electron transport. i. EGSnrc, the new EGS4 version. Med Phys 27 (2000) 485-498
    • (2000) Med Phys , vol.27 , pp. 485-498
    • Kawrakow, I.1
  • 52
    • 0034023628 scopus 로고    scopus 로고
    • Accurate condensed history Monte Carlo simulation of electron transport. ii. Application to ion chamber response simulations
    • Kawrakow I. Accurate condensed history Monte Carlo simulation of electron transport. ii. Application to ion chamber response simulations. Med Phys 27 (2000) 499-513
    • (2000) Med Phys , vol.27 , pp. 499-513
    • Kawrakow, I.1
  • 53
    • 0003797114 scopus 로고    scopus 로고
    • The EGSnrc Code System: Monte Carlo simulation of electron and photon transport
    • PIRS-701. Ionizing Radiation Standards National Research Council of Canada
    • Kawrakow I, Rogers D. The EGSnrc Code System: Monte Carlo simulation of electron and photon transport, NRCC report PIRS-701. Ionizing Radiation Standards National Research Council of Canada.
    • NRCC report
    • Kawrakow, I.1    Rogers, D.2
  • 54
    • 0038644818 scopus 로고    scopus 로고
    • DOSXYZnrc users manual
    • PIRS-794revB. Ionizing Radiation Standards National Research Council of Canada
    • Walters B, Kawrakow I, Rogers D. DOSXYZnrc users manual, NRCC report PIRS-794revB. Ionizing Radiation Standards National Research Council of Canada.
    • NRCC report
    • Walters, B.1    Kawrakow, I.2    Rogers, D.3
  • 55
    • 0026750950 scopus 로고
    • A pencil beam model for photon dose calculation
    • Ahnesjö A., Saxner M., and Trepp A. A pencil beam model for photon dose calculation. Med Phys 19 (1992) 263-273
    • (1992) Med Phys , vol.19 , pp. 263-273
    • Ahnesjö, A.1    Saxner, M.2    Trepp, A.3
  • 57
    • 0024424112 scopus 로고
    • Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media
    • Ahnesjö A. Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media. Med Phys 16 (1989) 577-592
    • (1989) Med Phys , vol.16 , pp. 577-592
    • Ahnesjö, A.1
  • 58
    • 65649086615 scopus 로고    scopus 로고
    • Patient dose computation for photon beams
    • Mayles P., Nahum A., and Rosenwald J.-C. (Eds), Taylor & Francis [chapter 26]
    • Rosenwald J.C., Rosenberg I., and Shentall G. Patient dose computation for photon beams. In: Mayles P., Nahum A., and Rosenwald J.-C. (Eds). Handbook of radiotherapy physics - theory and practice (2007), Taylor & Francis 559-585 [chapter 26]
    • (2007) Handbook of radiotherapy physics - theory and practice , pp. 559-585
    • Rosenwald, J.C.1    Rosenberg, I.2    Shentall, G.3
  • 59
    • 33845220491 scopus 로고    scopus 로고
    • Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning
    • Bragg C.M., and Conway J. Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning. Radiother Oncol 81 (2006) 315-323
    • (2006) Radiother Oncol , vol.81 , pp. 315-323
    • Bragg, C.M.1    Conway, J.2
  • 60
    • 34247211685 scopus 로고    scopus 로고
    • Determination of parameters for a multiple-source model of megavoltage photon beams using optimization methods
    • Tillikainen L., Siljamäki S., Helminen H., Alakuijala J., and Pyyry J. Determination of parameters for a multiple-source model of megavoltage photon beams using optimization methods. Phys Med Biol 52 (2007) 1441-1467
    • (2007) Phys Med Biol , vol.52 , pp. 1441-1467
    • Tillikainen, L.1    Siljamäki, S.2    Helminen, H.3    Alakuijala, J.4    Pyyry, J.5
  • 61
    • 33846609500 scopus 로고    scopus 로고
    • Technical note: overprediction of dose with default PRESTA-I boundary crossing in DOSXYZnrc and BEAMnrc
    • Walters B.R., and Kawrakow I. Technical note: overprediction of dose with default PRESTA-I boundary crossing in DOSXYZnrc and BEAMnrc. Med Phys 34 (2007) 647-650
    • (2007) Med Phys , vol.34 , pp. 647-650
    • Walters, B.R.1    Kawrakow, I.2
  • 62
    • 0004286738 scopus 로고    scopus 로고
    • BEAMnrc users manual
    • PIRS-0509(a)revK. Ionizing Radiation Standards National Research Council of Canada
    • Rogers D, Walters B, Kawrakow I. BEAMnrc users manual. NRCC report PIRS-0509(a)revK. Ionizing Radiation Standards National Research Council of Canada.
    • NRCC report
    • Rogers, D.1    Walters, B.2    Kawrakow, I.3
  • 63
    • 0348227877 scopus 로고    scopus 로고
    • Development and validation of a BEAMnrc component module for accurate Monte Carlo modelling of the varian dynamic millennium multileaf collimator
    • Heath E., and Seuntjens J. Development and validation of a BEAMnrc component module for accurate Monte Carlo modelling of the varian dynamic millennium multileaf collimator. Phys Med Biol 48 (2003) 4045-4063
    • (2003) Phys Med Biol , vol.48 , pp. 4045-4063
    • Heath, E.1    Seuntjens, J.2
  • 64
    • 7244223252 scopus 로고    scopus 로고
    • Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting
    • Kawrakow I., Rogers D.W., and Walters B.R. Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting. Med Phys 31 (2004) 2883-2898
    • (2004) Med Phys , vol.31 , pp. 2883-2898
    • Kawrakow, I.1    Rogers, D.W.2    Walters, B.R.3
  • 65
    • 0034036210 scopus 로고    scopus 로고
    • Converting absorbed dose to medium to absorbed dose to water for monte carlo based photon beam dose calculations
    • Siebers J.V., Keall P.J., Nahum A.E., and Mohan R. Converting absorbed dose to medium to absorbed dose to water for monte carlo based photon beam dose calculations. Phys Med Biol 45 (2000) 983-995
    • (2000) Phys Med Biol , vol.45 , pp. 983-995
    • Siebers, J.V.1    Keall, P.J.2    Nahum, A.E.3    Mohan, R.4
  • 66
    • 0034851923 scopus 로고    scopus 로고
    • Tumour control probability modelling: basic principles and applications in treatment planning
    • Nahum A.E., and Sanchez-Nieto B. Tumour control probability modelling: basic principles and applications in treatment planning. Phys Med 17 Suppl. 2 (2001) 13-23
    • (2001) Phys Med , vol.17 , Issue.SUPPL. 2 , pp. 13-23
    • Nahum, A.E.1    Sanchez-Nieto, B.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.