메뉴 건너뛰기




Volumn 9, Issue 3, 2008, Pages 90-98

Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film

Author keywords

Dosimetry; GafChromic; IMRT; MLC penumbra; MLC transmission; mMLC

Indexed keywords


EID: 57049150104     PISSN: None     EISSN: 15269914     Source Type: Journal    
DOI: 10.1120/jacmp.v9i3.2802     Document Type: Article
Times cited : (44)

References (20)
  • 1
    • 0032951630 scopus 로고    scopus 로고
    • Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery
    • Cosgrove VP, Jahn U, Pfaender M, Bauer S, Budach V, Wurm RE. Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery. Radiother Oncol. 1999;50(3):325-336.
    • (1999) Radiother Oncol , vol.50 , Issue.3 , pp. 325-336
    • Cosgrove, V.P.1    Jahn, U.2    Pfaender, M.3    Bauer, S.4    Budach, V.5    Wurm, R.E.6
  • 2
    • 16444368473 scopus 로고    scopus 로고
    • Development of a Monte Carlo model for the Brainlab microMLC
    • Belec J, Patrocinio H, Verhaegen F. Development of a Monte Carlo model for the Brainlab microMLC. Phys Med Biol. 2005;50(5):787-799.
    • (2005) Phys Med Biol , vol.50 , Issue.5 , pp. 787-799
    • Belec, J.1    Patrocinio, H.2    Verhaegen, F.3
  • 3
    • 0036917861 scopus 로고    scopus 로고
    • A Monte Carlo based phase space model for quality assurance of intensity modulated radiotherapy incorporating leaf specific characteristics
    • Aaronson RF, DeMarco JJ, Chetty IJ, Solberg TD. A Monte Carlo based phase space model for quality assurance of intensity modulated radiotherapy incorporating leaf specific characteristics. Med Phys. 2002;29(12):2952-2958.
    • (2002) Med Phys , vol.29 , Issue.12 , pp. 2952-2958
    • Aaronson, R.F.1    DeMarco, J.J.2    Chetty, I.J.3    Solberg, T.D.4
  • 5
    • 33645394174 scopus 로고    scopus 로고
    • Commissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT
    • Wang L, Li J, Paskalev K, et al. Commissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT. J Appl Clin Med Phys. 2006;7(1):21-34.
    • (2006) J Appl Clin Med Phys , vol.7 , Issue.1 , pp. 21-34
    • Wang, L.1    Li, J.2    Paskalev, K.3
  • 6
    • 0038030868 scopus 로고    scopus 로고
    • Commissioning of mini-multi-leaf-collimator (MMLC) for stereotactic radiosurgery and radiotherapy
    • Mardirossian G, Urie M, Fitzgerald TJ, Mayr N, Montebello J, Lo YC. Commissioning of mini-multi-leaf-collimator (MMLC) for stereotactic radiosurgery and radiotherapy. J X Ray Sci Technol. 2003;11(1):21-31.
    • (2003) J X Ray Sci Technol , vol.11 , Issue.1 , pp. 21-31
    • Mardirossian, G.1    Urie, M.2    Fitzgerald, T.J.3    Mayr, N.4    Montebello, J.5    Lo, Y.C.6
  • 7
    • 33746611132 scopus 로고    scopus 로고
    • Measurement of high energy x-ray beam penumbra with Gafchromic EBT radiochromic film
    • Cheung T, Butson MJ, Yu PKM. Measurement of high energy x-ray beam penumbra with Gafchromic EBT radiochromic film. Med Phys. 2006;33(8):2912-2914.
    • (2006) Med Phys , vol.33 , Issue.8 , pp. 2912-2914
    • Cheung, T.1    Butson, M.J.2    Yu, P.K.M.3
  • 8
    • 0033795118 scopus 로고    scopus 로고
    • The value of the PinPoint ionization chamber for characterization of small field segments used in intensity-modulated radiotherapy
    • Martens C, Wagter C De, Neve W De. The value of the PinPoint ionization chamber for characterization of small field segments used in intensity-modulated radiotherapy. Phys Med Biol. 2000;45(9):2519-2530.
    • (2000) Phys Med Biol , vol.45 , Issue.9 , pp. 2519-2530
    • Martens, C.1    Wagter, C.D.2    Neve, W.D.3
  • 9
    • 34447321620 scopus 로고    scopus 로고
    • Evaluation of GafChromic EBT film for CyberKnife dosimetry
    • Wilcox EE, Daskalov GM. Evaluation of GafChromic EBT film for CyberKnife dosimetry. Med Phys. 2007;34(6):1967-1974.
    • (2007) Med Phys , vol.34 , Issue.6 , pp. 1967-1974
    • Wilcox, E.E.1    Daskalov, G.M.2
  • 10
    • 0008796868 scopus 로고    scopus 로고
    • Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55
    • Niroomand-Rad A, Blackwell CR, Coursey BM, et al. Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55. Med Phys. 1998;25(11):2093-2115.
    • (1998) Med Phys , vol.25 , Issue.11 , pp. 2093-2115
    • Niroomand-Rad, A.1    Blackwell, C.R.2    Coursey, B.M.3
  • 11
    • 33646461862 scopus 로고    scopus 로고
    • Evaluation of GafChromic EBT prototype B for external beam dose verification
    • Todorovic M, Fischer M, Cremers F, Thom E, Schmidt R. Evaluation of GafChromic EBT prototype B for external beam dose verification. Med Phys. 2006;33(5):1321-1328.
    • (2006) Med Phys , vol.33 , Issue.5 , pp. 1321-1328
    • Todorovic, M.1    Fischer, M.2    Cremers, F.3    Thom, E.4    Schmidt, R.5
  • 12
    • 21344465230 scopus 로고    scopus 로고
    • Absorption spectra variations of EBT radiochromic film radiation exposure
    • Butson MJ, Cheung T, Yu PK. Absorption spectra variations of EBT radiochromic film radiation exposure. Phys Med Biol. 2005;50(13):N135-N140.
    • (2005) Phys Med Biol , vol.50 , Issue.13
    • Butson, M.J.1    Cheung, T.2    Yu, P.K.3
  • 13
    • 33846582373 scopus 로고    scopus 로고
    • Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter
    • Rink A, Vitkin IA, Jaffray DA. Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter. Med Phys. 2007;34(2):458-463.
    • (2007) Med Phys , vol.34 , Issue.2 , pp. 458-463
    • Rink, A.1    Vitkin, I.A.2    Jaffray, D.A.3
  • 14
    • 33750197979 scopus 로고    scopus 로고
    • Scanning orientation effects on GafChromic EBT film dosimetry
    • Butson MJ, Cheung T, Yu PK. Scanning orientation effects on GafChromic EBT film dosimetry. Australas Phys Eng Sci Med. 2006;29(3):281-284.
    • (2006) Australas Phys Eng Sci Med , vol.29 , Issue.3 , pp. 281-284
    • Butson, M.J.1    Cheung, T.2    Yu, P.K.3
  • 15
    • 85024366896 scopus 로고    scopus 로고
    • Bethesda, Maryland, USA
    • Rasband WS. ImageJ, U. S. National Institutes of Health, Bethesda, Maryland, USA, http://rsb.info.nih.gov/ij/, 1997-2007.
    • (1997)
    • Rasband, W.S.1    ImageJ, U.S.2
  • 17
    • 4644295382 scopus 로고    scopus 로고
    • Dosimetric properties of improved GafChromic films for seven different digitizers
    • Devic S, Seuntjens J, Hegyl G, et al. Dosimetric properties of improved GafChromic films for seven different digitizers. Med Phys. 2004;31(9):2392-2401.
    • (2004) Med Phys , vol.31 , Issue.9 , pp. 2392-2401
    • Devic, S.1    Seuntjens, J.2    Hegyl, G.3
  • 18
    • 60449113534 scopus 로고    scopus 로고
    • International Specialty Products, White Paper [Electronic version, 2007
    • International Specialty Products, GafChromic EBT Self-Developing Film for Radiotherapy Dosimetry, White Paper [Electronic version], 2007, http://www.ispcorp.com/products/dosimetry/content/gafchromic/content/ products/ebt/pdfs/EBTwhitepaper.pdf
    • Self-Developing Film for Radiotherapy Dosimetry
    • GafChromic, E.B.T.1
  • 20
    • 22544440841 scopus 로고    scopus 로고
    • Precise radiochromic film dosimetry using a flat-bed document scanner
    • Devic S, Seuntjens J, Sham E, Podgorsak EB, et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys. 2005;32(7):2245-2253.
    • (2005) Med Phys , vol.32 , Issue.7 , pp. 2245-2253
    • Devic, S.1    Seuntjens, J.2    Sham, E.3    Podgorsak, E.B.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.