메뉴 건너뛰기




Volumn 35, Issue 12, 2008, Pages 5360-5365

Determination of absorbed dose in water at the reference point D (r 0, θ0) for an 192Ir HDR brachytherapy source using a Fricke system

Author keywords

192Ir; Audit; Dose inter comparison; Fricke dosimetry; HDR; Radial dose; Well type chamber

Indexed keywords

CYLINDERS (SHAPES); IRRADIATION; ONCOLOGY; RADIOACTIVE PROSPECTING; RADIOACTIVITY; RADIOTHERAPY; RATE CONSTANTS; UNCERTAINTY ANALYSIS;

EID: 56749158507     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.2996178     Document Type: Article
Times cited : (18)

References (30)
  • 1
    • 0028919251 scopus 로고
    • Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine
    • R. Nath, L. L. Anderson, G. Luxton, K. A. Weaver, J. F. Williamson, and A. S. Meigooni, " Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.," Med. Phys. 22, 209-234 (1995).
    • (1995) Med. Phys. , vol.22 , pp. 209-234
    • Nath, R.1    Anderson, L.L.2    Luxton, G.3    Weaver, K.A.4    Williamson, J.F.5    Meigooni, A.S.6
  • 3
    • 0025922323 scopus 로고
    • Calibration of Ir192 high-dose-rate afterloading systems
    • S. J. Goetsch, F. H. Attix, D. W. Pearson, and B. R. Thomadsen, " Calibration of Ir192 high-dose-rate afterloading systems.," Med. Phys. 18, 462-467 (1991).
    • (1991) Med. Phys. , vol.18 , pp. 462-467
    • Goetsch, S.J.1    Attix, F.H.2    Pearson, D.W.3    Thomadsen, B.R.4
  • 4
    • 0035996434 scopus 로고    scopus 로고
    • Calibration of new high dose rate Ir192 sources
    • K. E. Stump, L. A. Dewerd, J. A. Micka, and D. R. Anderson, " Calibration of new high dose rate Ir192 sources.," Med. Phys. 29, 1483-1488 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 1483-1488
    • Stump, K.E.1    Dewerd, L.A.2    Micka, J.A.3    Anderson, D.R.4
  • 5
    • 33748294452 scopus 로고    scopus 로고
    • On the accuracy of techniques for obtaining the calibration coefficient N(K) of Ir192 HDR brachytherapy sources
    • E. Mainegra-Hing and D. W. Rogers, " On the accuracy of techniques for obtaining the calibration coefficient N(K) of Ir192 HDR brachytherapy sources.," Med. Phys. 33, 3340-3347 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 3340-3347
    • Mainegra-Hing, E.1    Rogers, D.W.2
  • 7
    • 36649026668 scopus 로고    scopus 로고
    • An absorbed dose to water standard for HDR Ir192 brachytherapy sources based on water calorimetry: Numerical and experimental proof-of-principle
    • A. Sarfehnia, K. Stewart, and J. Seuntjens, " An absorbed dose to water standard for HDR Ir192 brachytherapy sources based on water calorimetry: numerical and experimental proof-of-principle.," Med. Phys. 34, 4957-4961 (2007).
    • (2007) Med. Phys. , vol.34 , pp. 4957-4961
    • Sarfehnia, A.1    Stewart, K.2    Seuntjens, J.3
  • 8
    • 0008700422 scopus 로고
    • Status of the primary standard of water absorbed dose for high energy photon and electron radiation at the PTB
    • (IAEA, Vienna)
    • M. Roos and K. Hohlfeld, " Status of the primary standard of water absorbed dose for high energy photon and electron radiation at the PTB.," Measurement Assurance in Dosimetry (IAEA, Vienna, 1994), pp. 25-33.
    • (1994) Measurement Assurance in Dosimetry , pp. 25-33
    • Roos, M.1    Hohlfeld, K.2
  • 9
    • 0034879566 scopus 로고    scopus 로고
    • Comparison of dosimetric standards of Canada and France for photons at 60Co and higher energies
    • K. Shortt, C. Ross, J. Seuntjens, F. Delaunay, A. Ostrowsky, P. Gross, and E. Leroy, " Comparison of dosimetric standards of Canada and France for photons at 60Co and higher energies.," Phys. Med. Biol. 46, 2119-2142 (2001).
    • (2001) Phys. Med. Biol. , vol.46 , pp. 2119-2142
    • Shortt, K.1    Ross, C.2    Seuntjens, J.3    Delaunay, F.4    Ostrowsky, A.5    Gross, P.6    Leroy, E.7
  • 10
    • 0033856077 scopus 로고    scopus 로고
    • Comparison of dosimetry calibration factors at the NRCC and the NIST. National Research Council of Canada. National Institute of Standards and Technology
    • K. Shortt, J. Shobe, and S. Domen, " Comparison of dosimetry calibration factors at the NRCC and the NIST. National Research Council of Canada. National Institute of Standards and Technology.," Med. Phys. 27, 1644-1654 (2000).
    • (2000) Med. Phys. , vol.27 , pp. 1644-1654
    • Shortt, K.1    Shobe, J.2    Domen, S.3
  • 11
    • 34147174813 scopus 로고    scopus 로고
    • Quality assurance of HDR Ir192 sources using a Fricke dosimeter
    • C. Austerlitz, H. Mota, C. E. deAlmeida, R. Allison, and C. Sibata, " Quality assurance of HDR Ir192 sources using a Fricke dosimeter.," Med. Phys. 34, 1348-1353 (2007).
    • (2007) Med. Phys. , vol.34 , pp. 1348-1353
    • Austerlitz, C.1    Mota, H.2    Dealmeida, C.E.3    Allison, R.4    Sibata, C.5
  • 13
    • 33744798255 scopus 로고    scopus 로고
    • Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, Ir192, 103Pd, and 169Yb sources
    • C. S. Melhus and M. J. Rivard, " Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, Ir192, 103Pd, and 169Yb sources.," Med. Phys. 33, 1729-1737 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 1729-1737
    • Melhus, C.S.1    Rivard, M.J.2
  • 15
    • 23844508954 scopus 로고    scopus 로고
    • Measurement of dosimetric parameters for the Alpha-Omega high-dose-rate iridium-192 source
    • R. Muller-Runkel, " Measurement of dosimetric parameters for the Alpha-Omega high-dose-rate iridium-192 source.," Med. Dosim. 30, 139-142 (2005).
    • (2005) Med. Dosim. , vol.30 , pp. 139-142
    • Muller-Runkel, R.1
  • 19
    • 0032800444 scopus 로고    scopus 로고
    • Fricke dosimetry: The difference between G (Fe3+) for 60Co gamma-rays and high-energy x-rays
    • N. V. Klassen, K. R. Shortt, J. Seuntjens, and C. K. Ross, " Fricke dosimetry: the difference between G (Fe3+) for 60Co gamma-rays and high-energy x-rays.," Phys. Med. Biol. 44, 1609-1624 (1999).
    • (1999) Phys. Med. Biol. , vol.44 , pp. 1609-1624
    • Klassen, N.V.1    Shortt, K.R.2    Seuntjens, J.3    Ross, C.K.4
  • 22
    • 0037034652 scopus 로고    scopus 로고
    • Characterization of a high-dose-rate 90Sr-90Y source for intravascular brachytherapy by using the Monte Carlo code PENELOPE
    • J. Asenjo, J. M. Fernandez-Varea, and A. Sanchez-Reyes, " Characterization of a high-dose-rate 90Sr-90Y source for intravascular brachytherapy by using the Monte Carlo code PENELOPE.," Phys. Med. Biol. 47, 697-711 (2002).
    • (2002) Phys. Med. Biol. , vol.47 , pp. 697-711
    • Asenjo, J.1    Fernandez-Varea, J.M.2    Sanchez-Reyes, A.3
  • 23
    • 38149039528 scopus 로고    scopus 로고
    • Monte Carlo simulation of a realistic anatomical phantom described by triangle meshes: Application to prostate brachytherapy imaging
    • A. Badal, I. Kyprianou, A. Badano, and J. Sempau, " Monte Carlo simulation of a realistic anatomical phantom described by triangle meshes: Application to prostate brachytherapy imaging.," Radiother. Oncol. 86, 99-103 (2008).
    • (2008) Radiother. Oncol. , vol.86 , pp. 99-103
    • Badal, A.1    Kyprianou, I.2    Badano, A.3    Sempau, J.4
  • 24
    • 85024954819 scopus 로고    scopus 로고
    • Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
    • M. L. Rodriguez and C. E. deAlmeida, " Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method.," J. Appl. Clin. Med. Phys. 3, 285-292 (2002).
    • (2002) J. Appl. Clin. Med. Phys. , vol.3 , pp. 285-292
    • Rodriguez, M.L.1    Dealmeida, C.E.2
  • 26
    • 1142287542 scopus 로고    scopus 로고
    • Benchmark of PENELOPE code for low-energy photon transport: Dose comparisons with MCNP4 and EGS4
    • S. J. Ye, I. A. Brezovich, P. Pareek, and S. A. Naqvi, " Benchmark of PENELOPE code for low-energy photon transport: dose comparisons with MCNP4 and EGS4.," Phys. Med. Biol. 49, 387-397 (2004).
    • (2004) Phys. Med. Biol. , vol.49 , pp. 387-397
    • Ye, S.J.1    Brezovich, I.A.2    Pareek, P.3    Naqvi, S.A.4
  • 27
    • 0038033849 scopus 로고    scopus 로고
    • The design of a new insert for calibration of LDR and HDR Ir192 sources in a well-type ionization chamber
    • A. E. Pineda, E. P. Alcon, and C. E. deAlmeida, " The design of a new insert for calibration of LDR and HDR Ir192 sources in a well-type ionization chamber.," Med. Phys. 30, 1566-1568 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 1566-1568
    • Pineda, A.E.1    Alcon, E.P.2    Dealmeida, C.E.3
  • 28
    • 33845310036 scopus 로고    scopus 로고
    • Monte Carlo and experimental dosimetry of an 1251 brachytherapy seed
    • J. Dolan, Z. Lia, and J. F. Williamson, " Monte Carlo and experimental dosimetry of an 1251 brachytherapy seed.," Med. Phys. 33, 4675-4684 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 4675-4684
    • Dolan, J.1    Lia, Z.2    Williamson, J.F.3
  • 30
    • 0014454847 scopus 로고
    • G-value determination of the Fricke dosemeter for 50 kVp x-rays
    • J. Geisselsoder and C. J. Karzmark, " G-value determination of the Fricke dosemeter for 50 kVp x-rays.," Phys. Med. Biol. 14, 61-71 (1969).
    • (1969) Phys. Med. Biol. , vol.14 , pp. 61-71
    • Geisselsoder, J.1    Karzmark, C.J.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.