Biventricular implantable cardioverter defibrillator right ventricle pace-sense ring electrode failure: Lead switch fix

Author keywords

Biventricular; Defibrillation ICD; Iead failure; Inappropriate shock

Indexed keywords

AGED; ANAMNESIS; ARTICLE; CARDIOMYOPATHY; CASE REPORT; CLINICAL FEATURE; DEFIBRILLATOR; ELECTRODE; HEART FAILURE; HEART RIGHT VENTRICLE; HUMAN; MALE; THORAX RADIOGRAPHY; ELECTRIC INJURY; ELECTRIC SHOCK; ELECTROCARDIOGRAPHY; EQUIPMENT; HEART VENTRICLE; INSTRUMENTATION;

EID: 55249102770     PISSN: 01478389     EISSN: 15408159     Source Type: Journal    
DOI: 10.1111/j.1540-8159.2008.01107.x     Document Type: Article
Times cited : (3)

References (7)
  • 2
    • 34547952842 scopus 로고    scopus 로고
    • Do small (6.6 Fr.) active and passive fixation defib- rillation leads perform as well as larger sized leads? A multi-centre analysis
    • Epub Jun 26
    • Kupper B, Yee R, O'Hara G, Simpson CS, Della Bella P, Workel M, Hull M, et al. Do small (6.6 Fr.) active and passive fixation defib- rillation leads perform as well as larger sized leads? A multi-centre analysis. Europace 2007; 9:657-661. Epub 2007 Jun 26.
    • (2007) Europace 2007 , vol.9 , pp. 657-661
    • Kupper, B.1    Yee, R.2    O'Hara, G.3    Simpson, C.S.4    Della Bella, P.5    Workel, M.6    Hull, M.7
  • 3
    • 34247186286 scopus 로고    scopus 로고
    • Danik SB, Mansour M, Singh J, Reddy VY, Ellinor PT, Milan D, Heist EK, et al. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator. Heart Rhythm 2007; 4:439-442. Epub 2007 Jan 7.
    • Danik SB, Mansour M, Singh J, Reddy VY, Ellinor PT, Milan D, Heist EK, et al. Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator. Heart Rhythm 2007; 4:439-442. Epub 2007 Jan 7.
  • 4
    • 64549163681 scopus 로고    scopus 로고
    • Urgent Medical Device Information: Sprint Fidelis Lead Patient Management Recommendations
    • October 15
    • Groves R. Urgent Medical Device Information: Sprint Fidelis Lead Patient Management Recommendations. Industry Dear Doctor Com munication October 15, 2007.
    • (2007) Industry Dear Doctor Com munication
    • Groves, R.1
  • 5
    • 33644749613 scopus 로고    scopus 로고
    • Use of a coronary sinus lead and biventricular ICD to correct a sensing abnormal ity in a patient with arrhythmogenic right ventricular dysplasia/cardiomyopathy
    • Bilchick KC, Judge DP, Calkins H, Marine JE. Use of a coronary sinus lead and biventricular ICD to correct a sensing abnormal ity in a patient with arrhythmogenic right ventricular dysplasia/cardiomyopathy. J Cardiovasc Electrophysiol 2006; 17:317-320.
    • (2006) J Cardiovasc Electrophysiol , vol.17 , pp. 317-320
    • Bilchick, K.C.1    Judge, D.P.2    Calkins, H.3    Marine, J.E.4
  • 6
    • 37549063635 scopus 로고    scopus 로고
    • Pacemaker/implantable cardioverter- defibrillator problem: Right ventricular lead mis- sensing in a biventricular implantable cardioverter-defibrillator
    • Otten RF, Foreman LD, Groh WJ. Pacemaker/implantable cardioverter- defibrillator problem: Right ventricular lead mis- sensing in a biventricular implantable cardioverter-defibrillator. Heart Rhythm 2008; 5:158-159.
    • (2008) Heart Rhythm , vol.5 , pp. 158-159
    • Otten, R.F.1    Foreman, L.D.2    Groh, W.J.3
  • 7
    • 64549159345 scopus 로고    scopus 로고
    • Moore H, Xu W, Karasik P, Lewis B, Singh S, Fletcher R. ICD lead failure management strategies. Abstract: Heart Rhythm 2008; 5:PO6-87, S389.
    • Moore H, Xu W, Karasik P, Lewis B, Singh S, Fletcher R. ICD lead failure management strategies. Abstract: Heart Rhythm 2008; 5:PO6-87, S389.


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.