Volumn 43, Issue 6, 2008, Pages 374-381

Effect of x-ray tube parameters, iodine concentration, and patient size on image quality in pulmonary computed tomography angiography: A chest-phantom-study

Author keywords

Chest CT; Contrast to noise ratio; Phantom study; Tube current; Tube energy

Indexed keywords

CONTRAST MEDIUM; IODINE;

EID: 44449089365     PISSN: 00209996     EISSN: None     Source Type: Journal    
DOI: 10.1097/RLI.0b013e3181690042     Document Type: Article
Times cited : (75)

References (24)
  • 1
    • 4143089129 scopus 로고    scopus 로고
    • Abdominal aortic aneurysms at multi-detector row helical CT: Optimization with interactive determination of scanning delay and contrast medium dose
    • Ho LM, Nelson RC, Thomas J, et al. Abdominal aortic aneurysms at multi-detector row helical CT: optimization with interactive determination of scanning delay and contrast medium dose. Radiology. 2004;232:854-859.
    • (2004) Radiology , vol.232 , pp. 854-859
    • Ho, L.M.1    Nelson, R.C.2    Thomas, J.3
  • 2
    • 33644981802 scopus 로고    scopus 로고
    • IV contrast selection for MDCT: Current thoughts and practice
    • Johnson PT, Fishman EK. IV contrast selection for MDCT: current thoughts and practice. AJR Am J Roentgenol. 2006;186:406-415.
    • (2006) AJR Am J Roentgenol , vol.186 , pp. 406-415
    • Johnson, P.T.1    Fishman, E.K.2
  • 3
    • 0034924356 scopus 로고    scopus 로고
    • Infrarenal abdominal aortic aneurysms at multi-detector row CT angiography: Intravascular enhancement without a timing acquisition
    • Macari M, Israel GM, Berman P, et al. Infrarenal abdominal aortic aneurysms at multi-detector row CT angiography: intravascular enhancement without a timing acquisition. Radiology. 2001;220:519-523.
    • (2001) Radiology , vol.220 , pp. 519-523
    • Macari, M.1    Israel, G.M.2    Berman, P.3
  • 4
    • 33748998438 scopus 로고    scopus 로고
    • MDCT angiography of the pulmonary arteries: Influence of body weight, body mass index, and scan length on arterial enhancement at different iodine flow rates
    • Schoellnast H, Deutschmann HA, Berghold A, et al. MDCT angiography of the pulmonary arteries: influence of body weight, body mass index, and scan length on arterial enhancement at different iodine flow rates. AJR Am J Roentgenol. 2006;187:1074-1078.
    • (2006) AJR Am J Roentgenol , vol.187 , pp. 1074-1078
    • Schoellnast, H.1    Deutschmann, H.A.2    Berghold, A.3
  • 5
    • 20044382437 scopus 로고    scopus 로고
    • MDCT angiography of the pulmonary arteries: Influence of iodine flow concentration on vessel attenuation and visualization
    • Schoellnast H, Deutschmann HA, Fritz GA, et al. MDCT angiography of the pulmonary arteries: influence of iodine flow concentration on vessel attenuation and visualization. AJR Am J Roentgenol. 2005;184:1935-1939.
    • (2005) AJR Am J Roentgenol , vol.184 , pp. 1935-1939
    • Schoellnast, H.1    Deutschmann, H.A.2    Fritz, G.A.3
  • 6
    • 0034788075 scopus 로고    scopus 로고
    • Individually adapted examination protocols for reduction of radiation exposure in chest CT
    • Wildberger JE, Mahnken AH, Schmitz-Rode T, et al. Individually adapted examination protocols for reduction of radiation exposure in chest CT. Invest Radiol. 2001;36:604-611.
    • (2001) Invest Radiol , vol.36 , pp. 604-611
    • Wildberger, J.E.1    Mahnken, A.H.2    Schmitz-Rode, T.3
  • 7
    • 0033735138 scopus 로고    scopus 로고
    • Technique factors and image quality as functions of patient weight at abdominal CT
    • Huda W, Scalzetti EM, Levin G. Technique factors and image quality as functions of patient weight at abdominal CT. Radiology. 2000;217:430-435.
    • (2000) Radiology , vol.217 , pp. 430-435
    • Huda, W.1    Scalzetti, E.M.2    Levin, G.3
  • 8
    • 33645562681 scopus 로고    scopus 로고
    • CT dose reduction and dose management tools: Overview of available options
    • McCollough CH, Bruesewitz MR, Kofler JM Jr. CT dose reduction and dose management tools: overview of available options. Radiographics. 2006;26:503-512.
    • (2006) Radiographics , vol.26 , pp. 503-512
    • McCollough, C.H.1    Bruesewitz, M.R.2    Kofler Jr, J.M.3
  • 9
    • 44449151276 scopus 로고
    • The formation of a radiographic image
    • East Norwalk, CO: Appleton and Lange
    • Wolbarst A. The formation of a radiographic image. In: Physics of Radiology. East Norwalk, CO: Appleton and Lange, 1993:147:205.
    • (1993) Physics of Radiology , vol.147 , pp. 205
    • Wolbarst, A.1
  • 10
    • 44449124505 scopus 로고    scopus 로고
    • Application of automatic vertical positioning software to reduce radiation exposure in multi-detector row computed tomography of the chest
    • Li J, Udayasankar UK, Toth TL, et al. Application of automatic vertical positioning software to reduce radiation exposure in multi-detector row computed tomography of the chest. Invest Radiol. 2008;43:447-452.
    • (2008) Invest Radiol , vol.43 , pp. 447-452
    • Li, J.1    Udayasankar, U.K.2    Toth, T.L.3
  • 11
    • 0036068722 scopus 로고    scopus 로고
    • Standard-dose and 50%-reduced-dose chest CT: Comparing the effect on image quality
    • Prasad SR, Wittram C, Shepard JA, McLoud T, Rhea J. Standard-dose and 50%-reduced-dose chest CT: comparing the effect on image quality. AJR Am J Roentgenol. 2002;179:461-465.
    • (2002) AJR Am J Roentgenol , vol.179 , pp. 461-465
    • Prasad, S.R.1    Wittram, C.2    Shepard, J.A.3    McLoud, T.4    Rhea, J.5
  • 12
    • 19844367466 scopus 로고
    • CT in the diagnosis of pulmonary embolism
    • author reply, 2005
    • Ravenel JG, Costello P, Schoepf UJ. CT in the diagnosis of pulmonary embolism. AJR Am J Roentgenol. 2005;184:1707; author reply 1707-1708.
    • (1707) AJR Am J Roentgenol , vol.184 , pp. 1707
    • Ravenel, J.G.1    Costello, P.2    Schoepf, U.J.3
  • 13
    • 33846695725 scopus 로고    scopus 로고
    • CT angiography of pulmonary arteries to detect pulmonary embolism: Improvement of vascular enhancement with low kilovoltage settings
    • Schueller-Weidekamm C, Schaefer-Prokop CM, et al. CT angiography of pulmonary arteries to detect pulmonary embolism: improvement of vascular enhancement with low kilovoltage settings. Radiology. 2006;241:899-907.
    • (2006) Radiology , vol.241 , pp. 899-907
    • Schueller-Weidekamm, C.1    Schaefer-Prokop, C.M.2
  • 14
    • 1642389455 scopus 로고    scopus 로고
    • Low-kilovoltage multi-detector row chest CT in adults: Feasibility and effect on image quality and iodine dose
    • Sigal-Cinqualbre AB, Hennequin R, Abada HT, et al. Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology. 2004;231:169-174.
    • (2004) Radiology , vol.231 , pp. 169-174
    • Sigal-Cinqualbre, A.B.1    Hennequin, R.2    Abada, H.T.3
  • 15
    • 0031978536 scopus 로고    scopus 로고
    • Low-dose spiral computed tomography of the thorax: Comparison with the standard-dose technique
    • Takahashi M, Maguire WM, Ashtari M, et al. Low-dose spiral computed tomography of the thorax: comparison with the standard-dose technique. Invest Radiol. 1998;33:68-73.
    • (1998) Invest Radiol , vol.33 , pp. 68-73
    • Takahashi, M.1    Maguire, W.M.2    Ashtari, M.3
  • 16
    • 22544435185 scopus 로고    scopus 로고
    • Comparison of different body size parameters for individual dose adaptation in body CT of adults
    • Menke J. Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology. 2005;236:565-571.
    • (2005) Radiology , vol.236 , pp. 565-571
    • Menke, J.1
  • 17
    • 4444309418 scopus 로고    scopus 로고
    • Effect of area x-ray beam equalization on image quality and dose in digital mammography
    • Wong J, Xu T, Husain A, et al. Effect of area x-ray beam equalization on image quality and dose in digital mammography. Phys Med Biol. 2004;49:3539-3557.
    • (2004) Phys Med Biol , vol.49 , pp. 3539-3557
    • Wong, J.1    Xu, T.2    Husain, A.3
  • 18
    • 34250743798 scopus 로고    scopus 로고
    • CTA contrast enhancement of the aorta and pulmonary artery: The effect of saline chase injected at two different rates in a canine experimental model
    • Lee CH, Goo JM, Bae KT, et al. CTA contrast enhancement of the aorta and pulmonary artery: the effect of saline chase injected at two different rates in a canine experimental model. Invest Radiol. 2007;42:486-490.
    • (2007) Invest Radiol , vol.42 , pp. 486-490
    • Lee, C.H.1    Goo, J.M.2    Bae, K.T.3
  • 19
    • 0037742555 scopus 로고    scopus 로고
    • Low-dose multislice CT of the thorax in follow-up of malignant lymphoma and extrapulmonary primary tumors
    • Dinkel HP, Sonnenschein M, Hoppe H, Vock P. Low-dose multislice CT of the thorax in follow-up of malignant lymphoma and extrapulmonary primary tumors. Eur Radiol. 2003;13:1241-1249.
    • (2003) Eur Radiol , vol.13 , pp. 1241-1249
    • Dinkel, H.P.1    Sonnenschein, M.2    Hoppe, H.3    Vock, P.4
  • 21
    • 3042530832 scopus 로고    scopus 로고
    • Iodine delivery rate of different concentrations of iodine-containing contrast agents with rapid injection
    • Knollmann F, Schimpf K, Felix R. Iodine delivery rate of different concentrations of iodine-containing contrast agents with rapid injection. Rofo. 2004;176:880-884.
    • (2004) Rofo , vol.176 , pp. 880-884
    • Knollmann, F.1    Schimpf, K.2    Felix, R.3
  • 22
    • 0023829891 scopus 로고
    • In vitro study of high-pressure catheters and various contrast agents
    • Rees CR, Merchun G, Becker GJ, et al. In vitro study of high-pressure catheters and various contrast agents. Radiology. 1988;166:53-56.
    • (1988) Radiology , vol.166 , pp. 53-56
    • Rees, C.R.1    Merchun, G.2    Becker, G.J.3
  • 23
    • 15344342736 scopus 로고    scopus 로고
    • Improved vascular opacification in cerebral computed tomography angiography with 80 kVp
    • Bahner ML, Bengel A, Brix G, et al. Improved vascular opacification in cerebral computed tomography angiography with 80 kVp. Invest Radiol. 2005;40:229-234.
    • (2005) Invest Radiol , vol.40 , pp. 229-234
    • Bahner, M.L.1    Bengel, A.2    Brix, G.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.