Clinical evaluation of the CA530-VET hematology analyzer for use in veterinary practice

Author keywords

CA530 VET; Cat; Dog; Hematology analyzer; Horse; Method comparison

Indexed keywords

HEMOGLOBIN;

EID: 34250336930     PISSN: 02756382     EISSN: None     Source Type: Journal    
DOI: 10.1111/j.1939-165X.2007.tb00202.x     Document Type: Article
Times cited : (21)

References (25)
  • 1
    • 0014516374 scopus 로고
    • Recommended scheme for the evaluation of instruments for automatic analysis in the clinical biochemistry laboratory
    • Broughton PM, Buttolph MA, Gowenlock AH, Neill DW, Skentelbery RG. Recommended scheme for the evaluation of instruments for automatic analysis in the clinical biochemistry laboratory. J Clin Pathol. 1969;22:278-284.
    • (1969) J Clin Pathol , vol.22 , pp. 278-284
    • Broughton, P.M.1    Buttolph, M.A.2    Gowenlock, A.H.3    Neill, D.W.4    Skentelbery, R.G.5
  • 2
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 3
    • 0016240332 scopus 로고
    • A revised scheme for the evaluation of automatic instruments for use in clinical chemistry
    • Broughton PM, Gowenlock AH, McCormack JJ, Neill DW. A revised scheme for the evaluation of automatic instruments for use in clinical chemistry. Ann Clin Biochem. 1974;11:207-218.
    • (1974) Ann Clin Biochem , vol.11 , pp. 207-218
    • Broughton, P.M.1    Gowenlock, A.H.2    McCormack, J.J.3    Neill, D.W.4
  • 5
    • 0001572239 scopus 로고    scopus 로고
    • Federal Medical Council]. Richtlinie der Bundesärztekammer zur Qualitätssicherung quantitativer laboratoriumsmedizinischer Untersuchungen [Guideline of the Federal Medical Council for quality assurance for quantitative laboratory medical examinations]
    • Bundesärztekammer [Federal Medical Council]. Richtlinie der Bundesärztekammer zur Qualitätssicherung quantitativer laboratoriumsmedizinischer Untersuchungen [Guideline of the Federal Medical Council for quality assurance for quantitative laboratory medical examinations]. Dt. Artzebl. 2001;98:A2747-2759.
    • (2001) Dt. Artzebl , vol.98
    • Bundesärztekammer1
  • 6
    • 0025169461 scopus 로고
    • Performance goals for internal quality control of multichannel haematology analysers
    • Klee GG. Performance goals for internal quality control of multichannel haematology analysers. Clin Lab Haematol. 1990;12:65-74.
    • (1990) Clin Lab Haematol , vol.12 , pp. 65-74
    • Klee, G.G.1
  • 7
    • 0347720012 scopus 로고    scopus 로고
    • Evaluation des Hämatologiegerätes Vet ABC/8P bei Hund, Katze und Pferd [Evaluation of the haematology-system Vet ABC in dogs, cats and horses]
    • Neuerer F, Hirschberger J. Evaluation des Hämatologiegerätes Vet ABC/8P bei Hund, Katze und Pferd [Evaluation of the haematology-system Vet ABC in dogs, cats and horses]. Prakt Tierarzt. 1999;80: 584-594.
    • (1999) Prakt Tierarzt , vol.80 , pp. 584-594
    • Neuerer, F.1    Hirschberger, J.2
  • 8
    • 34250376450 scopus 로고
    • Nytt Analyssystem för Veterinärmedicinsk Hematologi [A new analyzing system for veterinary hematology]
    • Sprangfors P, Kineberger B, Holmen A. Nytt Analyssystem för Veterinärmedicinsk Hematologi [A new analyzing system for veterinary hematology]. Sve Vet Tid. 1990;42:527-531.
    • (1990) Sve Vet Tid , vol.42 , pp. 527-531
    • Sprangfors, P.1    Kineberger, B.2    Holmen, A.3
  • 9
    • 84984080025 scopus 로고
    • Evaluation of a haematological analyser for its use in canine clinical pathology
    • Lavin SF, Mora J, Monreal L, Vinas L. Evaluation of a haematological analyser for its use in canine clinical pathology. J Vet Med. A 1991;A38:702-709.
    • (1991) J Vet Med. A , vol.A38 , pp. 702-709
    • Lavin, S.F.1    Mora, J.2    Monreal, L.3    Vinas, L.4
  • 10
    • 0346248864 scopus 로고    scopus 로고
    • Evaluation of a hematology analyser with canine and feline blood
    • Pastor J, Cuenca R, Velarde R, Vinas L, Lavin S. Evaluation of a hematology analyser with canine and feline blood. Vet Clin Pathol. 1997;26:138-147.
    • (1997) Vet Clin Pathol , vol.26 , pp. 138-147
    • Pastor, J.1    Cuenca, R.2    Velarde, R.3    Vinas, L.4    Lavin, S.5
  • 12
    • 34250305101 scopus 로고    scopus 로고
    • Suchfort C. Die automatisierte Blutzellzählung und -differenzierung mit dem Technicon H*1 bei der Katze unter besonderer Berücksichtigung der Thrombozyten und der Erythrozytenmorphologie bei Anämien [The automatic blood cell count and differentiation with the Technicon H*1 in cats with special consideration of the platelet and erythrocyte morpholgy in anemic patients][dissertation]. Justus-Liebig-Universität: Gießen, Germany; 1998.
    • Suchfort C. Die automatisierte Blutzellzählung und -differenzierung mit dem Technicon H*1 bei der Katze unter besonderer Berücksichtigung der Thrombozyten und der Erythrozytenmorphologie bei Anämien [The automatic blood cell count and differentiation with the Technicon H*1 in cats with special consideration of the platelet and erythrocyte morpholgy in anemic patients][dissertation]. Justus-Liebig-Universität: Gießen, Germany; 1998.
  • 14
    • 0011044214 scopus 로고
    • Zelldifferenzierung in Blutausstrichen: Variabilität der Ergebnisse [Differentiating cells using blood smears: variability of the results]
    • Rümke CL. Die Zelldifferenzierung in Blutausstrichen: Variabilität der Ergebnisse [Differentiating cells using blood smears: variability of the results]. Triangel. 1960;4:154-158.
    • (1960) Triangel , vol.4 , pp. 154-158
    • Die, R.C.L.1
  • 15
    • 0026096493 scopus 로고
    • Monocyte counting: Discrepancies in results obtained with different automated instruments
    • Goossens W, Van-Hove L, Verwilghen RL. Monocyte counting: discrepancies in results obtained with different automated instruments. J Clin Pathol. 1991;44:224-227.
    • (1991) J Clin Pathol , vol.44 , pp. 224-227
    • Goossens, W.1    Van-Hove, L.2    Verwilghen, R.L.3
  • 16
    • 0017654180 scopus 로고
    • A delineation of performance criteria for the differentiation of leukocytes
    • Koepke M. A delineation of performance criteria for the differentiation of leukocytes. Am J Clin Pathol. 1977;68:202-206.
    • (1977) Am J Clin Pathol , vol.68 , pp. 202-206
    • Koepke, M.1
  • 17
    • 34250332647 scopus 로고    scopus 로고
    • Automatische Zellzählung und Zelldifferenzierung mit dem Technicon H*1 bei Blutproben der Katze [Validation of the Technicon H*1 for the automated cell count and cell differentiation of feline blood samples]
    • Liedl R, Hirschberger J. Automatische Zellzählung und Zelldifferenzierung mit dem Technicon H*1 bei Blutproben der Katze [Validation of the Technicon H*1 for the automated cell count and cell differentiation of feline blood samples]. Kleintierpraxis. 1997;42: 683-774.
    • (1997) Kleintierpraxis , vol.42 , pp. 683-774
    • Liedl, R.1    Hirschberger, J.2
  • 18
    • 0023653874 scopus 로고
    • Modification and evaluation of a multichannel blood cell counting system for blood analysis in veterinary hematology
    • Weiser MG. Modification and evaluation of a multichannel blood cell counting system for blood analysis in veterinary hematology. J Am Vet Med Assoc. 1987;190:411-415.
    • (1987) J Am Vet Med Assoc , vol.190 , pp. 411-415
    • Weiser, M.G.1
  • 19
    • 2742548876 scopus 로고    scopus 로고
    • Evaluation of a haematological analyser (Sysmex F-800) with equine blood
    • Pastor J, Cuenca R, Velarde R, Marco I, Vinas L, Lavin S. Evaluation of a haematological analyser (Sysmex F-800) with equine blood. J Vet Med. A 1997;A45:119-126.
    • (1997) J Vet Med. A , vol.A45 , pp. 119-126
    • Pastor, J.1    Cuenca, R.2    Velarde, R.3    Marco, I.4    Vinas, L.5    Lavin, S.6
  • 20
    • 0029396754 scopus 로고
    • Untersuchung zur automatischen Zellzählung von Katzenblut [Investigation on the automatic cell counting in feline blood]
    • Mischke R, Deniz A, Weiss J. Untersuchung zur automatischen Zellzählung von Katzenblut [Investigation on the automatic cell counting in feline blood]. Dtsch Tierärztl Wschr. 1995;102:435-440.
    • (1995) Dtsch Tierärztl Wschr , vol.102 , pp. 435-440
    • Mischke, R.1    Deniz, A.2    Weiss, J.3
  • 21
    • 0034667191 scopus 로고    scopus 로고
    • Evaluation of an in-house centrifugal hematology analyser for use in veterinary practice
    • Bienzle D, Stanton JB, Embry JM, Bush SE, Mahaffey E. Evaluation of an in-house centrifugal hematology analyser for use in veterinary practice. J Am Vet Med Assoc. 2000;217:1195-1200.
    • (2000) J Am Vet Med Assoc , vol.217 , pp. 1195-1200
    • Bienzle, D.1    Stanton, J.B.2    Embry, J.M.3    Bush, S.E.4    Mahaffey, E.5
  • 23
    • 34250359879 scopus 로고    scopus 로고
    • Boule-Medical. User Manual V3.7 CA620/530-VET. Stockholm, Sweden: Boule-Medical AB; 2003.
    • Boule-Medical. User Manual V3.7 CA620/530-VET. Stockholm, Sweden: Boule-Medical AB; 2003.
  • 24
    • 0346890138 scopus 로고    scopus 로고
    • Thrombozytenzählung bei der Katze [Platelet count in the cat]
    • Moritz A, Hoffmann C. Thrombozytenzählung bei der Katze [Platelet count in the cat]. Tierärztl Prax. 1997;25:695-700.
    • (1997) Tierärztl Prax , vol.25 , pp. 695-700
    • Moritz, A.1    Hoffmann, C.2
  • 25
    • 0035618736 scopus 로고    scopus 로고
    • Prevalence of low automated platelet counts in cats: Comparison with prevalence of thrombocytopenia based on blood smear estimation
    • Norman EJ, Barron RCJ, Nash AS, Clampitt RB. Prevalence of low automated platelet counts in cats: comparison with prevalence of thrombocytopenia based on blood smear estimation. Vet Clin Pathol. 2001;30:137-140.
    • (2001) Vet Clin Pathol , vol.30 , pp. 137-140
    • Norman, E.J.1    Barron, R.C.J.2    Nash, A.S.3    Clampitt, R.B.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.