메뉴 건너뛰기




Volumn 40, Issue 6, 2007, Pages 591-599

Uncertainty evaluation of humidity sensors calibrated by saturated salt solutions

Author keywords

Calibration; Humidity sensor; Uncertainty

Indexed keywords

CALIBRATION; INVERSE KINEMATICS; SATURATION (MATERIALS COMPOSITION); TEMPERATURE MEASUREMENT; UNCERTAINTY ANALYSIS;

EID: 34249720641     PISSN: 02632241     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.measurement.2006.09.012     Document Type: Article
Times cited : (76)

References (19)
  • 2
    • 0030108676 scopus 로고    scopus 로고
    • The CMA humidity standard
    • Heinonen M. The CMA humidity standard. Measurement 17 3 (1996) 183-188
    • (1996) Measurement , vol.17 , Issue.3 , pp. 183-188
    • Heinonen, M.1
  • 3
    • 2942711883 scopus 로고    scopus 로고
    • Uncertainty of humidity sensors testing by means of divided-flow generator
    • Su P., and Wu R. Uncertainty of humidity sensors testing by means of divided-flow generator. Measurement 36 (2004) 21-27
    • (2004) Measurement , vol.36 , pp. 21-27
    • Su, P.1    Wu, R.2
  • 4
    • 0030258634 scopus 로고    scopus 로고
    • Calibration method of flow meters for a divided flow humidity generator
    • Takahashi C., and Kitano H. Calibration method of flow meters for a divided flow humidity generator. Sensor Actuator B 35-36 (1996) 522-527
    • (1996) Sensor Actuator B , vol.35-36 , pp. 522-527
    • Takahashi, C.1    Kitano, H.2
  • 5
    • 0017437920 scopus 로고
    • Humidity fixed points of binary saturated aqueous solutions
    • Greenspan L. Humidity fixed points of binary saturated aqueous solutions. J. Res. Nat. Bur. Stand. A 81A 1 (1977) 89-96
    • (1977) J. Res. Nat. Bur. Stand. A , vol.81 A , Issue.1 , pp. 89-96
    • Greenspan, L.1
  • 6
    • 0037394345 scopus 로고    scopus 로고
    • Uncertainty evaluation of shore hardness testers
    • Mohamed M.I., and Aggag G.A. Uncertainty evaluation of shore hardness testers. Measurement 33 (2003) 251-257
    • (2003) Measurement , vol.33 , pp. 251-257
    • Mohamed, M.I.1    Aggag, G.A.2
  • 7
    • 0035452674 scopus 로고    scopus 로고
    • Application of statistics to oil density measurement
    • Park J.T., and Chang K. Application of statistics to oil density measurement. Measurement 30 (2001) 95-103
    • (2001) Measurement , vol.30 , pp. 95-103
    • Park, J.T.1    Chang, K.2
  • 8
    • 0034214639 scopus 로고    scopus 로고
    • Assessment of uncertainty in calibration of a gas mass flow meter
    • Kruh D. Assessment of uncertainty in calibration of a gas mass flow meter. Accredit. Qual. Assur. 5 (2000) 280-284
    • (2000) Accredit. Qual. Assur. , vol.5 , pp. 280-284
    • Kruh, D.1
  • 9
    • 34249723253 scopus 로고    scopus 로고
    • ISO GUM, Guide to the expression of uncertainty in measurement, 1993.
  • 10
    • 34249716411 scopus 로고    scopus 로고
    • OMIL, The scale of relative humidity of air certified against saturated salt solutions, Organization Internationale De Metrologie Legale OMIL R 121, 1996.
  • 11
    • 84946632251 scopus 로고
    • Classical and inverse regression methods of calibration
    • Krutchkoff R.G. Classical and inverse regression methods of calibration. Technometrics 9 3 (1967) 425-439
    • (1967) Technometrics , vol.9 , Issue.3 , pp. 425-439
    • Krutchkoff, R.G.1
  • 12
    • 0003062633 scopus 로고    scopus 로고
    • Inverse calibration predicts better than classical calibration
    • Centner V., Massart D.L., and de Jong S. Inverse calibration predicts better than classical calibration. Fresenius J. Anal. Chem. 361 (1998) 2-9
    • (1998) Fresenius J. Anal. Chem. , vol.361 , pp. 2-9
    • Centner, V.1    Massart, D.L.2    de Jong, S.3
  • 13
    • 0008804583 scopus 로고    scopus 로고
    • Inverse vs. classical calibration for small data sets
    • Tellinghuisen J. Inverse vs. classical calibration for small data sets. Fresenius J. Anal. Chem. 368 (2000) 585-588
    • (2000) Fresenius J. Anal. Chem. , vol.368 , pp. 585-588
    • Tellinghuisen, J.1
  • 14
    • 0001008644 scopus 로고    scopus 로고
    • Relation between prediction errors of inverse and classical calibration
    • Grientschnig D. Relation between prediction errors of inverse and classical calibration. Fresenius J. Anal. Chem. 367 (2000) 497-498
    • (2000) Fresenius J. Anal. Chem. , vol.367 , pp. 497-498
    • Grientschnig, D.1
  • 15
    • 0036629357 scopus 로고    scopus 로고
    • Evaluation of measurement uncertainty for analytical procedures using a linear calibration function
    • Bruggemann L., and Wennrich R. Evaluation of measurement uncertainty for analytical procedures using a linear calibration function. Accredit. Qual. Assur. 7 (2002) 269-273
    • (2002) Accredit. Qual. Assur. , vol.7 , pp. 269-273
    • Bruggemann, L.1    Wennrich, R.2
  • 17
    • 4143110209 scopus 로고    scopus 로고
    • Validation steps for traceability of linear calibrated chemical measurements
    • Bruggemann L., and Wennrich R. Validation steps for traceability of linear calibrated chemical measurements. Accredit. Qual. Assur. 9 (2004) 493-498
    • (2004) Accredit. Qual. Assur. , vol.9 , pp. 493-498
    • Bruggemann, L.1    Wennrich, R.2
  • 18
    • 34249748453 scopus 로고    scopus 로고
    • M. Stevens, T. Collier, R. Harris, Comparison of instruments for measuring relative humidities, NPL Report CBTLM 6, National Physical Laboratory, UK, 2000.
  • 19
    • 34249732661 scopus 로고    scopus 로고
    • Eurachem, EURACHEM/CITAC Guide Quantifying uncertainty in analytical measurement, Eurachem, 2000.


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.