Volumn 21, Issue 3, 2007, Pages 333-340

Accuracy of GDx variable corneal compensation polarization measurements in normal human eyes: Effect of accommodation, cycloplegia, focus, pupil size, and eye selection on reproducibility

Author keywords

[No Author keywords available]

Indexed keywords

ACCOMMODATION PARALYSIS; ADULT; ARTICLE; CONTROLLED STUDY; EYE REFRACTION; HUMAN; NORMAL HUMAN; POLARIMETER; PUPIL; PUPILLOMETER; REFRACTOMETER; REPRODUCIBILITY; STATISTICAL ANALYSIS; VARIABLE CORNEAL COMPENSATION; VISUAL SYSTEM PARAMETERS;

EID: 33847621616     PISSN: 0950222X     EISSN: 14765454     Source Type: Journal    
DOI: 10.1038/sj.eye.6702187     Document Type: Article
Times cited : (5)

References (24)
  • 1
    • 0001188112 scopus 로고
    • Spatially resolved birefringence of the retinal nerve fiber layer assessed with a retinal ellipsometer
    • Dreher AW, Reiter K, Weinreb RN. Spatially resolved birefringence of the retinal nerve fiber layer assessed with a retinal ellipsometer. Appl Opt 1992; 31: 3730-3749.
    • (1992) Appl Opt , vol.31 , pp. 3730-3749
    • Dreher, A.W.1    Reiter, K.2    Weinreb, R.N.3
  • 2
    • 0029044003 scopus 로고
    • Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes
    • Weinreb RN, Shakiba S, Zangwill L. Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol 1995; 119: 627-636.
    • (1995) Am J Ophthalmol , vol.119 , pp. 627-636
    • Weinreb, R.N.1    Shakiba, S.2    Zangwill, L.3
  • 3
    • 0036293309 scopus 로고    scopus 로고
    • Individualized compensation of anterior segment birefringence during scanning laser polarimetry
    • Zhou Q, Weinreb RN. Individualized compensation of anterior segment birefringence during scanning laser polarimetry. Invest Ophthalmol Vis Sci 2002; 43: 2221-2228.
    • (2002) Invest Ophthalmol Vis Sci , vol.43 , pp. 2221-2228
    • Zhou, Q.1    Weinreb, R.N.2
  • 4
    • 0036310419 scopus 로고    scopus 로고
    • Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry
    • Weinreb RN, Bowd C, Greenfield DS, Zangwill LM. Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry. Arch Ophthalmol 2002; 120: 901-906.
    • (2002) Arch Ophthalmol , vol.120 , pp. 901-906
    • Weinreb, R.N.1    Bowd, C.2    Greenfield, D.S.3    Zangwill, L.M.4
  • 5
    • 0036158184 scopus 로고    scopus 로고
    • Analytical model of scanning laser polarimetry for retinal nerve fiber layer assessment
    • Knighton RW, Huang XR, Greenfield DS. Analytical model of scanning laser polarimetry for retinal nerve fiber layer assessment. Invest Ophthalmol Vis Sci 2002; 43: 383-392.
    • (2002) Invest Ophthalmol Vis Sci , vol.43 , pp. 383-392
    • Knighton, R.W.1    Huang, X.R.2    Greenfield, D.S.3
  • 7
    • 0242668741 scopus 로고    scopus 로고
    • Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments as determined by scanning laser polarimetry
    • Choplin NT, Zhou Q, Knighton RW. Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments as determined by scanning laser polarimetry. Ophthalmology 2003; 110: 719-725.
    • (2003) Ophthalmology , vol.110 , pp. 719-725
    • Choplin, N.T.1    Zhou, Q.2    Knighton, R.W.3
  • 8
    • 0842310446 scopus 로고    scopus 로고
    • Variable corneal compensation improves discrimination between normal and glaucomatous eyes with the scanning laser polarimeter
    • Tannenbaum DP, Hoffman D, Lemij HG, Garway-Heath DF, Greenfield DS, Caprioli J. Variable corneal compensation improves discrimination between normal and glaucomatous eyes with the scanning laser polarimeter. Ophthalmology 2004; 111: 259-264.
    • (2004) Ophthalmology , vol.111 , pp. 259-264
    • Tannenbaum, D.P.1    Hoffman, D.2    Lemij, H.G.3    Garway-Heath, D.F.4    Greenfield, D.S.5    Caprioli, J.6
  • 9
    • 13444249777 scopus 로고    scopus 로고
    • Operator learning effect and interoperator reproducibility of the scanning laser polarimeter with variable corneal compensation
    • Frenkel S, Slonim E, Horani A, Molcho M, Barzel I, Blumenthal EZ. Operator learning effect and interoperator reproducibility of the scanning laser polarimeter with variable corneal compensation. Ophthalmology 2005; 112(2): 257-261.
    • (2005) Ophthalmology , vol.112 , Issue.2 , pp. 257-261
    • Frenkel, S.1    Slonim, E.2    Horani, A.3    Molcho, M.4    Barzel, I.5    Blumenthal, E.Z.6
  • 12
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 13
    • 0347125022 scopus 로고    scopus 로고
    • Eccentricity and measurement variability and repeatability with the retinal thickness analyser
    • Gilmore ED, Hudson C. Eccentricity and measurement variability and repeatability with the retinal thickness analyser. Br J Ophthalmol 2004; 88: 62-65.
    • (2004) Br J Ophthalmol , vol.88 , pp. 62-65
    • Gilmore, E.D.1    Hudson, C.2
  • 14
    • 0004148316 scopus 로고
    • (El Hage SG, Translator). Springer-Verlag: New York
    • Le Grand Y. Physiological Optics (El Hage SG, Translator). Springer-Verlag: New York, 1980, pp 85-86.
    • (1980) Physiological Optics , pp. 85-86
    • Le Grand, Y.1
  • 15
    • 0015453126 scopus 로고
    • An analysis of retinal orientation. IV Center of the entrance pupil and the center of convergence of orientation and directional sensitivity
    • Enoch JM, Hope GM. An analysis of retinal orientation. IV Center of the entrance pupil and the center of convergence of orientation and directional sensitivity. Invest Ophthalmol 1972; 11: 1017-1021.
    • (1972) Invest Ophthalmol , vol.11 , pp. 1017-1021
    • Enoch, J.M.1    Hope, G.M.2
  • 16
    • 0036293815 scopus 로고    scopus 로고
    • Pupil location under mesopic, photopic, and pharmacologically dilated conditions
    • Yang Y, Thompson K, Burns SA. Pupil location under mesopic, photopic, and pharmacologically dilated conditions. Invest Ophthalmol Vis Sci 2002; 43: 2508-2512.
    • (2002) Invest Ophthalmol Vis Sci , vol.43 , pp. 2508-2512
    • Yang, Y.1    Thompson, K.2    Burns, S.A.3
  • 17
    • 0000094764 scopus 로고    scopus 로고
    • Effect of pupillary dilation on retinal nerve fiber layer thickness as measured by scanning laser polarimetry in eyes with and without cataract
    • Hoh ST, Greenfield DS, Liebmann JM, Hillenkamp J, Ishikawa H, Mistlberger A et al. Effect of pupillary dilation on retinal nerve fiber layer thickness as measured by scanning laser polarimetry in eyes with and without cataract. J Glaucoma 1999; 8: 159-163.
    • (1999) J Glaucoma , vol.8 , pp. 159-163
    • Hoh, S.T.1    Greenfield, D.S.2    Liebmann, J.M.3    Hillenkamp, J.4    Ishikawa, H.5    Mistlberger, A.6
  • 18
    • 33847668516 scopus 로고    scopus 로고
    • The effect of pupil dilation on scanning laser polarimetry with variable cornea compensation
    • (E-abstract #3311; April 27)
    • Blumenthal EZ, Horani A, Frenkel S. The effect of pupil dilation on scanning laser polarimetry with variable cornea compensation. Invest Ophthalmol Vis Sci ARVO (E-abstract #3311; April 27, 2004).
    • (2004) Invest Ophthalmol Vis Sci ARVO
    • Blumenthal, E.Z.1    Horani, A.2    Frenkel, S.3
  • 19
    • 15244360641 scopus 로고    scopus 로고
    • Influence of pupillary diameter, ciliary muscle tone, and ambient light on nerve fiber layer measurements with scanning laser polarimetry
    • Paranhos Jr A, Silvestre AM, Passerotti CC, Martins EMR, de Arruda Mello PA, Prata Jr JA et al. Influence of pupillary diameter, ciliary muscle tone, and ambient light on nerve fiber layer measurements with scanning laser polarimetry. J Glaucoma 2005; 14: 124-127.
    • (2005) J Glaucoma , vol.14 , pp. 124-127
    • Paranhos Jr., A.1    Silvestre, A.M.2    Passerotti, C.C.3    Martins, E.M.R.4    de Arruda Mello, P.A.5    Prata Jr., J.A.6
  • 20
    • 33847689973 scopus 로고    scopus 로고
    • Personal communication
    • Laser Diagnostic Technologies Inc.: San Diego, CA
    • Sinai M. Personal communication. Laser Diagnostic Technologies Inc.: San Diego, CA.
    • Sinai, M.1
  • 21
    • 0036139716 scopus 로고    scopus 로고
    • Linear birefringence of the central human cornea
    • Knighton RW, Huang X. Linear birefringence of the central human cornea. Invest Ophthalmol Vis Sci 2002; 43: 82-86.
    • (2002) Invest Ophthalmol Vis Sci , vol.43 , pp. 82-86
    • Knighton, R.W.1    Huang, X.2
  • 22
    • 20444487701 scopus 로고    scopus 로고
    • Quantitative thickness measurement of retinal layers imaged by optical coherence tomography
    • Shahidi M, Wang Z, Zelkha R. Quantitative thickness measurement of retinal layers imaged by optical coherence tomography. Am J Ophthalmol 2005; 139(6): 1056-1061.
    • (2005) Am J Ophthalmol , vol.139 , Issue.6 , pp. 1056-1061
    • Shahidi, M.1    Wang, Z.2    Zelkha, R.3
  • 23
    • 0030470612 scopus 로고    scopus 로고
    • Retinal nerve fiber layer thickness in normal human eyes
    • (MedlineGDx Operation Manual. Version 5.3. Laser Diagnostic Technologies Inc.: San Diego, CA, pp 2-7)
    • Varma R, Skaf M, Barron E. Retinal nerve fiber layer thickness in normal human eyes. Ophthalmology 1996; 103(12): 21 (MedlineGDx Operation Manual. Version 5.3. Laser Diagnostic Technologies Inc.: San Diego, CA, pp 2-7).
    • (1996) Ophthalmology , vol.103 , Issue.12 , pp. 21
    • Varma, R.1    Skaf, M.2    Barron, E.3
  • 24
    • 33847658472 scopus 로고    scopus 로고
    • Laser Diagnostic Technologies Inc.: San Diego, CA
    • GDx Operation Manual Version 5.3. Laser Diagnostic Technologies Inc.: San Diego, CA, pp 2-7.
    • GDx Operation Manual Version 5.3 , pp. 2-7


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.