Volumn 98, Issue 1, 2007, Pages 45-52

Non-invasive metabolic monitoring of patients under anaesthesia by continuous indirect calorimetry - An in vivo trial of a new method

Author keywords

Measurement techniques, Fick principle; Metabolism, oxygen consumption; Monitoring, oxygen

Indexed keywords

OXYGEN;

EID: 33845434137     PISSN: 00070912     EISSN: 14716771     Source Type: Journal    
DOI: 10.1093/bja/ael310     Document Type: Article
Times cited : (12)

References (31)
  • 1
    • 0038781169 scopus 로고    scopus 로고
    • Recent advances in gas exchange measurement in intensive care patients
    • Walsh T. Recent advances in gas exchange measurement in intensive care patients. Br J Anaesth 2003; 91: 120-31
    • (2003) Br J Anaesth , vol.91 , pp. 120-131
    • Walsh, T.1
  • 2
    • 0036308817 scopus 로고    scopus 로고
    • Measurement of systemic oxygen uptake during low-flow anaesthesia with a standard technique vs. a novel method
    • Leonard IE, Weitkamp B, Jones K, Aittomaki J, Myles PS. Measurement of systemic oxygen uptake during low-flow anaesthesia with a standard technique vs. a novel method. Anaesthesia 2002; 57: 654-8
    • (2002) Anaesthesia , vol.57 , pp. 654-658
    • Leonard, I.E.1    Weitkamp, B.2    Jones, K.3    Aittomaki, J.4    Myles, P.S.5
  • 3
    • 85044010430 scopus 로고    scopus 로고
    • Correlation between oxygen consumption calculated using Fick's method and measured with indirect calorimetry in critically ill patients
    • Marson F, Martins MA, Coletto FA, Campos AD, Basile-Filho A. Correlation between oxygen consumption calculated using Fick's method and measured with indirect calorimetry in critically ill patients. Arq Bras Cardiol 2003; 82: 77-81
    • (2003) Arq Bras Cardiol , vol.82 , pp. 77-81
    • Marson, F.1    Martins, M.A.2    Coletto, F.A.3    Campos, A.D.4    Basile-Filho, A.5
  • 4
    • 0029941330 scopus 로고    scopus 로고
    • The use of oxygen consumption and delivery as endpoints for resuscitation in critically ill patients
    • Durham R, Neunaber K, Mazuski J, Shapiro M, Baue A. The use of oxygen consumption and delivery as endpoints for resuscitation in critically ill patients. J Trauma 1996; 41: 32-9
    • (1996) J Trauma , vol.41 , pp. 32-39
    • Durham, R.1    Neunaber, K.2    Mazuski, J.3    Shapiro, M.4    Baue, A.5
  • 5
    • 0032007225 scopus 로고    scopus 로고
    • A formula to calculate oxygen uptake during low flow anesthesia based on FIO2 measurement
    • Biro P. A formula to calculate oxygen uptake during low flow anesthesia based on FIO2 measurement. J Clin Monit Comput 1998; 14: 141-4
    • (1998) J Clin Monit Comput , vol.14 , pp. 141-144
    • Biro, P.1
  • 6
    • 0005534353 scopus 로고
    • The administration of nitrous oxide-oxygen anesthesia in closed systems
    • Foldes FF, Ceravolo AJ, Carpenter SL. The administration of nitrous oxide-oxygen anesthesia in closed systems. Ann Surg 1952; 136: 978-81
    • (1952) Ann Surg , vol.136 , pp. 978-981
    • Foldes, F.F.1    Ceravolo, A.J.2    Carpenter, S.L.3
  • 7
    • 2942611292 scopus 로고    scopus 로고
    • Accuracy of the Foldes-Biro equation for measurement of oxygen uptake during anaesthesia: A laboratory simulation
    • Stuart-Andrews C, Peyton P, Robinson GJB, Lithgow B. Accuracy of the Foldes-Biro equation for measurement of oxygen uptake during anaesthesia: A laboratory simulation. Anaesthesia 2004; 59: 541-4
    • (2004) Anaesthesia , vol.59 , pp. 541-544
    • Stuart-Andrews, C.1    Peyton, P.2    Robinson, G.J.B.3    Lithgow, B.4
  • 8
    • 32544453696 scopus 로고    scopus 로고
    • Continuous indirect calorimetry - A laboratory simulation of a new method for non-invasive metabolic monitoring of patients under anaesthesia or in critical care
    • Stuart-Andrews C, Peyton P, Robinson G, Lithgow B. Continuous indirect calorimetry - a laboratory simulation of a new method for non-invasive metabolic monitoring of patients under anaesthesia or in critical care. Physiol Meas 2006; 27: 155-64
    • (2006) Physiol Meas , vol.27 , pp. 155-164
    • Stuart-Andrews, C.1    Peyton, P.2    Robinson, G.3    Lithgow, B.4
  • 9
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-10
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.1    Altman, D.2
  • 11
    • 0025193542 scopus 로고
    • Evaluation of metabolic measuring instruments for use in critically ill patients
    • Makita K, Nunn JF, Royston B. Evaluation of metabolic measuring instruments for use in critically ill patients. Crit Care Med 1990; 18: 638-44
    • (1990) Crit Care Med , vol.18 , pp. 638-644
    • Makita, K.1    Nunn, J.F.2    Royston, B.3
  • 12
    • 0023156833 scopus 로고
    • Continuous measurement of oxygen uptake. Evaluation of the Engstrom metabolic computer and clinical experiences
    • Behrendt W, Weiland C, Kalff J, Giani G. Continuous measurement of oxygen uptake. Evaluation of the Engstrom metabolic computer and clinical experiences. Acta Anaesthesiol Scand 1987; 31: 10-4
    • (1987) Acta Anaesthesiol Scand , vol.31 , pp. 10-14
    • Behrendt, W.1    Weiland, C.2    Kalff, J.3    Giani, G.4
  • 13
    • 0024423390 scopus 로고    scopus 로고
    • Measurement of gas exchange in intensive care: Laboratory and clinical validation of a new device
    • Takala J, Keinanen O, Vaisanen P, Kari A. Measurement of gas exchange in intensive care: Laboratory and clinical validation of a new device. Crit Care Med 1998; 17: 1041-7
    • (1998) Crit Care Med , vol.17 , pp. 1041-1047
    • Takala, J.1    Keinanen, O.2    Vaisanen, P.3    Kari, A.4
  • 14
    • 1342268938 scopus 로고    scopus 로고
    • A new method for measurement of gas exchange during anaesthesia using an extractable marker gas
    • Peyton PJ, Robinson GJB, Terry D, Junor PA. A new method for measurement of gas exchange during anaesthesia using an extractable marker gas. Physiol Meas 2004; 25: 1-14
    • (2004) Physiol Meas , vol.25 , pp. 1-14
    • Peyton, P.J.1    Robinson, G.J.B.2    Terry, D.3    Junor, P.A.4
  • 15
    • 4344677815 scopus 로고    scopus 로고
    • Continuous measurement of gas uptake and elimination in anesthetized patients using an extractable marker gas
    • Robinson G, Peyton P, Terry D, Malekzadeh S, Thompson B. Continuous measurement of gas uptake and elimination in anesthetized patients using an extractable marker gas. J Appl Physiol 2004; 97: 960-6
    • (2004) J Appl Physiol , vol.97 , pp. 960-966
    • Robinson, G.1    Peyton, P.2    Terry, D.3    Malekzadeh, S.4    Thompson, B.5
  • 16
    • 0016639110 scopus 로고
    • Totally closed circuit nitrous oxide/oxygen anaesthesia
    • Barton F, Nunn JF. Totally closed circuit nitrous oxide/oxygen anaesthesia. Br J Anaesth 1975; 47: 350-7
    • (1975) Br J Anaesth , vol.47 , pp. 350-357
    • Barton, F.1    Nunn, J.F.2
  • 18
    • 0036652512 scopus 로고    scopus 로고
    • Development and in vitro validation of a device for measuring non-shunt cardiac output by nitrous oxide throughflow
    • Vartuli GM, Burfoot RB, Robinson GJB, Peyton PJ, Junor PA. Development and in vitro validation of a device for measuring non-shunt cardiac output by nitrous oxide throughflow. Med Biol Eng Comput 2002; 40: 415-22
    • (2002) Med Biol Eng Comput , vol.40 , pp. 415-422
    • Vartuli, G.M.1    Burfoot, R.B.2    Robinson, G.J.B.3    Peyton, P.J.4    Junor, P.A.5
  • 19
    • 0027447134 scopus 로고
    • Errors in the measurement of cardiac output by thermodilution
    • Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 1993; 40: 142-53
    • (1993) Can J Anaesth , vol.40 , pp. 142-153
    • Nishikawa, T.1    Dohi, S.2
  • 20
    • 29844448951 scopus 로고    scopus 로고
    • Comparison of invasive and less-invasive techniques of cardiac output measurement under different haemodynamic conditions in a pig model
    • Bajorat J, Hofmockel R, Vagts DA, Pohl B, Beck C, Noeldge-Schomburg G. Comparison of invasive and less-invasive techniques of cardiac output measurement under different haemodynamic conditions in a pig model. Eur J Anaesth 2006; 23: 23-30
    • (2006) Eur J Anaesth , vol.23 , pp. 23-30
    • Bajorat, J.1    Hofmockel, R.2    Vagts, D.A.3    Pohl, B.4    Beck, C.5    Noeldge-Schomburg, G.6
  • 21
    • 0025102157 scopus 로고
    • Frequently repeated Fick cardiac output measurements during anesthesia
    • Doi M, Morita K, Ikeda K. Frequently repeated Fick cardiac output measurements during anesthesia. J Clin Monit 1990; 6: 107-12
    • (1990) J Clin Monit , vol.6 , pp. 107-112
    • Doi, M.1    Morita, K.2    Ikeda, K.3
  • 22
    • 0030799387 scopus 로고    scopus 로고
    • Overestimation of low cardiac output measured by thermodilution
    • Tournadre J, Chassard D, Muchada R. Overestimation of low cardiac output measured by thermodilution. Br J Anaesth 1997; 79: 514-6
    • (1997) Br J Anaesth , vol.79 , pp. 514-516
    • Tournadre, J.1    Chassard, D.2    Muchada, R.3
  • 25
    • 0031841771 scopus 로고    scopus 로고
    • Oxygen consumption in the intensive care unit: Indirect calorimetry is the way to go, but where?
    • deBoisblanc B, McClarity E, Lord K. Oxygen consumption in the intensive care unit: Indirect calorimetry is the way to go, but where? Crit Care Med 1998; 26: 1153-4
    • (1998) Crit Care Med , vol.26 , pp. 1153-1154
    • deBoisblanc, B.1    McClarity, E.2    Lord, K.3
  • 26
    • 0031838159 scopus 로고    scopus 로고
    • A comparison between the Fick method and indirect calorimetry for determining oxygen consumption in patients with fulminant hepatic failure
    • Walsh T, Hopton P, Lee A. A comparison between the Fick method and indirect calorimetry for determining oxygen consumption in patients with fulminant hepatic failure. Crit Care Med 1998; 26: 1200-7
    • (1998) Crit Care Med , vol.26 , pp. 1200-1207
    • Walsh, T.1    Hopton, P.2    Lee, A.3
  • 27
    • 13544254552 scopus 로고    scopus 로고
    • Measured pulmonary oxygen consumption: Difference between systemic oxygen uptake measured by the reverse Fick method and indirect calorimetry in cardiac surgery
    • Peyton PJ, Robinson GJB. Measured pulmonary oxygen consumption: difference between systemic oxygen uptake measured by the reverse Fick method and indirect calorimetry in cardiac surgery. Anaesthesia 2005; 60: 146-50
    • (2005) Anaesthesia , vol.60 , pp. 146-150
    • Peyton, P.J.1    Robinson, G.J.B.2
  • 28
    • 0030807596 scopus 로고    scopus 로고
    • Calculated versus measured oxygen consumption during and after cardiac surgery. Is it possible to estimate lung oxygen consumption?
    • Keinanen O, Takala J. Calculated versus measured oxygen consumption during and after cardiac surgery. Is it possible to estimate lung oxygen consumption? Acta Anaesthesiol Scand 1997; 41: 803-9
    • (1997) Acta Anaesthesiol Scand , vol.41 , pp. 803-809
    • Keinanen, O.1    Takala, J.2
  • 29
    • 0028784190 scopus 로고
    • Comparison between oxygen consumption calculated by Fick's principle using a continuous thermodilution technique and measured by indirect calorimetry
    • Bizouarn P, Blanloeil Y, Pinaud M. Comparison between oxygen consumption calculated by Fick's principle using a continuous thermodilution technique and measured by indirect calorimetry. Br J Anaesth 1995; 75: 719-23
    • (1995) Br J Anaesth , vol.75 , pp. 719-723
    • Bizouarn, P.1    Blanloeil, Y.2    Pinaud, M.3
  • 30
    • 0023217547 scopus 로고
    • Continuous measurement of coronary sinus oxygen saturation after cardiac surgery
    • Kumon K, Kuwabara M, Hirata T, et al. Continuous measurement of coronary sinus oxygen saturation after cardiac surgery. Crit Care Med 1987; 15: 595-7
    • (1987) Crit Care Med , vol.15 , pp. 595-597
    • Kumon, K.1    Kuwabara, M.2    Hirata, T.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.