메뉴 건너뛰기




Volumn 2, Issue 11, 2005, Pages 1253-1255

Inappropriate therapy due to triple counting of the ventricular electrogram in a patient with implantable cardioverter-defibrillator

Author keywords

Implantable cardioverter defibrillator; Inappropriate therapy; Triple counting

Indexed keywords

ADULT; ARTICLE; CARDIOVERSION; CASE REPORT; CONSCIOUSNESS DISORDER; DEFIBRILLATOR; DYSPNEA; ELECTROCARDIOGRAM; FATALITY; HEART FAILURE; HEART LEFT BUNDLE BRANCH BLOCK; HEART LEFT VENTRICLE FUNCTION; HEART VENTRICLE TACHYCARDIA; HOLTER MONITORING; HUMAN; HYPOTENSION; ISCHEMIC HEART DISEASE; LUNG EDEMA; MALE; PRIORITY JOURNAL; QRS COMPLEX; RR INTERVAL; SINUS RHYTHM; THERAPEUTIC ERROR;

EID: 27744492189     PISSN: 15475271     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.hrthm.2005.08.004     Document Type: Article
Times cited : (4)

References (6)
  • 2
    • 7544220287 scopus 로고    scopus 로고
    • Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter defibrillators. PainFREE Rx II trial results
    • for the PainFREE Rx II Investigators
    • Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, Canby RC, Khalighi K, Machado C, Rubenstein DS, Volosin JK, for the PainFREE Rx II Investigators. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter defibrillators. PainFREE Rx II trial results. Circulation 2004;110:2592-2596
    • (2004) Circulation , vol.110 , pp. 2592-2596
    • Wathen, M.S.1    DeGroot, P.J.2    Sweeney, M.O.3    Stark, A.J.4    Otterness, M.F.5    Adkisson, W.O.6    Canby, R.C.7    Khalighi, K.8    Machado, C.9    Rubenstein, D.S.10    Volosin, J.K.11
  • 4
    • 0036162035 scopus 로고    scopus 로고
    • Inappropriate tachycardia detection by a biventricular implantable cardioverter defibrillator
    • Garcia-Moran E, Mont L, Brugada J. Inappropriate tachycardia detection by a biventricular implantable cardioverter defibrillator. Pacing Clin Electrophysiol 2002;25:123-124
    • (2002) Pacing Clin. Electrophysiol. , vol.25 , pp. 123-124
    • Garcia-Moran, E.1    Mont, L.2    Brugada, J.3
  • 5
    • 0042200233 scopus 로고    scopus 로고
    • Double counting of the ventricular electrogram in biventricular pacemakers and ICDs
    • Barold SS, Herweg B, Gallardo I. Double counting of the ventricular electrogram in biventricular pacemakers and ICDs. Pacing Clin Electrophysiol 2003;26:1645-1648
    • (2003) Pacing Clin. Electrophysiol. , vol.26 , pp. 1645-1648
    • Barold, S.S.1    Herweg, B.2    Gallardo, I.3
  • 6
    • 2442482427 scopus 로고    scopus 로고
    • Inappropriate therapy from a defibrillator complicating transcoronary ablation of septal hypertrophy in a patient with hypertrophic obstructive cardiomyopathy
    • Shah P, Harris L, Hill A, Schwartz L, Cameron D. Inappropriate therapy from a defibrillator complicating transcoronary ablation of septal hypertrophy in a patient with hypertrophic obstructive cardiomyopathy. Pacing Clin Electrophysiol 2004;27:677-680
    • (2004) Pacing Clin. Electrophysiol. , vol.27 , pp. 677-680
    • Shah, P.1    Harris, L.2    Hill, A.3    Schwartz, L.4    Cameron, D.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.