메뉴 건너뛰기




Volumn 129, Issue 5, 2004, Pages 398-399

Application of selected reaction monitoring tandem mass spectrometry to the quantitative determination of an arsenic-containing nucleoside in a crude biological extract

Author keywords

[No Author keywords available]

Indexed keywords

5' DIMETHYLARSINOYLADENOSINE; ARSENIC; NUCLEOSIDE; UNCLASSIFIED DRUG;

EID: 2442658951     PISSN: 00032654     EISSN: None     Source Type: Journal    
DOI: 10.1039/b402876g     Document Type: Article
Times cited : (20)

References (12)
  • 5
    • 2442658535 scopus 로고    scopus 로고
    • note
    • -1.
  • 6
    • 2442699689 scopus 로고    scopus 로고
    • note
    • An electrospray triple quadrupole mass spectrometer (TSQ Quantum, Thermo Electron) in combination with a HPLC pump and an autosampler (Surveyor, Thermo Electron) was used throughout this study.
  • 7
    • 2442678487 scopus 로고    scopus 로고
    • note
    • The CID breakdown curves are available as electronic supplementary material.†.
  • 8
    • 2442692261 scopus 로고    scopus 로고
    • note
    • -1; the effluent was split post-column with 16% being directed to the ES MS. Injection volumes were 20 μl.
  • 9
    • 2442638376 scopus 로고    scopus 로고
    • note
    • Limit of detection was determined with mobile phase containing 3% methanol, and defined as the analyte concentration giving a signal equal to the blank plus three standard deviations of the blank. The HPLC window of 3.5-6.0 min was used to measure blank signals for 7 consecutive runs.
  • 10
    • 2442710884 scopus 로고    scopus 로고
    • note
    • 2,3 Extracts were prepared by weighing a portion (17.8 mg) of the fine powder into an Eppendorf tube, adding 1.00 mL of 20 mM ammonium bicarbonate, pH 10.0 (HPLC stock mobile phase), and shaking for about two min. The mixture was centrifuged (13,000 rpm, 5 min) and the clear, brown supernatant was analysed directly or after diluting with mobile phase (1 + 1).
  • 11
    • 2442643417 scopus 로고    scopus 로고
    • note
    • -1 of standard 5′- dimethylarsinoyladenosine.
  • 12
    • 2442648451 scopus 로고    scopus 로고
    • note
    • A minimum value based on complete extraction of the target compound. We have not optimised the extraction procedure, but our previous experience with water-soluble arsenicals indicates high extraction efficiency with this simple procedure.


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.