Optimizing the tensile properties of polyvinyl alcohol hydrogel for the construction of a bioprosthetic heart valve stent

Author keywords

Bioprosthetic heart valve; Freeze thaw; Polyvinyl alcohol hydrogel; Stent; Stress relaxation; Tensile

Indexed keywords

BIOCOMPATIBILITY; CALCIFICATION (BIOCHEMISTRY); HEMODYNAMICS; HYDROGELS; OPTIMIZATION; POLYVINYL ACETATES; POLYVINYL ALCOHOLS; STRESS RELAXATION; TENSILE PROPERTIES;

EID: 1842836414     PISSN: 00219304     EISSN: None     Source Type: Journal    
DOI: 10.1002/jbm.10333     Document Type: Article
Times cited : (202)

References (35)
  • 2
    • 0026550699 scopus 로고
    • Durability of porcine valves at fifteen years in a representative North American patient population
    • Burton TA et al. Durability of porcine valves at fifteen years in a representative North American patient population. J Thorac Cardiovasc Surg 1992;103:238-252.
    • (1992) J Thorac Cardiovasc Surg , vol.103 , pp. 238-252
    • Burton, T.A.1
  • 4
    • 0022623739 scopus 로고
    • Comparative study of primary tissue valve failure between Ionescu-Shiley pericardial and Hancock porcine valves in the aorta position
    • Nistal F, Garcia SE, Artinano E, Duran C, Gallo I. Comparative study of primary tissue valve failure between Ionescu-Shiley pericardial and Hancock porcine valves in the aorta position. American J Cardiol 1986;57:161-164.
    • (1986) American J Cardiol , vol.57 , pp. 161-164
    • Nistal, F.1    Garcia, S.E.2    Artinano, E.3    Duran, C.4    Gallo, I.5
  • 5
    • 0020660663 scopus 로고
    • Long term failure rate and morphologic correlations in porcine bioprosthetic heart valves
    • Schoen FJ, Collins JJ, Cohn LH. Long term failure rate and morphologic correlations in porcine bioprosthetic heart valves. Am J Cardiol 1983;51:957-964.
    • (1983) Am J Cardiol , vol.51 , pp. 957-964
    • Schoen, F.J.1    Collins, J.J.2    Cohn, L.H.3
  • 9
    • 0029117077 scopus 로고
    • Longitudinal and radial distensibility of the porcine aortic root
    • Hansen B, Menkis AH, Vesely I. Longitudinal and radial distensibility of the porcine aortic root, Ann Thorac Surg 1995;60:384-390.
    • (1995) Ann Thorac Surg , vol.60 , pp. 384-390
    • Hansen, B.1    Menkis, A.H.2    Vesely, I.3
  • 10
    • 0025499734 scopus 로고
    • A computerized system for video analysis of the aortic valve
    • Vesely I, Menkis AH, Campbell G. A computerized system for video analysis of the aortic valve. IEEE Trans Biomed Eng 1990;37:925-929.
    • (1990) IEEE Trans Biomed Eng , vol.37 , pp. 925-929
    • Vesely, I.1    Menkis, A.H.2    Campbell, G.3
  • 12
    • 0016334717 scopus 로고
    • Dimensions and geometric relationships of the human aortic valve as a function of pressure
    • Swanson WM, Clark RE. Dimensions and geometric relationships of the human aortic valve as a function of pressure. Circ Res 1974;35:871-882.
    • (1974) Circ Res , vol.35 , pp. 871-882
    • Swanson, W.M.1    Clark, R.E.2
  • 13
    • 0028413699 scopus 로고
    • An optimal mounting frame to reduce flexural stresses of bioprosthetic heart valves
    • Vesely I, Krucinski S, Dokainish MA, Campbell G. An optimal mounting frame to reduce flexural stresses of bioprosthetic heart valves. ASAIO J 1994;199-205.
    • (1994) ASAIO J , pp. 199-205
    • Vesely, I.1    Krucinski, S.2    Dokainish, M.A.3    Campbell, G.4
  • 18
    • 0002770914 scopus 로고
    • Synthetic hydrogels for biomedical applications
    • Andrade JD, editor, ACS Symposium Series, American Chemical Society, Washington, DC
    • Ratner BD, Hoffman AS. Synthetic hydrogels for biomedical applications. In Andrade JD, editor, Hydrogels for Medical and Related Applications, ACS Symposium Series, American Chemical Society, Washington, DC; 1985. Vol. 31, p 14.
    • (1985) Hydrogels for Medical and Related Applications , vol.31 , pp. 14
    • Ratner, B.D.1    Hoffman, A.S.2
  • 19
    • 0015669545 scopus 로고
    • PVA hydrogel for synthetic articular cartilage material
    • Bray JC, Marty EW. PVA hydrogel for synthetic articular cartilage material. J Biomed Mater Res 1973;7:431.
    • (1973) J Biomed Mater Res , vol.7 , pp. 431
    • Bray, J.C.1    Marty, E.W.2
  • 20
    • 0001925374 scopus 로고
    • Anomalous increase of the elastic modulus of frozen poly(vinyl alcohol) gels
    • Watase M, Nishinari K, Nambu M. Anomalous increase of the elastic modulus of frozen poly(vinyl alcohol) gels. Cryo-Letters 1983;4:197-200.
    • (1983) Cryo-Letters , vol.4 , pp. 197-200
    • Watase, M.1    Nishinari, K.2    Nambu, M.3
  • 21
    • 0025885279 scopus 로고
    • Reinforced uncrosslinked poly(vinyl alcohol) gels produced by cyclic freezing-thawing processes: A short review
    • Peppas NA. Stauffer SR. Reinforced uncrosslinked poly(vinyl alcohol) gels produced by cyclic freezing-thawing processes: A short review. J Controlled Release 1991;16:305-310.
    • (1991) J Controlled Release , vol.16 , pp. 305-310
    • Peppas, N.A.1    Stauffer, S.R.2
  • 22
    • 0001664446 scopus 로고
    • Poly(vinyl alcohol) hydrogels prepared by freezing-thawing cyclic processing
    • Stauffer SR, Peppas NA. Poly(vinyl alcohol) hydrogels prepared by freezing-thawing cyclic processing, Polymer 1992;33: 3932-3936.
    • (1992) Polymer , vol.33 , pp. 3932-3936
    • Stauffer, S.R.1    Peppas, N.A.2
  • 25
    • 33751268423 scopus 로고
    • The reason for the shape of the distensibility curves of arteries
    • Roach MR, Burton AC. The reason for the shape of the distensibility curves of arteries, Can J Biochem Physiol 1957;35:681.
    • (1957) Can J Biochem Physiol , vol.35 , pp. 681
    • Roach, M.R.1    Burton, A.C.2
  • 26
    • 0024427911 scopus 로고
    • Effects of age on the anisotropy of the descending human thoracic aorta determined by uniaxial tensile testing and digestion by NaOH under load
    • Sherebrin MH, Hegney JE, Roach MR. Effects of age on the anisotropy of the descending human thoracic aorta determined by uniaxial tensile testing and digestion by NaOH under load. Can J Physiol Pharmacol 1989;67:871.
    • (1989) Can J Physiol Pharmacol , vol.67 , pp. 871
    • Sherebrin, M.H.1    Hegney, J.E.2    Roach, M.R.3
  • 27
    • 0022386915 scopus 로고
    • Mechanical properties of human pericardium. Differences in viscoelastic response when compared with canine pericardium
    • Lee JM, Boughner DR. Mechanical properties of human pericardium. Differences in viscoelastic response when compared with canine pericardium. Circ Res 1985;55:475-481.
    • (1985) Circ Res , vol.55 , pp. 475-481
    • Lee, J.M.1    Boughner, D.R.2
  • 28
    • 0024673537 scopus 로고
    • The bovine pericardial xenograft: I Effect of fixation in aldehydes without constraint on the tensile viscoelastic properties of bovine pericardium
    • Lee JM, Haberer SA, Boughner DR. The bovine pericardial xenograft: I Effect of fixation in aldehydes without constraint on the tensile viscoelastic properties of bovine pericardium. J Biomed Mater Res 1989;23:457-475.
    • (1989) J Biomed Mater Res , vol.23 , pp. 457-475
    • Lee, J.M.1    Haberer, S.A.2    Boughner, D.R.3
  • 29
    • 0028870202 scopus 로고
    • Porcine pulmonary and aortic valves: A comparison of their tensile viscoelastic properties at physiological strain rates
    • Leeson-Dietrich J, Boughner DR, Vesely I. Porcine pulmonary and aortic valves: A comparison of their tensile viscoelastic properties at physiological strain rates. J Heart Valve Diseases 1995;4:88-94.
    • (1995) J Heart Valve Diseases , vol.4 , pp. 88-94
    • Leeson-Dietrich, J.1    Boughner, D.R.2    Vesely, I.3
  • 30
    • 0021787582 scopus 로고
    • The mechanical response of glutaraldehyde fixed bovine pericardium to uniaxial load
    • Trowbridge EA, Black MM, Daniel CL. The mechanical response of glutaraldehyde fixed bovine pericardium to uniaxial load. J Mater Sci 1985;20:114-140.
    • (1985) J Mater Sci , vol.20 , pp. 114-140
    • Trowbridge, E.A.1    Black, M.M.2    Daniel, C.L.3
  • 32
    • 0018425356 scopus 로고
    • Experimental evaluation of aortic homograft valves mounted on flexible support frames and comparison with glutaraldehyde-treated porcine valves
    • Pohlner PG, Thomson FJ, Hjelms E, Barratt-Boyes BG. Experimental evaluation of aortic homograft valves mounted on flexible support frames and comparison with glutaraldehyde-treated porcine valves. J Thorac Cardiovasc Surg 1979;77:287-293.
    • (1979) J Thorac Cardiovasc Surg , vol.77 , pp. 287-293
    • Pohlner, P.G.1    Thomson, F.J.2    Hjelms, E.3    Barratt-Boyes, B.G.4
  • 33
    • 0019493944 scopus 로고
    • Creep and flexibility measurements on plastic frames for tissue heart valves
    • Drury PJ, Wain WH, Richards BW, Ross DN. Creep and flexibility measurements on plastic frames for tissue heart valves. Thorac Cardiovasc Surg 1981;29:105-107.
    • (1981) Thorac Cardiovasc Surg , vol.29 , pp. 105-107
    • Drury, P.J.1    Wain, W.H.2    Richards, B.W.3    Ross, D.N.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.