A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement

Author keywords

Body composition; Densitometry; Dual energy X ray absorptiometry; DXA; Fat mass; Lean mass

Indexed keywords

ADULT; AGED; ARTICLE; BODY COMPOSITION; CALIBRATION; DUAL ENERGY X RAY ABSORPTIOMETRY; FEMALE; HUMAN; MALE; NORMAL HUMAN; PHANTOM; SOFT TISSUE;

EID: 17544391129     PISSN: 09543007     EISSN: None     Source Type: Journal    
DOI: 10.1038/sj.ejcn.1600400     Document Type: Article
Times cited : (37)

References (47)
  • 1
    • 0025780560 scopus 로고
    • Intersite comparison of the Hologic QDR-1000 dual energy x-ray bone densitometer
    • Blake GM, Tong CM & Fogelman I (1991): Intersite comparison of the Hologic QDR-1000 dual energy x-ray bone densitometer. Br. J. Radiol. 64, 440-446.
    • (1991) Br. J. Radiol. , vol.64 , pp. 440-446
    • Blake, G.M.1    Tong, C.M.2    Fogelman, I.3
  • 2
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM & Altman DGM (1986): Statistical methods for assessing agreement between two methods of clinical measurement. The. Lancet. 2, 307-310.
    • (1986) The. Lancet. , vol.2 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.M.2
  • 4
    • 0026778507 scopus 로고
    • Regional changes in body composition by time of year in healthy postmenopausal women
    • Dawson-Hughes B & Harris S. (1992): Regional changes in body composition by time of year in healthy postmenopausal women. Am. J. Clin. Nutr 56, 307-313.
    • (1992) Am. J. Clin. Nutr , vol.56 , pp. 307-313
    • Dawson-Hughes, B.1    Harris, S.2
  • 6
    • 0028053695 scopus 로고
    • Accuracy of dual-energy x-ray absorptiometry for body composition measurements in children
    • Ellis KJ, Shypailo RJ, Pratt J & Pond WG (1994): Accuracy of dual-energy x-ray absorptiometry for body composition measurements in children. Am. J. Clin. Nutr. 60, 660-665.
    • (1994) Am. J. Clin. Nutr. , vol.60 , pp. 660-665
    • Ellis, K.J.1    Shypailo, R.J.2    Pratt, J.3    Pond, W.G.4
  • 7
    • 0026748952 scopus 로고
    • Four-component model for the assessment of body composition in humans: Comparison with alternative methods, and evaluation of the density and hydration of fat-free mass
    • Fuller NJ, Jebb SA, Laskey MA, Coward WA & Elia M (1992): Four-component model for the assessment of body composition in humans: comparison with alternative methods, and evaluation of the density and hydration of fat-free mass. Clin. Sci. 82, 687-693.
    • (1992) Clin. Sci. , vol.82 , pp. 687-693
    • Fuller, N.J.1    Jebb, S.A.2    Laskey, M.A.3    Coward, W.A.4    Elia, M.5
  • 8
    • 0026552865 scopus 로고
    • Assessment of the composition of major body regions by dual-energy x-ray absorptiometry (DEXA) with special reference to limb muscle
    • Fuller NJ, Laskey MA & Elia M (1992): Assessment of the composition of major body regions by dual-energy x-ray absorptiometry (DEXA) with special reference to limb muscle. Clin. Physiol. 12, 253-256.
    • (1992) Clin. Physiol. , vol.12 , pp. 253-256
    • Fuller, N.J.1    Laskey, M.A.2    Elia, M.3
  • 10
    • 0022613416 scopus 로고
    • Measurement of lean body mass and total fat using dual photon absorptiometry
    • Gotfredsen A, Jensen J, Borg J & Christiansen C (1986): Measurement of lean body mass and total fat using dual photon absorptiometry. Metabolism 35, 88-93.
    • (1986) Metabolism , vol.35 , pp. 88-93
    • Gotfredsen, A.1    Jensen, J.2    Borg, J.3    Christiansen, C.4
  • 11
    • 0025787190 scopus 로고
    • Validation of body composition by dual energy X-ray absorptiometry (DEXA)
    • Haarbo J, Gotfredsen A, Hassager C, Christiansen C (1991): Validation of body composition by dual energy X-ray absorptiometry (DEXA). Metabolism. 11, 331-341.
    • (1991) Metabolism , vol.11 , pp. 331-341
    • Haarbo, J.1    Gotfredsen, A.2    Hassager, C.3    Christiansen, C.4
  • 12
    • 0025355265 scopus 로고
    • Influence of fat on bone measurements with dual-energy absorptiometry
    • Hangartner TN & Johnston CC ( 1990): Influence of fat on bone measurements with dual-energy absorptiometry. Bone. and. Mineral. 9, 71-81.
    • (1990) Bone. And. Mineral. , vol.9 , pp. 71-81
    • Hangartner, T.N.1    Johnston, C.C.2
  • 13
    • 0024345735 scopus 로고
    • Dual-photon absorptiometry: Comparison of bone mineral and soft tissue mass measurements in vivo with established methods
    • Heymsfield SB, Wang J, Heshka S, Keyayias JJ & Pierson RN (1989): Dual-photon absorptiometry: comparison of bone mineral and soft tissue mass measurements in vivo with established methods. Am. J. Clin. Nutr. 49, 1283-1289.
    • (1989) Am. J. Clin. Nutr. , vol.49 , pp. 1283-1289
    • Heymsfield, S.B.1    Wang, J.2    Heshka, S.3    Keyayias, J.J.4    Pierson, R.N.5
  • 15
    • 0020213666 scopus 로고
    • Photon mass attenuation and energy-absorption coefficients from 1 keV to 20 MeV
    • Hubbell JH (1982): Photon mass attenuation and energy-absorption coefficients from 1 keV to 20 MeV. Int. J. Appl. Rad. Isot. 33, 1269-1290.
    • (1982) Int. J. Appl. Rad. Isot. , vol.33 , pp. 1269-1290
    • Hubbell, J.H.1
  • 16
    • 0028949386 scopus 로고
    • Dual-energy X-ray absorptiometry measurements of body composition: Effects of depth and tissue thickness, including comparisons with direct analysis
    • Jebb AS, Goldberg GR, Jennings G & Elia M (1995): Dual-energy X-ray absorptiometry measurements of body composition: effects of depth and tissue thickness, including comparisons with direct analysis. Clin. Sci. 88, 319-324.
    • (1995) Clin. Sci. , vol.88 , pp. 319-324
    • Jebb, A.S.1    Goldberg, G.R.2    Jennings, G.3    Elia, M.4
  • 17
    • 0027259473 scopus 로고
    • Assessment of body composition with use of Dual-Energy X-ray Absorptiometry: Evaluation and comparison with other methods
    • Jensen MD, Kanaley JA, Roust LR, OBrien PC, Braun JS, Dunn WL & Wahner HW (1993): Assessment of body composition with use of Dual-Energy X-ray Absorptiometry: Evaluation and comparison with other methods. Mayo. Clin. Proc. 68, 867-873.
    • (1993) Mayo. Clin. Proc. , vol.68 , pp. 867-873
    • Jensen, M.D.1    Kanaley, J.A.2    Roust, L.R.3    Obrien, P.C.4    Braun, J.S.5    Dunn, W.L.6    Wahner, H.W.7
  • 18
    • 0027402954 scopus 로고
    • Determination of body composition - A comparison of dual-energy x-ray absorptiometry and hydrodensitometry
    • Johansson AG, Forslund A, Sjodin A, Mallmin H, Hambraeus L & Ljunghall S (1993): Determination of body composition - a comparison of dual-energy x-ray absorptiometry and hydrodensitometry. Am. J. Clin. Nutr. 57, 323-326.
    • (1993) Am. J. Clin. Nutr. , vol.57 , pp. 323-326
    • Johansson, A.G.1    Forslund, A.2    Sjodin, A.3    Mallmin, H.4    Hambraeus, L.5    Ljunghall, S.6
  • 19
    • 0026012481 scopus 로고
    • Precision and stability of Dual-Energy X-ray Absorptiometry Measurements
    • Johnson J & Dawson-Hughes B (1991): Precision and stability of Dual-Energy X-ray Absorptiometry Measurements. Calc. Tiss. Inter. 49, 174-178.
    • (1991) Calc. Tiss. Inter. , vol.49 , pp. 174-178
    • Johnson, J.1    Dawson-Hughes, B.2
  • 21
    • 26344457136 scopus 로고
    • Assessment of body composition by dual x-ray absorptiometry using a simultaneously scanned calibration standard
    • abstr.
    • Kelly TL, Richardson TL, Berger N, Steiger P & Stein, JA (1991): Assessment of body composition by dual x-ray absorptiometry using a simultaneously scanned calibration standard. Med. Sci. Sports. Exerc. 23, Suppl. (4), S90 (abstr.).
    • (1991) Med. Sci. Sports. Exerc. , vol.23 , Issue.4 SUPPL.
    • Kelly, T.L.1    Richardson, T.L.2    Berger, N.3    Steiger, P.4    Stein, J.A.5
  • 23
    • 0026503118 scopus 로고
    • The influence of tissue depth and composition on the performance of the Lunar dual-energy X-ray absorptiometer whole-body scanning mode
    • Laskey MA, Lyttle KD, Flaxman ME & Barber RW (1992): The influence of tissue depth and composition on the performance of the Lunar dual-energy X-ray absorptiometer whole-body scanning mode. Eur. J. Clin. Nutr 46, 39-45.
    • (1992) Eur. J. Clin. Nutr , vol.46 , pp. 39-45
    • Laskey, M.A.1    Lyttle, K.D.2    Flaxman, M.E.3    Barber, R.W.4
  • 25
    • 0025021940 scopus 로고
    • Precision of regional bone mineral measurements obtained from total-body scans
    • LeBlanc AD, Schneider VS, Engelbretson DA & Evans HJ (1990): Precision of regional bone mineral measurements obtained from total-body scans. J. Nucl. Med. 31, 43-45.
    • (1990) J. Nucl. Med. , vol.31 , pp. 43-45
    • Leblanc, A.D.1    Schneider, V.S.2    Engelbretson, D.A.3    Evans, H.J.4
  • 26
    • 0025303078 scopus 로고
    • Dual-energy x-ray absorptiometry for total-body and regional bone-mineral and soft-tissue composition
    • Mazess R, Barden HS, Bisek JP & Hanson J (1990): Dual-energy x-ray absorptiometry for total-body and regional bone-mineral and soft-tissue composition. Am. J. Clin. Nutr. 51, 1106-1112.
    • (1990) Am. J. Clin. Nutr. , vol.51 , pp. 1106-1112
    • Mazess, R.1    Barden, H.S.2    Bisek, J.P.3    Hanson, J.4
  • 29
    • 0023855527 scopus 로고
    • Long term precision of dual photon absorptiometry in the lumbar spine in clinical settings
    • Nilas L, Hassager C & Christiansen C (1988): Long term precision of dual photon absorptiometry in the lumbar spine in clinical settings. Bone. and. Mineral. 3, 305-315.
    • (1988) Bone. And. Mineral. , vol.3 , pp. 305-315
    • Nilas, L.1    Hassager, C.2    Christiansen, C.3
  • 31
    • 0026034394 scopus 로고
    • Longitudinal precision of Dual-Energy X-ray Absorptiometry in a multicenter study
    • Orwoll E, Oviatt S and the Nafarelin bone study group (1991): Longitudinal precision of Dual-Energy X-ray Absorptiometry in a multicenter study. J. Bone. Miner. Res. 6, 191-197.
    • (1991) J. Bone. Miner. Res. , vol.6 , pp. 191-197
    • Orwoll, E.1    Oviatt, S.2
  • 33
    • 0019868386 scopus 로고
    • Total body bone mineral and lean body mass by dual-photon absorptiometry. Theory and measurement procedure
    • Peppler WW & Mazess RB (1981): Total body bone mineral and lean body mass by dual-photon absorptiometry. Theory and measurement procedure. Calc. Tiss. Inter. 33, 353-359.
    • (1981) Calc. Tiss. Inter. , vol.33 , pp. 353-359
    • Peppler, W.W.1    Mazess, R.B.2
  • 34
    • 0025908745 scopus 로고
    • Measuring body fat: Calibrating the rulers. Intermethod comparisons in 389 normal Caucasian subjects
    • Endocrinol. Metab. 24
    • Pierson RN, Wang J, Heymsfield SB, Russell-Aulet M & Mazariegos M (1991): Measuring body fat: calibrating the rulers. Intermethod comparisons in 389 normal Caucasian subjects. Am. J. Physiol. 261, (Endocrinol. Metab). 24, E103-E108.
    • (1991) Am. J. Physiol. , vol.261
    • Pierson, R.N.1    Wang, J.2    Heymsfield, S.B.3    Russell-Aulet, M.4    Mazariegos, M.5
  • 36
    • 0025908786 scopus 로고
    • In Vitro comparability of dual energy X-ray Absorptiometry (DEXA) bone densitometers
    • Rencken ML, Murano R, Drinkwater BL & Chestnut III, CH (1991): In Vitro comparability of dual energy X-ray Absorptiometry (DEXA) bone densitometers. CaIc. Tiss. Inte. 48, 245-248.
    • (1991) CaIc. Tiss. Inte. , vol.48 , pp. 245-248
    • Rencken, M.L.1    Murano, R.2    Drinkwater, B.L.3    Chestnut III, C.H.4
  • 37
    • 0028273565 scopus 로고
    • The four-compartment models in body composition data from a study with dual-energy x-ray absorptiometry and near-infrared interactance on 815 normal subjects
    • Rico H, Revilla M, Villa L, Ruiz-Contreras D, Hernandez ER & Alverez de Buergo M (1994): The four-compartment models in body composition data from a study with dual-energy x-ray absorptiometry and near-infrared interactance on 815 normal subjects. Metabolism. 43, 417-422.
    • (1994) Metabolism , vol.43 , pp. 417-422
    • Rico, H.1    Revilla, M.2    Villa, L.3    Ruiz-Contreras, D.4    Hernandez, E.R.5    Alverez De Buergo, M.6
  • 38
    • 0027373635 scopus 로고
    • Use of dual-energy x-ray absorptiometry in body composition studies: Not yet a gold standard
    • Roubenoff R, Kehayias JJ, Dawson-Hughes B & Heymsfield SB (1993): Use of dual-energy x-ray absorptiometry in body composition studies: not yet a gold standard. Am. J. Clin. Nutr. 58, 589-591.
    • (1993) Am. J. Clin. Nutr. , vol.58 , pp. 589-591
    • Roubenoff, R.1    Kehayias, J.J.2    Dawson-Hughes, B.3    Heymsfield, S.B.4
  • 39
    • 0004282518 scopus 로고
    • SAS Institute Inc.: Cary, NC
    • SAS (1990): SAS users guide: statistics, SAS Institute Inc.: Cary, NC
    • (1990) SAS Users Guide: Statistics
  • 40
    • 0025782934 scopus 로고
    • Precision of Total-Body and Regional Bone Mineral Measurement by Dual-Energy X-ray Absorptiometry
    • Seto H, Kageyama M, Nomura K, Kakishita M & Tonami S (1991): Precision of Total-Body and Regional Bone Mineral Measurement by Dual-Energy X-ray Absorptiometry. Radiat. Med. 9, 110-113.
    • (1991) Radiat. Med. , vol.9 , pp. 110-113
    • Seto, H.1    Kageyama, M.2    Nomura, K.3    Kakishita, M.4    Tonami, S.5
  • 42
    • 0027225470 scopus 로고
    • Accuracy of measurements of body composition by dual-energy x-ray absorptiometry in vivo
    • Svendsen OL, Haarbo J, Hassager C & Christiansen C (1993): Accuracy of measurements of body composition by dual-energy x-ray absorptiometry in vivo. Am. J. Clin. Nutr. 57, 605-608.
    • (1993) Am. J. Clin. Nutr. , vol.57 , pp. 605-608
    • Svendsen, O.L.1    Haarbo, J.2    Hassager, C.3    Christiansen, C.4
  • 43
    • 0025753066 scopus 로고
    • Measurement of body fat in elderly subjects by dual-energy x-ray absorptiometry, bioelectrical impedance, and anthropometry
    • Svendsen OL, Haarbo J, Heitmann BL, Gotfredsen A & Christiansen C (1991): Measurement of body fat in elderly subjects by dual-energy x-ray absorptiometry, bioelectrical impedance, and anthropometry. Am. J. Clin. Nutr. 53, 1117-1123.
    • (1991) Am. J. Clin. Nutr. , vol.53 , pp. 1117-1123
    • Svendsen, O.L.1    Haarbo, J.2    Heitmann, B.L.3    Gotfredsen, A.4    Christiansen, C.5
  • 45
    • 0028364532 scopus 로고
    • Quality control of bone densitometry in a national healthy survey (NHANES III) using three mobile examination centers
    • Wahner HW, Looker A, Dunn WL, Walters LC, Hauser MF & Novak C (1994): Quality control of bone densitometry in a national healthy survey (NHANES III) using three mobile examination centers. J. Bone. Miner. Res. 9, 951-960.
    • (1994) J. Bone. Miner. Res. , vol.9 , pp. 951-960
    • Wahner, H.W.1    Looker, A.2    Dunn, W.L.3    Walters, L.C.4    Hauser, M.F.5    Novak, C.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.