메뉴 건너뛰기




Volumn 1, Issue 6, 2004, Pages 728-731

Absent ventricular tachycardia detection in a biventricular implantable cardioverter-defibrillator due to intradevice interaction with a rate smoothing pacing algorithm

Author keywords

Biventricular pacing; Device interaction; Implantable cardioverter defibrillator; Pacing; Rate smoothing; Ventricular tachycardia

Indexed keywords

ADULT; ALGORITHM; ARTICLE; BRADYCARDIA; CARDIOVERSION; CASE REPORT; CONGESTIVE HEART FAILURE; DEFIBRILLATOR; DISEASE ASSOCIATION; ELECTROCARDIOGRAPHY; HEART ATRIUM PACING; HEART HEMODYNAMICS; HEART LEFT VENTRICLE FUNCTION; HEART VENTRICLE FIBRILLATION; HEART VENTRICLE TACHYCARDIA; HUMAN; ISCHEMIC HEART DISEASE; MALE; PRIORITY JOURNAL; QRS COMPLEX;

EID: 11144262281     PISSN: 15475271     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.hrthm.2004.08.013     Document Type: Article
Times cited : (22)

References (6)
  • 2
    • 0033762014 scopus 로고    scopus 로고
    • Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection
    • K. Shivkumar Z. Feliciano N.G. Boyle I. Wiener Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection J Cardiovasc Electrophysiol 11 2000 1285-1288
    • (2000) J. Cardiovasc. Electrophysiol. , vol.11 , pp. 1285-1288
    • Shivkumar, K.1    Feliciano, Z.2    Boyle, N.G.3    Wiener, I.4
  • 3
    • 1542379958 scopus 로고    scopus 로고
    • Delayed detection of stable ventricular tachycardia in a dual-chamber implantable cardioverter defibrillator: What is the mechanism?
    • B. Strohmer C. Schernthaner M. Pichler Delayed detection of stable ventricular tachycardia in a dual-chamber implantable cardioverter defibrillator what is the mechanism? J Cardiovasc Electrophysiol 15 2004 244-246
    • (2004) J. Cardiovasc. Electrophysiol. , vol.15 , pp. 244-246
    • Strohmer, B.1    Schernthaner, C.2    Pichler, M.3
  • 4
    • 0036107976 scopus 로고    scopus 로고
    • Impaired detection of ventricular tachyarrhythmias by a rate-smoothing algorithm in dual-chamber implantable defibrillators: Intradevice interactions
    • M. Glikson A.L. Beeman D.M. Luria D.L. Hayes P.A. Friedman Impaired detection of ventricular tachyarrhythmias by a rate-smoothing algorithm in dual-chamber implantable defibrillators intradevice interactions J Cardiovasc Electrophysiol 13 2002 312-318
    • (2002) J. Cardiovasc. Electrophysiol. , vol.13 , pp. 312-318
    • Glikson, M.1    Beeman, A.L.2    Luria, D.M.3    Hayes, D.L.4    Friedman, P.A.5
  • 5
    • 0347482416 scopus 로고    scopus 로고
    • Ventricular rate stabilization algorithm of ICD causing dual chamber pacing during ventricular tachycardia
    • S.S. Barold Ventricular rate stabilization algorithm of ICD causing dual chamber pacing during ventricular tachycardia J Interv Card Electrophysiol 9 2003 397-400
    • (2003) J. Interv. Card. Electrophysiol. , vol.9 , pp. 397-400
    • Barold, S.S.1
  • 6
    • 0036910817 scopus 로고    scopus 로고
    • Ventricular rate stabilization for the prevention of pause dependent ventricular tachyarrhythmias: Results from a prospective study in 309 ICD recipients
    • on behalf of Worldwide Gem DR Study Group
    • G.C. Grönefeld C.W. Israel V. Padmanabhan J. Koehler A. Cuijpers S.H. Hohnloser on behalf of Worldwide Gem DR Study Group Ventricular rate stabilization for the prevention of pause dependent ventricular tachyarrhythmias results from a prospective study in 309 ICD recipients Pacing Clin Electrophysiol 25 2002 1708-1714
    • (2002) Pacing Clin. Electrophysiol. , vol.25 , pp. 1708-1714
    • Grönefeld, G.C.1    Israel, C.W.2    Padmanabhan, V.3    Koehler, J.4    Cuijpers, A.5    Hohnloser, S.H.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.