Volumn 63, Issue 6, 2003, Pages 431-440

Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus

Author keywords

Finger vs. ear; Sampling site; Sampling variation

Indexed keywords

GLUCOSE;

EID: 0142062122     PISSN: 00365513     EISSN: None     Source Type: Journal    
DOI: 10.1080/00365510310002590     Document Type: Article
Times cited : (15)

References (12)
  • 1
    • 0031964320 scopus 로고    scopus 로고
    • Screening for type 2 diabetes
    • American Diabetes Association. Screening for type 2 diabetes. Diabetes care 1998; 21: S20-2.
    • (1998) Diabetes Care , vol.21
  • 2
    • 17744400689 scopus 로고    scopus 로고
    • Optimization of preanalytical conditions and analysis of plasma glucose: Impact of the new WHO and ADA recommendations on diagnosis of diabetes mellitus (1)
    • Stahl M, Jørgensen LGM, Hyltoft Petersen P, Brandslund I, Olivarius de F, Borch-Johnsen K. Optimization of preanalytical conditions and analysis of plasma glucose: impact of the new WHO and ADA recommendations on diagnosis of diabetes mellitus (1). Scand J Clin Lab Invest 2001; 61: 169-80.
    • (2001) Scand J Clin Lab Invest , vol.61 , pp. 169-180
    • Stahl, M.1    Jørgensen, L.G.M.2    Hyltoft Petersen, P.3    Brandslund, I.4    De Olivarius, F.5    Borch-Johnsen, K.6
  • 4
    • 84896707223 scopus 로고    scopus 로고
    • Accu-Chek Inform, kit insert
    • Roche Diagnostic. Accu-Chek Inform, kit insert.
    • Roche Diagnostic
  • 6
    • 34250935832 scopus 로고
    • Die enzymatische Bestimmung von Glucose and Fructose nebeneinander
    • Schmidt FH. Die enzymatische Bestimmung von Glucose and Fructose nebeneinander. Klin Wschr 1961; 39: 1244.
    • (1961) Klin Wschr , vol.39 , pp. 1244
    • Schmidt, F.H.1
  • 7
    • 84875308663 scopus 로고    scopus 로고
    • Gaithersburg, MD 20899: NIST, December
    • National Institute of Standards and Technology. Standard reference material 965. Gaithersburg, MD 20899: NIST, December 1996.
    • (1996) Standard Reference Material 965
  • 8
    • 0032884749 scopus 로고    scopus 로고
    • Measuring agreement in method comparison studies
    • Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 135-60.
    • (1999) Stat Methods Med Res , vol.8 , pp. 135-160
    • Bland, J.M.1    Altman, D.G.2
  • 9
    • 0028173754 scopus 로고
    • Comparison of tests for glycated haemoglobin and fasting and two hour plasma glucose concentrations as diagnostic methods for diabetes
    • McCance R, Hanson RL, Charles M-A, Jacobsson LTH, Pettitt DJ, Bennett PH, Knowler WC. Comparison of tests for glycated haemoglobin and fasting and two hour plasma glucose concentrations as diagnostic methods for diabetes. Br Med J 1994; 38: 1323-8.
    • (1994) Br Med J , vol.38 , pp. 1323-1328
    • McCance, R.1    Hanson, R.L.2    Charles, M.-A.3    Jacobsson, L.T.H.4    Pettitt, D.J.5    Bennett, P.H.6    Knowler, W.C.7
  • 11
    • 0035002473 scopus 로고    scopus 로고
    • Evaluation of systemic and random factors in measurements of fasting plasma glucose as the basis for analytical quality specifications in the diagnosis of diabetes. 3. Impact of the new WHO and ADA recommendations on diagnosis of diabetes mellitus
    • Hyltoft Petersen P, Brandslund I, Jørgensen L, Stahl M, de Fine Olivarius N, Borch Johnsen K. Evaluation of systemic and random factors in measurements of fasting plasma glucose as the basis for analytical quality specifications in the diagnosis of diabetes. 3. Impact of the new WHO and ADA recommendations on diagnosis of diabetes mellitus. Scand J Clin Lab Invest 2001; 61: 191-204.
    • (2001) Scand J Clin Lab Invest , vol.61 , pp. 191-204
    • Hyltoft Petersen, P.1    Brandslund, I.2    Jørgensen, L.3    Stahl, M.4    De Fine Olivarius, N.5    Borch Johnsen, K.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.