Volumn 57, Issue 8, 2003, Pages 977-987

Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers - Bone mineral density and body composition studies

Author keywords

Body composition; Bone mineral; Cross calibration; Dual energy X ray absorptiometers; Fan beam; Pencil beam

Indexed keywords

ADULT; ANALYTICAL EQUIPMENT; ARTICLE; BODY COMPOSITION; BODY FAT; BONE DENSITY; BONE MINERAL; CALIBRATION; CONTROLLED STUDY; DUAL ENERGY X RAY ABSORPTIOMETRY; FEMALE; FEMUR NECK; HUMAN; IN VITRO STUDY; IN VIVO STUDY; INTERMETHOD COMPARISON; LUMBAR SPINE; MALE; PHANTOM; RADIODENSITOMETRY;

EID: 0043025553     PISSN: 09543007     EISSN: None     Source Type: Journal    
DOI: 10.1038/sj.ejcn.1601633     Document Type: Article
Times cited : (61)

References (36)
  • 1
    • 0029123724 scopus 로고
    • Cross calibration of QDR-2000 and QDR-1000 dual-energy X-ray densitometers for bone mineral and soft-tissue measurements
    • Abrahamsen B, Gram J, Hansen TB & Beck-Nielsen H (1995): Cross calibration of QDR-2000 and QDR-1000 dual-energy X-ray densitometers for bone mineral and soft-tissue measurements. Bone 16, 385-390.
    • (1995) Bone , vol.16 , pp. 385-390
    • Abrahamsen, B.1    Gram, J.2    Hansen, T.B.3    Beck-Nielsen, H.4
  • 3
    • 0027504762 scopus 로고
    • Dual X-ray absorptiometry: A comparison between fan beam and pencil beam scans
    • Blake GM, Parker JC, Buxton FMA & Fogelman I (1993). Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans. Br. J. Radial. 66, 902-906.
    • (1993) Br. J. Radiol. , vol.66 , pp. 902-906
    • Blake, G.M.1    Parker, J.C.2    Buxton, F.M.A.3    Fogelman, I.4
  • 5
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM & Altman DG (1986): Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1, 307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 6
    • 0029666110 scopus 로고    scopus 로고
    • Cross-calibration of a fanbeam X-ray densitometer with a pencil-beam system
    • Bouyoucef SE, Cullum ID & Ell PJ (1996): Cross-calibration of a fanbeam X-ray densitometer with a pencil-beam system. Br. J. Radiol. 69, 522-531.
    • (1996) Br. J. Radiol. , vol.69 , pp. 522-531
    • Bouyoucef, S.E.1    Cullum, I.D.2    Ell, P.J.3
  • 9
    • 0028077289 scopus 로고
    • Switching from DXA pencil-beam to fan-beam. II: Studies in vivo
    • Eiken P, Kolthoff N, Barenholdt O, Hermansen F & Pors NS (1994): Switching from DXA pencil-beam to fan-beam. II: studies in vivo. Bone 15, 667-670.
    • (1994) Bone , vol.15 , pp. 667-670
    • Eiken, P.1    Kolthoff, N.2    Barenholdt, O.3    Hermansen, F.4    Pors, N.S.5
  • 10
    • 0028043666 scopus 로고
    • Switching from DXA pencil-beam to fan-beam. I: Studies in vitro at four centres
    • Eiken P, Barenholdt O, Bjorn JL, Gram J & Pors NS (1994): Switching from DXA pencil-beam to fan-beam. I: studies in vitro at four centres. Bone 15, 671-676.
    • (1994) Bone , vol.15 , pp. 671-676
    • Eiken, P.1    Barenholdt, O.2    Bjorn, J.L.3    Gram, J.4    Pors, N.S.5
  • 11
    • 0031709713 scopus 로고    scopus 로고
    • Bone mineral and body composition measurements: Cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers
    • Ellis KJ & Shypailo RJ (1998): Bone mineral and body composition measurements: cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers. J. Bone Miner. Res. 13, 1613-1619.
    • (1998) J. Bone Miner. Res. , vol.13 , pp. 1613-1619
    • Ellis, K.J.1    Shypailo, R.J.2
  • 12
    • 0027532266 scopus 로고
    • Cross-calibration of DXA equipment: Upgrading from a Hologic QDR 1000/W to a QDR 2000
    • Faulkner KG, Gluer C-C, Estilo M & Genant HK (1993): Cross-calibration of DXA equipment: upgrading from a Hologic QDR 1000/W to a QDR 2000. Calcif. Tissue Int. 52, 79-84.
    • (1993) Calcif. Tissue Int. , vol.52 , pp. 79-84
    • Faulkner, K.G.1    Gluer, C.-C.2    Estilo, M.3    Genant, H.K.4
  • 13
    • 0028956926 scopus 로고
    • Evaluation of dual-energy X-ray absorptiometry bone mineral measurement - Comparison of a single-beam and fan-beam design: The effect of osteophytic calcification on spine bone mineral density
    • Franck H, Munz M & Scherrer M (1995): Evaluation of dual-energy X-ray absorptiometry bone mineral measurement - comparison of a single-beam and fan-beam design: the effect of osteophytic calcification on spine bone mineral density. Calcif. Tissue Int. 56: 192-195.
    • (1995) Calcif. Tissue Int. , vol.56 , pp. 192-195
    • Franck, H.1    Munz, M.2    Scherrer, M.3
  • 15
    • 0031020430 scopus 로고    scopus 로고
    • Correcting the magnification error of fan beam densitometers
    • Griffiths MR, Noakes KA & Pocock NA (1997): Correcting the magnification error of fan beam densitometers. J. Bone Miner. Res. 12, 119-123.
    • (1997) J. Bone Miner. Res. , vol.12 , pp. 119-123
    • Griffiths, M.R.1    Noakes, K.A.2    Pocock, N.A.3
  • 16
    • 0029080928 scopus 로고
    • The European spine phantom - A tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT
    • Kalender WA, Felsenberg D, Genant HK, Fischer M, Dequeker J & Reeve J (1995): The European spine phantom - a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT. Eur. J. Radial. 20, 83-92.
    • (1995) Eur. J. Radiol. , vol.20 , pp. 83-92
    • Kalender, W.A.1    Felsenberg, D.2    Genant, H.K.3    Fischer, M.4    Dequeker, J.5    Reeve, J.6
  • 17
    • 0033784694 scopus 로고    scopus 로고
    • Follow-up of individual patients on two DXA scanners of the same manufacturer
    • Kolta S, Ravaud P, Fechtenbaum J, Dougados M & Roux C (2000): Follow-up of individual patients on two DXA scanners of the same manufacturer. Osteoporos Int 11, 709-713.
    • (2000) Osteoporos. Int. , vol.11 , pp. 709-713
    • Kolta, S.1    Ravaud, P.2    Fechtenbaum, J.3    Dougados, M.4    Roux, C.5
  • 19
    • 0031469665 scopus 로고    scopus 로고
    • Evalution of the European spine phantom in a multi-centre clinical trial
    • Lees B, Garland SW, Walton C & Stevenson JC (1997): Evalution of the European spine phantom in a multi-centre clinical trial. Osteoporos. Int. 7, 570-574.
    • (1997) Osteoporos. Int. , vol.7 , pp. 570-574
    • Lees, B.1    Garland, S.W.2    Walton, C.3    Stevenson, J.C.4
  • 20
    • 0033800567 scopus 로고    scopus 로고
    • Evaluation of differences between fan-beam and pencil-beam densitometers
    • Mazess RB, & Barden HS (2000): Evaluation of differences between fan-beam and pencil-beam densitometers. Calcif. Tissue Int. 67, 291-296.
    • (2000) Calcif. Tissue Int. , vol.67 , pp. 291-296
    • Mazess, R.B.1    Barden, H.S.2
  • 21
    • 0025303078 scopus 로고
    • Dual-energy X-ray absorptiometry for total-body and regional bone-mineral and soft-tissue composition
    • Mazess RB, Barden HS, Bisek JP & Hanson J. (1990): Dual-energy X-ray absorptiometry for total-body and regional bone-mineral and soft-tissue composition. Am. J. Clin. Nutr. 51, 1106-1112.
    • (1990) Am. J. Clin. Nutr. , vol.51 , pp. 1106-1112
    • Mazess, R.B.1    Barden, H.S.2    Bisek, J.P.3    Hanson, J.4
  • 23
    • 0034108621 scopus 로고    scopus 로고
    • Axial and total-body bone densitometry using a narrow-angle fan-beam
    • Mazess RB, Hanson JA, Payne R, Nord R & Wilson M (2000): Axial and total-body bone densitometry using a narrow-angle fan-beam. Osteoporos. Int. 11, 158-166.
    • (2000) Osteoporos. Int. , vol.11 , pp. 158-166
    • Mazess, R.B.1    Hanson, J.A.2    Payne, R.3    Nord, R.4    Wilson, M.5
  • 24
    • 4244140313 scopus 로고    scopus 로고
    • Anthropomorphic quality assurance phantoms for bone densitometry
    • 14th International Bone Densitometry Workshop, Warnemünde, Germany
    • Milner R, Berry E, Marsden AJ, Smith AH & Smith MA (2000): Anthropomorphic quality assurance phantoms for bone densitometry. 14th International Bone Densitometry Workshop, Warnemünde, Germany. Osteoporos. Int. 11(Suppl 3), S27.
    • (2000) Osteoporos. Int. , vol.11 , Issue.SUPPL. 3
    • Milner, R.1    Berry, E.2    Marsden, A.J.3    Smith, A.H.4    Smith, M.A.5
  • 25
    • 0042096662 scopus 로고    scopus 로고
    • A new hydroxyapatite spine phantom
    • 12th International Bone Densitometry Workshop, Crieff, Scotland
    • Nord RH, Bisek JP & Miller CG (1997): A new hydroxyapatite spine phantom. 12th International Bone Densitometry Workshop, Crieff, Scotland. Osteoporos. Int. 7: 287.
    • (1997) Osteoporos. Int. , vol.7 , pp. 287
    • Nord, R.H.1    Bisek, J.P.2    Miller, C.G.3
  • 26
    • 0034046337 scopus 로고    scopus 로고
    • Evaluation of a new DXA fan-beam instrument for measuring body composition
    • Nord RH, Homuth JR, Hanson JA & Mazess RB (2000): Evaluation of a new DXA fan-beam instrument for measuring body composition. Ann. N. Y. Acad. Sci. 904, 118-125.
    • (2000) Ann. N. Y. Acad. Sci. , vol.904 , pp. 118-125
    • Nord, R.H.1    Homuth, J.R.2    Hanson, J.A.3    Mazess, R.B.4
  • 27
    • 0032079364 scopus 로고    scopus 로고
    • A total body phantom for use with Lunar dual-energy X-ray absorptiometers
    • Oldroyd B, Milner R, Smith AH & Smith MA (1998a): A total body phantom for use with Lunar dual-energy X-ray absorptiometers. Appl. Radiat. Isotopes. 49, 525-526.
    • (1998) Appl. Radiat. Isotopes , vol.49 , pp. 525-526
    • Oldroyd, B.1    Milner, R.2    Smith, A.H.3    Smith, M.A.4
  • 30
    • 0033925848 scopus 로고    scopus 로고
    • Differences between dual X-ray absorptiometry using pencil beam and fan beam modes and their determinants in vivo and in vitro
    • Ruetsche AG, Lippuner K, Jaeger P & Casez J-P (2000): Differences between dual X-ray absorptiometry using pencil beam and fan beam modes and their determinants in vivo and in vitro. J. Clin. Densit. 3, 157-166.
    • (2000) J. Clin. Densit. , vol.3 , pp. 157-166
    • Ruetsche, A.G.1    Lippuner, K.2    Jaeger, P.3    Casez, J.-P.4
  • 32
    • 0027986512 scopus 로고
    • Comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers and other techniques used for whole-body soft tissue measurements
    • Tothill P, Avenell A, Love J & Reid DM (1994): Comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers and other techniques used for whole-body soft tissue measurements. Eur. J. Clin. Nutr. 48, 781-794.
    • (1994) Eur. J. Clin. Nutr. , vol.48 , pp. 781-794
    • Tothill, P.1    Avenell, A.2    Love, J.3    Reid, D.M.4
  • 33
    • 0035084650 scopus 로고    scopus 로고
    • Comparisons between a pencil beam and two fan beam dual energy X-ray absorptiometers used for measuring total body bone and soft tissue
    • Tothill P, Hannan WJ & Wilkinson S (2001): Comparisons between a pencil beam and two fan beam dual energy X-ray absorptiometers used for measuring total body bone and soft tissue. Br. J. Radiol. 74: 166-176.
    • (2001) Br. J. Radiol. , vol.74 , pp. 166-176
    • Tothill, P.1    Hannan, W.J.2    Wilkinson, S.3
  • 34
    • 0026535327 scopus 로고
    • Body composition assessment: Dual energy X-ray absorptiometry (DEXA) compared to reference methods
    • Van Loan MD & Mayclin PL (1992): Body composition assessment: dual energy X-ray absorptiometry (DEXA) compared to reference methods. Eur. J. Clin. Nutr. 46, 125-130.
    • (1992) Eur. J. Clin. Nutr. , vol.46 , pp. 125-130
    • Van Loan, M.D.1    Mayclin, P.L.2
  • 35
    • 0023727897 scopus 로고
    • Comparison of dual-energy X-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine
    • Wahner HW, Dunn WL, Brown ML, Morin RL & Riggs L (1988): Comparison of dual-energy X-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine. Mayo Clin. Proc. 63, 1075-1084.
    • (1988) Mayo Clin. Proc. , vol.63 , pp. 1075-1084
    • Wahner, H.W.1    Dunn, W.L.2    Brown, M.L.3    Morin, R.L.4    Riggs, L.5
  • 36
    • 0030003105 scopus 로고    scopus 로고
    • Influence of changes in peritoneal fluid on body composition measurements by dual-energy X-ray absorptiometry in patients receiving continuous ambulatory peritoneal dialysis
    • Woodrow G, Oldroyd B, Turney JH & Smith MA (1996): Influence of changes in peritoneal fluid on body composition measurements by dual-energy X-ray absorptiometry in patients receiving continuous ambulatory peritoneal dialysis. Am. J. Clin. Nutr. 64, 237-241.
    • (1996) Am. J. Clin. Nutr. , vol.64 , pp. 237-241
    • Woodrow, G.1    Oldroyd, B.2    Turney, J.H.3    Smith, M.A.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.