메뉴 건너뛰기




Volumn 113, Issue 5, 2003, Pages 853-858

Stapes prosthesis attachment: The effect of crimping on sound transfer in otosclerosis surgery

Author keywords

Laser Doppler; Otosclerosis; Prosthesis; Stapedotomy; Stapes surgery

Indexed keywords

ACOUSTICS; ADULT; AGED; ALGORITHM; ARTICLE; CONTROLLED STUDY; ELECTROCHEMICAL ANALYSIS; ENDOSCOPY; EXPERIMENTAL MODEL; FEMALE; HEARING ACUITY; HUMAN; HUMAN TISSUE; INCUS; INTERFEROMETRY; LASER DOPPLER FLOWMETRY; MALE; MASTOIDECTOMY; OTOSCLEROSIS; PHOTOGRAPHY; PRIORITY JOURNAL; PROSTHESIS; SCANNING ELECTRON MICROSCOPY; SOUND TRANSMISSION; STAPEDECTOMY; STAPES; SURGICAL APPROACH; SURGICAL TECHNIQUE; TEMPORAL BONE; TREATMENT FAILURE;

EID: 0038642097     PISSN: 0023852X     EISSN: None     Source Type: Journal    
DOI: 10.1097/00005537-200305000-00015     Document Type: Article
Times cited : (67)

References (16)
  • 2
    • 0023834908 scopus 로고
    • Thirty years of stapes surgery
    • Shea JJ Jr. Thirty years of stapes surgery. J Laryngol Otol 1988;102:14-19.
    • (1988) J Laryngol Otol , vol.102 , pp. 14-19
    • Shea J.J., Jr.1
  • 3
    • 84971697868 scopus 로고
    • Hearing improvement after stapedectomy: Up to 19 years' follow-up period
    • Birch L, Elbrond O, Pedersen U. Hearing improvement after stapedectomy: up to 19 years' follow-up period. J Laryngol Otol 1986;100:1-7.
    • (1986) J Laryngol Otol , vol.100 , pp. 1-7
    • Birch, L.1    Elbrond, O.2    Pedersen, U.3
  • 4
    • 0014115357 scopus 로고
    • Factors influencing results in stapes surgery: A long-term evaluation
    • Shambaugh G. Factors influencing results in stapes surgery: a long-term evaluation. Ann Otol Rhinol Laryngol 1967;76:599-602.
    • (1967) Ann Otol Rhinol Laryngol , vol.76 , pp. 599-602
    • Shambaugh, G.1
  • 6
    • 0019406710 scopus 로고
    • The loose wire syndrome
    • McGee TM. The loose wire syndrome. Laryngoscope 1981;91:1478-1483.
    • (1981) Laryngoscope , vol.91 , pp. 1478-1483
    • McGee, T.M.1
  • 7
    • 0035191289 scopus 로고    scopus 로고
    • Malleo-stapedotomy in revision surgery for otosclerosis
    • Fisch U, Oezbilen G, Huber AM. Malleo-stapedotomy in revision surgery for otosclerosis. Otol Neurotol 2001;22:776-785.
    • (2001) Otol Neurotol , vol.22 , pp. 776-785
    • Fisch, U.1    Oezbilen, G.2    Huber, A.M.3
  • 8
    • 0001923985 scopus 로고    scopus 로고
    • 2 laser
    • Carrasco VN, Pillsbury HC III, eds. New York: Thieme
    • 2 laser. In: Carrasco VN, Pillsbury HC III, eds. Revision Otologic Surgery. New York: Thieme, 1997:3-21.
    • (1997) Revision Otologic Surgery , pp. 3-21
    • Lesinski, S.G.1
  • 10
    • 0035106722 scopus 로고    scopus 로고
    • Evaluation of eardrum laser Doppler interferometry as a diagnostic tool
    • Huber A, Schwab C, Linder T. Evaluation of eardrum laser Doppler interferometry as a diagnostic tool. Laryngoscope 2001;111:501-507.
    • (2001) Laryngoscope , vol.111 , pp. 501-507
    • Huber, A.1    Schwab, C.2    Linder, T.3
  • 12
    • 0032936194 scopus 로고    scopus 로고
    • Analysis of the best site on the stapes footplate for ossicular chain reconstruction
    • Asai M, Huber A, Goode R. Analysis of the best site on the stapes footplate for ossicular chain reconstruction. Acta Otolaryngol (Stockh) 1999;119:356-361.
    • (1999) Acta Otolaryngol (Stockh) , vol.119 , pp. 356-361
    • Asai, M.1    Huber, A.2    Goode, R.3
  • 13
    • 0023108520 scopus 로고
    • Measurement of the ossicular vibration ratio in human temporal bones by use of a video measuring system
    • Gyo k, Aritomo H, Goode RL. Measurement of the ossicular vibration ratio in human temporal bones by use of a video measuring system. Acta Otolaryngol (Stockh) 1987;103:87-95.
    • (1987) Acta Otolaryngol (Stockh) , vol.103 , pp. 87-95
    • Gyo, K.1    Aritomo, H.2    Goode, R.L.3
  • 14
    • 0028875644 scopus 로고
    • Sound pressure gain produced by the human middle ear
    • Kurokawa H, Goode RL. Sound pressure gain produced by the human middle ear. Otolaryngol Head Neck Surg 1995;113:349-355.
    • (1995) Otolaryngol Head Neck Surg , vol.113 , pp. 349-355
    • Kurokawa, H.1    Goode, R.L.2
  • 15
    • 0034848213 scopus 로고    scopus 로고
    • Comparative electron microscopic study of the surface structure of gold, Teflon and titanium stapes prostheses
    • Kwok P, Fisch U, Strutz J, Jacob P. Comparative electron microscopic study of the surface structure of gold, Teflon and titanium stapes prostheses. Otol Neurotol 2001;22:608-613.
    • (2001) Otol Neurotol , vol.22 , pp. 608-613
    • Kwok, P.1    Fisch, U.2    Strutz, J.3    Jacob, P.4
  • 16
    • 0036255447 scopus 로고    scopus 로고
    • Stapes surgery: How precisely do different prostheses attach to the long process of the incus with different instruments and different surgeons?
    • Kwok P, Fisch U, Strutz J, May J. Stapes surgery: how precisely do different prostheses attach to the long process of the incus with different instruments and different surgeons? Otol Neurotol 2002;23:289-295.
    • (2002) Otol Neurotol , vol.23 , pp. 289-295
    • Kwok, P.1    Fisch, U.2    Strutz, J.3    May, J.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.