Evaluation of the i-STAT Portable Clinical Analyzer for point-of-care blood testing in the intensive care units of a University Children's Hospital
a a a a a a a |
Author keywords
Neonatal pediatric intensive care, clinical chemistry assays; Point of care testing
|