Volumn 37, Issue 12, 2000, Pages 1075-1081

Determination of gamma-ray efficiency curves for volume samples by the combination of monte carlo simulations and point source calibration

Author keywords

Calibration; Efficiency; Gamma ray; Ge semi conductor detector; Monte Carlo method; Self absorption; Simulation; Source preparation; Standard point source; Volume sample

Indexed keywords

COMPUTER SIMULATION; DENSITY (SPECIFIC GRAVITY); DETECTORS; GAMMA RAYS; GERMANIUM; MONTE CARLO METHODS;

EID: 0034472759     PISSN: 00223131     EISSN: None     Source Type: Journal    
DOI: 10.1080/18811248.2000.9714994     Document Type: Article
Times cited : (21)

References (8)
  • 1
    • 85010562934 scopus 로고
    • Characteristics of prepared gamma-ray calibration sources for radioactivity measurement of environmental and radiation control samples
    • Samat, S. B., et al: Characteristics of prepared gamma-ray calibration sources for radioactivity measurement of environmental and radiation control samples, JAERI-Tech 95-100, (1995).
    • (1995) Jaeri-Tech , pp. 95-100
    • Samat, S.B.1
  • 2
    • 85047574801 scopus 로고
    • The 7 ray peak efficiency determination for volume samples using a radioactive point source
    • [in Japanese]
    • Honda, T., et al: The 7 ray peak efficiency determination for volume samples using a radioactive point source, Radioisotopes, 27, 367 (1978), [in Japanese].
    • (1978) Radioisotopes , vol.27 , pp. 367
    • Honda, T.1
  • 3
    • 0029035145 scopus 로고
    • Hpge detector absolute-peak-efficiency calibration by using the esolan program
    • Wang, T. K., et al: HPGe detector absolute-peak-efficiency calibration by using the ESOLAN program, Appl Radiat. Isot., 46, 933 (1995).
    • (1995) Appl Radiat. Isot , vol.46 , pp. 933
    • Wang, T.K.1
  • 4
    • 85010562927 scopus 로고    scopus 로고
    • Correction methods of 7-ray self-absorption in bulk sample
    • [in Japanese]
    • Noguchi, M., et al: Correction methods of 7-ray self-absorption in bulk sample, Radioisotopes, 49, 189 (2000), [in Japanese].
    • (2000) Radioisotopes , vol.49 , pp. 189
    • Noguchi, M.1
  • 5
    • 0001009932 scopus 로고
    • Efficiency calibration of semiconductor detectors by primary standard sources and monte carlo calculations
    • Debertin, K., Grosswendt, B.: Efficiency calibration of semiconductor detectors by primary standard sources and Monte Carlo calculations, Nucl. Instrum. Methods, 203, 343 (1982).
    • (1982) Nucl. Instrum. Methods , vol.203 , pp. 343
    • Debertin, K.1    Grosswendt, B.2
  • 6
    • 0002011999 scopus 로고
    • Monte carlo calculation of peak efficiencies of ge(Li) and ge detectors to volumi-nal sources and comparison with environmental radioactivity measurement
    • Nakamura, T., Suzuki, T.: Monte Carlo calculation of peak efficiencies of Ge(Li) and Ge detectors to volumi-nal sources and comparison with environmental radioactivity measurement, Nucl Instrum. Methods, 205, 211 (1983).
    • (1983) Nucl Instrum. Methods , vol.205 , pp. 211
    • Nakamura, T.1    Suzuki, T.2
  • 7
    • 24544438274 scopus 로고    scopus 로고
    • A calibration technique for radioactive gas monitor with a built-in germanium detector
    • Hiroshima, Japan, 2000, P-3a-151
    • Yoshida, M., et al: A calibration technique for radioactive gas monitor with a built-in germanium detector, Proc. 10th Int. Congress of the Int. Radiat. Prot. Assoc. (IRPA-10), Hiroshima, Japan, 2000, P-3a-151 (2000).
    • (2000) Proc. 10Th Int. Congress of the Int. Radiat. Prot. Assoc.
    • Yoshida, M.1
  • 8
    • 0003395844 scopus 로고    scopus 로고
    • A general monte carlo n-particle transport code, version 4b-manual
    • Briesmeister, J. F. (ed.): A general Monte Carlo N-Particle transport code, version 4B-Manual, LA-12625-M, (1997).
    • (1997) LA
    • Briesmeister, J.F.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.