Volumn 88, Issue 6, 2000, Pages 2143-2150

Analysis of the mechanisms of expiratory asynchrony in pressure support ventilation: A mathematical approach

Author keywords

Mathematical modeling; Mechanical ventilation; Patient ventilator synchrony

Indexed keywords

AIRFLOW; ARTICLE; ARTIFICIAL VENTILATION; BREATHING MECHANICS; LUNG VENTILATION; MATHEMATICAL MODEL; PRIORITY JOURNAL;

EID: 0033933102     PISSN: 87507587     EISSN: None     Source Type: Journal    
DOI: 10.1152/jappl.2000.88.6.2143     Document Type: Article
Times cited : (102)

References (21)
  • 1
    • 0000238869 scopus 로고    scopus 로고
    • Partial ventilatory support
    • edited by Marini JJ and Slutsky AS. New York: Marcel Dekker
    • Aslanian P and Brochard LJ. Partial ventilatory support. In: Physiological Basis of Ventilatory Support, edited by Marini JJ and Slutsky AS. New York: Marcel Dekker, 1998, p. 817-846.
    • (1998) Physiological Basis of Ventilatory Support , pp. 817-846
    • Aslanian, P.1    Brochard, L.J.2
  • 2
    • 0020617352 scopus 로고
    • Respiratory mechanics during halothane anesthesia and anesthesia-paralysis in humans
    • Behrakis PK, Higgs BD, Baydur A, Zin WA, and Milic-Emili J. Respiratory mechanics during halothane anesthesia and anesthesia-paralysis in humans. J Appl Physiol 55: 1085-1092, 1983.
    • (1983) J Appl Physiol , vol.55 , pp. 1085-1092
    • Behrakis, P.K.1    Higgs, B.D.2    Baydur, A.3    Zin, W.A.4    Milic-Emili, J.5
  • 3
    • 0032890595 scopus 로고    scopus 로고
    • Effects of pressure ramp slope values on the work of breathing during pressure support ventilation in restrictive patients
    • Bonmarchand G, Chevron V, Menard J-F, Girault C, Morits-Berthelot F, and Pasquis P, and Leroy J. Effects of pressure ramp slope values on the work of breathing during pressure support ventilation in restrictive patients. Crit Care Med 27: 715-722, 1999.
    • (1999) Crit Care Med , vol.27 , pp. 715-722
    • Bonmarchand, G.1    Chevron, V.2    Menard, J.-F.3    Girault, C.4    Morits-Berthelot, F.5    Pasquis, P.6    Leroy, J.7
  • 4
    • 0032469807 scopus 로고    scopus 로고
    • Pressure support ventilation, patient-ventilator synchrony, and ventilator algorithm
    • Branson RD and Campbell RS. Pressure support ventilation, patient-ventilator synchrony, and ventilator algorithm. Respir Care 43: 1045-1047, 1998.
    • (1998) Respir Care , vol.43 , pp. 1045-1047
    • Branson, R.D.1    Campbell, R.S.2
  • 5
    • 0003288954 scopus 로고
    • New generation of microprocessor-based ventilators
    • edited by Tobin MJ. New York: McGraw-Hill
    • Branson RD and Chatburn RL. New generation of microprocessor-based ventilators. In: Principles and Practice of Mechanical Ventilation, edited by Tobin MJ. New York: McGraw-Hill, 1994, p. 1247-1256.
    • (1994) Principles and Practice of Mechanical Ventilation , pp. 1247-1256
    • Branson, R.D.1    Chatburn, R.L.2
  • 6
    • 0030892323 scopus 로고    scopus 로고
    • Performance characteristics of bilevel pressure ventilators: A lung model study
    • Bunburaphong T, Imanaka H, Nishimura M, Hess D, and Kacmarek RM. Performance characteristics of bilevel pressure ventilators: a lung model study. Chest 111: 1050-1060, 1997.
    • (1997) Chest , vol.111 , pp. 1050-1060
    • Bunburaphong, T.1    Imanaka, H.2    Nishimura, M.3    Hess, D.4    Kacmarek, R.M.5
  • 7
    • 0031468718 scopus 로고    scopus 로고
    • Patient-ventilator trigger asynchrony in prolonged mechanical ventilation
    • Chao DC, Scheinhorn DJ, and Stearn-Hassenpflug M. Patient-ventilator trigger asynchrony in prolonged mechanical ventilation. Chest 112:1592-1599, 1997.
    • (1997) Chest , vol.112 , pp. 1592-1599
    • Chao, D.C.1    Scheinhorn, D.J.2    Stearn-Hassenpflug, M.3
  • 8
    • 0029014749 scopus 로고
    • An analysis of desynchronization between the spontaneous breathing and ventilator during inspiratory pressure support
    • Fabry B, Guttmann J, Eberhard L, Bauer T, Haberthur C, and Wolff G. An analysis of desynchronization between the spontaneous breathing and ventilator during inspiratory pressure support. Chest 107: 1387-1394, 1995.
    • (1995) Chest , vol.107 , pp. 1387-1394
    • Fabry, B.1    Guttmann, J.2    Eberhard, L.3    Bauer, T.4    Haberthur, C.5    Wolff, G.6
  • 9
    • 33750837483 scopus 로고    scopus 로고
    • Incidence and types of desynchronization between spontaneous breaths and ventilator assistance with pressure support during routine weaning from mechanical ventilation in postoperative patients
    • Garcia-Raimundo M, Fraga R, Saz T, Aguilar G, Belda FJ, and Maruenda A. Incidence and types of desynchronization between spontaneous breaths and ventilator assistance with pressure support during routine weaning from mechanical ventilation in postoperative patients (Abstract). Crit Care Med 27: S335, 1999.
    • (1999) Crit Care Med , vol.27
    • Garcia-Raimundo, M.1    Fraga, R.2    Saz, T.3    Aguilar, G.4    Belda, F.J.5    Maruenda, A.6
  • 10
    • 0029121830 scopus 로고
    • Variability of patient-ventilator interaction with pressure support ventilation in patients with chronic obstructive pulmonary disease
    • Jubran A, Van de Graaff WB, and Tobin MJ. Variability of patient-ventilator interaction with pressure support ventilation in patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 152: 129-136, 1995.
    • (1995) Am J Respir Crit Care Med , vol.152 , pp. 129-136
    • Jubran, A.1    Van De Graaff, W.B.2    Tobin, M.J.3
  • 11
    • 0022612038 scopus 로고
    • Respiratory function during pressure support ventilation
    • MacIntyre NR. Respiratory function during pressure support ventilation. Chest 89: 677-683, 1986.
    • (1986) Chest , vol.89 , pp. 677-683
    • MacIntyre, N.R.1
  • 12
    • 0023874216 scopus 로고
    • Weaning from mechanical ventilator support: Volume assisting intermittent breaths vs. pressure assisting every breath
    • MacIntyre NR. Weaning from mechanical ventilator support: volume assisting intermittent breaths vs. pressure assisting every breath. Respir Care 33: 121-125, 1988.
    • (1988) Respir Care , vol.33 , pp. 121-125
    • MacIntyre, N.R.1
  • 13
    • 0003175498 scopus 로고
    • Relationship between neuromuscular respiratory drive and ventilatory output
    • Bethesda, MD: Am. Physiol. Soc., sect. 3
    • Milic-Emili J and Zin WA. Relationship between neuromuscular respiratory drive and ventilatory output. In: Handbook of Physiology. The Respiratory System. Bethesda, MD: Am. Physiol. Soc., 1986, sect. 3, vol. III, pt. 2, p. 631-646.
    • (1986) Handbook of Physiology. The Respiratory System , vol.3 , Issue.PT. 2 , pp. 631-646
    • Milic-Emili, J.1    Zin, W.A.2
  • 14
    • 0031033608 scopus 로고    scopus 로고
    • Patient-ventilator interaction and inspiratory effort during pressure support ventilation in patients with different pathologies
    • Nava S, Bruschi C, Fracchia C, Braschi A, and Rubini F. Patient-ventilator interaction and inspiratory effort during pressure support ventilation in patients with different pathologies. Eur Respir J 10: 177-183, 1997.
    • (1997) Eur Respir J , vol.10 , pp. 177-183
    • Nava, S.1    Bruschi, C.2    Fracchia, C.3    Braschi, A.4    Rubini, F.5
  • 17
    • 0032469970 scopus 로고    scopus 로고
    • Effects of different pressure support termination on patient-ventilator synchrony
    • Yamada Y and Du H-L. Effects of different pressure support termination on patient-ventilator synchrony. Respir Care 43: 1048-1057, 1998.
    • (1998) Respir Care , vol.43 , pp. 1048-1057
    • Yamada, Y.1    Du, H.-L.2
  • 18
    • 0027946170 scopus 로고
    • Respiratory muscle pressure analysis in pressure-support ventilation
    • Yamada Y, Shigeta M, Suwa K, and Hanaoka K. Respiratory muscle pressure analysis in pressure-support ventilation. J Appl Physiol 77: 2237-2243, 1994.
    • (1994) J Appl Physiol , vol.77 , pp. 2237-2243
    • Yamada, Y.1    Shigeta, M.2    Suwa, K.3    Hanaoka, K.4
  • 19
    • 0000596815 scopus 로고
    • Proportional assist ventilation and pressure support ventilation: Similarities and differences
    • edited by Marini JJ and Roussos C. Berlin: Springer
    • Younes M. Proportional assist ventilation and pressure support ventilation: similarities and differences. In: Ventilatory Failure, edited by Marini JJ and Roussos C. Berlin: Springer, 1992, p. 361-380.
    • (1992) Ventilatory Failure , pp. 361-380
    • Younes, M.1
  • 20
    • 0027323905 scopus 로고
    • Patient-ventilator interaction with pressure-assisted modalities of ventilatory support
    • Younes M. Patient-ventilator interaction with pressure-assisted modalities of ventilatory support. Semin Respir Med 14: 299-322, 1993.
    • (1993) Semin Respir Med , vol.14 , pp. 299-322
    • Younes, M.1
  • 21
    • 0019972004 scopus 로고
    • Single-breath method for measurement of respiratory mechanics in anesthetized animals
    • Zin WA, Pengelly LD, and Milic-Emili J. Single-breath method for measurement of respiratory mechanics in anesthetized animals. J Appl Physiol 52: 1266-1271, 1982.
    • (1982) J Appl Physiol , vol.52 , pp. 1266-1271
    • Zin, W.A.1    Pengelly, L.D.2    Milic-Emili, J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.