Volumn 5, Issue 6, 1997, Pages 419-435

Magnetic resonance imaging and the reduction of motion artifacts: Review of the principles

Author keywords

Artifacts; Correction; Magnetic resonance imaging; Motion; Post processing

Indexed keywords

ARTIFACT REDUCTION; BLOOD FLOW VELOCITY; COMPUTER ASSISTED TOMOGRAPHY; DIAGNOSTIC IMAGING; DIAGNOSTIC TEST; ECHOGRAPHY; FOURIER TRANSFORMATION; IMAGE ANALYSIS; IMAGE QUALITY; NUCLEAR MAGNETIC RESONANCE IMAGING; PRIORITY JOURNAL; REVIEW; SIGNAL NOISE RATIO;

EID: 0031397501     PISSN: 09287329     EISSN: None     Source Type: Journal    
DOI: 10.3233/thc-1997-5602     Document Type: Review
Times cited : (33)

References (65)
  • 3
    • 0026116322 scopus 로고
    • A respiratory motion artifact reduction method in magnetic resonance imaging of the chest
    • E. Atalar and L. Onural, A respiratory motion artifact reduction method in magnetic resonance imaging of the chest, IEEE Transactions on Medical Imaging 10(1) (1991), 11-24.
    • (1991) IEEE Transactions on Medical Imaging , vol.10 , Issue.1 , pp. 11-24
    • Atalar, E.1    Onural, L.2
  • 4
    • 0021741851 scopus 로고
    • Blood flow effects in magnetic resonance imaging
    • L. Axel, Blood flow effects in magnetic resonance imaging, American Journal of Radiology 143 (1984), 1157-1166.
    • (1984) American Journal of Radiology , vol.143 , pp. 1157-1166
    • Axel, L.1
  • 5
    • 0023890468 scopus 로고
    • The RF coil as a sensitive motion detector for magnetic resonance imaging
    • D. Buikman, T. Helzel and P. Röschmann, The RF coil as a sensitive motion detector for magnetic resonance imaging, Magnetic Resonance Imaging 6(2) (1988), 281-289.
    • (1988) Magnetic Resonance Imaging , vol.6 , Issue.2 , pp. 281-289
    • Buikman, D.1    Helzel, T.2    Röschmann, P.3
  • 7
    • 0021152307 scopus 로고
    • A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: Dependency on tissue type, NMR frequency, temperature, species, excision and age
    • P.A. Bottomley, T.H. Forster, R.E. Argersinger and L.M. Pfeifer, A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: Dependency on tissue type, NMR frequency, temperature, species, excision and age, Medical Physics 11 (1994), 425-448.
    • (1994) Medical Physics , vol.11 , pp. 425-448
    • Bottomley, P.A.1    Forster, T.H.2    Argersinger, R.E.3    Pfeifer, L.M.4
  • 10
    • 0022739029 scopus 로고
    • Optimal control solutions to the magnetic resonance selective excitation problem
    • S. Conolly, D. Nishimura and A. Macovski, Optimal control solutions to the magnetic resonance selective excitation problem, IEEE Transactions on Medical Imaging 5(2) (1986), 106-115.
    • (1986) IEEE Transactions on Medical Imaging , vol.5 , Issue.2 , pp. 106-115
    • Conolly, S.1    Nishimura, D.2    Macovski, A.3
  • 15
    • 0024453449 scopus 로고
    • Adaptive technique for high-definition MR imaging of moving structures
    • R.L. Ehman and J.P. Felmlee, Adaptive technique for high-definition MR imaging of moving structures, Radiology 173 (1989), 255-263.
    • (1989) Radiology , vol.173 , pp. 255-263
    • Ehman, R.L.1    Felmlee, J.P.2
  • 16
    • 0022592172 scopus 로고
    • Rapid NMR imaging of dynamic processes using the FLASH technique
    • J. Frahm, A. Haase and D. Matthau, Rapid NMR imaging of dynamic processes using the FLASH technique, Magnetic Resonance in Medicine 3(2) (1986), 321-327.
    • (1986) Magnetic Resonance in Medicine , vol.3 , Issue.2 , pp. 321-327
    • Frahm, J.1    Haase, A.2    Matthau, D.3
  • 19
    • 0027258745 scopus 로고
    • Use of a projection reconstruction method to decrease motion sensitivity in diffusion-weighted MRI
    • A.F. Gmitro and A.L. Alexander, Use of a projection reconstruction method to decrease motion sensitivity in diffusion-weighted MRI, Magnetic Resonance in Medicine 29 (1993), 835-838.
    • (1993) Magnetic Resonance in Medicine , vol.29 , pp. 835-838
    • Gmitro, A.F.1    Alexander, A.L.2
  • 20
    • 0023013859 scopus 로고
    • Reducing motion artifacts in two-dimensional Fourier transform imaging
    • E.M. Haacke and J.L. Patrick, Reducing motion artifacts in two-dimensional Fourier transform imaging, Magnetic Resonance Imaging 4 (1986), 359-376.
    • (1986) Magnetic Resonance Imaging , vol.4 , pp. 359-376
    • Haacke, E.M.1    Patrick, J.L.2
  • 21
    • 0023217321 scopus 로고
    • Improving MR image quality in the presence of motion by using rephasing gradients
    • E.M. Haacke and G.W. Lenz, Improving MR image quality in the presence of motion by using rephasing gradients, American Journal of Radiology 148 (1987), 1251-1258.
    • (1987) American Journal of Radiology , vol.148 , pp. 1251-1258
    • Haacke, E.M.1    Lenz, G.W.2
  • 22
    • 0023285728 scopus 로고
    • Pseudo-gating: Elimination of periodic motion artifacts in magnetic resonance imaging without gating
    • E.M. Haacke, G.W. Lenz and A.D. Nelson, Pseudo-gating: Elimination of periodic motion artifacts in magnetic resonance imaging without gating, Magnetic Resonance in Medicine 4(2) (1987), 162-174.
    • (1987) Magnetic Resonance in Medicine , vol.4 , Issue.2 , pp. 162-174
    • Haacke, E.M.1    Lenz, G.W.2    Nelson, A.D.3
  • 23
    • 0025213880 scopus 로고
    • Snapshot FLASH MRI. Applications to T1, T2 and chemical-shift imaging
    • A. Haase, Snapshot FLASH MRI. Applications to T1, T2 and chemical-shift imaging, Magnetic Resonance in Medicine 13 (1990), 77-89.
    • (1990) Magnetic Resonance in Medicine , vol.13 , pp. 77-89
    • Haase, A.1
  • 25
    • 0026688914 scopus 로고
    • Motion artifact suppression: A review of post-processing techniques
    • M. Hedley and H. Yan, Motion artifact suppression: a review of post-processing techniques, Magnetic Resonance Imaging 10 (1992), 627-635.
    • (1992) Magnetic Resonance Imaging , vol.10 , pp. 627-635
    • Hedley, M.1    Yan, H.2
  • 27
    • 0025040013 scopus 로고
    • Adaptive technique for three-dimensional MR imaging of moving structures
    • H.W. Korin, J.P. Felmlee, R.L. Ehman and S.J. Riederer, adaptive technique for three-dimensional MR imaging of moving structures, Radiology 177 (1990), 217-221.
    • (1990) Radiology , vol.177 , pp. 217-221
    • Korin, H.W.1    Felmlee, J.P.2    Ehman, R.L.3    Riederer, S.J.4
  • 30
    • 0021327514 scopus 로고
    • Cardiac imaging using gated magnetic resonance
    • P. Lanzer, E.H. Botvinick and N.B. Schriller, Cardiac imaging using gated magnetic resonance, Radiology 150(1) (1984), 121-127.
    • (1984) Radiology , vol.150 , Issue.1 , pp. 121-127
    • Lanzer, P.1    Botvinick, E.H.2    Schriller, N.B.3
  • 31
    • 33746803555 scopus 로고
    • Image formation by induced local interactions: Examples employing nuclear magnetic resonance
    • P.C. Lauterbur, Image formation by induced local interactions: Examples employing nuclear magnetic resonance, Nature 242 (1973), 190-191.
    • (1973) Nature , vol.242 , pp. 190-191
    • Lauterbur, P.C.1
  • 32
    • 0027426040 scopus 로고
    • Generalized k-space analysis and correction of motion effects in MR imaging
    • M.L. Lauzon and B.K. Rutt, Generalized k-space analysis and correction of motion effects in MR imaging, Magn. Reson. Med. 30 (1993), 438-446.
    • (1993) Magn. Reson. Med. , vol.30 , pp. 438-446
    • Lauzon, M.L.1    Rutt, B.K.2
  • 33
    • 0024469533 scopus 로고
    • Retrospective cardiac gating: A review of technical aspects and future directions
    • G.W. Lenz, E.M. Haacke and R.D. White, Retrospective cardiac gating: A review of technical aspects and future directions, Magnetic Resonance Imaging 7(5) (1989), 445-455.
    • (1989) Magnetic Resonance Imaging , vol.7 , Issue.5 , pp. 445-455
    • Lenz, G.W.1    Haacke, E.M.2    White, R.D.3
  • 34
    • 0022542905 scopus 로고
    • Comparison of respiratory triggering and gating techniques for the removal of respiratory artifacts in MR imaging
    • C.E. Lewis, F.S. Prato, D.J. Drost and R.L. Nicholson, Comparison of respiratory triggering and gating techniques for the removal of respiratory artifacts in MR imaging, Radiology 160(3) (1986), 803-810.
    • (1986) Radiology , vol.160 , Issue.3 , pp. 803-810
    • Lewis, C.E.1    Prato, F.S.2    Drost, D.J.3    Nicholson, R.L.4
  • 36
    • 0022340998 scopus 로고
    • A systematic design procedure for selective pulses in NMR imaging
    • D.J. Lurie, A systematic design procedure for selective pulses in NMR imaging, Magnetic Resonance Imaging 3 (1985), 235-243.
    • (1985) Magnetic Resonance Imaging , vol.3 , pp. 235-243
    • Lurie, D.J.1
  • 40
    • 0023742854 scopus 로고
    • Motion artifact reduction in MR imaging of the abdomen: Gradient moment nulling versus respiratory-sorted phase encoding
    • D.G. Mitchell, S. Vinitski, D.L. Burk, D.W. Levy and M.D. Rifkin, Motion artifact reduction in MR imaging of the abdomen: Gradient moment nulling versus respiratory-sorted phase encoding, Radiology 169(1) (1988), 155-160.
    • (1988) Radiology , vol.169 , Issue.1 , pp. 155-160
    • Mitchell, D.G.1    Vinitski, S.2    Burk, D.L.3    Levy, D.W.4    Rifkin, M.D.5
  • 47
    • 0028095056 scopus 로고
    • Magnetic resonance velocity imaging using a fast spiral phase contrast sequence
    • G.B. Pike, C.H. Meyer, T.J. Brosnan and N.J. Pelc, Magnetic resonance velocity imaging using a fast spiral phase contrast sequence, Magnetic Resonance in Medicine 32(4) (1994), 476-483.
    • (1994) Magnetic Resonance in Medicine , vol.32 , Issue.4 , pp. 476-483
    • Pike, G.B.1    Meyer, C.H.2    Brosnan, T.J.3    Pelc, N.J.4
  • 48
    • 0029144252 scopus 로고
    • Analytical resolution and noise characteristics of linearly reconstructed magnetic resonance data with arbitrary k-space sampling
    • J.G. Pipe and J.L. Duerk, Analytical resolution and noise characteristics of linearly reconstructed magnetic resonance data with arbitrary k-space sampling, Magnetic Resonance in Medicine 34 (1995), 170-178.
    • (1995) Magnetic Resonance in Medicine , vol.34 , pp. 170-178
    • Pipe, J.G.1    Duerk, J.L.2
  • 49
    • 0029952524 scopus 로고    scopus 로고
    • Three-dimensional respiratory-gated MR angiography of coronary arteries: Comparison with conventional coronary angiography
    • J.C. Post, A.C. van Rossum, M.B. Hofman, J. Valk and C.A. Visser, Three-dimensional respiratory-gated MR angiography of coronary arteries: comparison with conventional coronary angiography, American Journal of Roentgenology 166(6) (1996), 1399-1404.
    • (1996) American Journal of Roentgenology , vol.166 , Issue.6 , pp. 1399-1404
    • Post, J.C.1    Van Rossum, A.C.2    Hofman, M.B.3    Valk, J.4    Visser, C.A.5
  • 50
    • 0345937757 scopus 로고
    • Maximizing signal-to-noise and contrast-to-noise ratios in spin-echo imaging using non-standard flip angles
    • T.J. Provost and R.E. Hendrick, Maximizing signal-to-noise and contrast-to-noise ratios in spin-echo imaging using non-standard flip angles, Magnetic Resonance Imaging 4(2) (1986), 105-110.
    • (1986) Magnetic Resonance Imaging , vol.4 , Issue.2 , pp. 105-110
    • Provost, T.J.1    Hendrick, R.E.2
  • 52
    • 0030425801 scopus 로고    scopus 로고
    • Quantification and improvement of the signal-to-noise ratio in a NMR image acquisition procedure
    • J. Sijbers, P. Scheunders, N. Bonnet, E. Raman and D. Van Dyck, Quantification and improvement of the signal-to-noise ratio in a NMR image acquisition procedure, Magnetic Resonance Imaging 14(10) (1996), 1157-1163.
    • (1996) Magnetic Resonance Imaging , vol.14 , Issue.10 , pp. 1157-1163
    • Sijbers, J.1    Scheunders, P.2    Bonnet, N.3    Raman, E.4    Van Dyck, D.5
  • 54
    • 0023275610 scopus 로고
    • Motion artifact reduction with fast spin-echo imaging
    • D.D. Stark, R.E. Hendrick, P.F. Hahn and J.T. Ferrucci, Motion artifact reduction with fast spin-echo imaging, Radiology 164 (1987), 183-191.
    • (1987) Radiology , vol.164 , pp. 183-191
    • Stark, D.D.1    Hendrick, R.E.2    Hahn, P.F.3    Ferrucci, J.T.4
  • 55
    • 0021166164 scopus 로고
    • Cranio-caudal movements of the liver, pancreas and kidneys in respiration
    • I. Suramo, M. Paivansalo and V. Myilyla, Cranio-caudal movements of the liver, pancreas and kidneys in respiration, Acta Radiologica 25(2) (1984), 129-131.
    • (1984) Acta Radiologica , vol.25 , Issue.2 , pp. 129-131
    • Suramo, I.1    Paivansalo, M.2    Myilyla, V.3
  • 59
    • 0030026702 scopus 로고    scopus 로고
    • Navigator-echo-based real-time respiratory gating and triggering for reduction of respiratory effects in three-dimensional coronary MR angiography
    • Y. Wang, P.J. Rossman, R.C. Grimm, S.J. Riederer and R.L. Ehman, Navigator-echo-based real-time respiratory gating and triggering for reduction of respiratory effects in three-dimensional coronary MR angiography, Radiology 198(1) (1996), 55-60.
    • (1996) Radiology , vol.198 , Issue.1 , pp. 55-60
    • Wang, Y.1    Rossman, P.J.2    Grimm, R.C.3    Riederer, S.J.4    Ehman, R.L.5
  • 61
    • 0022995073 scopus 로고
    • Suppression of respiratory motion artifacts in magnetic resonance imaging
    • M.L. Wood and R.M. Henkelman, Suppression of respiratory motion artifacts in magnetic resonance imaging, Medical Physics 13(6) (1986), 794-805.
    • (1986) Medical Physics , vol.13 , Issue.6 , pp. 794-805
    • Wood, M.L.1    Henkelman, R.M.2
  • 63
    • 0029184205 scopus 로고
    • Planar-motion correction with use of k-space data acquired in Fourier MR imaging
    • M.L. Wood, M.J. Shivji and P.L. Stanchev, Planar-motion correction with use of k-space data acquired in Fourier MR imaging, Journal of Magnetic Resonance Imaging 5(1) (1995), 57-64.
    • (1995) Journal of Magnetic Resonance Imaging , vol.5 , Issue.1 , pp. 57-64
    • Wood, M.L.1    Shivji, M.J.2    Stanchev, P.L.3
  • 65
    • 0030045301 scopus 로고    scopus 로고
    • Automatic motion correction for breast MR imaging
    • C.S. Zuo, A. Jiang, B.L. Buff, T.G. Mahon and T.Z. Wong, Automatic motion correction for breast MR imaging, Radiology 198(3) (1996), 903-906.
    • (1996) Radiology , vol.198 , Issue.3 , pp. 903-906
    • Zuo, C.S.1    Jiang, A.2    Buff, B.L.3    Mahon, T.G.4    Wong, T.Z.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.