Volumn 15, Issue 2, 1997, Pages 147-156

A new photo-oscillometric method employing the delta-algorithm for accurate blood pressure measurement

Author keywords

Auscultatory method; Blood pressure determination; Indirect blood pressure; Intra arterial blood pressure

Indexed keywords

ADULT; ALGORITHM; ARTICLE; AUSCULTATION; BLOOD PRESSURE MEASUREMENT; BRACHIAL ARTERY; CLINICAL ARTICLE; CLINICAL TRIAL; CONTROLLED CLINICAL TRIAL; CONTROLLED STUDY; CUFF; DIAGNOSTIC ACCURACY; DIASTOLIC BLOOD PRESSURE; FEMALE; HUMAN; INTERMETHOD COMPARISON; MALE; OSCILLATION; OSCILLOMETRY; PRIORITY JOURNAL; PULSE WAVE; SENSOR; SYSTOLIC BLOOD PRESSURE; TIME;

EID: 0031059674     PISSN: 02636352     EISSN: None     Source Type: Journal    
DOI: 10.1097/00004872-199715020-00005     Document Type: Article
Times cited : (14)

References (29)
  • 1
    • 84973130223 scopus 로고    scopus 로고
    • The centenary of blood pressure measurement: A tribute to Scipione Riva-Rocci
    • Zanchetti A, Mancia G: The centenary of blood pressure measurement: a tribute to Scipione Riva-Rocci. J Hypertens 1996, 14:1-12.
    • (1996) J Hypertens , vol.14 , pp. 1-12
    • Zanchetti, A.1    Mancia, G.2
  • 2
    • 0023176777 scopus 로고
    • An examination of the differences that occur between direct and indirect blood pressure measurement
    • Rebenson-Piano M, Holm K, Powers M: An examination of the differences that occur between direct and indirect blood pressure measurement. Heart Lung 1987, 26:285-294.
    • (1987) Heart Lung , vol.26 , pp. 285-294
    • Rebenson-Piano, M.1    Holm, K.2    Powers, M.3
  • 4
    • 0014183591 scopus 로고
    • Recommendations for human blood pressure determination by sphygmomanometers
    • Kirkendall WM, Burton AC, Epstein FH, Freis ED: Recommendations for human blood pressure determination by sphygmomanometers. Circulation 1967, 36:980-988.
    • (1967) Circulation , vol.36 , pp. 980-988
    • Kirkendall, W.M.1    Burton, A.C.2    Epstein, F.H.3    Freis, E.D.4
  • 9
    • 0019523842 scopus 로고
    • Direct blood pressure measurement - Dynamic response requirements
    • Gardner RM: Direct blood pressure measurement - dynamic response requirements. Anesthesiology 1981, 54:227-236.
    • (1981) Anesthesiology , vol.54 , pp. 227-236
    • Gardner, R.M.1
  • 10
    • 8044251127 scopus 로고
    • Physical response requirements of pressure transducers for the reproduction of physiological phenomena
    • Wood EH: Physical response requirements of pressure transducers for the reproduction of physiological phenomena. Trans Am Inst Electrical Engineers [I] 1956, 75:32-40.
    • (1956) Trans Am Inst Electrical Engineers [I] , vol.75 , pp. 32-40
    • Wood, E.H.1
  • 11
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JA, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement Lancet 1986, i:307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.A.1    Altman, D.G.2
  • 12
    • 0001055284 scopus 로고
    • The accuracy of clinical measurements of arterial blood pressure
    • Ragan C, Bordley J III: The accuracy of clinical measurements of arterial blood pressure. Bull Johns Hopk Hosp 1941, 69:504-528.
    • (1941) Bull Johns Hopk Hosp , vol.69 , pp. 504-528
    • Ragan, C.1    Bordley III, J.2
  • 13
    • 0000421330 scopus 로고
    • Measurement of blood-pressure: Comparison of intra-arterial and cuff values
    • Holland WW, Humerfelt S: Measurement of blood-pressure: comparison of intra-arterial and cuff values. BMJ 1962, 2:1241-1243.
    • (1962) BMJ , vol.2 , pp. 1241-1243
    • Holland, W.W.1    Humerfelt, S.2
  • 14
    • 0008208941 scopus 로고
    • Sphygmomanometer cuff size and the accuracy of indirect measurement of blood pressure
    • Karvonen MJ, Telivuo LJ, Järvinenen EJK: Sphygmomanometer cuff size and the accuracy of indirect measurement of blood pressure. Am J Cardiol 1964, 13:688-693.
    • (1964) Am J Cardiol , vol.13 , pp. 688-693
    • Karvonen, M.J.1    Telivuo, L.J.2    Järvinenen, E.J.K.3
  • 15
    • 85047677668 scopus 로고
    • The indirect method of recording blood pressure
    • Raftery EB, Ward AP: The indirect method of recording blood pressure. Cardiovasc Res 1968, 2:210-218.
    • (1968) Cardiovasc Res , vol.2 , pp. 210-218
    • Raftery, E.B.1    Ward, A.P.2
  • 16
    • 0014870189 scopus 로고
    • Validity of blood pressure measurement with cuff in the arm and forearm
    • Forsberg SÅ, De Guzman M, Berlind S: Validity of blood pressure measurement with cuff in the arm and forearm. Acta Med Scand 1970, 188:389-396.
    • (1970) Acta Med Scand , vol.188 , pp. 389-396
    • Forsberg, S.Å.1    De Guzman, M.2    Berlind, S.3
  • 17
    • 0021021728 scopus 로고
    • The accuracy of auscultatory blood pressure measurement in the elderly
    • Nielsen PE: The accuracy of auscultatory blood pressure measurement in the elderly. Acta Med Scand Suppl 1983, 676:39-51.
    • (1983) Acta Med Scand Suppl , vol.676 , pp. 39-51
    • Nielsen, P.E.1
  • 19
    • 0025008045 scopus 로고
    • Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations
    • Gravlee GP, Brockschmidt JK: Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations. J Clin Monit 1990, 6:284-298.
    • (1990) J Clin Monit , vol.6 , pp. 284-298
    • Gravlee, G.P.1    Brockschmidt, J.K.2
  • 20
    • 0025187383 scopus 로고
    • Pseudohypertension in hypertensive patients on multiple drug therapy
    • Weisser B, Veiling P, Geller C, Kraft K, Göbel B, Vetter H, et al.: Pseudohypertension in hypertensive patients on multiple drug therapy. J Hypertens 1990, 8 (suppl 4):579-581.
    • (1990) J Hypertens , vol.8 , Issue.4 SUPPL. , pp. 579-581
    • Weisser, B.1    Veiling, P.2    Geller, C.3    Kraft, K.4    Göbel, B.5    Vetter, H.6
  • 21
    • 0025049455 scopus 로고
    • Direct and indirect blood pressure in critical care patients
    • Byra-Cook CJ, Dracup KA, Lazik AJ: Direct and indirect blood pressure in critical care patients. Nurs Res 1990, 39:285-288.
    • (1990) Nurs Res , vol.39 , pp. 285-288
    • Byra-Cook, C.J.1    Dracup, K.A.2    Lazik, A.J.3
  • 22
    • 0021986832 scopus 로고
    • A comparison of two automated indirect arterial blood pressure meters: With recordings from a radial arterial catheter in anesthetized surgical patients
    • Nystrom E, Reid KH, Bennett R, Conture L, Edmonds HL Jr: A comparison of two automated indirect arterial blood pressure meters: with recordings from a radial arterial catheter in anesthetized surgical patients. Anesthesiology 1985, 62:526-530.
    • (1985) Anesthesiology , vol.62 , pp. 526-530
    • Nystrom, E.1    Reid, K.H.2    Bennett, R.3    Conture, L.4    Edmonds Jr., H.L.5
  • 23
    • 0021920324 scopus 로고
    • Invasive v. noninvasive measurement of arterial pressure
    • Van Egmond J, Hasenbos M, Crul JH: Invasive v. noninvasive measurement of arterial pressure. Br J Anaesth 1985, 57:434-444.
    • (1985) Br J Anaesth , vol.57 , pp. 434-444
    • Van Egmond, J.1    Hasenbos, M.2    Crul, J.H.3
  • 24
    • 0023940672 scopus 로고
    • Wideband external pulse recording during cuff deflation: A new technique for evaluation of the arterial pressure pulse and measurement of blood pressure
    • Blank SG, West JE, Müller FB, Cody RJ, Harshfield GA, Pecker MS, et al.: Wideband external pulse recording during cuff deflation: a new technique for evaluation of the arterial pressure pulse and measurement of blood pressure. Circulation 1988, 77:1297-1305.
    • (1988) Circulation , vol.77 , pp. 1297-1305
    • Blank, S.G.1    West, J.E.2    Müller, F.B.3    Cody, R.J.4    Harshfield, G.A.5    Pecker, M.S.6
  • 25
  • 26
    • 0022869969 scopus 로고
    • Noninvasive measurement of baroreflex sensitivity index using an indirect and continuous blood-pressure recorder
    • Tochikubo O, Kaneko Y, Yukinari Y, Takeda I: Noninvasive measurement of baroreflex sensitivity index using an indirect and continuous blood-pressure recorder. Jpn Heart J 1986, 27:849-857.
    • (1986) Jpn Heart J , vol.27 , pp. 849-857
    • Tochikubo, O.1    Kaneko, Y.2    Yukinari, Y.3    Takeda, I.4
  • 27
    • 0004661140 scopus 로고
    • Arrival time and calibrated contour of the pulse wave determined indirectly from recordings of arterial compression sounds
    • Rodbard S, Rubinstein HM, Rosenblum S: Arrival time and calibrated contour of the pulse wave determined indirectly from recordings of arterial compression sounds. Am Heart J 1957, 53:205-212.
    • (1957) Am Heart J , vol.53 , pp. 205-212
    • Rodbard, S.1    Rubinstein, H.M.2    Rosenblum, S.3
  • 28
    • 0029609347 scopus 로고
    • Factors influencing validation of ambulatory blood pressure measuring devices
    • O'Brien E, Atkins N, Staessen J: Factors influencing validation of ambulatory blood pressure measuring devices. J Hypertens 1995, 13:1235-1240.
    • (1995) J Hypertens , vol.13 , pp. 1235-1240
    • O'Brien, E.1    Atkins, N.2    Staessen, J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.