Volumn 13, Issue 1, 1994, Pages 176-185

Performance of QPET, a High-Resolution 3D PET Imaging System for Small Volumes

Author keywords

[No Author keywords available]

Indexed keywords

CORRELATION DETECTORS; GEOMETRY; IMAGE RECONSTRUCTION; MATHEMATICAL MODELS; MEDICAL IMAGING; OPTICAL RESOLVING POWER; PERFORMANCE; POSITRON EMISSION TOMOGRAPHY; SAMPLING; SENSITIVITY ANALYSIS; SIGNAL TO NOISE RATIO; THREE DIMENSIONAL;

EID: 0028393332     PISSN: 02780062     EISSN: 1558254X     Source Type: Journal    
DOI: 10.1109/42.276156     Document Type: Article
Times cited : (11)

References (31)
  • 3
    • 0025416371 scopus 로고
    • High resolution positron emission tomography with a prototype camera based on solid state scintillation detectors
    • R. Lecomte, J. Cadorette, M. Hon, D. Rouleau, and G. Gauthier, “High resolution positron emission tomography with a prototype camera based on solid state scintillation detectors,” IEEE Trans. Nucl. Sci., vol. 37, pp. 805-811, 1990.
    • (1990) IEEE Trans. Nucl. Sci. , vol.37 , pp. 805-811
    • Lecomte, R.1    Cadorette, J.2    Hon, M.3    Rouleau, D.4    Gauthier, G.5
  • 7
    • 0025418432 scopus 로고
    • Detector, shielding and geometric design factors for a high-resolution PET system
    • W. M. Digby, M. Dáhlbom, and E.J. Hoffman, “Detector, shielding and geometric design factors for a high-resolution PET system, “ IEEE Trans. Nucl. Sci.,. vol. 37, pp. 664-670, 1990.
    • (1990) IEEE Trans. Nucl. Sci. , vol.37 , pp. 664-670
    • Digby, W.M.1    Dáhlbom, M.2    Hoffman, E.J.3
  • 8
    • 0026136787 scopus 로고
    • Dynamic high resolution imaging of rats: Design considerations
    • R. S. Miyaoka, T. K. Lewellen, and A. N. Bice, “Dynamic high resolution imaging of rats: Design considerations,” IEEE Trans. Nucl. Sci., vol. 38, pp: 670-677, 1991.
    • (1991) IEEE Trans. Nucl. Sci. , vol.38 , pp. 670-677
    • Miyaoka, R.S.1    Lewellen, T.K.2    Bice, A.N.3
  • 11
    • 0024373225 scopus 로고
    • The performance of a multiwire proportional chamber positron camera for clinical use
    • P. K. Marsden, R. J. Ott, J. E. Bateman, S. R. Cherry, M. A. Flower, and S. Webb, “The performance of a multiwire proportional chamber positron camera for clinical use,” Phys. Med. Biol., vol. 34, pp. 1034-1062, 1989.
    • (1989) Phys. Med. Biol. , vol.34 , pp. 1034-1062
    • Marsden, P.K.1    Ott, R.J.2    Bateman, J.E.3    Cherry, S.R.4    Flower, M.A.5    Webb, S.6
  • 12
    • 0026074175 scopus 로고
    • Image reconstruction for a 3D PET system using a minimum norm constraint
    • M. J. Hogan, P. J. Harvey, D. C. Howse, and B. T. A. McKee, “Image reconstruction for a 3D PET system using a minimum norm constraint,” Phys. Med. Biol., vol. 36, pp. 35-44, 1991.
    • (1991) Phys. Med. Biol. , vol.36 , pp. 35-44
    • Hogan, M.J.1    Harvey, P.J.2    Howse, D.C.3    McKee, B.T.A.4
  • 14
    • 0007539291 scopus 로고
    • A deconvolution scatter correction for a 3-D PET system
    • B. T. A. McKee, A. T. Gurvey, P. J. Harvey, and D. C. Howse, “A deconvolution scatter correction for a 3-D PET system,” IEEE Trans. Med. Imaging, vol. 11, no. 4, pp. 560-569, 1992.
    • (1992) IEEE Trans. Med. Imaging , vol.11 , Issue.4 , pp. 560-569
    • McKee, B.T.A.1    Gurvey, A.T.2    Harvey, P.J.3    Howse, D.C.4
  • 15
    • 0021715108 scopus 로고
    • Increased sensitivity and field of view for a rotating positron camera
    • R. Clack, D. Townsend, and A. Jeavons, “Increased sensitivity and field of view for a rotating positron camera,” Phys. Med. Biol., vol. 29, pp. 1421-1431, 1984.
    • (1984) Phys. Med. Biol. , vol.29 , pp. 1421-1431
    • Clack, R.1    Townsend, D.2    Jeavons, A.3
  • 16
    • 0020088574 scopus 로고
    • The high-density avalanche chamber for positron emission tomography
    • A. Jeavons, K. Hood, G. Herlin, and C. Parkman, “The high-density avalanche chamber for positron emission tomography,” IEEE Trans. Nucl. Sci., vol. 30, pp. 640-644, 1983.
    • (1983) IEEE Trans. Nucl. Sci. , vol.30 , pp. 640-644
    • Jeavons, A.1    Hood, K.2    Herlin, G.3    Parkman, C.4
  • 17
    • 0022111983 scopus 로고
    • Proportional chambers for very high counting rates based on gas mixtures of CF4 with hydrocarbons
    • J. Fischer, A. Hrisoho, V. Radeka, and P. Rehak, “Proportional chambers for very high counting rates based on gas mixtures of CF4 with hydrocarbons,” Nucl. Instr. and Meth., vol. A238, pp. 249-264, 1985.
    • (1985) Nucl. Instr. and Meth. , vol.A238 , pp. 249-264
    • Fischer, J.1    Hrisoho, A.2    Radeka, V.3    Rehak, P.4
  • 18
    • 44949271659 scopus 로고
    • Data acquisition for a medical imaging MWPC detector
    • B. T. A. McKee, P. J. Harvey, and J. D. MacPhail, “Data acquisition for a medical imaging MWPC detector,” Nucl. Inst. & Meth., vol. A310, pp. 545-551, 1991.
    • (1991) Nucl. Inst. & Meth. , vol.A310 , pp. 545-551
    • McKee, B.T.A.1    Harvey, P.J.2    MacPhail, J.D.3
  • 19
    • 0022669524 scopus 로고
    • The effect of live time in components of a positron tomograph on image quantification
    • C. G. Thompson and E. Meyer, “The effect of live time in components of a positron tomograph on image quantification,” IEEE Trans. Nucl. Sci., vol. 33, pp. 337-343, 1987.
    • (1987) IEEE Trans. Nucl. Sci. , vol.33 , pp. 337-343
    • Thompson, C.G.1    Meyer, E.2
  • 20
    • 0024330187 scopus 로고
    • Image quantification with a large area multiwire proportional chamber positron camera (MUP-PET)
    • S. R. Cherry, P. K. Marsden, R. J. Ott, M. A. Flower, S. Webb, and J. W. Babich, “Image quantification with a large area multiwire proportional chamber positron camera (MUP-PET),” European J. Nucl. Med., vol. 15, 694-700, 1989.
    • (1989) European J. Nucl. Med. , vol.15 , pp. 694-700
    • Cherry, S.R.1    Marsden, P.K.2    Ott, R.J.3    Flower, M.A.4    Webb, S.5    Babich, J.W.6
  • 21
    • 0020618677 scopus 로고
    • Predicting count loss in modem PET systems
    • T. J. Holmes, “Predicting count loss in modem PET systems,” IEEE Trans. Nucl. Sci., vol. 30, pp. 723-728, 1983.
    • (1983) IEEE Trans. Nucl. Sci. , vol.30 , pp. 723-728
    • Holmes, T.J.1
  • 22
    • 0025414253 scopus 로고
    • Measuring PET scanner sensitivity: Relating countrates to image signal-to-noise ratios using noise equivalent counts
    • S. C. Strother, M. E. Casey, and E. J. Hoffman, “Measuring PET scanner sensitivity: Relating countrates to image signal-to-noise ratios using noise equivalent counts,” IEEE Trans. Nucl. Sci., vol. 37, pp. 783-788, 1990.
    • (1990) IEEE Trans. Nucl. Sci. , vol.37 , pp. 783-788
    • Strother, S.C.1    Casey, M.E.2    Hoffman, E.J.3
  • 23
    • 0018939380 scopus 로고
    • Method for optimising side shielding in positron emission tomographs and for comparing detector materials
    • S. E. Derenzo, “Method for optimising side shielding in positron emission tomographs and for comparing detector materials,” J. Nucl. Med., vol. 21, pp. 971-977, 1980.
    • (1980) J. Nucl. Med. , vol.21 , pp. 971-977
    • Derenzo, S.E.1
  • 24
    • 0020911246 scopus 로고
    • Performance study of single-slice positron emission tomography scanners by Monte Carlo techniques
    • L. R. Lupton and N. A. Keller, “Performance study of single-slice positron emission tomography scanners by Monte Carlo techniques,” IEEE Trans. Med. Imaging, vol. 2, pp. 154-168, 1983.
    • (1983) IEEE Trans. Med. Imaging , vol.2 , pp. 154-168
    • Lupton, L.R.1    Keller, N.A.2
  • 25
    • 0024069138 scopus 로고
    • Compton scattering in a largeaperture positron imaging system
    • B. T. A. McKee, M. J. Hogan, and D. C. N. Howse, “Compton scattering in a largeaperture positron imaging system,” IEEE Trans. Med. Imaging, vol. 7, 198-202, 1988.
    • (1988) IEEE Trans. Med. Imaging , vol.7 , pp. 198-202
    • McKee, B.T.A.1    Hogan, M.J.2    Howse, D.C.N.3
  • 26
    • 84944380772 scopus 로고
    • System characterization of the high resolution positron emission tomography scanner at Queens University
    • Queens University, Kingston, Ontario, Canada
    • A. W. Dickson, “System characterization of the high resolution positron emission tomography scanner at Queens University,” M.Sc. thesis, Queens University, Kingston, Ontario, Canada, 1991.
    • (1991) M.Sc. thesis
    • Dickson, A.W.1
  • 27
    • 0026922961 scopus 로고
    • Effect of detector weighting functions on the point spread function of high-resolution PET tomographs: a simulation
    • B. Karuta and R. Lecomte, “Effect of detector weighting functions on the point spread function of high-resolution PET tomographs: a simulation,” IEEE Trans. Med. Imaging, vol. 11, pp. 379-385, 1992.
    • (1992) IEEE Trans. Med. Imaging , vol.11 , pp. 379-385
    • Karuta, B.1    Lecomte, R.2
  • 28
    • 0024612188 scopus 로고
    • Compensation for crystal penetration in high resolution positron tomography
    • R. H. Huesman, E. M. Salmeron, and J. R. Baker, “Compensation for crystal penetration in high resolution positron tomography,” IEEE Trans. Nucl. Sci., vol. 36, pp. 1100-1107, 1989.
    • (1989) IEEE Trans. Nucl. Sci. , vol.36 , pp. 1100-1107
    • Huesman, R.H.1    Salmeron, E.M.2    Baker, J.R.3
  • 29
    • 0024610834 scopus 로고
    • PET system calibrations and corrections for quantitative and spatially accurate images
    • E. J. Hoffman, T. M. Guerrero, G. Germano, W. M. Digby, and M. Dahlbom, “PET system calibrations and corrections for quantitative and spatially accurate images,” IEEE Trans. Nucl. Sci., vol. 36, pp. 1108-1112, 1989.
    • (1989) IEEE Trans. Nucl. Sci. , vol.36 , pp. 1108-1112
    • Hoffman, E.J.1    Guerrero, T.M.2    Germano, G.3    Digby, W.M.4    Dahlbom, M.5
  • 30
    • 0026505062 scopus 로고
    • Removing the interference error between attenuation correction and scatter subtraction in 3D PET imaging
    • P. J. Harvey, D. C. Howse, and B.T.A. McKee, “Removing the interference error between attenuation correction and scatter subtraction in 3D PET imaging,” Phys. Med. Biol., vol. 37, pp. 767-777, 1992.
    • (1992) Phys. Med. Biol. , vol.37 , pp. 767-777
    • Harvey, P.J.1    Howse, D.C.2    McKee, B.T.A.3